• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to ve

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to ve see related information
Date Initiated by Firm February 05, 2013
Date Posted March 27, 2013
Recall Status1 Terminated 3 on May 08, 2015
Recall Number Z-1002-2013
Recall Event ID 64461
510(K)Number K002519  K062154  
Product Classification System, image processing, radiological - Product Code LLZ
Product Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and 6.3


The software is a medical diagnostic system that allows the processing, review, analysis, communication, and media interchange of multidimensional digital images acquired from a variety of imaging devices. It provides the ability to review digital images from original DICOM data for multiple modalities, including CT, MR, D, CR, XA, US, NM, PET, etc. The software is not meant for primary image interpretation in mammography.
Code Information Serial Number 5080109343 5080109769 5080109726 5080123222 5080108401 5080107812 5080107804 5080107790 5080108428 5070933221 5070933248 5070933264 5070933272 5070933256 5080123109 5050411467 5080103442 5080103418 5090110913 5070934023 5080103434 5070931695 5080110775 5080113081  5080102977 5080103272 5080113340 5080121009 5080121122 5080121092 5080121106 5080121114 5080120827 5080111631 5080111755 5080116595 5080116609 5080116617 5080112883 5080122439 5080117214 5080119764 5080119772 5080119780 5080123095 5080121769 5080121750 5090102805 5070222244 5060213870 5070931709 5070216643 5090113203 5090103763 5090103771 5090104697 5090104700 5090104719 5090105030 5090105049 5090105057 5090106045 5090106053 5090106568 5090106584 5090106592 5090106606 5090110344 5090110352 5090112800 5060217876 5060217884 5070220144 5070218417 5070931687 5090113181 5090113173 5080103426 5060214044 5080119276 5080119284 5080110783 5080117206 5080119144 5080115483 5080121017 5080117532 5070219472 5080118911 5080108320 5090113548 5080107375 5020918865 5020919055 5080107405 5090113092 5080121963 540349 5050911664 5050911672 5070120134  5070120117  5050410843 5080119454 5080107367 5060218317 5050415063 5050910463 5080122404 5030334610 5080107499 3090900764 5090106460 5090106479 5090109648 457925 5030132252 5080121084 5060219208 5060218228 5060214206 5080116196 5070219405 5080121165 5030859349 5080115572 5080112905 5090111189 CRB1206112 5040916809 5090104298 5020915556 5090111804 5090111847 CRB1209129 5080114444 5080110198 5080108509 5080110465 5080105461 5080105860 5080119039 5080107391 5080108002 5080117559 5080109246 5080107502 5080107383 5080107464 5080107480 5080122781 5030549595 5080121041 540914 5070223372 5030857389 5060214184 5090113084 CRB1212157 5080109610 5080114568 5080112840 5080102519 5080121971 540444  VESADLCC1109026 5080109181 5080110309 5080115556 5080114681 5060219216 5090109567 5090111103 CRB1203086 CRB1206111 5080107758 5040224826 5060219240 5060215296 5060220273 5050410886 5021022792 5080112743 5080114657 5031216581   5030334971 5070120151 5080121203  5050414342 UNK127684  5050910501 5020919640 5070219448  5050413141 5050910390 5050910404 5070934996 5060219224 5060214192 VAU0703006 VAU0703007 5070219421 5070219456 5070222635 5050414679 UNK128481 UNK128485 5050411106 5040913893 5073058666 5070216627 5080118040 5080115168 540660 5090102309 540974 5040230508 5073058631 5090110581 5090104999 5090109214 5060217809  5060220257 5060220265 5020918644 5090110182 CRB1206099 5090110441 5090110441 CRB1206105 CRB1206105 5090110433 CRB1206104 5060220303 5090109354 5080109653 5060219232 5080112832 5080104503 5080119314 5070219464 5080109513 5080102705 5090111081 5080109823   5031218703 5031218711 5070935135 5070935143 5080121998 5090110301 CRB1206102 5080120495 5080120118 5080107324 5080107359 5060214214 5080121955 5080113367 5090107009 CRB1201053   5080101555   5031216271 5080119446 5090111529 5020510637 5070219413 5031216697 5090112207 5030858415 5040229933 5090105685  5050913403  5020920053 50330334971 5021021273 5050311551 5050311578 5050311586 5060218457 5080102268 5080111801 5021020404 5080111283  5060211886   5060213757 5080114304  5050414016 5080121475 5070223461 5080114525 5080105003 5080110473 5050913322 5050410827 5050410835 5020713007 5090106665 CRB1112047  5080117451 5050414547 5070932128 5050910714 5060220346 5090112118 5090112126 5090112134 5090112142 5090112150 5090112169 5090112177 5090112185 5080118180 5080104511 5080114711 5080114738 5080114746 5080114754 5090102929 5090102937 5090102953 5090102945 5090106339 541001 5050413303 5080110252 5090102201 5080119268 540321  5050414237 5070934430 5080108371 5041211868 5080122676 541387 5031214902 5050311381 5050311403 5041215936 5050911702  5090106630 CRB1112043 5050312116 5050310849 5030859659 5080116625 542002 5040510268 5050913446 5040911793 5090112916 CRB1212150 5090113033 5050312027  5090103798  5080121777 5080115580 5090112762 CRB1211146 5041215189 5090109664 5080104872 5070221698 5050413966 5050311187 5080119683 540269 5080117192 540305 5040911963 5080114096 5080123427 541413 5080105771 5080120282 5080104597 5080110724 5080110732 5090105618 5080105755 5080119977 5080121416 5080118121 5060219267 5090106274 5090109443 5060212823 5060212831 5080123257 5070220179 5080116293 5080116285  5080105798 5080108088 5090105677 5090106177 5090106193 5080106603 5080104155 5080111445 5090112002 5090110328 5080114401 5080102071 5090104638 5080122730 5090103054 5080112808 5080102098 5080107413 5070930273 5070930281 5080114398 5090105561 5090105588 5090105626 5090109990 5090106150 5090106282 5090106061 5090106231 5080111399 5090107394 5090106347 5090105197 5080111402 5090104670 5090102554 5070219057 5070219065 5080119985 5090104565 5080119217 5080106581 5090104891 5090105596 5080107316 5080118776 5090104581 5090106320 5090106169 5080120274 5080122714 5090105014 5090105006 5090105669 5080111461 5080111410 5080104198 5090106312 5090106207 5090109451 5090110743 5090110638 5080111372 5090104166 5070220187 5080111453 5080104139 5080102101 5080118148 5080120037 5080105747 5090106215 5090104441 5080111437 5080118156 5080111380 5090104468 5080111429 5080109963 5090106088 5060212378 5060214176 5090103747 5080123265 5050910730 5050910749 5070934325 5080121181 5090104883 5070220152 5070222597 5090111855 5090111863 5080121408 5090102147 5070932675 5080120894 540467 5050912385 5060212858 5080109483 5080116315 5080116323 5080102128 5080102136 5080104171   5080118164 5090112509 5080105321 5090102325 5090111987 5080105763 5070218891 5090112517 5090113513 5080118482 5090102163 5080120002 5070220187 5090112010 5090112592 5090110018 5080110597 5090111154 5090109796 5070220160 5090112606 5070934414 5090109249 5090102546 5080102772 5080108096   5090112614 5090104190 5090111995 5080119950 5090109788 5080104163 5080104295 5080110856 5080110791 5080104791 5030542922 5080112670  UNK164622 5030858466  5080105224 5060218325 5080120401 5050411580 5050413982 5080106913 5080112190 5060218511 5080108703 5060218503 5060220419 5060220427 5070933760 5090102236 5080101482 7090212968 5060220397 5090107335  5050311217 5080104740 5080106360 5080106379 5080115068    5070933809 5070933817 5080101474  5050913535 5050913608 5080117966 5080112395 5080112549 5080102241 5021126016 5080102942 5080110023 5060215040 5070934228 5050413974 5050414997 5050412129 5090111405 CRB1209123  5080122684 541334 5070220918 5080116390   5060215059 5080102764 UNK116490 UNK116501 N/A 5050311470 5060215598 5050410487 5050910919 5080116439 5080101466 5050410495  5040913141 5050410479 5050411599 5070935062 5080102144 5080116102 5080116110 5090103267 5040913125 5060215571 5080103353 5080109238 5080111607 5020915246 5040913613 5070220969 5060215024 5050310121 UNK147126 5060215032 5090108730 CRU1203032 5070932802 N/A N/A N/A N/A N/A 5040911777 5050411556 5050411564 5080108681 5090103186   5070220896 5080108835 5070221558 5060214036 5080118784 5060218267 5090111316 5070932810 5090109761 5090111138 7090212975 5080121300 5080121319 5080117796 5050913586 5090102465 5090102473 5070932829 5090103801 5060218279 5080114266 5050414040 5080123052 5080123036 5080117818 5090113130 5090109729 5080123044 5080108061 5090113122 5080119691 5040916876 5090109532 5080114274 5080122358 5070222678 5070219545  5080104228 5090107610  5090102198 5080119098 5050410428 5080118814 5090103399 5080109351 5090102414 5040915489 5040916787 5070218514 5080109408 5080118822 5050415195 5080111887 5090110689 5080113758 5080113766 5080113774 5090103933 5090102880 5080110031 5050411513 5070216724 5050913527 5080123028 5090109702 5090109710 5080121564 5080121572 5060215644 5080115742 5080115769 5080119381 541177 5080119624  5090108048 5080116137 5090107270 5090108056 5090108064 5080118997 5080121858 541109 5080114975 5090102112 5020713597 5090107629 5080120304 5080107928 5090107289 5080107936 5080120339 5090108072 5090108080 5050913519 5080115750 5080115777 5090109311 5080123370 5080111615 5080115238 5080115246  5080117605 5030334203 5090108013 5080118903 5080119136 5090110530 5090110549 5090110557  5060215628 5031214848 5080122854 5080105666 5080109386 5080107952 5080108770 5070933965 5080105976 5080118024 5090103356 5070932241 5070120814 5080112573 5080112581 5080105984 5080110295 5090108021 5070219529 5050913675 5090102449 5060215652 5050914485 5090107637 5080113162 5050414474 5050414482 5080115939 5050410851 5040913249 5050414709 5080105453 5060215636 5090105790 5090105871 5040914407 5080113731 5090108722 5090102686 5080108045 5080107456 5080112360 5080114479 5090106495 5090111162 5080120223 5080111852 5090109591 5080107944 5080107960 5080112042 5090109907 5080122811 5080122919 357535 357542 543858 5090104956 5070223623 5080121882 5090106509 5090111456 5090110603 5050414989 5080116854 5080116862 5080118083 5050910773 5080116730 5080116749 5080120258 5070932063 5080114452 5080117311 5080104635 5070219596 5070219502 5080102578 5080107448 5080102292 5070120627 5080112352 5090102961 5090102988 5031219378 5041215650 5060215660 5080111844 5090110514 5090110522 5080114487 5080116722 5080120568 5070934570 5080110686 5080104473 5050411610 5080113308 5090105820 5090105839 5080122641 5080111860 5070120882 5041215634 5080116846 5090108226 5080105267 5070932284 5080122668 5080104317 5090108692 5090109842 5040917023 5080115521 5080115548 5080122390 5080112557 5080112379 5080102837 5090107602 5090110905 5090111170 5080109602 5080108762 5070933973 5080122838 5090107688 5080117699 5090109958 5040916442 5070933639 5080107022 5060215857 5050413109 5080115173 5090107580 5080118792 5070222481 5080119128 5090109923 5080119101 5080117338 5070219634 5080110384 5090110662 5020916722 5070932276 5070222473 5070222503 5070933981 5090105324 5080110864 5080119594 5080117370 5090107645 5090102678 5050911419 5060214648 5090110891 5070216546  5040514034 5070219510 5080116870 5090110204 5060214753 5090102430 5090102422 5080121874 5070930303 5090107599 5073058828 5080120959 5090105847 5080117362 5090102996 5060218155 5080105968 5070934112  5050413117 5080120908 5050911427 5080120916 5020713570 5090105804  5090109834   5090105731 5090112029 5090109176 5090106029 5090106037 5080112212 5080112220  5031219394 5090109184 5080119047 5090110360 5070934597 5070120746 5090110298 5090110379 5080119535 5080119489 541218  5090110387 5040230001 5090109583 5080111879 5090102317 5090107025 5080112662 5090109915 5090109931 5080119543 5080120061 5080120088 5080120096 5090109206 5070934589 5080109394 5080123389 5080115688 5080105275 5080105305 5080117346 5080117354 5070932268 5080109599 5080109912 5070934090 5090109753 5080113154 5080120584 5080104481 5080113316 5050414059 5060214052 5080117583 5080117591 5080115009 5090102600 5050914035 5090111448 5080121866 5080116714 5080120789 5050910471  5080109378 5090105812 5080105291 5080116757 5080116765 5080109637  5080105941 5090110271 5090107661 5080121033 5080122471 5030549722 5030549889 5060213862 5080116706 5080109920 5080107030 5050415144 5080123117 5070934104  5090109850 CRB1203089  5070932233 5041215839 5070220284 5090111146 5090107653 5090112908 5090113505 5090112975 5090112355 5090112991 5090113009 5090111421 5090112584 5090110611 5090112460 5090112657  5090112347 5090112649 5090112223 5080105445 5070220853 5080123141 5050911842 5090108927 5060214729 5080107901 5040914431 5090112894 5060217426 5080117664 5060219291 5080123125 5080122161 5090108935 5080107863 5090108943 5060217434 5070120729 5090109575 5060219275 5080107898 5070933914 5060219283 5080102454 5080109270 5090104387 5070933191 5080123168 5070930265 5050911834 5080107871 5080113170 5050911826 5080121238 5080123133 5080121246 5080121254 5070933205 5080105631 5080105623 5080105615 5080105038 5080104767 CZC6481N94 5080104708 5080118245 5070935011 5080102152 5080108908 5080121904 5040914342 5080103477 5080115564 5080117389 5080110511 5080113383 5080113804 5080104376 5080117753 5080105135 5080120525 5080109629 5070931393 5050914418 5080113812 5080112875 5050414199 5070222805 5080104406 5070218409 5070933957 5080121440 5070216767 5080122005 541308 5040229011 5040229046 5050913950 5090111359 CRB1209118 5050911001 5090105405 541480 5080117249 5090111901 CRB1209134 5040916361 5080112689 5090111340 CRB1209117 5080117184 541105 VESADLCC1206032 5050412099 5050414202 5050913683 5070120236 5070120253 5070935151 5080104783 5080110082 5080110090 5080110104 5090102295 5090113351 CRB1212162 5090111332 CRB1209116 5060219925 5060219933 5080108029 5080103876 5090109133 5060216268  5060211436 5080113634 5080116579 5050910382 5060215679 5060219550 5060219542 5080112913 5080110554 5080116404 5080117168 5070935178 5060214893 5070934031 5070934058 5080112085 5060215695 5060217817 5050413877 5070222589 5050412374  5080103736 5080103744 5080118857 5070935186 5070219618 5070219626 5080107731 5080106956  5080118059 5080122072  5070221531 5060220435 5060220443 5070930052 5080112611 5090109087 CRB1203076 507028387 507028506 21-262470 5080934651 5080934678 5090102694 CRU1107014 5060213765 5060213773 5080113642  5060214559 5050913942 5050910102 5080108614 5080109440 5080110058 5080112700 5090104921 5090112479  5020916781 5070222538 5080111259 540736     5060211312 5060211320 5050913470 5050913489 5050913497 5070216732 5070222708 1025056 5080110163  5031212470 5031218371 5080106832 5050414571 5050414598 5020712590 5030132767 5090102856 CRU1107016 5021021400 5040229917 5080110422 5080105402    5040225113 5040225180 5060213854 5090111812 CRB1209130 1024944 5050414431 5050414458 5090106657 CRB1112044 5080103817 5050413621 5050413648 5050413656 5040914725 5090103429 CRU1108018 5090104549 5080115157 540654 VESADLCC1210039 5050910927 5050913543 5080119608 540488 1026037 1024089 1024444 5070216783 5090104433 540872 5060220605 5060220613 5090106754 CRB1112045  5030334041 5080104848 5090104417 5060215237 5020713341 5080123311 5070222783 5070222791 5030851720 5030852220 5030859535 5031216948 5040225032 5040914873 5080104082 5080104090 5090111324 CRB1209115 5050415349 5090106517 CRB1112041 5060218384 5060218392 5090111464 CRB1209124   5080121157 540380 VESADLCC1201030 5031217928 5031218320 5050415039 5050415047 5060216497 5070932837 5080103930 5080116889 5090106576 541113 CRB1112042 5060220095 5060220109 5060220117 5060220125 5070932594 5070932608 5070932616 5050413087  5021021907  5050914396 5080103981 5050413419 5030131639 5080118210 5070200143 5090104425 CRU1109021 5080123346 5080109092 5080115211 541056 5060218538 5060220508 5090109109 5090109745 CRB1203078 CRB1203088 5080103493 5080110430 5080111828 5020510750 5060220664 5090109672 1026036 5080117982 541107 1026518 5050910943 5080117990 542000 5030856927 5070222279 5090109737 CRB1203087  5080111682 5080118172 5070931768 1027831 5080114010 5080115877 5090105715 5090108455 CRB1111032 CRB1203065 5040916469 5070219391 5050311276 5050311314 5060212653 5060212661 5060212688 5020712612 5030855718 5080117419 5040224907 5040229895 5090103437 5090112789 5020915084 5080107596 5030851763 5050910441 5030334459 5050414008 5080111895 1027159 5080119616 540405 5060220990 5040228872 5070222252 5070222260 5090109117 CRB1203079 5090102783 CRU1107015 5080104007 5080104015 5050913969 5020511560 5030862277 5030862315 5080102853 5031217693 5080123354 5020713163 5030855904 5080118199 5090103674 CRB1108019 5090108323 CRB1203061 5090104395  5080113480 5060220559 5080102349 5090103631 5080118830 5080108754 5080108746 5080115424 5060215814 505041245 5020511528 5050913977 5070934546 5030542965    5070934333 5090113157 CRB1212160 5060220532 5050914019 5090105022 5030549374 5050311195 5080103302 5050411254 5090111383 CRB1209121 5080104384 5080104333 5040914814  5040229291 5030549536 5050415187 5040917260 5040229089 5030856900  5080115904 5080113901 5060218376 5020917931 5070934473 5040914741 5041211779 5050910978 5070934457 5080115599 5080110929 5070219979 5060220524 5080103280 5080107170 5060218422 5030855440 5050913411 5080102888 5060215725  5080114258  5060218414 UNK115786 UNK115789 5090101973 540555 5070223526 5021022660 5050414180 5090112436 5060214508 5060214516 5090106355 541493 CRB1112035 5050913314 5070934643  5080113677 2100021453 5080123400 CRU1103012 5080119721 5050414229 5090111111 CRB1206110 5080114541 5060218449 5060213730 5040228821 5050910323 5080114932 5060213579 5050413664 5050413672 5090103968 5090103968 CRB1109022 CRB1109022 5021022563 5020511544 5090113521 5060219917 5060220575 5050413273 5050410541 5090108439 5030855920 5080104368 5060219429 5080114118 5080109211 5060218430 5070934074 5060214567 5080108355 5070216686 5030128174 5060214400 5070934937 5080111933 5050910722 5060215792 5070217690 5030859950 5080103973 5030859888 UNK164666 UNK164683 5080115831 5050911877 505091986 5080102357 5080108886 5080107545 5060217361 5020918717 5080111984 5070221620 5070222775 5050911435 5040917031 5060215474 5070216678 5080115432 5040225350 5090111391 CRB1209122 5060220494 5080102624  5070221701 5070931504 5070216775 5070216902 5080120312 5021025732 5080117443  5090113149 CRB1212159 5080116986 542055 5030857478 5080114142 5050414695 5070221647  5090112819 CRB1211147 5080109114 5060217906 5040225393 5060218740 5070934341 5080107200 5060215229 5060214494  5090106134 CRB1112034 5060216489    5070223445 5021022148 5040916736  5080105836 5080105844 5050311306 5080113359 5040225245 5090111251 5090111294 5090111278 5090111308 5080121947 5090111286 5090111243 5060220451 5080114347 5070932330 5070932349 5080109289 5070220861 5070223534 5080111364 5070223488 5080111674 5080113626 5080111356 5080111240 21-26245 5070221612 5060216470 5080111348 5080112247 5080109106 5080101806 5080122846 5090103844 5090103828 5090112940 5090112959 CRB1212151 CRB1212152 5021024426 5041215375 5070933027 5080113219 5080120878 5090108897 5090108900 5090108919 540415 CRB1112036      5090110840 5090110859 5080102322 5080102330 5050413826 5080106352 5080107766 5080107774 5080110376 5070218557 5070218565 5070218573 5080112204   5080110368   5070932098 5070932101 5040912757 5030862463 5090112363 CRB1209139 5020714968  5040916965 5080121025 5070218212 5070932020 5080115661 5070219642 5060217698 5080122595 5080108053 5080102969 5080109645 5080117907 5050413079 5070220942 5080103264 5080104546 5090102104 5070930087 5090103658 5080119012 5080117141 5080122188 5050413060 5080117842 5060213331 5080102497 5080102500 5080118709 5080120177 5090104514 5060220656 5080120150 5060215520 5080115122 5060215539 5080115130 5080112654 5070219731 5090105316 5090103623 5080116218 5080113294 5080119578 5080119586 5080119640 5090104778 5060213315 5070119729 5060220486 5080104414 5080104422 5080112034 5080103299 5090104727 CRU1109022 5080121831 540241   424044 5080115351 N/A N/A 5070216775   5090106096 5090106118 5080122536 5080112263 5050411955 5050913462 5070222767 5090109303 5070216910 5060215768 5080109084 5060217787 5080119160 5080112603 5070222759 5080110066 5050914000 5080107081 5080107103 5080110244 5090111049 5070223585 5060220001 5080117133 5080105062 5070221728 5080112271 5080112727  5050414539 5080113650 5080113669 5080110805 5080108584 5060219461 5050913306 5060214613 5060220079 5080110546 5050910951 5080117567 540431  5060220982  5090110778 5080121297 5080119179 5080105054 5080105089 5080108789 5080106301 5090109982 5090110719 5090110735 5090110727 5080118253 5070223496  5050414024  5080112646 5070223518 5080108916 5050414504 5080121505 5070223550 5080102403 5080105380 5080102985 5080123087 5080108800 5080108991 5070934287  5080117125 5030852107 5060211835 5050413885 5080120851  5090112541 5060212300 5070217712 5040916957 5060220583 5060220591 5050913993 5040916981   5070220888 5073058658 5050913934  5080111704 5090110956   5060215768 5080117435 5060219488 5080118962 5030859853 5090104948 5090105480 5050913934 5060215733 5050414601 5080122528 5080103965 5020714577 5080121289 5080122749   5030130993  5070931784 5080117397 5021022709 5031212012 5050910854 5050410282 5040916019 5073058658 5050414644 5050910528 5080116781 5080121939 5060211843 5070934716 5080104392 5080102470 5041215952 5060213412 5021020455 5030852042 5050410894 5060215776  5070932209 5090109826 5030855521 5070935046 5080102810  5020917877 5080120517  5080117400 5060215784 5030333916 5090103739 5080108541 5080108967 5060216454 5040230125 5050414466 1027804 5080107510 5080101563  5050415071 5050414512  5050412404 5041212295 5080112182 5070934481 5080121270 5070221604 5040918275 5080121483 5080114495 5080103957 5080104341 5050911451 5080105011 5090104115 5021022709 5031212012 5080102845 5030858245 5050410908    5031218576 5030549439 5080116412 5080109734 5080105097 5080105119 5080105100    5020919128         5021021540         5080117885 5020916455 5080110155  5090102597 5080113782 5090104875 5080116420 5080116447 5090102287 5080113820 5080113839 5080113847 5090108749 5070222740 5080102411 5080112719 5090103690 5090105286 5080104805 5060215245 5090104603 5080117672 5090106886 5090104611 5090106878 5090111588 5080123176 5080123184 5060214125 5090111537 5070219723 5090111545 5090108153 5090108218 5090104646 5090108129 5080101598 5090108188 5080101601 5080122498 5090104573 5060219259 5080121823 540516 5070223615  5090110832 J2934811703  5090105553 5080123397 5090112665  5090108137 5080118946 5070219553 5080114126 5090108161 5080119659 5080119829 541153 5060213935 5080119373 541129 5080118954 5080120215 540934 5080117575 5080114460 5080102934 5080103833 5090108145 5090104654 5080121513 5070218840 5070219499 5090108196 5060216624 5080115300 5090103909 5080107421 541230 5090111553 5060218295 5090103895 5070216635 5090111561 5080123451 5090101949 5090101957 5090101965 541443 5060213358 5060220036 5080101628 5080117680 5080116307 5050415411 5070218646 5090102627 5090113068 5070220136 5050913829 5080109742  5060218724 5090106967 21-26984 5070931741 5050413540 5090107130 5080111992 5050413559 5090110751 5060220087 5030858350 5080102233 5070933620 5080117745 5050410878 5090110166 5090112622 5050914523 5080104554 5080104562 5080102683 5080105577 5090102767 5050913845 5080102659 5080102667 5080102691 5050311616 5050913853 5090108099 5080106344 5080113405 5080115734 5090103194  5080118067 5070223275 5050910757 5060219569 5090103941 5040916485 5080106336 5050911869 5080112093 5070221523 5080105739 5070120559 5070120559 5080118350 5050910838 5060213943 5080104600 5080115181 5080114983  5060214761 5060214788 5080114819 5080111054 5080111216 5070930192 5050912008 5080114991 5080118431 5080110201 5080118423 5080118415 5080114827 5060216403 5090110174 5080111003 5090105464 5050912016 5080118458 5080111038 5050910846 5070931512 5080111097 5080111070 5080111135 5080111224 5080111127 5080111194 5080111151 5080110988 5080111100 5080110996 5080111046 5060213943 5080110961 5080111011 5060219534 5080111178 5080111119 5080111089 5080111143 5080111186 5080118466 5080111062 5080111208 5090105456 5090113262 5070931032 5090107483 5080118636 5090113270 5090113289 5090113297 5090113300 5090113319 5090113327 5090113335 5090113343 5090102589 5060212467 5080123362 5080122889 5090105383 5080122900 5090105375 5070934635 5090110867 5020918822 5050311306 5080114134  5090106649 CRU1112028 5050415314 5080102926    326827 5070935127 543839 5060215741 5090105200 5090108951 CRB1203075 5090106541 CRU1112027 5040912897 5050411963 5040917619 5060213986 5060213994 5060218368 5080102276 5080102284 5090105936 5090105936 UNK261679 UNK261687 UNK261688 UNK261690 UNK261691 UNK380082 5060214001 5070932772 5070932780 5040914644 5030549471 5070216570 5070216589 5050910889 5090108641 CRB1203072 5070933906 5070216651 5030546553 5060215806 5031218452 5080122978     5080121467  5080113898 542146 5050913349 5080107332 540862 5020920169 5021021303 5080123419 541430 5080109149 5060214028 5080108843 5080115041 5090102864 5090105898 5090111219 5090110875 5090111227 5080109750 5090111200 50308062528 5080114886 5080122773 5070217763 5080120363 5080108118 5080108126 5080108134 5080108142 5080108150 5080108169 5080108177 5080108185 5080108193 5080108207 5080108215 5080108223 5080108231 5080108258 5080108266 5050913926 5070223542 5030132376 5030132392 5030539298 5070218875 5070218883 5070218867 5070934503 5030545417 5060215253 5070218182 5070934392 5070934406 5080120436 5080120444 5060215717 5050913918 5080121327 5080121335 5031212276 5031212292 5080107065 5080107073 5050910862 5050910870 5080104104 7090813710 7090813717 5070217895 5070217909 5090109222 CRB1203081 VDU0703008 VDU0703007 VDU0703006  5080116633 5080116641 5080116668 5080116676 5080116684 5060216438 5080112239  7090813703   5070217941 5080116803 541028 5070217798 5080110287 5080107111 5050910439 5080109556 5090102406 5080109580 5080114665 5080109564 5080109572 5080109548 5090102244 5090104077 5090103992 5090103976 5090104026 5090104042 5090104034 5090104050 5090103984 5090104018 5060213307 5080114843 5080109521 5080114673 5090102279 5090102708 5090102724 5060216578 5050911680 5080110759 5080110767 5090102333 5070932985 5090104212 5070932624 5070931938 5040226071 5040226101 5080107162 5080118881 5080104201 5090111197 CRB1206113 5080106476 5090112568 5080105399 5080102748 5080117478 5070932497 5090103011 5090103348 5080107227 5021021621 5080108932 5070931172 5080109041 5080119152 5050911796 5050911818 5050412102 5050910412 5050910420 5050910897 5050913985  5080118733 5080110619 5080110627 5060213692 5060213722 5060214486 5080114894 5080114908 5060216608 5090111731 5080123303 5090109362 5080119438 5080119500 5080110406 5090104530 5050913691 5070932489 5070932446 5070932454 5070932462 5090111650 5090111669 5090105073 5090105081 5090112274 5080115602 5080115610 5050911710 5060213838 5080114517 5080116013 5050913438 5080116994 5070216740 5080108622 5080113391 5080110635 5090105472 CRU1111023 5030857605 5070931776 5060218406 5080110880 5050413257 392039 5080118490 542061 5070120780 5070933876 5080102756 5050911478 5090103410 CRU1108017 5080112751 5080121920  5080102861 5070221248 5050413699   J2924511641 5080115394 5060214540 5031218355 5080123192 5090102619 5020915122 5070934449 5070219960 5030334157 5080113472  5031217170  UNK157892 5060219038 5060220745 UNK145207 UNK145210 5050311780  5070222287 5070222295 5090109168 5080103590 5080116226 5080122633 5080115106 540664 5090112444 5090103917 5080114878 5080118873 5090104409 5090107017 5060211282 5090109974 5060213676 5080106964 5090105367 5090111006 5090106258 5090111782 5080106514 5070223437 5070931881 5090103313 5090105952 5090105189 5060213781 5080106727 5080114371 5080105143 5090103062 5090103496 5090105170 5090108633 5080104430 5080106611 5060215849 5090104824 5060213587 5090104840 5050910048 5080111712 5080112018 5080112514 5080112417 5080112522 5080112956 5080112948 5090102910 5090109338 5060213846 5070931482 5090105227 5090105243 5090105251 5090104247 5090108250 5090105995 5070217720 5090110123 5090110131 5090104913 5060220133 5090105340 5090111685 5090111693 5090105499 5070931458 5070931466 5080106638 5090103089 5090108617 5090107491 5090109427 5090109052 5090109060 5090108765 5080112964 5090108382 5090107963 5070932527 5070935054 5080104058 5080106395 5080106409 5080106417 5080107715 5080117850 5080120398 5080120819 5080120975 5080120983 5090105502 5070931873 5070934198 5070934201 5080104449 5080120142 5090103569 5080107685 5090103461 5090108277 5090109818 5090104344 5090107467 5080120460 5090103100 5090103526 5090104328 5090103135 5090107505 5090103577 5090108412 5090103585 5090107955 5090103275 5090105219 5090104336 5090108293 5090103550 5090109044 5090103097 5090104220 5080119802 5080120347 5090104352 5090107475 5090103232 5090103542 5090103127 5090103119 5090108234 5090103240 5090108242 5090104360 5090103283 5090111790 5090107971 5090108595 5090111774 5080106387 5080108983 5080113979 5080113987 5080113995 5080114002 5080118938 5080105046 5080115858 5080113928 5080113936 5080113944 5080117826 5080114320 5080114321 5080114339 5080110260 5080117729 5080113499 5080113502 5080113510 5080113529 5080113545 5080118695 5080115327 5080114029 5080114037 5080114045 5080114053 5080119322 5080119330 5090110794 5090108781 5090107866 5090108978 5090107874 5090107882 5090107890 5090107912 5090107718 5090107734 5090108854 5090108986 5090108994 5090109001 5090109028 5090109036 5090108773 5090107939 5090107777 5090108846 5090107785 5090107793 5090107920 5090107750 5090107807 5090107815 5090107823 5090107726 5090107769 5090107831 5090107858 5090108838 5090108811 5090108803 5090105154 5090104255 5050910994 5050913861 5080113855 5070931040 5060219496  5090110506 5080106840 5080106646 5080106654 5080102551 5090107548 5090104263 5090108625 5070931423 5070932411 5070931407 5070932470 5080122412 5090107513 5090104271 5090107521 5070932365 5070223356 5090104301 5070931903 5090111707 5090103860 5080106735 5090105235 5090105391 5090106002 5080106662 5080118806 5080120924 5090103488 5090108862 5090104085 5090103208 5090103518 5090108315 5090110670 5060213781 506021383 5080106794 5080116005 5090109273 5090102899 5090109869 5090112770 5020713228 5070930184 5080105372 5080105070 5080110279 5080104988 5090106185 5080102438 5080103787 5080103795 5080104864 5080104694 5080105186 5080104953 5080106441 5080105178 5080106557 5080106549 5080105208 5080106867 5080106492 5080106875 5080106530 5080106859 5080106506 5080106883 5080107146 5080110120 5080104309 5080120355 5080107782 5070933574 5080107618 5080107529 5080103345 5080103396 5080105534 5080113189 5080103485 5080104945 5080106565 5080113197 5070221655 5080120320 5080108940 5080111747 5080111720 5080112026 5060215482 5090108390 5070932373 5090107904 5090104239 5070932381 5090105294 5090107033 5090105979 5080105216 5080105356 5060220974 5080111836 5070932535 5070932543 5070932551 5070932578 5070932586 5080105607 5090103224 5090103445 5090108269 5070932403 5060213560 5080110872 5080110643 5080107626 5080107634 5080118261 5080118288 5080118296 5080118318 5080118326 5080118334 5080118342 5080120835 5080116153 5080116161 5080116188 5050414652 5070930060 5070932519 5090108609 5060217655 5090104905 5090104832 5090104557 5090103925 5080106298 5080113537 5020713198 5080113332 5080112166 5080112131 5070931857 5070931865 5080106743 5080106522 5080106433 5080106468 5080106689 5080106697 5080106700 5080106719 5080107537 5080109319 5090105138 5070932357 5080107219 5080112824 5090103305 5070932322 5080113715 5070220535 5070221663 5080112972 5090110921 5090107742 5020916684 5090103070 5090103046 5060215458 5070218395 5080105151 5090105308 5090103143 5090103453 5090108404 5080120169 5080106816 5060213811 5080120533 5090110115 5090110700 5090108358 5090111715 5090105510 5070221574 5070221582 5090105642 5090105162 5080118741 5080105194 5080105674 5080111925 5080112123 5080112530 5080111577 5090104859 5080106484 5080106778 5070931806 5070931814 5070221671 5070931474 5080106786 5080118202 5090103321 5090105065 5090107556 5080120134 5080120592 5090106010 5090106673 5090111766 5090107149 5070932438 5090110786 5080112816 5090111758 5090110492 5070223348 5090105359 5060221008 5080106808 5080104856 5080114355 5090105421 5090105448 5060216462 5080110112 5080104651 5090103593 5080110538 5080120770 5080113200 5080114533 5060213641 5060213668 5080122544 5020712892 5080107138 5080107154 5080111739 5090105278 5090105146 5090103216 5090108285 5090105650 5090104174 5090103852 5080105364 5070931911 5080116366 5080106751 5090107947 5090106266 5090105960 5070930001 5070930028 5080110414 5080108630 5090103259 5090108307 5090105987 5090103291 5080120800 5080104880 5080104899 5080104902 5080104910 5080104929 5080116331 5080116358 5090112371 5090112320 5090112266 5090111723 5090112681 5090112282 5090109435 5090109419 5090112258 5090111979 5090112088 5090112339 5090112037 5090112045 5090112428 5090112401 5080102055 5070218204 5060220044 5090102732 541467 5050913616 5080116250 5080108010 5080107669 5080107677 5070933310 5080104112 5070933280 5090110395 CRB1206103 5060219437 5080106972 5080116242 5080112174 5080108037 5090110220 CRB1206101 5090105537 5090105529 5090112487 CRB1211140 5080108568 5080112867 5070223631 5060211274 5070120661 5060220478 5070120576 5080107049 5080107057 5060213706 5060213714 5080120541 5080120967 5090106975 5080111941 5080113871 5080113863  5070222643 5070222651  5060210510 5060210529 5060210537 5060210545 5090107459 CRB1201058 5050412765 5070223097 5050312000 5050412757 5070935070 5090105758 5090105766 5080120126  5060210561 5060210596 5080108592 5080108606 5080121262 5070223089  5090107351 CRB1201057 5080121521 5090108536 CRB1203069 5080107553 5080118237 5080102373 5080121491 5080121912 5080103469 5060211304 5080110341 5070931490 5080123281 5080120428 5080107197 5070930206 5050415020 5070930036 5050914140 5080116773 542175 5060217663 5070221515 5080109831 5060215342 5060215504 5080118512 5080118520 5080114940 5090105634 5060218589 5080109866 5060216551 5080117516 5060216543 5050410533 5090109400 5080109858 5050413923 5090111413   5090102384 5060218562 5060218570 5080114959 5060219992 5050910765 5090111014 5050912997 5060215490 5080109874 5050414717 5090102376  5050415179  5050415152 5080118970 5060219518 5060219976 5070120542 5060219984 5080112999  5080121890 5080113618 5070934244 5070934252 5070931717 5070934600 5070934619 5080115475 5070933752 5090112797 5080109122 UNK-29241 5060220540 5070932500 5070933825 5080110228 5090109370 5060219526 5090112290 5070221639 5080111968 5080115815 5070222716 5080107588 5021022598 5080102896 5080110716 5090104131 CRU1109020 5090104816 5070931377 5070931385 5080119497  5070216554 5070216562 5060215261 5070216805 5070216864 5070216694 5030862749 5040911270 5040912854    5040912870  5090104123 CRU1109019 5060211290 5060218732 5050411173 5070223321 5070931369 5060213439 5060220516 5060211851 5070933833 5070933841  5060217779 5030851984 5080110236 5070932217 5070932225 5090106681 5020510769 5050910900   5080109033   5080107723 5050143230 5050413583 5021020463 5030547479 5041215332 5041215359 5090106304  5060215822 5060215830 5080103809 5080116234 5041212635 5080110392 5050414210  5080119306 5070930214 5070930222 5070930230  5080104279 5080104287 5070931326 5080106328 5060220516 5070216759 5070222724 5070222732 5070219383 5070931415  5021021060 5080102543 5070931725  5030651500 5070931954 5070931962   5080121459 5070223364   5070220365 5080119055 5080104821 5070219383 5040913036 5080122110 5060215601 5080109947 5080108649 5090109508 CRU1205034 5080104643 5080120185 5090106533 CRU1112026  540540 UNK427081 5070934511 5070934538 5090107432 5080110325 5021126628 5021126644 5070217739  5080112638 5031217219 5031217820 5080112409 5090111839 CRA1109018 CRB1209132 5080111658 5080111666 5080122587 5080123273 541401 5050913454 5080122552 5080111798 5080104074  5090106819 CRU1112029 5073051114 5090108552 CRU1203030 5060218333 5080102365 5050910145 5070931059 5030852603 5080113960 5080108533 5080107987 5080107189 5080108975 5080122501 5070222546 5070220209 5090111235 CRB1209114 5070220217 5080117095 541956 5060212785 5060212793 5080117974 542089   5070217682 5050410517 5080115092 531936 VESADLCC1204031 5030338012 5050414660 5090103003 5070934465 5090109079  5080115203 540656 VESADLCC1109027 5090109095 CRB1203077 5030337458 5030651616 5050910161 5050910188 5050910196 5050910218 5080112107 5090111065 5080121173 5050415357 5090102562 5050413281 5070217801 5070934422 5070934740 5050413958 5060218546 5080122463 5090102570 5080118393 5080118407 5050410819 5070217828 5070217836 5090111057 5080112387 5090111952 5080119187 5070220845 5040917651 5080112115 5060214249 5090103615 5080114703 5080115785 5060212459 5060213978 5080118369 5070222627 5070931024 5070934236 5070934759 5070932292 5070930079 5080102063 5070932306 5070932314 5080107855 5080111593 5050911095  5060215709 5080104538 5050913179 5070934732 5060213366 5080109165 5080117710 5060218775 5080121580 5080110694 5080110708 5080115998 5080119799 5080104066 5080122129 5080109327 5080109335 5080115416 5080111909 5060218104 5090102503 5080122706 5080102713 5080102721 5080118091 5080122323 5080122048 5080115947 5080121076 5080114282 5080114290 5080119225 5080104260 5080118008 5080121661 540589 5080113553 5080119837 541236 5080115378 5080104031 5080122102 5080107243 5080107235 5070217879 5070217887 5070217879 5090105782 5080113413 5080113421 5080115955 5080115963 5090105855 5080107251 5080120479 5080123435 543307 5080122382 5080118601 5080118628 5080115920 5090108102 5080122137 5080122021 5060215393 5080107286 5080107294 5080113448 5080113456 5080113464 5080113561 5090102759 5090104506 540578 5080122897 5080115971 5090102740 5070934562 5080114797 542157 5090107327 5080108339 542119 5080122757 5080111267 5070223380 5070223399 5080107472 5090108463 CRB1203066 5090104689 5080103000 5090104476 5040918348 5050413842 5050413850 5080119519 5090103380 5070219537 5060214141 5060215148 5060215156 5050413834 5080109696 5080109718 5030334564 5080104627 5080107839 5080110953 5050411092 5080104120 5070930249 5080104147 5050913101 5050913128 5070932705 5070932713 5090104484 5080105488 5080105496 5080105518 5060219445 5070934368 5070934376 5070934384 CZC6161RZZ 5090112576 5060215121 5070930257 5080111976 5080114363 5080105658 5070932799  5080102632 5080102640 5060214168 5080105526 5090106983 5090106991 5090112711 CRB1211142 5090110816 5090105103 5060217388 5070220586 5070931822 5080119748 5070931733 5080107561 5070931792 5090105111 5070931431 5070935038 5070931849 5080105127 5090110425 5080102039 5090103607 5090111642 5090111634 5080106425 5080102020 5090110417 5090112738 VRB1211143 5060220060 5070933930 5080102802 5090102902 5070931350 5090104522 5080106670 5090109605 5090109613 5090109621 5090112452 5090111677 5070120610 5070120287 5090111375 CRB1209120 5090104204 5080120509 5080118229 5090110808 5090102813 5090104662 5090112878 5090108757 5070932071 5090112215 N/A 5080115386 5080119942 5080114061 5050413605 5050413613 5050414636 5080108363 5080108819 5040916701 5060220206 5060220214 5060220222 5080109432 5080116587 5031217022 5031217049 5031217561 5050310032 5080109424 5090108110 5080115467 5080115440 5080107820 5070216937 5070216945 5070216953 5070934007 5070934015 1023597 5070120593 5080109068 5080109076 5080121637 5080115823 5050910935 5050912245 5090109141 CRU1203033 5030128492 5080120371 5080109130 5050913365 5050913381 5090102635 5090113114 CRA1106012 CRB1212158 5080115319  5090106622 5090109125 CRB1203080 5080121815 541014  5070217968 5070217976 5031217936 5040914849 5050911443 5070221221 5080108894 5080115459 5060220567 5090113378 CRB1212163 5080102918 5080108657 5080108665 5080118849 5080115866 5050414520 5050414628 5080122625 5050913888 5050913896 5060215466 5050414555 5050414563 5080103884 543837 5090107696 5090111820 CRB1209131 5070218654 5070218662   5040912676 5040912919 5090108471 CRB1203067 5041211051 5041213305 5090108889 CRB1203074 5080106824 5080115262 VSU0910914 5080108797 5090110050 CRB1205096 5080119071 5080103752 5080103760 5080103779 5080108517 5080108525 5080108959 5090107440 5080105852 5080106271 5070223569 5070223577 5050911788 5070120644 5090105774  5060214575 5060214583 5090104492 CRB1109026 5060219941 5060219968 5090106614 5070216961 5070216988 5080103310 5080103329 5080103337   5080115890 5060213323 5070935089 5070934155 5070934163 5030334491 5070223593 5070930176 5031212365 5040916345 5050913500 5060213722  5070933868 5070223453 5080114150 5040917090 5070934295 5070934309 5070934317 5070934120 5070934139 5080121807 541038  5050310016 5050913330 5080109017 5080109025 5080102950 5040916396 5080103949 5070220373 5070220381 5070220403 5070220411 5070222694 5060214524 5060214532 5080115491 5080115653 5031218541 5041215073 5080106891 5080106905 5050412110 5080110481 5080122579 5070217755 5070217844 5070216848 5070216856 5031212055 5031212241 5040230109 5070934147 5090107343 5080108878 5080109904 UNK296407 UNK296410 5030310024 5080121343 540886  5080114509 5080110503 5090113386 CRB1212164 5070933884 5070933892 5080122692 5090113408 CRB1212165 5080109459 5080123338 5080108924 5090108374 5090108374 CRB1203063 5080102004 5080102012 5070217813 5090112851 CRB1211149 5021127411 5080115505 5090111618 CRB1209128 5080112050 5080112069 5030132201 5080121602 541276 5030131655 5080115882 5080109009 5080113006 5080113014 5080115254 5070216872 5070216880 5070216899 5080110570 5090102791 541455 5080118032 5070934260 5070934279  5070933647 5080113111 386173 5080118377 5080118385 5080109297 5080111291 5080121629 541292 5070216597 5090103402 541264 5050414822 N/A N/A 5090110107 CRB1208098 5080114436 UNK54329 UNK54266 N/A 5020915599 5040914970 5060218309 5040228457 5041215553 5070218255 5070218263 5070218271 5070218298 5070218301 5070218328 5060216136 5060220176 5050912202 5040224125 5070217100 5080106042 5090106916 5090107416 5090108366 5030548904 5050912199 5070218018 58465 58482 5040919336 5031163070 5050911060 5090111472 5090111480 CRB1209125 5040919301 5040918402 5060212343 5060212351 5060215113 5070217399 UNK124361 N/A N/A N/A N/A 5050414407 5070220659 5050910277 5080103647 5070930702 5050911524 5050912962   5050911486  5040915020 5070217496 5050911648 5050911656 5073050797 5040224745 5040224869 N/A N/A 5080105828 5090102457 5070932640 5070932659 5050911540  5080119004 540739 VESADLCC1201028 5020512044 5030859608 5080101814 5080109300 5030548343 5030856277 UNK55487 N/A 59013 59016 59019 59023 59029   UNK55940 N/A 5031216832 5040229100 5080101571 5070933515 UNK56143 5050910625 UNK206119  5050311349 5050912652 5050912393 5050912407 5050912415 5021021516 5070930427 5050912067 5041215294 5031212233 5070934856 5080101873  5040225288 5070218115 5050410622 5050410703 5050912725 N/A 5050913659 5070217259 5030539263 5020917354 5041211019 5060211592 5050410754 5050910358 5050910366 5060218015 5060218023 5060214109 5060219186 5060219194 5090103704 541722 VESADLCC1206033 UNK145735 5040913109 5030859861 5070124986 5050912237 5060216535 5080102217 UNK136191 5050911222 5060215865 N/A N/A1 5060213544 5060213552 5080117869 5090111960 540837 CRU1210038 5030859845 5040514921 5050311721 5050410460 5073050703 UNK126759 5060211371 UNK157693 5090106827 5090106835 5090106843 5090106851 CRB1201050 5041212309 N/A 5070219987 5080118539 5080118725 540846 5070221884 5040224648 5080106204 5070222309 5050910544 5050912563 5050412854 5050412862 5090112746 CRB1211144 5050910560 5040919395 5090102775 CRB1107015 5080121351 5050415454 5021022180 5050912857 UNK123193 5070222848 UNK126130 5090112193 CRB1209138 N/A 5050412773 5020511323 97812 5060216020 5040221363 5050914426 5050410762 5050413052  5060218716 5031212578 5073051262 UNK150410  5040918089 5070930494 UNK157534  N/A UNK87859 5080108711  UNK155031  5070934767 5080105240 5080117230 5080117656 540638 5040913435 5070930796 5070223259 5021127128 5060216586 5060220826 5070219480 N/A 5041215219 5080121653 541783 5070124587 5070124625 5080107847  5060217213 5080118865 5070931989 5090104808 UNK55398 5080119713 540900 5020715050 5070933477 5040223994 5060213471 5090113394 542125 5030131299 UNK60285  UNK58815 UNK56841 5080110449  5050310083 UNK56972 5060217183 506014672 5060214656 5060214664 5050913551 5050913578 N/A 5060213005 5080103450 5090112304 5050411823 50504131435 5080117257 5090110263 CRB1206107 5090108560 CRB1203070 5080110945 5060213293 5080106212 5060216926 5060216934 5050414369 5041215340 5070930095 5060219801 5060219828 5060219836 5060219844 5060219860 5060219879 50702199552 5070220985 5070220993 5070221000 5070221019 5070221027 5070221035 5070221043 5070221051 5070221078 5070221086 5070221094 5070221108 5070221116 5070221124 5070221132 5070221140 5070221159 5070221167 5070221175 5070221183 5070221205 5070221205   N/A 5070220764 5021021494 5080117079 541142 5021021478 5060211452 5050913055 N/A N/A 5050912253 N/A 5080120991 UNK173823 5080117923 541888 UNK77245 UNK77246 UNK77247 5040915012 5090107386 5090108587  5041210675 5080104759 5080105429 5041215030 5080106069 5073056191 5020714674  5070934880 N/A N/A N/A 5060216918 5050412951 5070931997 5070932004 5070932012 5080104457  VCU0709017 5050414326 5090102821 5031212500 5070932977 5060220818 5090102058 5040918194 5040918216 5040223757 5050414415 5070932934 5080109262  5080112158 N/A UNK142231 5021021583 UNK154865 5060218961 5060219178 5030542817 5070930737  5050913640 5041210500 5060219909 5050910021 5040919867 5060215210 5060215385 5060216942 5060216950 5060215415 5060216640 5040919719 5050414490 5040911653 5040224737 5080119063  5080103221 5073056817 186549 UNK142378 5070220802 5060212556 5061212564 5060214346 5090103720 5090104980 CRB1110030 345221 5080115025 5060211398 5060211401 5060211614 5060211622 5080107650 542178 525608 5070930699 5080105720 VCU0711030 UNK144390 5030546154 5030130837 5030539557 5060218147 5050914213 5070933132 5050913063 5050912015 5070932845 5073056043 5030334017 5030131299   5041215847 5060211037 5080101784  UNK145188 5030539069 5080118768 541923 5041215804  UNK106140 5080101660 N/A N/A N/A 5080102586 5080102594 5080102608 5080102616 5020713767    5030660509 5040230354 5060218120 5080103515 5060214699 5060214702 5060214710 5050414873 5050414881 5050414903 UNK132965 UNK132968 5070222899 5080121068 540308 5080110589 N/A 5050412900 5090103755 CRB1108021 5080103922   5040915446 UNK151161 UNK151164 UNK151166 UNK124480 5090107300 CRB1201055  5090110948 5080105348 VCU0711029 5080116463 541350 5080115033 N/A 50441215766 5080117834 540484 5090102848 CRA1107015 CRA1107016 5070930389 5080106999 N/A N/A N/A 5080105550 5070930818 5080116382 540935     5060210359 5070218069 5020511293 5070217372 5080108738 540821  5080121696 N/A 5070223607 5080109890 541383 UNK153928 5030856102 5050911621 5080101695 5080117621 5090102007 540819 5060215172 5060215180 5060215199 N/A N/A N/A 5080107642 542113 5070217518 UNK161929  5050910226 5050912636 5080109254 5060219313  5080101687 UNK158426 5080117931 541870 5070930664 5060212963 5040919816 5050910455 5080114967 541875 5080114088 541171 N/A 5070217585 5070219243 5070220950 VDU0703013 5070930478 5080123079 5090106290 541698 5080119470 540676 5060217574 5060217582 5050910536 5060218112 5073050509 5040913192 5060214680 5041215677 5070217577 5080119233 540970 5050912083  N/A N/A N/A N/A 502051172 5020512079 5020512109 5080110937 5060220737 5070930591 UNK155712 5070217283  UNK134482 5050414814 5050911281 5050911303 5050911311 5050911338 5050911346 5050911354 5050911362 5050911370 5080116269 540864       5080105933 542193 UNK104509 UNK104510 UNK104511 5030131329 5060216802 UNK104737 5050912075 5070930516 5070930524 5070930532 5050415160 5060220729 UNK123336    5060213633 UNK126413 5080119020 540351 5090106444 CRB1112039  5080103892 5080103906 5080103914 5050912814    5080106190 5040918380 5060218767 N/A 5080119195 5080122455 541179  5060212270 5080101776 5050415373 5050415381 5060220834 5060214087 5080113057 540249 5050414954 5080111917 5080117265 540795 N/A N/A N/A 5070216619 5070218743 5070218751 VCU0612004 VDU0703011 5070216600 5070218719 5060218341 5020713937 UNK60966 5020510416 5070223003 5080119349 5040919832 5040919905 5050912911 5060215288 5040221436 5040514271 5050412889 5041212198  5070221922 5041215324 5041215871 5080106182 5090109877 CRB1203091 5080115793 540466 5060218058 5050911257 5050911265 5050911273 5050913705 5050913713 5060213145 5060213625 5041215251 5060211509 5050311632 5040228287 5070216791 5070933485 5080109807 5080113596 5040917635 5040515464  5070932152 5070932160 5041211159 5041215901 131046 131047 131048 131049 131050 131051 131052 131053 131054 131055 131056 131057 131058 131059 131060 5070218123 5080109491 5080101857 5080113685 541331 5030852492 5030852492 5080118644 542177 5030856196 5040919778 5090103666 5090103682 CRB1108020  5060212920 5060218465 5060218473 5070221450 5070221469 5070221477 5070221485 5070221507 5080221493 5090112835 5090112843 CRB1211148 5040919433 5040919352 5060216004 5050412749 5030334084 5021127721 5030651594 5031163259 5080112344 540913  5041212287 5090109508 5090109516 CRB1203085  5060212629     5060213498     UNK387576 5040224664 5040224702 5070217992 5060211878 UNK127094 5060220230 5090102252 5090102260 CRB1105014 VESADLCC1209037 5050914353 5060219151 5090102155 5090102171 542732 CRB1104013 5060216047 5090112096 CRB1209137  5030334742 5080109203 5040224834 5070220756 5070934910 VDU0703012 5060216969 5050411882 5041212222 5040912986 5080109688 UNK55119 5080120940 5030860576 5080117117 541173 UNK55800 5080106263 5090107998 5090108005 512669 5060219011 UNK206057  5030856269 5070933337 5080113693 UNK91969 5041215693 5060215962 5050911591 N/A 5080101733 5070222953 N/A N/A 5040913427 5070932179 5060219607 5060215873 5070222236 UNK138716 5060219321 5060220338 5040917007 5040917082 5050410738 5070220314 UNK103488 UNK103491  UNK157896 5020712809 5070930400 5070930508 5070217380 UNK140658 5070222406 5080103612 5030858822 5040515 UNK157884 5050914310 5080103248 5060213102 UNK141880 5050912830 5030131094 5030860150 UNK142025 5060213250 5030545328 5040913494 5060211479 5080122765 5080108487 542077 UNK143029 5080103655 5080119896 5080122560 UNK157623 5070221779 5070221787 5070221795 5070221809 5070221817 VDU0706020 373677 5080117486 UNK144353 5080113952 5090104751 5090112053 5090112061 9111100001 5050414377 5060219399 5050311993 5050311977 5050415462 5070933140 5070933329 5060215911 503059409 5090102481 5090109559 5030856161 5070220780 5090108447 5070120797 5041211159 5080112565 5050914531 5050914558 5070930583 UNK150862 5080108460 UNK145738 UNK149841 5060219739 5070222570 5070934872 5050911605 5050415497 5040223463 UNK149785 5050410614 UNK153828 5070220675 5070934929  UNK153947 5040223839 5031163070 5070223267 5070217488 UNK153927 5030856048 5040224931 5040913516 5041215790 5050311934 5050311942 5050412242 5050911753 5050913632 5080101792 5080105569 5080105801 5070219049 5070930109 5040810016 5070221213  5041212112 5060210944  5060214389 5070218913 VDU0703015 5050912474 UNK287547 UNK287551 5080115718 5040223544 UNK132728 5070217291 UNK156698    5080102179 5021125761 N/A 5030545344   5041214727 5041215731 5050912164 5070933078 5070933086 5070933094 5080112735 540803 N/A 5020511552 5070930354 5090111022 CRU1208035    5080122935 5060217841 5090103372 5050910706 5080121734 5050912695 5050912601 UNK72569 UNK72951 5070933442  5070217364 5050311268 5080117915 541538 5040514263 5060213234 5060220915 5090110697 CRB1206109 5080120614 5080120622 5080120630 5080120649 5080120657 5080120665 5080120673 5050410681 5080119934 540621 UNK130591 UNK85321 5080117958 540409 5090104069 541633 5030545565 5060210650 5020916862  5070217224 5020511137 5080118016 541631 5060210405 5080105879 5080101679 5050910013 5060219372 5060219380 5060219720 5040910169 410345 5080120576 5080112077 UNK153698 5090105332 541985 5040223595 UNK307190 5050912687  5080115149 60637 5060213242 5040911688  5050910609 5060211584 5070220624 5050311861 5040919824 5031214716   5070930680  5090107122 CRB1201054 N/A  5090106126 541578  5070220632 5090113483 5090113491 CRB1212169 5050912776 5040220189 5060214958 5070217054 UNK72212 5070933396   5020915289 5080111321 5060213455 5073050835 5070934686 5050411920 5070222422 5070222872 5080113251 542109 5060218937 5060219895  5080109416 5031217960 UNK141286 5060212998 5080118547 UNK142487 5080104775 540922 5060217205 5080101849 UNK145147 5090113025 CRB1212155 5030855815 5080120797 5090104093 540552 UNK211904 5040223315 5060212262 UNK89760 5060211118 UNK106181 5080103698 5070221302 5030862439 5080114762 71308 UNK71296 5041211795 5080109955 542082 5031216875 UNK156211 5060212505 UNK157743 5080121718 5050910552 5050414784 5070217321 UNK138736  5070933418 5060219747 5060220842 N/A N/A 5050912628 5090110247 5090113416 5090113424 5090113432 CRB1212166  5060216225 5070218034 5070222007 UNK102738 VVU0703009 5041215782 5040220162 5060214478 5060216233 UNK27225   5080122951 540240 N/A N/A 5040224087 5080109815 542065 5090106924 5090106932 UNK56674 5070217437 5070219197 5020712477 5031218606 5060218600 5060218619 5080103582 5090109540 5070930729 5050412897 5090110336 5031218592 5020510130 5041211760  5070222996 5050910110 5080123060 UNK206187 5041215545  5050410673 5080101539 5030337237 5030539344 5040912781 5030132031 UNK136527 5060214443 5031218479 5050910285 UNK122418 5060210383 5080107979 542057 5031212624 5070933434 5090103836 541873 5070217267 5090109885 CRB1203092 5060219062 5080123443 5021022199 5070932861 5080119241 541090 VESADLCC1206035 5080106018 5030859403 5031218436 5050912938 5080119853 5060216071 UNK140543 5050410770  5040221266 5021127012 5090103879 5070221744 5060213021 5060220753 5040913087 5020917060 5070932993 5070933000 5070933019 UNK159977 5090104735 5090104743 CRB1109027 N/A N/A N/A N/A 5090112673 N/A N/A N/A N/A N/A 58746 5080107014 541924 5060216217 5040910908 UNK149669 5090106908 541828 5070930419 5041215367 5080115807 UNK153563 5070219189 5060210995 5070217569 N/A N/A UNK152291 UNK159605 103021 103022 5070930613 5080103507 UNK153962 5040911297 5060212890 5060212904 5070217275 5060218872 5060218880 5060219089 5040910886 5040913362 5060215970 5080122862 5040226268 5050912881 5070223011 5041213437 UNK22264 5060211045 5060211096  5060213277 5060217124 5060217221 5060217248 5060217256 5060217264 506027272  5030852441  5050411890 5060216314 5050911559 UNK91867 5060216276 5060218945 5080117494 540428 5050912873 5031217812 N/A 5060211606 5080120029 541808 VESADLCC1211040 5090103712 UNK84189 UNK84189 5090109389 CRB1203082 5080105909 5090106452 CRB1112040 5060211738 5080110899 5080102195 5080111569 541304 5030859780 501024825 5070220012 28920A 28920B 5090112533 CRU1211039 5090108676 CRB1203073 UNK143278 5060212939  N/A N/A N/A 5070222317  5040917104 5050910099 UNK149779 5050910234 5090102074 5070930117 5073050940  5031218584 5040917074 5073056825 5060211517 5080101865 5060216292 5060216306 5090106894 5080117893 541823 5040913230 UNK154036 5060217949 5030333975 UNK154876  N/A N/A N/A 127217A 127217B 5073051122  5080112301 5020713880   5070222856 5050912768 UNK61175 5050910293 5050411912   5070222228 5070223046 5041210969 5030547967 5040911858 5070217046 5040919840 5060211231 5060211258 5060216632 5060215326 5060215334 5060216179 UNK132129 UNK145283 5031218665 5070932896 5060216187 5050914388 5060216888 5060216896 5090110026 CRB1203093 5070218425 5070217038 5030335188 5041214603 5041214638 5041215014 5060216527 N/A N/A  5050912482 5080102160 5050411904 UNK138296 5050912733  UNK140537 5050914639 5050914647 5050914655 5050914663 5050914671 5050914698 5050914701 5050914728 5050914744 5050914752 5050914760 5050914779 5050914787 5050914795 5050914809  5080106174 5040223366 5090103887 5030859713 5080116048 UNK130565 5070222929 5050410711 5070220640 UNK144271 5030539514 5080117508 540436   5041215081 5070222902 111867 111876  UNK146188 5050912520 5050914337 5090112495 CRB1211141 5030862293 5070222910 5080121726 5070222864 5080108827 5031212543 5060213196 5060213137 5070218778 5050413575 5050914361  5080104252 5080121645 541618 5030852557 5060213536 5060211673 5090108706 CRU1203031  5060213013 5050415519 5030131078 5040224192 5021021702 5060220680 5080122609 542148 5030859918 5060219712 5050414725 N/A N/A  5070222813 5070217410 5070217526 5050311748 5050311888 5050910668 5050912865 UNK206572 5060212475 5080101741 5080101768 5021126342 5080116471 5090106363 5090106371 5090110093 5090110980 5060220699 UNK149580 50402223552 5070932195 5090111510 CRU1209037    5041212260 5040910851 5070934627 5050912903 5060211029 5040919417 5060211487 5080103167 5030860444 N/A 5040223730 5090106487 CRU1112024 5020510351 5030856153 5050411815 5080102381 5090112312 CRU1107013 UNK111355  111347 5080103019 5090111871 5090111898 CRB1209133 UNK56064 5060220249  5080108495 5080110848 542118 5090111073 71259 71260 71261 CRU1209036 5040221231 5040610017  UNK56907 5050411246 UNK73986 UNK74002 5080113065 542085 5070222333 5080103671 5080110007 5070218166 5090109265 541541 59492 59495 59498 5070932187 5030859014 5060214362 UNK129889 UNK129899 5080120290 541782 5070221388 5070221396 N/A N/A  5021021923 5080111550   149990 5070216996 50702193753 5030545344 5060220354 5060220362 N/A N/A 50300131353 5041210373 N/A N/A N/A 5080103663 5050410606 5060213382 5060213390     5070219200 5070933779 UNK155765 5080116080 5080121211 540719 N/A N/A N/A 5041214646 5030338942 5030338969  5060213463 5070932721 5040514166 5070220799 5060217159 5090107319 CRB1201056 5080113286 542143 5021022741 5020511102 5050311284 5070219707 5070220438 5070220446 5070220454 5070220462 5070220470 5070220489 5070220497 5070220500 5090110999 UNK130224 UNK174993 5031217448 5040224060 5040919859 5050413001 5080103388 5090111626 UNK174992 UNK206013  5090107378 5080101881 5060220885 5080102446 5080120207 5090108714 5070218093 5080104244 5080105313 5080106050 5080107278 5080115912 542007 5070930311 5040513453 5060215954 5060216160 N/A N/A 5041212147 5050415098 5050415101 5050415128 5050415136 5050311853 N/A  UNK156283 5080105917 UNK119180 UNK119181 UNK119182 N/A 5070217453 5030130098 5070930710 5080118474 540384 5090110042 CRB1203095 5030539700 5030650482 5031217790 UNK92472 5050415055 5060218163   5030855998 5060218236 5060218244 5041212236 5080110457 541036 5080121432 540781 5050415500 5031217405 5080103132 5070932969 5050414768 5050910579 5080105925 542093 5031218622 5060214427 5050910595 5040913478 5080108436 5030545360  5031217359 5070217593  5080119551 UNK142179 UNK142194 359922 5090109699 5040410018 5040913273 5070217313 5080117524 540704 5070221264 5070221272 5070933523  5070217232 5030334343 5070221965 5050911583 5020917370 5080118652 541343 17722A 17722B 5080103086 5060212726 5060212734 5040223803 5040224257 5041210683 5080120843 5080113227 58742 5070223143 5050913020 5050913039 5090111928 CRB1209135 5030335390 5030335455 5050413346 5070220667 5031163070 5070933361 UNK159847 5050912539 5070222821 5090110972 5080112328 5080113324 540693 5080103574 UNK125282 5060213110 5070217194 5060211576 5041211965 5041215413 5021022490 5050414741 5060217167 5060220877 5070930567 5080106166 131241A 131241B 5070222414 5070933566 UNK93835 UNK154107 58633 58636 5040510152 5080113030 5080118075 540658 541603  5020916811 5050311608 5050410959 5070216708 5080114800 540641      5040226195 5080102209 5070120202 5090111502 CRB1209126 5070218832 5041210942 5050912369 5080116692 5090106770 5090108870 541030 CRB1203090 5060211339 UNK91980  5021125745 UNK91872 5080103620 5090106398 5090106401 5090106428 CRB1112037 UNK71868 5080116064 541428 5090104972  5050414911 5090109486 5090109494 CRB1203084  UNK92393 UNK92396 UNK92399 96453 5070933531 5050912156  5080115343 540322  5050410592  5080104619 N/A  5040223358     5070217143  5050912970 535271 UNK350099 UNK368118 UNK370252 UNK387927 UNK388050 UNK422332 UNK422365 UNK422370 UNK422375 UNK447117 UNK447121 UNK447125 UNK447129 UNK447133 UNK462061 UNK462065 UNK146261  5050411033 5080115076 541902 UNK158709 5080108452 542095 VESADLSV1212004 5080101547 5030132295 N/A  UNK158405 5080116498 5080122803 5090102651 5090107408 CRA1107014 CRB1103012 ENP1106001 EOP1106001 VESADLCC1106023 5041215170 N/A 5090105928 CRB1111033 5090106800 CRB1201049 UNK155514 UNK176409    N/A 5040911718 5040911734 5040911750 5040911777 5040911831 5070223402 5070223410   UNK58223 5060212297 5080119810 540601  5070223062 5070930486 5080113707 541289 5040225229 5060216519 UNK57125 5080122374 5090110069 541954 VESADLCC1109025 5080122986 5090109281 542126  5080118687 5050910676 UNK130680 UNK137951 5070934791 5080109785 5030860193 5030860428 5080120193 540376 5060212645 N/A N/A  5070217216 5020510017 5040230036 5040230052 5040230184 5050413788 5050413796 5050413818 5060215318 5060215350 5060215369 5060215407 UNK145177 5020916382 5021126229 5060217027 5060217035 UNK208921 UNK98992 5021127594 5060219046  5040917198 5050910013 UNK26476 5060217132 5050914264 UNK157819 5080101652 5080119632 541064 5050912121 5090106762 CRB1112046  5041211760 5090102511 5050410797 5080105682 UNK156792 5080112255 5060212599 5060212602 5060212610 5090102120 5090102139  5040919875 5040919891 N/A N/A N/A N/A N/A UNK159132 5041214662 5050414172 5020916412 5040229402 5021126482 5090107424 5090109893 5021127039 5070217348 5060220907 5030853650 5050911133 5050911141 5050911168 UNK106949 UNK106950 5050311837 5060217353 UNK139386 5090113165 CRU1212041    5080109998 5050913004 5050913012  5090106746 541890 5070930621 5050914574 5050914582 5070218050 5050912598 N/A N/A N/A 5080106026 5060217914 5090113254 CRU1212043 5060218694 5060218708 UNK138663 5090107297 5070932748 26557 26560 5060214907 5060214923 5040917066 5020510688 5020714003 5080119845 5040223331 5050414393 5070218905 5090111030 5030544690 5050910064 5070223194 5060210839 5080105585 5080106980 5080120886 540506 5030860177 5070218085 5070217402 5021126792 5031214740 5080119705 540324 N/A   5030860363 5080108851  5040225385 5070930346 5030856145 5080105593 541340 5020511919 5020713651 5020713708 5030856110 5040915098 5050411017 5070217631   5080113723 540674 5060213889 5060213897 5073050843 5060213919 5070217011 5080123230 541843 5040915268 5050311594  5030334807 5060216500 76102 76114 76120 UNK140238 5060216837 5020510149 5080102462 5090104379 CRB1109025 5021021044 5080113049 541531 5090112754 CRB1211145 5070221310 5020915440 5020915572 5060215903  5080111232 541821 5030859330 5031212217 UNK129172 5031218096 5090102023 5090102023 UNK150847 5080113790 540651 5020919551 5080121149  5070930370 UNK158241 5040224710 5030132511   N/A N/A N/A   5040915136 5060215997  5060213153 5060213161 TEMP0041886 TEMP0041887 5030131558 5080110015 5050910072 5080107340 541997  UNK157083 5030128611 N/A N/A N/A N/A 5070223232  5090112703 5050415330 5060210758 5060210766 5060210774 5060210782 5070220721 5080122013 540286 UNK206357 5060212238 5060212246 N/A  N/A UNK71275  5040224877 5070219669 5050912113 5040220049 5060213188 5060211541 5060219674 5090105723 541672 5060214095 5060215431 UNK109702 UNK109705 5030545409     5060215075 5060216977 5060216985 UNK64668 59174 5060219054  5060217930 5050910617 5070218816 UNK59349 5080103094 5070221345 N/A 5070217607 5030337393 50409919484 5030542434 5080107707 5080106921 5041215138 5070217151   5040910770 5050911508 5060211525 UNK72553 5020714992  5040230419 5080106093 5080111623 UNK121980 UNK121983 5020510491-103450 5020510491-103453 84168 UNK140399 5060211711 UNK141043  5073050770 5040919468 5050412994 5040917120 N/A 5050911516 UNK136739 504125308 5050410789 5073056094 5080109661 5080109793 540799 5041215820 UNK24280 UNK24287 UNK157504 5030859896 5040223420 5050914078 5090106940 CRB1201051 5050914183 5050914302 5060217345 5060217973 5050412919   5080116056   5060219100  5060212572 5060212580 5070217240 5050912377 5090103364 5070120678 5070932764 UNK131956 5090106223 5070932691 UNK144373 UNK144380 5080113375 5060217000 N/A UNK156437 UNK126650 5041212252 5060219623 5060211665 5070217070 5070933612 N/A N/A 5073050649 5080111313 5050410983 5080116501 542009 5070223224 5080109777 UNK72516 5080101636 5090104158 5070218107 5070935097 5040910924 5050913098 5050312108 5060220710 UNK99150 5080119675 541748 UNK99191 5050914329 5070933469 5030549234 5060217191 5090108684 5073050967 5090106703 CRB1201048 5073050983 5040230303  5080120053 541767 5060212769 5060212777 5090104786 5090107564 5090107572 CRB1201059 CRB1203071 5060210022 5040510772 5040913176 UNK156721 5021020641 5050311756 5080113243 5080116129 541414 UNK75374 UNK75379 UNK75383 5080119403 540441 5040913303 5060214257 5060214281 5060214303 5060214311 5070219081 5040910029 5090108331 CRB1201060 5031216891 5020510432 5080113103 5090102716 540644 5041210888 5041210381 5070930834 5030855807 5050911974 5050911982 5050913195 5050913209 5050913217 5050913225 5050913233 5050913276 5050913284 5050913292 5060216101 5070933159 5080103108    5090110654 CRB1206108 5080102993 UNK142388 60707  5060214370  UNK130778 UNK167432 5070930788 5070934694 5080102047 5090110824 5040225253 5060220141 5030855866 5060219003 5080103191 5080104465 5040912625 5041215626 UNK207035 UNK207036 5070219685 5050311918 N/A  UNK123064 5050914906 5050911761 5050912644 5060211363 5070217208 5070935100 5080101512 5070200496 5080115408 541324  5041215669 UNK145344 5021127439 5080116544 5080119888 540493 5070221442 5080117648 540413  5080116374 5060215083 5021127152 5040915071 5040915101 5090113440 CRB1212167  5060219690 5030855742  5040913389 5041210608 5080119527 541940    5050311713 5040912846 5030131272 5070217615 5030542264 5080104716 5080106239 5080122927 5090102082 5090102090 540758 5080115696 5040220316 5050912784 5050410657 5050910129 N/A N/A UNK153525 5030114122 5060218139 5050410509 5040225334 5070933590 5070933604 5080119292 541077 508011327 542005 5040919751 5030855734 5090103151 5090103178 CRB1108018 N/A 5050414385 5080116072   5070934899 5090111936 5090111944 CRB1209136 5090113041 CRB1212156 5070223178  UNK156380 5080108479 540839 VESADLCC1209036 5040915586 5060219585  5080121793 541204 5060211681 5050913594 5041211760 5060219658 5060219771 5050912504 5020915432  5070930990 5070931008 5070931016 5070931288 5070931296 5070931318 5041215472 5060220168 5050912954 5050913144 UNK72854 5060217140 5040912811 UNK56054 5030852328 5080105690 5090106959 CRB1201052 5060212491 5080116277 540614 5080103027 5070930168 5070933426 5070930923 5070930931 5070930958 5070930966 5070930974 5070930982 5070931156 5070931164  N/A UNK61337 5060219356 5060219364 5030128085 5050414776 5031217499 5060220761 UNK72422 5070223186 5030859969 5020714127 5070932853 5020916315 5060211460 UNK138249 5060211126 5070930761 5070932756 5031218274 5070217356 5021022725 5021022725 UNK103407 UNK141192 5070930435 5060218007 5080111585 5020914789 5070931970 5050912059 5080106247 5090112932 UNK142604   5041210896 5060214796 5060214818 5060214826 5080112786 5060212947 UNK123232 5021021362 5070218522 5030548394 5090108420 CRB1203064 5060218988 5060219682 UNK143153   5040910797 5070217305 5090109346 541554 UNK144363 5070933353  5060217337 5060210952 5060210960 5060216098 5060217566 5070218158 5090108501 UNK123308 UNK123311 5050311764 5050311845 5060213447 5060214265 5060214273 5090108498 5050910137 5060217558 5060219615 5070217089 5060219097 5080117702 5090104867 CRB1110029 5073050924 5080111690 5080101709 5080101717 5090106789 5090106797 5040224028 5041212201 5070935003 UNK149828 5041212392  5070221892 5070221906 5030539204 UNK158419 5030338047 5050311535 5060213269 5070933035 5080112921 5080101954 5080115084 542224 5040224680 5040224893 5070219162 5070934821 5080108282 542063 UNK153738 5073050665 UNK161114 UNK205359  5070219170 5050913667 5090101981 541971 5041215537 5050410630 5073050819 5090112398 UNK130134 UNK204707 5020917079 5080123214 542046 UNK161240  UNK158446 71854 5060212327 N/A  5060218910 5021021486 5090113238 5030539522 5021025139 5080110317 540808 5060217833 UNK158258 5080103639 5060212637 5030856293 5040223919 5060210391 5050412803 5060211010 5060220702 5070930648 5090102538 5090105901 5073059875   5070220004 5060214842  5080108304 5090112886 5060218198 5060218201  UNK155749 N/A 5060216039  UNK55836 UNK55839 5080117222 5030862455 UNK59953 5041215685 5050311322 5021127144 UNK137996 5030338551 5070930842 5031218061 5070932055 58126 58129 5070218794 5070218808 5090112967 CRB1212153 3070900197 5050312051 5050312078 5070933124 5050910633 5090106711 541686 5021021869 5060219755 5080101725 5060219577 5080113588 5060220923 5060220931 5060220958 5060220966 5030852360 5031218630 5070222384 5041215383 5070120100 5030128557 5080121599 541233 5030132627 5030132627 5060217981 UNK32401 5030856242 UNK54962 307152 5080112859 UNK110771  5050411807 5070930605 5090110883 5060217965 5090106436 CRB1112038 5040223447 5060216152 5070217623 5070217666 5070217674 5080114924 540669 5070222015 5070222074 5070222082 5030545174 5080109467 540662 5070934953 5040510705 5060211061 5060214451 UNK19002  5040223749 5060220869 5070220934 5060220788 5080105437 5070222449  5080110902 540711 5070222430 5070934813 UNK206387  5080103051 UNK138837  5050910021 5070220748 UNK162493  5070220691   5031216816 5080106158 5080121742 UNK128961 UNK128969 UNK128972 5030860134 5070222392 5090112983 CRB1212154 5041210810 5040510810 UNK143102  5040916248 5073050568 5073050771 5070218689 5080115645 541917 5030855785 5080120231 541735 UNK144304  5080110678 540756  5070218360 UNK159980 5021126415 5040917058  5050910056 5060211746  UNK124079 5041215618 5041215618 5080118563  5090113459 5090113467 5090113475 CRB1212168   5040224141 5070934961 UNK130079 5080106948 5080110821 5030539468 5070120831 5080114851 541381 5090105863 5050415276 5050415284 UNK121922 UNK238923 5050912709 5080106255 540869 5080111496 5090102392 542197 5080122994 5060218953  5050910307 5070220772 5073057082 UNK160065 5070218026 UNK157031 5050411947 5060218481 5080117109 540648 5080108312 5080119756  UNK123342 5090109680 5030860304 5090103038 5090105545 UNK158799  5030860282 5050411572 5070219022 5070220268 5090104107 CRB1109023 5060216330 5060220028 5060218082  5050413028 5040911580 5020713899 5070933558 5060217922 5080101644 5080117303 542195  5090108579 5030132325    5080116536 541110 5070218824 N/A N/A 5090110077 541710 N/A 5060211495 5031311426 5080115335  5070934864  5070933582  5070217127 5080119462 542163 5050414792 UNK109308 5070217119 5050912261 5030336818 5030542426 5050912334 5080122722 541519 5040229550 5070930672 5070933493 5060216128 5020915726 5070932950 5050412846 5041215707 5060211134 5050412870 5070222376 5080119993 541131 N/A N/A 5030334270 5080116528 540487 5050410746 5030129294 5070222198 5080118105 5060212521 5090110190 541659 5060211002 N/A 5060211630 5080118571 540714  5040514298 5020915653 5060214435 5040913737   5060211649 5060219763  5080120452 541590 5050912660 5020511579 5020511757  UNK158432 5041215812 5050412927 5070220594 5070933507 5050910692 5041214689 5070932888 UNK126427 N/A N/A 5040225784 UNK156575 5080119926 5070934171 5060216063 5060216144 5040513305 5060215938 5060218902 UNK72557 5070934848 5041215510 5060210979 5050910250 5060219704 5050413044   5031217758 5021022555 5050914914 5080109939 UNK131384  5060219119 5080106085 5080109173 5080113138 5020919470 58391 5060210804 VCU0612003 VDU0612005 5080107995 542099 5050414334 5050910242 5070218077 5070222066 5060213056 UNK57244 5021125974 5021021249 5070222880 5070222945 5080103183  5050911494 5031217774 504091835 5060210340 5080114770 5070933787 5080110651 UNK147938 5070930133 N/A N/A 5070217429 5070222961 5060215547 5080112980 541835 UNK188821 5040917279 5070934902 5060216195 N/A N/A 5030859632 5041212120 5080101504 5050415004 5070218239 5070218247 5050910315 5060217043 5050912180 5041212244 VDU0612001 5070217461 5060213048 5050311357 349528 543841  5060219135 5090103534  50312121709 5020915688 5050910641 5060219143 5080103043 5050415209 5070217550 UNK140043 5090113246 CRU1212042   5070218352 5080102489 5021020943 5080111518 540699 5070934805 5060212319 N/A 5080120932 542204 5060212203 5050912555 5080122943 5060215946 5070934066      5070223216 5070932039 5070932047 5060217590 5040917228 5080108673 5041215774 5090109257 541903 5070930826 5080105232 5080106123 5090109656 CRB1205097 5031216859 5050311799 5090105413 CRB1111031   5070930575     5080106220  5040916973 5080105992 5070930397 5070932632 5080116811 5080116838 5080121386 5080121378 5040223528 N/A 5050912229 5050912490 5080111771 5050412978 5070934708 UNK147686 5041215642 5050411343 5060212416 5060212424 5060212432 5060212440 5060219402 5060219410  5040919786 5090110646 CRB1206106 5070220683 5060218856 5070223038 59063 UNK59057 5020919454 5030132465 5080117737 371035 543843 5031217952 5060211533 5080103566  5090106738 541855 5041213593 5080112298 5080112697 5070218638 5080103825 542115 5021022288 5080115629 5080119861 540407 5040510721 5080103205 N/A 5070930745 5050914280 5050914205 5090108528 CRB1203068 5040225067 5080103841 542012 5050412811 5070933388 5080116145 UNK164410 5070933116 5070933345 5080114789 5090109966 5090110085  5040913346 5031214813 5080109882  5040913397 5080113022 542141 5030131299 5080122617 540267 5050914191 N/A 5020510955 5080115297 N/A 5040228708 5080103078 UNK157224 5090112525 CRU1211040 N/A N/A 5030853278 5040919271 5050310970 5080103175 5080122420 5090102228 541357 5070930141 5040914369 UNK141322 5020511986 5080119918  5070932667 5070933655 5070933663 5070933671 5070933698 5070933701 5070933728 5070933736 5070933744 5080115270 540771 5031218533 5060216829 5080101970 5090102015 5090102031 5020916668 5070930656 5080103035 N/A 5090111596 CRB1209127 5030855750 5030859691 5060213900 5080112778 5021126113 5080103701 5090104182 CRB1109024 UNK21930 5050415470 5090113106 5050412722 5060217825 N/A N/A 5040918291 N/A 5070934775 5050415306 N/A N/A 5050414857 5040514050 UNK154382 5070931520 5040913257 5080118504 541919 5041212570 5050414849 5060213374 5090113211 CRB1212161 UNK59977 UNK59980 5080119411 541645 5031218614 5040223951  5021122837 5040510756 5050914175 5050914221 5050312086 5050312094 5020713406 UNK91788 5080103124 5080117613 UNK61590 5070933450  UNK130171 5050311772 5060214338  5060211444 5090110034 58075 58078 CRB1203094 5070220705 5080117060 541922 5070933922 5040911947 5050413893 5050413907 5060211347 5070222058 5080106131 5080119209 5050412943 5021021168  5050413427 5050414423 5070930443   5031212527 5041211175 5070220713 5080121688 542020 5060217108 5060217116 5060219127 5060219631 UNK153800 5040919255 UNK159124 5080112336 5080107693 50602118635 5060218627 N/A N/A 5080103361 5080106107 5080103523 5050912679 5040913443 5060211568 5060212289 5070217062 UNK110905 5060210987  N/A N/A N/A 5073050754 N/A N/A N/A N/A 5030862285  UNK157075 5050912717 5050414164 5080104325 542111 5080115289 542161 5070222937  5080104937 542199 5070217178 5060214419 5050914345 5030539190 5050910587 UNK72918 5080118660 5070217178 5060219666 UNK71954 UNK71957 5080106077 UNK160249  UNK58639 5030548300 5090109397 CRB1203083  VESADLCC1209038 5080108398 UNK55900 5080111488 5080114428 542087 5080103213 UNK56151 5050911613 5070217186  5080101989 5030862412 5050411025 5040223854 5070934783 5080118598 5090106142 541242 ECP1201001  5040223498 UNK145973  5040510012 5080118679 540678 5070217542 UNK116727 5041215499     5070218581 5070218603 5070218611 5020714607 5020510483 5020920177 5090102066   5070932926 5060220850 5050910269 5070935119 N/A 5070932918 5040913370 5060215989 5050910080 5080119667 541355 5041212074 5050912350 5020510378 5080103140 5080109971 541862  5050912342 5020916552 5070930125 5050411858 5070220357 5040916914 UNK154870 UNK154916 5050911737 5050910781 5050910803 5050910811 5050914566 5060216241 5090104794 CRB1110028 5060212513 5040912420 5080123206 541566 N/A N/A N/A N/A N/A VESADLCC1206034 5070219227 5080118989 541605 5070223240 UNK157727 5040910223 5030131043 5041210012 5060218597 5070220608 5050413478 5050912024 5050912032 5050912040 2511 5070223208 5060216055 5020915815 5080118113 UNK142571 5060211088 5060212971 5020713910 5030218821 5060218805 5060218813 5060218848 5060218899 5080103531 UNK130075 UNK155017 5050413494 5050413508 5050413516 5050413524 5080115637 5080114835 5090109478 542201   5060216373 5060216381 5070222201 5070934988  5060219348 5060220672 5031218495 UNK207135 5040918186   5070222325 5080103116 5080117087 540702 N/A  5090112924 5090112231     5080122447 5080107308 5080107006 5090107181 5090107203 5080117761 5090107157 5090107238 5080112891 5090107173 5080120010 5080116560 5080119365 5090107254 5090107262 5090110255 5070220292 5080106573 5080116552 5080117788 5090102368 5080119969 5090107246 5090107211 5090110212 5090109230 5080119357 5090104964  5090102341 5080121556 5090110964 5090105693 5090105707 5080120266 5070934554 55NFP1S 5050415403 5080102675 5080121548 5050911850 5060218651 5090112827 5090112630 5080111275 5080109505 5080112794  UPDATED:  5070930818 27880 5070222198 5080118105 5090110816    
Recalling Firm/
Manufacturer
Vital Images, Inc.
5850 Opus Pkwy Ste 300
Minnetonka MN 55343-4411
For Additional Information Contact
952-487-9500
Manufacturer Reason
for Recall
Vital Images has found a potential error in measurement generated through the software which affects all versions of Vitrea Enterprise Suite, Vitrea, VitreaAdvanced, VitreaCore (VitalConnect) and Vitrea fX prior to version 6.3.3, except 6.1.6, 6.2.3 and 6.3. This error occurs when images are rotated at the time of reconstruction by the scanner in non-90 degree increments of the transverse/axia
FDA Determined
Cause 2
Software design
Action Vital Images sent an "Urgent Field Safety Notice' letter dated February 5, 2013, to all affected customers. The letter identified the probelm the product and the action needed to be taken by the customer. Please immediately inform all potential users of Vitrea of the information contained in this letter. We recommend that all measurements associated with the error described in this letter be verified against results from other technologies. You will be contacted by Vital Images Customer Support or by the customer support department of your distributor to arrange for installation of this patch on your system. We recommend that you implement the patch. If you have an earlier version, Vital Images Customer Support or your distributor will contact you to discuss your options for replacement. Please sign this Notice in the space below to verify that you have received and understood it, and return your signed Notice by email to fieldnotices@vitalimages.com. If you have questions, direct customers of Vital Images should contact Vital Images Customer Support at 1-800-208-3005 or by e-mail to support@vitalimages.com; customers through a distributor of Vital Images products should contact that distributor. Updated 5/28/2013: A second consignee Vital Images, Inc "Urgent Field safety Notice" letter dated March 26, 2013 was sent to consignees on 3/25/13. The follow-up letter was sent via e-mail and mail to 4,616 already-notified consginees, who had received affected versions 6.3.0, 6.2.3, or 6.1.6. The letter described the problem and the affected product being recalled and provided a number of steps to follow which included to retun the Effectiveness Check form.
Quantity in Commerce 7649 (Expnaded 5/28/13 to 7654)
Distribution Worldwide Distribution - All states in continental USA including DC, PR and the countries of ALGERIA, ARGENTINA, ANGOLA, ARMENIA, AUSTRALIA, AUSTRIA, BENGLADESH, BELARUS, BELGIUM, BOLIVIA, BOTSWANA, BRAZIL BULAGARIA, CAMBODIA, CANADA, CENTRAL AFRICAN REPUBLIC, CHAD, CHILE, CHINA , COLOMBIA, CONGO, CROATIA, CYRPRUS, CZECH REPUBLIC, DENMARK, DJIBOUTI, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GREAT BRITAIN, GEORGIA, GERMANY, GHANA, GREECE, GUAM, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAPAN, JORDAN, KAZAKHASTAN, KOREA, LEBANON, LITHUANIA, MACEDONIA, MALYASIA, MALI, MAURITIUS, MEXICO, MALDOVA, MOSCOW, NYNMAR, NAMBIA, NAMIRGIA, NEPAL, NETHERLANDS, NEW ZEALAND, NIGER, NIGERIA, NORTHERN IRELAND, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILLIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, RUSSIAN FEDERATION, SAUDI ARABIA, SENEGAL, SERBIA, SEBIA AND MONROBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOMALIA, SOUTH AFRICA, SPAIN, SRI LANKA, SUDAN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, THE NETHERLANDS, TIBET, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBERKISTAN, VENEZUELA, VIETNAM and YEMEN.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = LLZ and Original Applicant = VITAL IMAGES, INC.
-
-