| | Class 2 Device Recall Dimension Vista(R) System Vista Drug 4 Calibrator |  |
| Date Initiated by Firm | February 21, 2013 |
| Date Posted | April 18, 2013 |
| Recall Status1 |
Terminated 3 on January 07, 2014 |
| Recall Number | Z-1151-2013 |
| Recall Event ID |
64596 |
| 510(K)Number | K093441 K112142 |
| Product Classification |
Enzyme immunoassay, digoxin - Product Code KXT
|
| Product | Siemens Dimension Vista(R) System Vista Drug 4 Calibrator. The device is used as an in vitro diagnostic product. |
| Code Information |
Catalog Number KC460, SMN 10472284, lot 2KD052, exp 5/1/14 and Catalog Number KC460A, SMN 10476471, lot 2KD053, exp 5/1/14. |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
| For Additional Information Contact | Aydee Crawford 302-631-6312 |
Manufacturer Reason for Recall | Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(R) Digoxin(DIGXN) Flex(R) reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL. |
FDA Determined Cause 2 | Process control |
| Action | The firm initiated their recall of this product by sending an Urgent Medical Device Recall letter dated February 2013 to all Dimension Vista Drug 4 Calibrator customers who received the affected lots. The letter notified them of the issue, the potential risk to health, and actions to be taken by customers. |
| Quantity in Commerce | 1,638 |
| Distribution | Distributed within the US to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KXT
|
|
|
|