  Class 3 Device Recall Chromovision High Definition (HD) Wireless Camera see related information
Date Initiated by Firm July 10, 2012
Date Posted August 06, 2012
Recall Status1 Terminated 3 on June 06, 2013
Recall Number Z-2148-2012
Recall Event ID 62683
Product Classification Camera, cine, surgical, without audio - Product Code FWH
Product Chromovision High Definition (HD) Wireless Camera, Pt # CC-9715900.

Visual recording of surgical procedures.
Code Information Serial numbers: 350111-S10017, 350111-S10018, 350111-S10019, 350111-S10020, 350111-S10021, 350111-S10022, 350111-S10023, 350111-S10024, 350111-S10025, 350111-S10026, 350111-S10027, 350111-S10028, 350111-S10030, 350111-S10031, 350111-S10032, 350111-S10033, 350111-S10034, 350111-S10035, 350111-S10036, 350111-S10037, 350111-S10038, 350111-S10039, 350111-S10040, 350111-S10041, 350111-S10043, 350111-S10044, 350111-S10045, 350111-S10046, 350111-S10047, 350111-S10048, 350111-S10049, 350111-S10050, 350111-S10051, 350111-S10052, 350111-S10053, 350111-S10054, 350111-S10055, 350111-S10056, 350111-S10060, 350111-S10061, 350111-S10062, 350111-S10063, 350111-S10064, 350111-S10065, 350111-S10066, 350111-S10067, 350111-S10068, 350111-S10069, 350111-S10070, 350111-S10071, 350111-S10072, 350111-S10073, 350111-S10074, 350111-S10075, 350111-S10076, 350111-S10077, 350111-S10078, 350111-S10079, 350111-S10080, 350111-S10081, 350111-S10082, 350111-S10083, 350111-S10084, 350111-S10085, 350111-S10086, 350111-S10087, 350111-S10088, 350111-S10089, 350111-S10090, 350111-S10091, 350111-S10092, 350111-S10093, 350111-S10094, 350111-S10095, 350111-S10096, 350111-S10097, 350111-S10098, 350111-S10099, 350111-S10100, 350111-S10101, 350111-S10102, 350111-S10103, 350111-S10104, 350111-S10105, 350111-S10106, 350111-S10107, 350111-S10108, 350111-S10109, 350111-S10110, 350111-S10111, 350111-S10112, 350111-S10113, 350111-S10114, 350111-S10115, 350111-S10116, 350111-S10117, 350111-S10118, 350111-S10119, 350111-S10120, 350111-S10121, 350111-S10122, 350111-S10123, 350111-S10124, 350111-S10125, 350111-S10126, 350111-S10127, 350111-S10128, 350111-S10129, 350111-S10130, 350111-S10131, 350111-S10132, 350111-S10133, 350111-S10134, 350111-S10135, 350111-S10136, 350111-S10137, 350111-S10138, 350111-S10139, 350111-S10140, 350111-S10141, 350111-S10142, 350111-S10143, 350111-S10144, 350111-S10145, 350111-S10146, 350111-S10147, 350111-S10148, 350111-S10149, 350111-S10150, 350111-S10151, 350111-S10152, 350111-S10153, 350111-S10154, 350111-S10155, 350111-S10156, 350111-S10157, 350111-S10158, 350111-S10159, 350111-S10160, 350111-S10161, 350111-S10162, 350111-S10164, 350111-S10165, 350111-S10166, 350111-S10167, 350111-S10168, 350111-S10169, 350111-S10170, 350111-S10171, 350111-S10172, 350111-S10173, 350111-S10176, 350111-S10177, 350111-S10178, 350111-S10179, 350111-S10180, 350111-S10181, 350111-S10192, 350111-S10193, 350111-S10194, 350111-S10195, 350111-S10196, 350111-S10197, 350111-S10198, 350111-S10199, 350111-S10200, 350111-S10201, 350111-S10202, 350111-S10203, 350111-S10204, 350111-S10205, 350111-S10206, 350111-S10207, 350111-S10208, 350111-S10209, 350111-S10210, 350111-S10211, 350111-S10212, 350111-S10213, 350111-S10215, 350111-S10216, 350111-S10218, 350111-S10219, 350111-S10220, 350111-S10221, 350111-S10222, 350111-S10223, 350111-S10224, 350111-S10225, 350111-S10226, 350111-S10227, 350111-S10228, 350111-S10229, 350111-S10230, 350111-S10231, 350111-S10232, 350111-S10233, 350111-S10234, 350111-S10235, 350111-S10236, 350111-S10237, 350111-S10238, 350111-S10239, 350111-S10248, 350111-S10249, 350111-S10250, 350111-S10251, 350111-S10252, 350111-S10253, 350111-S10254, 350111-S10255, 350111-S10256, 350111-S10257, 350111-S10258, 350111-S10259, 350111-S10260, 350111-S10261, 350111-S10262, 350111-S10263, 350111-S10265, 350111-S10266, 350111-S10267, 350111-S10268, 350111-S10269, 350111-S10270, 350111-S10273, 350111-S10274, 350111-S10276, 350111-S10277, 350111-S10278, 350111-S10279, 350111-S10280, 350111-S10281, 350111-S10282, 350111-S10283, 350111-S10284, 350111-S10285, 350111-S10286, 350111-S10287, 350111-S10288, 350111-S10289, 350111-S10290, 350111-S10291, 350111-S10292, 350111-S10293, 350111-S10295, 350111-S10296, 350111-S10297, 350111-S10298, 350111-S10300, 350111-S10301, 350111-S10302, 350111-S10303, 350111-S10305, 350111-S10306, 350111-S10307, 350111-S10309 and 350111-S10310.
Recalling Firm/
Manufacturer
Berchtold Corp.
1950 Hanahan Rd
N Charleston SC 29406-4878
For Additional Information Contact Chris Mannarino
843-569-6100
Manufacturer Reason
for Recall
Defective camera: There is a possibility that the connection of the wires to the connector may become loose, causing image distortion or loss.
FDA Determined
Cause 2
Nonconforming Material/Component
Action BERCHTOLD Corporation sent a letter dated July 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. BERCHTOLD will replace the customer's cameras at no cost to the customer. Customers with questions should call 843-569-6100, ext 322. For questions regarding this recall call 843-569-6100.
Quantity in Commerce 17 cameras
Distribution Nationwide Distribution including FL, MI, MS, and NH.