Class 2 Device Recall Layershield Adhesion Barrier Matrix |
|
Date Initiated by Firm |
March 28, 2013 |
Date Posted |
May 10, 2013 |
Recall Status1 |
Terminated 3 on June 03, 2015 |
Recall Number |
Z-1300-2013 |
Recall Event ID |
64778 |
Product Classification |
layershield adhesion - Product Code GXQ
|
Product |
Layershield Adhesion Barrier Matrix
Layershield Adhesion Barrier Matrix is an absorbable implant for use as an adhesion barrier for the reduction of peridural fibrosis. |
Code Information |
1125812 1125816 1126021 |
Recalling Firm/ Manufacturer |
Integra LifeSciences Corporation 105 Morgan Ln Plainsboro NJ 08536-3339
|
For Additional Information Contact |
609-275-2700
|
Manufacturer Reason for Recall |
Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
|
FDA Determined Cause 2 |
Process control |
Action |
The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: FCA@integralife.com or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product.
"However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol."
For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723. |
Quantity in Commerce |
215 boxes |
Distribution |
Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa. |
Total Product Life Cycle |
TPLC Device Report
|
|