• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall CoCr Head

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall CoCr Headsee related information
Date Initiated by FirmAugust 30, 2013
Date PostedOctober 25, 2013
Recall Status1 Terminated 3 on September 25, 2014
Recall NumberZ-0072-2014
Recall Event ID 65799
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
ProductCoCr Head (Not distributed in the United States) Used in total hip replacement.
Code Information Part Number Range:01.01012.405/408
Recalling Firm/
Manufacturer
Zimmer, Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information Contact
800-613-6131
Manufacturer Reason
for Recall
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
FDA Determined
Cause 2
Package design/selection
ActionAugust 30, 2013 Urgent Recall Notices were sent to Distributors, hospital risk managers and surgeons identifying the issue and directing consignees to make all affected personnel aware of the contents of the notification. The notice identified the products involved, the issue, possible adverse events associated with the issue and recommended that implants found with LDPE bag adhesion not use the product and return it to Zimmer sales representatives. Questions or concerns should be addressed to Zimmer customer call center at 1-877-946-2761 between the hours of 8 a.m. and 5 p.m. EST.
Quantity in Commerce604064 total
DistributionWorldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-