| Trade Name | OP-1 IMPLANT |
| Classification Name | filler, recombinant human bone morphogenetic protein, collagen scaffold, osteoinduction |
| Generic Name | filler, recombinant human bone morphogenetic protein, collagen scaffold, osteoinduction |
| Applicant |
| Olympus Biotech Corporation |
| 35 s. st. |
| hopkinton, MA 01748 |
|
| HDE Number | H010002 |
| Supplement Number | S027 |
| Date Received | 06/30/2009 |
| Decision Date | 02/24/2011 |
| Product Code | |
| Advisory Committee |
Orthopedic |
| Supplement Type | normal 75 day track |
| Supplement Reason | process change - manufacturer/sterilizer/packager/supplier |
| Expedited Review Granted? | No |
| Combination Product | Yes |
Approval Order Statement Approval for:1) an alternate collagen matrix manufacturing facility within the existing building 3 on thewest lebanon, nh campus;2) minor collagen matrix manufacturing process changes to accommodate a new suite andequipment;3) an alternate collagen matrix container closure;4) collagen matrix release specification changes; and5) use of both tibia and femur bones as starting material for the manufacture of the raw material for collagen matrix component. |