| Class 2 Device Recall MAYFIELD Base Unit |  |
Date Initiated by Firm | November 05, 2012 |
Date Posted | December 19, 2012 |
Recall Status1 |
Terminated 3 on February 21, 2014 |
Recall Number | Z-0558-2013 |
Recall Event ID |
63776 |
Product Classification |
Accessories, operating-room, table (kit) - Product Code FWZ
|
Product | MAYFIELD Composite Series Base Unit, Model A3101, a Swivel Adaptor (A3018) is an integral component of the Base Unit, Model A3101.
The MAYFIELD Base Unit (A3101) is an operating room table accessory intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used. A Swivel Adaptor (A3018) is an integral component of the Base Unit A3101. |
Code Information |
The recalle product is identified with the Lot Numbers: 097547 and 096546. |
Recalling Firm/ Manufacturer |
Integra LifeSciences Corporation 4900 Charlemar Dr Bldg A Cincinnati OH 45227-1402
|
For Additional Information Contact | Michael Molloy 513-533-7979 |
Manufacturer Reason for Recall | An investigation of an adverse trend of complaints conducted by Integra LifeSciences identified various product performance issues which could impact the functionality of the firm's Mayfield Composite Series Base Units and Mayfield Skull Clamps. In order to prevent additional complaint reports, the firm decided to recall and replace certain lot codes of the Base Units and Skull Clamps that were d |
FDA Determined Cause 2 | Under Investigation by firm |
Action | Integra issued an 'Urgent: Voluntary Medical Device Recall Notification with 'Recall Acknowledgment and Return Form' to all Integra Sales Representatives and Customers via visit and E-mail on November 5, 2012. The recall notice informs the customers of the firm's receipt of complaints pin-pointing quality issues with the MAYFIELD Base Units and MAYFIELD Skull Clamps which may impair their functionality. The customers are asked to sign and return the enclosed Recall Acknowledgement and Return Form so as to document their receipt of the recall notification and replacement products. The replacement products are to be delivered by the Integra NeuroSpecialist who is also responsible for returning the signed forms to Integra as documentation that the recalled devices have been replaced. Should the customer(s) have questions or concerns, they are directed to contact Integra Customer Service at: 1-877-444-1114,Option 3 for Technical Support. |
Quantity in Commerce | 5 base units were distributed |
Distribution | Nationwide Distribution including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA TN, TX, UT, VA, WA and WI. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|