• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Sedecal MobileDiagnost wDR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Sedecal MobileDiagnost wDR see related information
Date Initiated by Firm April 08, 2014
Date Posted December 16, 2014
Recall Status1 Terminated 3 on May 31, 2018
Recall Number Z-0508-2015
Recall Event ID 67869
510(K)Number K111725  
Product Classification System, x-ray, mobile - Product Code IZL
Product Intended for both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with the patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.
Code Information Serial Numbers: SN12000024, SN12000166 , SN13000081 , SN13000127 , SN13000140 , SN13000141 , SN13000143 , SN13000144 , SN13000145 , SN13000167 , SN13000168 , SN13000169 , SN13000196 , SN13000199 , SN13000200 , SN13000201 , SN13000202 , SN13000203 , SN13000204 , SN13000205 , SN14000002 , SN14000004 , SN14000005 , SN14000006 , SN14000008 , SN14000009 , SN14000010 , SN14000011 , SN14000013 , SN14000014 , SN14000015 , SN14000016 , SN14000017 , SN14000019 , SN14000022 , SN14000024 , SN14000026 , SN14000028 , SN14000030 , SN14000031 , SN14000032 , SN14000033 , SN14000038 , SN14000041 , SN14000046 , SN14000047 , SN14000048 ,  SN13000153 ,  SN13000171 ,  SN13000155 ,  SN13000165 ,  SN13000187 ,  SN13000170 ,  SN13000176 ,  SN13000139 ,  SN13000189 ,  SN7369686 ,  SN13000150 ,  SN13000186 ,  SN13000197 ,  SN13000152 ,  SN13000162 ,  SN13000163 ,  SN14000012 ,  SN11000018 ,  SN13000136 ,  SN13000137 ,  SN13000138 ,  SN13000147 ,  SN13000154 ,  SN13000156 ,  SN13000159 ,  SN13000160 ,  SN13000164 ,  SN13000172 ,  SN13000175 ,  SN13000181 ,  SN13000188 ,  SN13000191 ,  SN13000194 ,  SN14000027 ,  SN130000184 ,  SN13000132 ,  SN13000146 ,  SN13000148 ,  SN13000149 ,  SN13000151 ,  SN13000161 ,  SN13000174 ,  SN13000177 ,  SN13000179 ,  SN13000179 ,  SN13000180 ,  SN13000183 ,  SN13000185 ,  SN13000190 ,  SN13000193 ,  SN13000206 ,  SN14000003 ,  SN14000023 ,  SN14000036 , SN11000001 , SN11000002 , SN11000003 , SN11000004 , SN11000005 , SN11000006 , SN11000007 , SN11000008 , SN11000009 , SN11000010 , SN11000011 , SN11000012 , SN11000013 , SN11000014 , SN11000014 , SN11000015 , SN11000016 , SN11000017 , SN11000020 , SN11000021 , SN11000021 , SN11000022 , SN11000023 , SN11000024 , SN11000025 , SN11000026 , SN11000027 , SN11000028 , SN11000029 , SN11000030 , SN11000031 , SN11000032 , SN11000033 , SN11000034 , SN11000035 , SN11000036 , SN11000037 , SN11000039 , SN11000040 , SN11000041 , SN11000042 , SN11000044 , SN11000045 , SN11000046 , SN11000047 , SN11000048 , SN11000048 , SN11000049 , SN11000050 , SN11000051 , SN11000052 , SN11000054 , SN11000055 , SN11000056 , SN11000057 , SN11000058 , SN11000059 , SN11000060 , SN11000061 , SN11000062 , SN11000063 , SN11000064 , SN11000065 , SN11000066 , SN11000067 , SN11000068 , SN11000069 , SN11000070 , SN11000071 , SN11000072 , SN11000073 , SN11000074 , SN11000075 , SN11000076 , SN11000077 , SN11000078 , SN11000079 , SN11000080 , SN11000081 , SN11000082 , SN11000083 , SN11000084 , SN12000002 , SN12000003 , SN12000004 , SN12000005 , SN12000006 , SN12000007 , SN12000008 , SN12000009 , SN12000010 , SN12000012 , SN12000013 , SN12000014 , SN12000015 , SN12000016 , SN12000018 , SN12000019 , SN12000020 , SN12000021 , SN12000022 , SN12000023 , SN12000025 , SN12000026 , SN12000027 , SN12000028 , SN12000029 , SN12000030 , SN12000031 , SN12000032 , SN12000033 , SN12000034 , SN12000035 , SN12000036 , SN12000037 , SN12000038 , SN12000039 , SN12000040 , SN12000041 , SN12000042 , SN12000043 , SN12000044 , SN12000045 , SN12000046 , SN12000047 , SN12000048 , SN12000049 , SN12000050 , SN12000051 , SN12000052 , SN12000053 , SN12000054 , SN12000055 , SN12000056 , SN12000057 , SN12000058 , SN12000059 , SN12000060 , SN12000061 , SN12000062 , SN12000063 , SN12000064 , SN12000065 , SN12000066 , SN12000067 , SN12000068 , SN12000069 , SN12000070 , SN12000071 , SN12000072 , SN12000073 , SN12000074 , SN12000075 , SN12000076 , SN12000077 , SN12000078 , SN12000079 , SN12000080 , SN12000081 , SN12000082 , SN12000083 , SN12000083 , SN12000084 , SN12000085 , SN12000086 , SN12000087 , SN12000088 , SN12000089 , SN12000090 , SN12000091 , SN12000092 , SN12000093 , SN12000094 , SN12000095 , SN12000098 , SN12000099 , SN12000100 , SN12000101 , SN12000102 , SN12000103 , SN12000104 , SN12000105 , SN12000106 , SN12000107 , SN12000108 , SN12000109 , SN12000110 , SN12000111 , SN12000112 , SN12000113 , SN12000114 , SN12000115 , SN12000116 , SN12000117 , SN12000118 , SN12000119 , SN12000120 , SN12000121 , SN12000122 , SN12000123 , SN12000124 , SN12000125 , SN13000001 , SN13000002 , SN13000004 , SN13000005 , SN13000006 , SN13000007 , SN13000008 , SN13000010 , SN13000011 , SN13000012 , SN13000013 , SN13000014 , SN13000015 , SN13000016 , SN13000017 , SN13000018 , SN13000019 , SN13000020 , SN13000021 , SN13000022 , SN13000023 , SN13000025 , SN13000026 , SN13000027 , SN13000028 , SN13000029 , SN13000030 , SN13000031 , SN13000032 , SN13000033 , SN13000034 , SN13000035 , SN13000036 , SN13000037 , SN13000038 , SN13000039 , SN13000040 , SN13000041 , SN13000042 , SN13000043 , SN13000044 , SN13000045 , SN13000046 , SN13000047 , SN13000048 , SN13000049 , SN13000050 , SN13000051 , SN13000052 , SN13000053 , SN13000054 , SN13000055 , SN13000056 , SN13000057 , SN13000058 , SN13000059 , SN13000060 , SN13000061 , SN13000062 , SN13000063 , SN13000064 , SN13000065 , SN13000066 , SN13000067 , SN13000069 , SN13000070 , SN13000071 , SN13000072 , SN13000073 , SN13000074 , SN13000075 , SN13000076 , SN13000077 , SN13000078 , SN13000079 , SN13000080 , SN13000082 , SN13000083 , SN13000084 , SN13000085 , SN13000086 , SN13000087 , SN13000088 , SN13000089 , SN13000090 , SN13000091 , SN13000092 , SN13000093 , SN13000094 , SN13000095 , SN13000096 , SN13000097 , SN13000098 , SN13000099 , SN13000100 , SN13000101 , SN13000102 , SN13000103 , SN13000104 , SN13000105 , SN13000106 , SN13000107 , SN13000108 , SN13000109 , SN13000110 , SN13000111 , SN13000112 , SN13000113 , SN13000114 , SN13000115 , SN13000116 , SN13000117 , SN13000118 , SN13000119 , SN13000120 , SN13000121 , SN13000122 , SN13000123 , SN13000124 , SN13000124 , SN13000125 , SN13000126 , SN13000128 , SN13000129 , SN13000130 , SN13000131 , SN13000133 , SN13000134 , SN13000135 , SN13000157 , SN13000173 , SN13000178 , SN13000182 , SN13000192 , SN13000198 , SN14000007 , SN14000018 , SN14000021 , SN14000025. Four systems were delivered to foreign (non-US) countries that do not have serial numbers.
Recalling Firm/
Manufacturer
Philips Medical Systems, Inc.
3000 Minuteman Rd
Andover MA 01810-1032
For Additional Information Contact Customer Care Solutions
800-722-9377 Ext. 5
Manufacturer Reason
for Recall
In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur.
FDA Determined
Cause 2
Software design
Action Philips Healthcare sent an Important Electronic Product Radiation Warning letters dated April 8, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter alerts customers of the product defect and advises customers of Philips' plan to update the ATP software. A Philips Service Engineer will contact the customer when the Field Action Kit is released for installation. A customer notification letter which includes a statement that Philips Healthcare will without charge, remedy the defect or bring the product into compliance. Customers with questions about this issue can contact their local Philips representative or Philips Healthcare office, or call the Customer Care Solutions Center at 800-722-9377, Option 5. Callers should reference Field Correction Order FCO-71200109.
Quantity in Commerce 247 in US, 1 in Puerto Rico
Distribution Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI.,and the countries of : Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iraq, Ireland, Italy, Jordan, Liberia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Arab Emirates and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = IZL and Original Applicant = SEDECAL SA
-
-