Quick Links: Skip to main page content
Skip to Search
Skip to Topics Menu
Skip to Common Links
U.S. Food & Drug Administration
Follow FDA
En Español
Search FDA
Home
Food
Drugs
Medical Devices
Radiation-Emitting Products
Vaccines, Blood & Biologics
Animal & Veterinary
Cosmetics
Tobacco Products
TPLC - Total Product Life Cycle
FDA Home
medical devices
databases
-
510(k)
|
DeNovo
|
Registration & Listing
|
Adverse Events
|
Recalls
|
PMA
|
HDE
|
Classification
|
Standards
CFR Title 21
|
Radiation-Emitting Products
|
X-Ray Assembler
|
Medsun Reports
|
CLIA
|
TPLC
New Search
show TPLC since
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
2025
2026
Back to Search Results
Device
implantable pulse generator
Definition
Deliver Cardiac Contractility Modulation signals during the absolute refractory period
Product Code
QFV
Device Class
3
Premarket Approvals (PMA)
2021
2022
2023
2024
2025
2026
4
5
9
10
5
2
MDR Year
MDR Reports
MDR Events
2022
1
1
2023
24
24
2024
45
45
2025
31
31
2026
33
33
Device Problems
MDRs with this Device Problem
Events in those MDRs
Adverse Event Without Identified Device or Use Problem
42
42
Computer Software Problem
26
26
Failure to Charge
16
16
Prophylactic removal due to corrective action
5
5
Misassembly by Users
5
5
Aborted Charge
4
4
Insufficient Device Problem Information
4
4
Signal Artifact/Noise
2
2
Failure to Interrogate
2
2
Incorrect, Inadequate or Imprecise Result or Readings
2
2
Charging Problem
2
2
Under-Sensing
2
2
Failure to Deliver Energy
1
1
Interrogation Problem
1
1
High impedance
1
1
Mechanical Problem
1
1
Extracardiac stimulation
1
1
Defective Component
1
1
Battery Problem
1
1
Impedance Problem
1
1
Device Difficult to Setup or Prepare
1
1
Device Sensing Problem
1
1
Installation-Related Problem
1
1
Migration
1
1
Device Dislodged or Dislocated
1
1
Electromagnetic Interference
1
1
Improper or Incorrect Procedure or Method
1
1
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
Failure of Implant
37
37
No Clinical Signs, Symptoms or Conditions
30
30
Heart Failure/Congestive Heart Failure
8
8
Unspecified Infection
7
7
Chest Pain
4
4
Discomfort
4
4
Cardiovascular Dysfunction/Insufficiency
3
3
Tricuspid Valve Insufficiency/ Regurgitation
3
3
Medical device site infection
3
3
Pain
3
3
Bacteremia
3
3
Bacterial Infection
3
3
Electric Shock
2
2
Sepsis
2
2
Fungal Infection
2
2
Diaphoresis
2
2
Shock from Patient Lead(s)
2
2
Atrial Fibrillation
2
2
Post Operative Wound Infection
2
2
Insufficient Information
2
2
Endocarditis
2
2
Respiratory Insufficiency
2
2
Skin Infection
1
1
Gastrointestinal Hemorrhage
1
1
Cognitive Changes
1
1
Pericardial Effusion
1
1
Cramp(s) /Muscle Spasm(s)
1
1
Device Overstimulation of Tissue
1
1
Twiddlers Syndrome
1
1
Dehydration
1
1
Fall
1
1
Dysphagia/ Odynophagia
1
1
Tachycardia
1
1
Swelling/ Edema
1
1
Inadequate Pain Relief
1
1
Pocket Erosion
1
1
Recalls
Manufacturer
Recall Class
Date Posted
1
IMPULSE DYNAMICS (USA) INC
II
Feb-22-2024
-
-