| Trade Name | Wingspan Stent System with Gateway PTA Balloon Catheter |
| Classification Name | intracranial neurovascular stent |
| Generic Name | intracranial neurovascular stent |
| Applicant |
| Stryker Corp. |
| 47900 bayside pkwy. |
| fremont, CA 94538 |
|
| HDE Number | H050001 |
| Supplement Number | S082 |
| Date Received | 08/21/2017 |
| Decision Date | 09/15/2017 |
| Product Code | |
| Advisory Committee |
Neurology |
| Supplement Type | 30-day notice |
| Supplement Reason | process change - manufacturer/sterilizer/packager/supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
Approval Order Statement A manufacturing change to remove the in-process burst testing for the outer shaft of gateway pta balloon catheter. This change does not involve the wingspan® stent system and also does not impact the burst testing performed on the distal balloon portion of the gateway pta catheter. |