| Date Initiated by Firm | May 04, 2012 |
| Date Posted | July 19, 2012 |
| Recall Status1 |
Terminated 3 on February 19, 2013 |
| Recall Number | Z-2037-2012 |
| Recall Event ID |
62372 |
| 510(K)Number | K101888 |
| Product Classification |
System, x-ray, tomography, computed - Product Code JAK
|
| Product | Hitachi Scenaria CT System Software
Product Usage: The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering. |
| Code Information |
Serial #'s S5001 - S5011 |
| FEI Number |
1528028
|
Recalling Firm/ Manufacturer |
Hitachi Medical Systems America Inc 1959 Summit Commerce Park Twinsburg OH 44087
|
| For Additional Information Contact | Douglas J. Thistlethwaite 330-425-1313 Ext. 3720 |
Manufacturer Reason for Recall | Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data. |
FDA Determined Cause 2 | Software design |
| Action | Hitachi sent a "Device Correction Notice" dated May 4, 2012 to their customers. The notice identified the affected product, problem and actions to be taken. Customers were instructed to "not use the Reconstruction Status" function to suspend active reconstructions until the software problem can be corrected. They also informed their customers that a representative will visit their facility to install the software update to permanently resolve the issue at no cost to the facility. For further information regarding this notice call 800-800-3106 x3720. |
| Quantity in Commerce | Six (6) units |
| Distribution | US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JAK
|