| Trade Name | Wingspan Stent System with Gateway PTA Balloon Catheter |
| Classification Name | intracranial neurovascular stent |
| Generic Name | intracranial neurovascular stent |
| Applicant |
| Stryker Corp. |
| 47900 bayside pkwy. |
| fremont, CA 94538 |
|
| HDE Number | H050001 |
| Supplement Number | S081 |
| Date Received | 03/20/2017 |
| Decision Date | 04/18/2017 |
| Product Code | |
| Advisory Committee |
Neurology |
| Supplement Type | 30-day notice |
| Supplement Reason | process change - manufacturer/sterilizer/packager/supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
Approval Order Statement Updates to your vacuum pump activation settings used in sterilization chambers 9 and 10 at your boston scientific (bsc) coventry rhode island site for sterilization of your gateway pta balloon catheter. |