| Class 2 Device Recall OEC Uroview 2800 | |
Date Initiated by Firm | November 20, 2006 |
Date Posted | June 22, 2012 |
Recall Status1 |
Terminated 3 on June 25, 2012 |
Recall Number | Z-1865-2012 |
Recall Event ID |
62016 |
510(K)Number | K993687 |
Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code JAA
|
Product | OEC Uroview 2800, Model number A349855.
Designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical, and interventional procedures. |
Code Information |
All Serial Numbers: P4-0004-L , P4-0005-L , P4-0006-R-RC1 , P4-0008-R , P4-0011-L , P6-0654-R , P4-0012-R , P4-0013-L , P4-0014-L , P4-0015-L , P4-0016-L , P4-0017-L , P4-0018-L , P4-0019-R , P4-0020-L , P4-0021-L , P4-0022-R , P4-0023-L , P4-0024-R , P4-0025-R , P4-0026-L , P4-0027-R , P4-0028-R , P4-0029-R , P4-0030-R , P4-0031-L , P4-0680-L , P4-0032-R , P4-0033-R , P4-0034-R , P4-0035-R , P4-0036-R , P4-0037-R , P4-0038-L , P4-0039-R , P4-0042-R , P4-0044-L , P4-0045-R , P4-0046-L , P4-0047-L-RC , P4-0048-L , P4-0049-R , P4-0050-L , P4-0051-L , P4-0052-R , P4-0053-L , P4-0054-R , P4-0074-R , P4-0055-L , P4-0058-R , P4-0059-L , P4-0062-L , P4-0064-L , P4-0067-L , P4-0075-L , P4-0068-R , P4-0070-L , P4-0071-L , P4-0073-R , P4-0076-R , P4-0078-R , P4-0081-L , P4-0082-R , P4-0388-L , P4-0084-R , P4-0085-R , P4-0087-R , P4-0088-L , P4-0089-L , P4-0090-L , P4-0092-R , P4-0093-R , P4-0094-R , P4-0095-R , P4-0097-R , P4-0098-R , P4-0099-L , P4-0100-R , P4-0101-R , P4-0102-R , P4-0103-R , P4-0104-R , P4-0105-R , P4-0107-L , P4-0108-R , P4-0110-L , P4-0111-L , P4-0112-L , P4-0113-R , P4-0114-L , P4-0661-L , P4-0115-R , P4-0117-R , P4-0118-R , P4-0119-L , P4-0120-L , P4-0121-L , P4-0123-L , P4-0124-R , P4-0127-L , P4-0128-R , P4-0131-R , P4-0132-R , P4-0133-L , P4-0138-R , P4-0140-L , P4-0141-R , P4-0142-R, P4-0154-R , P4-0696-L , P4-0143-L , P4-0146-R , P4-0147-L , P4-0149-R , P4-0211-R , P4-0151-L , P4-0153-R , P4-0155-R , P4-0159-L , P4-0160-L , P4-0161-L , P4-0162-R , P4-0163-R , P6-0529-L , P4-0164-R , P4-0165-L , P4-0167-R , P4-0168-R , P4-0169-R , P4-0170-R , P4-0172-R , P4-0175-L , P4-0176-L , P4-0177-R , P4-0178-R , P4-0179-L , P4-0181-L , P4-0183-R , P4-0185-R , P4-0301-R , P4-0187-L , P4-0188-L , P4-0190-R , P4-0191-L , P4-0194-L , P4-0195-R , P4-0196-L , P4-0197-L , P4-0199-R , P4-0201-R , P4-0202-R , P4-0203-L , P4-0204-R , P4-0205-R , P4-0206-L , P4-0209-L , P4-0212-L , P4-0215-L , P4-0216-R , P4-0223-R , P4-0232-L , P4-0237-R , P4-0238-L , P4-0239-R , P4-0240-L , P4-0241-L , P4-0242-R , P4-0244-R , P4-0246-R , P4-0247-R , P4-0248-R , P4-0451-L , P4-0254-L , P4-0255-R , P4-0256-R , P4-0257-L , P4-0258-L , P4-0259-L , P4-0445-L , P4-0260-R , P4-0263-L , P4-0264-R , P4-0268-R , P4-0659-R , P4-0269-R , P4-0271-R , P4-0273-L , P6-0715-L , P4-0274-R , P4-0277-L , P4-0279-R , P4-0685-L , P4-0280-R , P4-0282-L , P4-0281-L , P4-0284-L , P4-0285-L , P4-0288-R , P4-0289-L , P4-0292-L , P4-0293-R , P4-0294-L , P4-0296-R , P4-0297-R , P4-0298-R , P4-0299-L , P4-0300-R , P4-0318-L , P4-0302-L , P4-0304-R , P4-0305-L , P4-0306-R , P4-0308-L , P4-0309-R , P4-0310-L , P4-0313-L , P4-0315-L , P4-0316-R , P4-0457-R , P4-0319-R , P4-0321-L , P4-0322-L , P4-0325-R , P4-0326-R , P4-0327-L , P4-0328-R , P4-0329-L , P4-0332-L , P4-0333-R , P4-0335-R , P4-0339-L , P4-0340-R , P4-0341-R , P4-0342-L , P4-0343-L , P4-0345-R , P4-0346-L , P4-0347-L , P4-0348-R , P4-0349-L , P4-0353-L , P4-0354-R , P4-0355-L , P4-0356-L , P4-0359-R , P4-0361-R , P4-0362-R , P4-0365-R , P4-0367-R , P4-0370-L , P4-0373-R , P4-0377-L , P4-0380-L , P4-0381-L , P4-0382-R , P4-0385-R , P4-0386-R , P4-0389-L , P4-0391-L , P4-0393-R , P4-0394-R , P4-0401-R , P4-0402-R , P4-0405-L , P4-0406-L , P4-0407-R , P4-0408-R , P4-0409-R , P4-0410-R , P4-0411-L , P4-0413-L , P4-0416-R , P4-0415-L , P4-0417-R , P4-0418-L-RC , P4-0419-L , P4-0422-L , P4-0424-R , P4-0425-R , P4-0426-L , P4-0428-R , P4-0431-R , P4-0436-R , P4-0440-R , P4-0442-L , P6-0472-R , P4-0443-R , P4-0444-R , P4-0446-R , P4-0447-L , P4-0448-L , P4-0449-L , P4-0450-L , P4-0452-L , P4-0453-R , P4-0454-L , P4-0455-L , P4-0456-R , P4-0459-R , P4-0460-R , P4-0463-L , P4-0464-L , P4-0465-R , P4-0466-R , P4-0467-L , P4-0469-R , P4-0471-R , P4-0473-R , P4-0477-L , P4-0478-L , P4-0482-R , P4-0483-L , P4-0487-R , P4-0489-L , P4-0491-L , P4-0492-L , P4-0495-L , P4-0496-L , P4-0498-L , P4-0500-L , P4-0503-R , P4-0504-L , P4-0505-L , P4-0506-R , P4-0507-R , P4-0508-L , P4-0511-R , P4-0517-L , P4-0523-R , P4-0524-L , P4-0531-R , P4-0532-R , P4-0537-L , P4-0539-L , P4-0540-L , P4-0541-L , P4-0545-L , P4-0548-L , P4-0550-L , P4-0551-L , P4-0560-R , P4-0561-L , P4-0566-L , P4-0569-R , P4-0570-R , P4-0573-R , P4-0574-R , P4-0575-L , P4-0576-L , P4-0581-R , P4-0588-R , P4-0589-L , P4-0590-R , P4-0591-R , P4-0592-R , P4-0594-L , P4-0595-L , P4-0596-L , P4-0597-R , P4-0602-L , P4-0603-R , P4-0605-R , P4-0607-L , P4-0608-R , P4-0611-L , P4-0613-R , P4-0615-L , P4-0616-R , P4-0626-R , P4-0628-L , P4-0631-L , P4-0633-L , P4-0636-R , P4-0637-R , P4-0638-R , P4-0639-R , P4-0640-R , P4-0641-R , P4-0642-R , P4-0644-L , P4-0645-R , P4-0648-L , P4-0650-L , P4-0652-L , P4-0653-L , P4-0655-R , P4-0656-R , P4-0657-L , P6-0598-L , P4-0662-L , P4-0666-L , P4-0668-R , P4-0669-R , P4-0670-R , P4-0679-R , P4-0686-L , P4-0694-R , P4-0701-R , P4-0702-R , P4-0717-L , P4-0723-R , P4-0730-L , P6-0040-R , P6-0043-R, P6-0214-R, P6-0219-L, P6-0220-L, P6-0221-R, P6-0222-R,P6-0225-R, P6-0227-R, P6-0230-L, P6-0249-R, P6-0251-L, P6-0252-L , P6-0253 , P6-0056-L , P6-0061-L-RC , P6-0066-R , P6-0069-R , P6-0072-R , P6-0077-R , P6-0079-L , P6-0080-L , P6-0083-L , P6-0091-R , P6-0096-R , P6-0106-R , P6-0109-L , P6-0116-R , P6-0122-L , P6-0125-R , P6-0126-L , P6-0582-L , P6-0129-R , P6-0130-R , P6-0134-R, P6-0144-L , P6-0182-L , P6-0135-L , P6-0136-R , P6-0137-R , P6-0139-R , P6-0145-L , P6-0148-R-RC , P6-0150-L , P6-0152-L , P6-0156-L , P6-0157-R , P6-0166-L , P6-0171-L , P6-0184-L , P6-0186-R , P6-0192-L , P6-0193-R , P6-0198-L , P6-0200-R , P6-0210-L , P6-0213-L , P6-0218-R , P6-0226-R , P6-0228-R , P6-0229-R , P6-0231-L , P6-0233-L , P6-0234-R , P6-0235-R , P6-0699-R , P6-0236-R , P6-0243-L , P6-0245-L , P6-0250-R , P6-0261-R , P6-0262-R , P6-0265-L , P6-0267-R , P6-0270-L-RC , P6-0709-R , P6-0275-L , P6-0276-L , P6-0278-R , P6-0283-R , P6-0286-R , P6-0287-L , P6-0290-R , P6-0291-L , P6-0295-R , P6-0303-L , P6-0307-L , P6-0311-R , P6-0312-L , P6-0314-L , P6-0317-R , P6-0320-R , P6-0324-R , P6-0330-L , P6-0336-L , P6-0338-L , P6-0344-R , P6-0351-L , P6-0704-L , P6-0352-R , P6-0358-R , P6-0360-L , P6-0363-L , P6-0364-L , P6-0366-R , P6-0371-R , P6-0372-R , P6-0375-L , P6-0378-L , P6-0383-R , P6-0387-L , P6-0392-R , P6-0397-L , P6-0398-L , P6-0399-L , P6-0412-R , P6-0414-R , P6-0420-R , P6-0421-R , P6-0423-R , P6-0427-L , P6-0432-R , P6-0435-R , P6-0437-R , P6-0438-R , P6-0439-L , P6-0461-L , P6-0462-R , P6-0468-L , P6-0470-R , P6-0479-R , P6-0480-R , P6-0486-R , P6-0493-R , P6-0494-R , P6-0497-R , P6-0499-R , P6-0502-R , P6-0522-R , P6-0509-L , P6-0510-L , P6-0512-L , P6-0513-R , P6-0514-R , P6-0515-L , P6-0516-R , P6-0518-R , P6-0519-R , P6-0520-L , P6-0521-R , P6-0530-R , P6-0533-L , P6-0534-L , P6-0538-R , P6-0542-R , P6-0543-R , P6-0544-R , P6-0546-L , P6-0549-R , P6-0552-L , P6-0553-R , P6-0554-R , P6-0555-L , P6-0558-L , P6-0559-L , P6-0562-R , P6-0563-R , P6-0564-R , P6-0565-R , P6-0567-R , P6-0571-L , P6-0572-L , P6-0577-R , P6-0578-R , P6-0579-R , P6-0583-R , P6-0584-R , P6-0585-L , P6-0586-L , P6-0587-L , P6-0599-L , P6-0600-R , P6-0612-R , P6-0614-L , P6-0617-L , P6-0618-R , P6-0619-L , P6-0620-L , P6-0621-R , P6-0622-L , P6-0623-R , P6-0624-L , P6-0625-R , P6-0627-L , P6-0629-L , P6-0632-R , P6-0634-R , P6-0635-L , P6-0643-R , P6-0646-L , P6-0647-R , P6-0658-R , P6-0660-R , P6-0663-R , P6-0672-L , P6-0674-L , P6-0676-L , P6-0677-L , P6-0682-L , P6-0683-L , P6-0687-L , P6-0688-R , P6-0689-R , P6-0690-L , P6-0691-L , P6-0692-L , P6-0693-L , P6-0695-R , P6-0698-L , P6-0700-R , P6-0703-L , P6-0705-L , P6-0706-L , P6-0708-R , P6-0710-L , P6-0712-R , P6-0713-R , P6-0714-L , P6-0716-R , P6-0720-R , P6-0718-R , P6-0719-L , P6-0721-L , P6-0724-R , P6-0726-L , P4-0007-R-RC , P6-0009-L, and P4-0481-L. |
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc 384 Wright Brothers Dr Salt Lake City UT 84116-2862
|
For Additional Information Contact | 801-536-4516 |
Manufacturer Reason for Recall | Recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mode" and "digital spot mode", and the system does not reboot. |
FDA Determined Cause 2 | Software design |
Action | On 11/20/06 two written notifications were sent to consignees addressed to the Facility Administrator and the Radiology Administrator. The notifications advised them that GE Healthcare had identified several intermittent potential safety issues that may occur with the product. Interim solutions were provided for each issue. The letter stated that GEHC OEC would contact customers and remedy the issues free of charge when a permanent solution became available. An Urgent Recall Notice, dated April 13, 2007, was also sent to customers. Customers were to contact the service team at 800-874-7378 for further information, if there were any questions or concerns. |
Quantity in Commerce | 717 units |
Distribution | Worldwide Distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAA
|
|
|
|