| | Class 2 Device Recall Active Renin IRMA Kit |  |
| Date Initiated by Firm | January 20, 2012 |
| Date Posted | October 05, 2012 |
| Recall Status1 |
Terminated 3 on October 05, 2012 |
| Recall Number | Z-0016-2013 |
| Recall Event ID |
63140 |
| 510(K)Number | K981607 |
| Product Classification |
Radioimmunoassay, angiotensin i and renin - Product Code CIB
|
| Product | Active Renin IRMA Kit, Part Number DSL25100
Product Usage:
The Renin IRMA; DSL25100 immunoradiometric assay kit provides materials for the quantitative measurement of Renin in plasma. This assay is intended for in vitro diagnostic use. |
| Code Information |
Lot # 110919D, 111017C, 111017D, 111114C, 111212C. |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd MS/E1.NE.02 Brea CA 92821-6232
|
| For Additional Information Contact | Clair K. O'Donovan, Ph.D. 714-961-4483 |
Manufacturer Reason for Recall | The recall was initiated because Beckman Coulter has confirmed that the Active Renin IRMA Test Kit may exhibit a decrease in stability of the coated tubes. As a result, quality control and patient sample results may be falsely elevated by up to 72%. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Beckman Coulter sent an Urgent Production Correction letter dated January 10, 2012 to all their customers who purchased the Manual Immunoassay Active Renin IRMA Test Kit. The letter provides the customers with an explanation of the problem and actions to be taken. Customers were instructed to discontinue use and discard all remaining affected inventory. Customers were asked to complete and return the enclosed Response Form with in 10 days. Customers with questions contact Customer Service Support at http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633. |
| Quantity in Commerce | 216 units total (6 units in the US) |
| Distribution | Worldwide Distribution - US Natiowide including the state of: OH, CT, CA and the state of Argentina, Austria, Belgium, Brazil, Canada, China, Columbia, Czech Republic, France, Germany, Greece, Hungary, Italy, Kuwait, Netherlands, Poland, Portugal, Slovakia, Spain, Switzerland, Turkey, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CIB
|
|
|
|