Date Initiated by Firm | August 08, 2012 |
Date Posted | January 09, 2013 |
Recall Status1 |
Terminated 3 on May 12, 2014 |
Recall Number | Z-0658-2013 |
Recall Event ID |
62950 |
510(K)Number | K011046 K022345 K032852 K040453 K063602 |
Product Classification |
Compressor, cardiac, external - Product Code DRM
|
Product | AutoPulse Resuscitation System Model 100,
Mfg by:Zoll Circulation
Sunnyvale, CA.
The AutoPulse Model 100 Automatic Mechanical Chest compressor is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse. |
Code Information |
all |
Recalling Firm/ Manufacturer |
Zoll Circulation, Inc. 650 Almanor Ave Sunnyvale CA 94085-3513
|
For Additional Information Contact | Sam Nanavati 408-419-1030 |
Manufacturer Reason for Recall | Complaints associated with an unexpected stop in AutoPulse compressions. |
FDA Determined Cause 2 | Employee error |
Action | Zoll sent Customer Notification letters dated August 8, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Zoll Technical Support Line at 1 (800) 348-9011 for further assistance. |
Quantity in Commerce | 3881 in US, 1846 ROW. |
Distribution | Worldwide distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DRM 510(K)s with Product Code = DRM
|