|
Class 2 Device Recall CelluTome |
 |
Date Initiated by Firm |
May 19, 2014 |
Date Posted |
September 29, 2014 |
Recall Status1 |
Terminated 3 on September 18, 2015 |
Recall Number |
Z-2717-2014 |
Recall Event ID |
68329 |
Product Classification |
Dermatome - Product Code GFD
|
Product |
CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50 |
Code Information |
Part number CT-H25 with the following lot numbers: 3244680000, 3446210000, 3246280000, 3246480000, 3246490000, 3246970000, 3247020000, 3247030000, 3247130000, 3247440000, 3247670000, 3247680000, 3247980000, 3248070000, 3248200000, 3248300000, 3249120000, 3249130000, 3249140000, 3249150000, 3249160000. Part number CT-H50 with the following lot numbers: 324782, 324796, 324797, 3245120000, 3245670000, 3245680000, 3246270000, 3246490000, 3246690000, 3246700000, 3246940000, 3246980000, 3247140000, 3247160000, 3247170000, 3247250000, 3247310000, 3247470000, 324780000, 3247810000, 3247820000, 3247960000, 3247970000, 3248080000, 3248190000, 3248230000, 3248300000, 3248380000, 3248960000, 3248970000, 3248980000, 3248990000, 3249000000, 3249010000, 3249020000, 3249030000, 3249040000, 3249050000, 3249060000, 3249070000, 3249080000, 3249090000, 3249100000, 3249110000, 3249390000, 3249580000. |
Recalling Firm/ Manufacturer |
KCI USA, Inc. 6203 Farinon Dr San Antonio TX 78249-3441
|
For Additional Information Contact |
Steven Jackson 800-275-4524 Ext. 3
|
Manufacturer Reason for Recall |
Blade can shift during shipping which makes the product ineffective.
|
FDA Determined Cause 2 |
Device Design |
Action |
KCI sent an Urgent - Voluntary Medical Device Field Notification letter dated June 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
The firm is sending Field Notifications to affected customers instructing them to complete appropriate actions.
Option I: Contact your local KCI Representative or KCI Customer Technical Support at 1-800-275-4524 and select Option 3 to arrange for return and replacement of the affected harvester(s).
Option 2: If you prefer, you may re-set the blade after removal from the packaging by gently tapping the curved end of the harvester against the palm of your hand, as shown in Figure 3, until the blade is no longer visible in the harvester top plate, as shown in Figure 4. |
Quantity in Commerce |
1,331 units |
Distribution |
US nationwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|