| Trade Name | WINGSPAN STENT SYSTEM WITH GATEWAY PTA BALLOON CATHETER |
| Classification Name | intracranial neurovascular stent |
| Generic Name | intracranial neurovascular stent |
| Applicant |
| Stryker Corp. |
| 47900 bayside pkwy. |
| fremont, CA 94538 |
|
| HDE Number | H050001 |
| Supplement Number | S080 |
| Date Received | 11/21/2016 |
| Decision Date | 12/19/2016 |
| Product Code | |
| Advisory Committee |
Neurology |
| Supplement Type | 30-day notice |
| Supplement Reason | process change - manufacturer/sterilizer/packager/supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
Approval Order Statement Acceptance for a change to the vendor of the linear low density polyethylene (lldpe) resin and a change to the vendor for the lldpe resin mixing/pelletization (base resin mixed with colorant) used in the bonding layer of the extruded tri-layer inner catheter shaft of the gateway pta balloon catheter. |