Date Initiated by Firm |
May 13, 2014 |
Date Posted |
August 26, 2014 |
Recall Status1 |
Terminated 3 on May 18, 2017 |
Recall Number |
Z-2280-2014 |
Recall Event ID |
68346 |
PMA Number |
P880081 |
Product Classification |
intraocular lens - Product Code HQL
|
Product |
AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002. Indicated for primary implantation for the visual correction of aphakia in adult patients. Tecnis CL Mono 13 mm with the following diopter sizes: 18.5-19.0
Product Usage: AMO TECNIS CL monofocal foldable IOLs with OptiEdge design are ultraviolet-absorbing posterior chamber intraocular lenses which compensate for corneal spherical aberrations. TECNIS CL Silicone Intraocular Lenses are indicated for primary implantation for the visual correction of aphakia in adult patients in whom a cataractous lens has been removed by extracapsular cataract extraction or phacoemulsification. These devices are intended to be placed in the capsular bag.
|
Code Information |
1001141312 1001151312 1001161312 1001171312 1001181312 1001191312 1001201312 1001211312 1001221312 1001231312 1001241312 1001251312 1001261312 1001271312 1001291312 1001301312 1001311312 1001321312 1001331312 1001341312 1001351312 1001361312 1001371312 1001381312 1001391312 1001401312 1001411312 1001421312 1001431312 1001441312 1001451312 1001461312 1001471312 1001481312 1001491312 1001501312 1001511312 1001521312 1001531312 1001541312 1001551312 1001571312 1001591312 1001611312 1001631312 1001651312 1001671312 1001691312 1001711312 1001731312 1001751312 1001771312 1001811312 1001831312 1001851312 1001871312 1001891312 1001911312 1001931312 1001951312 1001971312 1001991312 1002001312 1002021312 1002041312 1002061312 1002081312 1002101312 1002121312 1002131312 1002141312 1002151312 1002161312 1002171312 1002181312 1002201312 1002221312 1002231312 1005251309 1005261309 1005271309 1005281309 1005291309 1005301309 1005321309 1005341309 1005361309 1005381309 1005401309 1005421309 1005431309 1005441309 1005451309 1005461309 1005471309 1005491309 1005501309 1005511309 1005521309 1005531309 1005551309 1005561309 1005571309 1005581309 1005591309 1005601309 1005611309 1005621309 1005631309 1005641309 1005651309 1005671309 1005681309 1005691309 1005701309 1005711309 1005721309 1005731309 1005751309 1005771309 1005791309 1005811309 1005831309 1005851309 1005871309 1005891309 1005911309 1005931309 1005951309 1005971309 1005991309 1006011309 1006031309 1006051309 1006061309 1006071309 1006081309 1006091309 1006101309 1006111309 1006121309 1006131309 1006151309 1006171309 1006211309 1006231309 1006241309 1006251309 1006261309 1006271309 1006281309 1006291309 1006301309 1009131311 1009141311 1009151311 1009161311 1009171311 1009181311 1009191311 1009201311 1009211311 1009221311 1009231311 1009241311 1009251311 1009261311 1009271311 1009281311 1009291311 1009301311 1009321311 1009331311 1009341311 1009351311 1009361311 1009371311 1009381311 1009391311 1009401311 1009411311 1009421311 1009431311 1009441311 1009451311 1009461311 1009471311 1009481311 1009491311 1009501311 1009511311 1009531311 1009551311 1009571311 1009591311 1009691311 1009711311 1009731311 1009751311 1009771311 1009791311 1009811311 1009831311 1009851311 1009871311 1009891311 1009911311 1009921311 1009931311 1009941311 1009951311 1009961311 1009971311 1009981311 1009991311 1010001311 1010011311 1010021311 1010031311 1010041311 1010051311 1010061311 1010071311 1010081311 1010091311 1010101311 1010111311 1036291312 1036301312 1036311312 1036321312 1036331312 1036341312 1036351312 1036361312 1036371312 1036381312 1036391312 1036401312 1036411312 1036421312 1036431312 1036451312 1036471312 1036501312 1036521312 1036541312 1036551312 1036561312 1036571312 1036591312 1036611312 1036621312 1036631312 1036641312 1036651312 1036661312 1036671312 1036681312 1036691312 1036701312 1036711312 1036731312 1036751312 1036761312 1036771312 1036781312 1036791312 1036801312 1036811312 1036821312 1036831312 1036851312 1036881312 1036891312 1036901312 1036911312 1036921312 1036941312 1036961312 1036981312 1037001312 1037021312 1037041312 1037061312 1037081312 1037091312 1037101312 1037111312 1037121312 1037131312 1037141312 1037151312 1037161312 1037171312 1037191312 1037211312 1037221312 1037231312 1037241312 1037251312 1037261312 1037271312 1037281312 1037291312 1037301312 1037311312 1037321312 1037331312 1037341312 1037351312 1037361312 1037381312 1053351312 1053361312 1053371312 1053381312 1053391312 1053401312 1053411312 1053421312 1053431312 1053441312 1053451312 1053461312 1053471312 1053481312 1053491312 1053501312 1053511312 1053521312 1053531312 1053541312 1053551312 1053561312 1053571312 1053581312 1053591312 1053601312 1053611312 1053621312 1053631312 1053641312 1053651312 1053661312 1053671312 1053681312 1053691312 1053701312 1053711312 1053731312 1053741312 1053751312 1053761312 1053771312 1053781312 1053791312 1053811312 1053821312 1053831312 1053841312 1053851312 1053861312 1053881312 1053891312 1053901312 1053911312 1053921312 1053941312 1053961312 1053981312 1054001312 1054021312 1054041312 1054061312 1054081312 1054101312 1054121312 1054141312 1054161312 1054181312 1054201312 1054221312 1054261312 1054301312 1054321312 1054341312 1054361312 1054381312 1054401312 1064091311 1064111311 1064131311 1064151311 1064191311 1064671311 1064771311 1064791311 1064811311 1064831311 1064851311 1064871311 1064891311 1064911311 1069841310 1069851310 1069861310 1069871310 1069881310 1069891310 1069901310 1069911310 1069921310 1069931310 1069941310 1069951310 1069961310 1069971310 1069981310 1069991310 1070001310 1070011310 1070021310 1070031310 1070041310 1070051310 1070061310 1070071310 1070081310 1070091310 1070101310 1070111310 1070121310 1070131310 1070141310 1070151310 1070161310 1070171310 1070181310 1070191310 1070201310 1070211310 1070221310 1070231310 1070241310 1070251310 1070261310 1070271310 1070281310 1070291310 1070301310 1070311310 1070321310 1070331310 1070341310 1070351310 1070361310 1070371310 1070381310 1070391310 1070401310 1070411310 1070421310 1070431310 1070441310 1070451310 1070461310 1070471310 1070481310 1070491310 1070501310 1070511310 1070521310 1070531310 1070541310 1070551310 1070561310 1070571310 1070581310 1070591310 1070601310 1070611310 1070621310 1070631310 1070641310 1070651310 1070661310 1070671310 1070681310 1070691310 1070701310 1070711310 1081681310 1081701310 1081721310 1081741310 1081761310 1081781310 1081801310 1081841310 1081861310 1081881310 1081901310 1081921310 1081941310 1081961310 1081981310 1082001310 1082011310 1082021310 1082041310 1082051310 1082061310 1082081310 1082291310 1082311310 1082331310 1082351310 1082371310 1082391310 1082401310 1082411310 1082421310 1082441310 1082461310 1082481310 1082501310 1082521310 1082541310 1082561310 1082581310 1082601310 1082621310 1082641310 1082661310 1082681310 1082701310 1082721310 1082741310 1099671309 1099691309 1099711309 1099721309 1099731309 1099741309 1099751309 1099761309 1099771309 1099781309 1099791309 1099801309 1099811309 1099821309 1099831309 1099841309 1099851309 1099861309 1099871309 1099881309 1099891309 1099901309 1099911309 1099921309 1099931309 1099951309 1099961309 1099971309 1099981309 1099991309 1100001309 1100011309 1100021309 1100031309 1100041309 1100051309 1100061309 1100071309 1100081309 1100091309 1100101309 1100111309 1100121309 1100131309 1100141309 1100151309 1100161309 1100171309 1100181309 1100191309 1100201309 1100211309 1100221309 1100231309 1100241309 1100251309 1100261309 1100271309 1100281309 1100301309 1100311309 1100321309 1100331309 1100341309 1100361309 1100381309 1100401309 1100421309 1100431309 1100441309 1100451309 1100461309 1100471309 1100481309 1100501309 1100521309 1100541309 1100561309 1100581309 1100601309 1100611309 1100621309 1100641309 1100661309 1100681309 1100691309 1100711309 1100731309 1100741309 1100761309 1100781309 1101091309 1101111309 1101121309 1101131309 1101141309 1101151309 1101161309 1101171309 1101181309 1101191309 1101201309 1101211309 1101221309 1101231309 1101241309 1101251309 1101261309 1101271309 1101281309 1101291309 1101301309 1101311309 1101331309 1101351309 1101361309 1101371309 1101391309 1101401309 1101411309 1101421309 1101431309 1101451309 1101471309 1101481309 1101491309 1101501309 1101511309 1101521309 1101531309 1101541309 1101551309 1101561309 1101571309 1101581309 1101591309 1101601309 1101611309 1101621309 1101631309 1101641309 1101661309 1101681309 1101701309 1101721309 1101741309 1101761309 1101781309 1101801309 1101811309 1102951309 1102961309 1102971309 1102981309 1102991309 1103001309 1103011309 1103021309 1103031309 1103071309 1103081309 1103091309 1103101309 1103111309 1103121309 1103131309 1103141309 1103151309 1103161309 1103171309 1103181309 1103201309 1103211309 1103241309 1103261309 1103271309 1103281309 1103301309 1103321309 1103341309 1103361309 1103381309 1103391309 1103401309 1103421309 1103431309 1103441309 1103451309 1103461309 1103471309 1103481309 1103491309 1103501309 1103511309 1103521309 1103531309 1103541309 1103551309 1103561309 1103571309 1103581309 1103591309 1103601309 1103621309 1103631309 1103641309 1103661309 1103681309 1103701309 1103721309 1103731309 1103751309 1103771309 1103811309 1103821309 1103831309 1103841309 1103851309 1103861309 1103871309 1103881309 1103891309 1103901309 1103911309 1103921309 1103931309 1103941309 1103951309 1103961309 1103971309 1103981309 1103991309 1104001309 1104011309 1104021309 1104031309 1104071310 1104091310 1104111310 1104131310 1104291310 1104311310 1104331310 1104341310 1104351310 1104381310 1104401310 1104641310 1104661310 1104681310 1104701310 1104731310 1104751310 1104771310 1104791310 1104811310 1104831310 1104841310 1104851310 1104861310 1104871310 1104881310 1104891310 1104901310 1104911310 1115581311 1115601311 1115621311 1115641311 1115661311 1115681311 1115701311 1115721311 1115741311 1115761311 1115841311 1115861311 1115881311 1115901311 1115911311 1115921311 1115931311 1115941311 1115961311 1115981311 1115991311 1116001311 1116011311 1116021311 1116041311 1116061311 1116081311 1116101311 1116111311 1116121311 1116131311 1116141311 1116151311 1116161311 1116171311 1116181311 1116191311 1116201311 1116211311 1116231311 1116251311 1116271311 1116291311 1116311311 1116331311 1116351311 1116371311 1116411311 1116431311 1116451311 1116471311 1116591311 1117741311 1117751311 1117761311 1117771311 1117781311 1117791311 1117801311 1117811311 1117821311 1117831311 1117841311 1117851311 1117861311 1117871311 1117881311 1117891311 1117901311 1117911311 1117921311 1117931311 1117941311 1117951311 1117961311 1117971311 1117981311 1117991311 1118001311 1118011311 1118021311 1118031311 1118041311 1118051311 1118061311 1118071311 1118081311 1118091311 1118101311 1118111311 1118121311 1118131311 1118141311 1118151311 1118161311 1118171311 1118181311 1118191311 1118201311 1118211311 1118221311 1118231311 1118241311 1118251311 1118261311 1118271311 1118281311 1118291311 1118301311 1118311311 1118321311 1118331311 1118341311 1118351311 1118361311 1118371311 1118381311 1118391311 1118401311 1118411311 1118421311 1118431311 1118441311 1118451311 1118461311 1118471311 1118481311 1118491311 1118501311 1118511311 1118521311 1118531311 1118541311 1118551311 1118561311 1118571311 1118581311 1118591311 1118601311 1118611311 1118621311 1118631311 1118641311 1118651311 1118661311 1118671311 1118681311 1118691311 1118701311 1118711311 1118721311 1118731311 1118741311 1118751311 1118761311 1118771311 1118781311 1118791311 1123191311 1123331311 1123351311 1123371311 1123391311 1123411311 1123431311 1123451311 1123471311 1123491311 1123511311 1123531311 1123551311 1123561311 1123571311 1123581311 1123591311 1123601311 1123611311 1123621311 1123631311 1123641311 1123651311 1123661311 1123671311 1123681311 1123701311 1123711311 1123721311 1123731311 1123741311 1123751311 1123761311 1123771311 1123781311 1123801311 1123811311 1123821311 1123841311 1123861311 1123881311 1123901311 1123921311 1123931311 1123941311 1123951311 1123961311 1123971311 1123981311 1123991311 1124001311 1124011311 1124021311 1124031311 1124041311 1124051311 1124061311 1124071311 1124081311 1124091311 1124101311 1124111311 1124121311 1124131311 1124141311 1124151311 1124161311 1124171311 1124181311 1124191311 1125251312 1125271312 1125291312 1125311312 1125331312 1125351312 1125371312 1125391312 1125451312 1125461312 1125621312 1125641312 1125731312 1125751312 1125771312 1125791312 1125811312 1125831312 1125851312 1125871312 1125891312 1125911312 1125931312 1125941309 1125951309 1125951312 1125971309 1125971312 1125981309 1125991309 1125991312 1126001309 1126011309 1126011312 1126031309 1126031312 1126041309 1126051309 1126051312 1126061309 1126061312 1126071309 1126071312 1126081309 1126081312 1126091309 1126091312 1126101309 1126101312 1126111309 1126111312 1126121309 1126121312 1126131309 1126131312 1126141309 1126141312 1126151309 1126151312 1126161309 1126161312 1126171309 1126171312 1126181309 1126181312 1126191309 1126191312 1126201309 1126201312 1126211309 1126211312 1126221309 1126221312 1126231309 1126231312 1126241309 1126261309 1126271309 1126281309 1126291309 1126301309 1126311309 1126321309 1126331309 1126341309 1126351309 1126361309 1126371309 1126381309 1126391309 1126401309 1126411309 1126421309 1126431309 1126441309 1126451309 1126461309 1126471309 1126481309 1126491309 1126501309 1126511309 1126521309 1126531309 1126541309 1126551309 1126561309 1126571309 1126581309 1126591309 1126601309 1126611309 1126621309 1126631309 1126641309 1126651309 1126661309 1126681309 1126691309 1126701309 1126711309 1126721309 1126731309 1126741309 1126751309 1126761309 1126771309 1126781309 1126791309 1126801309 1126811309 1126821309 1126831309 1126841309 1126851309 1126861309 1126871309 1126881309 1126891309 1126901309 1126911309 1126921309 1126931309 1126941309 1126951309 1126961309 1126971309 1126981309 1126991309 1127001309 1127011309 1127021309 1127031309 1127041309 1127051309 1127061309 1127071309 1127081309 1127561311 1127571311 1127581311 1127591311 1127611311 1127621311 1127631311 1127641311 1127651311 1127661311 1127671311 1127681311 1127691311 1127701311 1127711311 1127721311 1127731311 1127741311 1127751311 1127761311 1127771311 1127781311 1127791311 1127801311 1127811311 1127821311 1127831311 1127841311 1127851311 1127861311 1127871311 1127881311 1127891311 1127901311 1127911311 1127921311 1127931311 1127941311 1127951311 1127961311 1127971311 1127981311 1127991311 1128001311 1128011311 1128021311 1128031311 1128041311 1128051311 1128061311 1128071311 1128081311 1128091311 1128101311 1128111311 1128121311 1128131311 1128141311 1128151311 1128161311 1128171311 1128181311 1128191311 1128201311 1128211311 1128221311 1128231311 1128241311 1128251311 1128261311 1128271311 1128281311 1128291311 1128301311 1128311311 1128321311 1128331311 1128341311 1128351311 1128361311 1128371311 1128381311 1128391311 1128401311 1128411311 1128421311 1128431311 1128441311 1128451311 1128461311 1128471311 1128481311 1128491311 1128501311 1128511311 1128521311 1128531311 1128541311 1128551311 1128571311 1128581311 1128591311 1128601311 1128611311 1128621311 1128631311 1128641311 1128651311 1128661311 1128671311 1128681311 1130691312 1130711312 1130941312 1130961312 1130981312 1131001312 1131021312 1131041312 1131051312 1131061312 1131081312 1131101312 1131121312 1131141312 1131151312 1131161312 1131181312 1131201312 1131211312 1131221312 1131231312 1131241312 1131251312 1131261312 1131271312 1131281312 1131291312 1131301312 1131311312 1131321312 1131331312 1131341312 1131351312 1131361309 1131381309 1131381312 1131391309 1131401309 1131401312 1131411309 1131421309 1131421312 1131431309 1131441309 1131451309 1131461309 1131471309 1131481309 1131491309 1131501309 1131511309 1131521309 1131531309 1131541309 1131551309 1131561309 1131571309 1131581309 1131591309 1131601309 1131611309 1131621309 1131631309 1131641309 1131651309 1131661309 1131671309 1131681309 1131691309 1131701309 1131711309 1131721309 1131741309 1131761309 1131781309 1131791309 1131801309 1131821309 1131841309 1131861309 1131881309 1131901309 1131921309 1131941309 1131981309 1132001309 1132021309 1132041309 1132061309 1132071309 1132081309 1132091309 1132101309 1132121309 1132131309 1132141309 1132151309 1132161309 1132171309 1132181309 1132191309 1132201309 1132211309 1132221309 1132231309 1132241309 1132261309 1132281309 1132301309 1132321309 1132351309 1132371309 1132391309 1132401309 1132411309 1132421309 1132431309 1132441309 1132451309 1132461309 1132471309 1132481309 1132491309 1133611310 1133621310 1133631310 1133651310 1133671310 1133691310 1133711310 1133731310 1133741310 1133751310 1133761310 1133771310 1133791310 1133811310 1133831310 1133851310 1133871310 1133881310 1133891310 1133901310 1133911310 1133921310 1133931310 1133941310 1133951310 1133961310 1133971310 1133981310 1133991310 1134001310 1134011310 1134021310 1134031310 1134041310 1134051310 1134061310 1134071310 1134081310 1134091310 1134101310 1134111310 1134121310 1134131310 1134141310 1134151310 1134161310 1134171310 1134181310 1134191310 1134201310 1134211310 1134221310 1134231310 1134241310 1134261310 1134281310 1134301310 1134321310 1134331310 1134341310 1134351310 1134361310 1134371310 1134381310 1134391310 1134401310 1134411310 1134421310 1134431310 1134441310 1134451310 1134461310 1134471310 1134481310 1134491310 1134501310 1134511310 1134531310 1134551310 1134571310 1134591310 1134681310 1134701310 1134721310 1135751310 1135761310 1135771310 1135781310 1135791310 1135801310 1135811310 1135821310 1135831310 1135841310 1135851310 1135861310 1135871310 1135881310 1135891310 1135901310 1135911310 1135921310 1135931310 1135941310 1135951310 1135961310 1135971310 1135981310 1135991310 1136001310 1136011310 1136021310 1136031310 1136041310 1136051310 1136061310 1136071310 1136081310 1136091310 1136101310 1136111310 1136121310 1136131310 1136141310 1136151310 1136161310 1136171310 1136181310 1136191310 1136201310 1136211310 1136221310 1136231310 1136241310 1136251310 1136261310 1136271310 1136281310 1136291310 1136301310 1136311310 1136321310 1136331310 1136341310 1136351310 1136361310 1136371310 1136381310 1136391310 1136411310 1136421310 1136431310 1136451310 1136461310 1136471310 1136491310 1136501310 1136511310 1136521310 1136531310 1136541310 1136551310 1136561310 1136571310 1136581310 1136591310 1136601310 1136611310 1136621310 1136631310 1136641310 1136651310 1136661310 1136671310 1136681310 1137841311 1137861311 1137881311 1137901311 1137921311 1137941311 1137961311 1137981311 1138001311 1138021311 1138041311 1138061311 1138081311 1138101311 1138111311 1138121311 1138131311 1138141311 1138161311 1138181311 1138201311 1138211311 1138221311 1138241311 1138251311 1138261311 1138271311 1138281311 1138291311 1138301311 1138311311 1138331311 1138351311 1138371311 1138391311 1138411311 1138431311 1138451311 1142011310 1142021310 1142031310 1142041310 1142051310 1142061310 1142071310 1142081310 1142091310 1142101310 1142111310 1142121310 1142131310 1142141310 1142151310 1142161310 1142171310 1142181310 1142191310 1142201310 1142211310 1142221310 1142231310 1142241310 1142251310 1142261310 1142271310 1142281310 1142291310 1142301310 1142311310 1142321310 1142331310 1142341310 1142351310 1142361310 1142371310 1142381310 1142391310 1142401310 1142411310 1142421310 1142431310 1142441310 1142451310 1142461310 1142471310 1142481310 1142491310 1142501310 1142511310 1142521310 1142531310 1142541310 1142551310 1142561310 1142571310 1142581310 1142591310 1142601310 1142611310 1142621310 1142631310 1142641310 1142651310 1142661310 1142671310 1142681310 1142691310 1142711310 1142731310 1142741310 1142751310 1142761310 1142771310 1142781310 1142791310 1142801310 1142811310 1142821310 1142831310 1142841310 1142851310 1142861310 1142871310 1142881310 1142891310 1142901310 1142911310 1142921310 1142931310 1142941310 1142951310 1142961310 1142971310 1142981310 1142991310 1143001310 1143011310 1143021310 1143031310 1143041310 1143051310 1143061310 1143071310 1143081310 1143091310 1143111310 1143131310 1148241311 1148261311 1148281311 1148301311 1148321311 1148341311 1148361311 1148381311 1148401311 1148421311 1148441311 1148461311 1148481311 1148501311 1148521311 1148541311 1148561311 1148581311 1148601311 1148621311 1148641311 1148651311 1148661311 1148671311 1148681311 1148691311 1148701311 1148711311 1148721311 1148731311 1148741311 1148751311 1148761311 1148771311 1148781311 1148791311 1148801311 1148811311 1148821311 1148831311 1148841311 1148851311 1148861311 1148871311 1148881311 1148901311 1148921311 1148941311 1148961311 1148981311 1148991311 1149001311 1149011311 1149021311 1149031311 1149041311 1149051311 1149061311 1149071311 1149081311 1149091311 1149101311 1149111311 1149121311 1149131311 1149141311 1149151311 1149161311 1149181311 1149191311 1149201311 1149211311 1149221311 1149231311 1149241311 1149251311 1149261311 1149271311 1149281311 1149291311 1149301311 1149311311 1149321311 1149331311 1149341311 1149351311 1176651310 1176661310 1176671310 1176681310 1176691310 1176701310 1176711310 1176721310 1176731310 1176741310 1176751310 1176761310 1176771310 1176781310 1176791310 1176801310 1176811310 1176821310 1176831310 1176841310 1176851310 1176861310 1176881310 1176901310 1176921310 1176941310 1176951310 1176961310 1176971310 1176981310 1176991310 1177001310 1177011310 1177021310 1177031310 1177041310 1177051310 1177061310 1177071310 1177081310 1177091310 1177101310 1177111310 1177121310 1177131310 1177141310 1177151310 1177161310 1177171310 1177181310 1177191310 1177201310 1177211310 1177221310 1177231310 1177241310 1177251310 1177261310 1177271310 1177281310 1177291310 1177301310 1177311310 1177321310 1177331310 1177341310 1177351310 1177361310 1177371310 1177381310 1177391310 1177401310 1177411310 1177421310 1177431310 1177441310 1177451310 1177461310 1177471310 1177481310 1177491310 1177501310 1177511310 1177521310 1177531310 1177541310 1177551310 1177561310 1177571310 1177581310 1177591310 1177601310 1177611310 1177621310 1177631310 1177641310 1177651310 1177661310 1177681310 1177691310 1177701310 1177711310 1177721310 1177731310 1177741310 1177751310 1177761310 1184531310 1184541310 1184551310 1184571310 1184581310 1184591310 1184611310 1184621310 1184631310 1184651310 1184661310 1184671310 1184691310 1184701310 1184711310 1184731310 1184741310 1184751310 1184771310 1184781310 1184791310 1184811310 1184821310 1184831310 1184851310 1184861310 1184871310 1184891310 1184901310 1184911310 1184921310 1184931310 1184941310 1184951310 1184961310 1184971310 1184981310 1184991310 1185001310 1185011310 1185021310 1185031310 1185041310 1185051310 1185061310 1185071310 1185081310 1185091310 1185101310 1185111310 1185121310 1185131310 1185141310 1185151310 1185161310 1185171310 1185181310 1185191310 1185201310 1185211310 1185221310 1185231310 1185241310 1185251310 1185261310 1185271310 1185281310 1185291310 1185301310 1185311310 1185321310 1185331310 1185341310 1185351310 1185361310 1185371310 1185381310 1185391310 1185401310 1185411310 1185421310 1185431310 1185441310 1185451310 1185461310 1185471310 1185481310 1185491310 1185501310 1185511310 1185521310 1185531310 1185541310 1185551310 1185561310 1185571310 1185581310 1185591310 1185601310 1185611310 1185621310 1185631310 1185641310 1185651310 1185661310 1185671310 1185681310 1209061309 1209071309 1209081309 1209091309 1209101309 1209111309 1209121309 1209131309 1209141309 1209151309 1209161309 1209171309 1209181309 1209191309 1209201309 1209211309 1209221309 1209231309 1209241309 1209251309 1209261309 1209281309 1209301309 1209321309 1209341309 1209361309 1209381309 1209401309 1209421309 1209441309 1209451309 1209461309 1209471309 1209481309 1209491309 1209501309 1209521309 1209531309 1209541309 1209551309 1209561309 1209571309 1209581309 1209591309 1209601309 1209611309 1209621309 1209631309 1209641309 1209661309 1209681309 1209701309 1209721309 1209741309 1209761309 1209781309 1209801309 1209821309 1209841309 1209861309 1209881309 1209901309 1209921309 1209941309 1209951309 1209961309 1209971309 1209981309 1209991309 1210001309 1210011309 1210021309 1210031309 1210041309 1210051309 1210061309 1210071309 1210081309 1210091309 1210101309 1210111309 1210121309 1210131309 1210141309 1210151309 1210161309 1210171309 1246681311 1246691311 1246701311 1246721311 1246731311 1246741311 1246751311 1246761311 1246771311 1246781311 1246791311 1246801311 1246811311 1246821311 1246831311 1246841311 1246851311 1246861311 1246871311 1246881311 1246891311 1246901311 1246911311 1246921311 1246931311 1246941311 1246951311 1246961311 1246971311 1246991311 1247001311 1247011311 1247021311 1247031311 1247041311 1247051311 1247061311 1247071311 1247081311 1247091311 1247101311 1247111311 1247121311 1247131311 1247151311 1247171311 1247191311 1247211311 1247231311 1247251311 1247271311 1247291311 1247311311 1247331311 1247351311 1247371311 1247391311 1247401311 1247411311 1247431311 1247441311 1247451311 1247471311 1247481311 1247491311 1247511311 1247521311 1247531311 1247541311 1247551311 1247561311 1247571311 1247581311 1247591311 1247601311 1247611311 1247621311 1247631311 1247641311 1247651311 1247661311 1247671311 1247681311 1247691311 1247701311 1247711311 1247721311 1247731311 1247741311 1247751311 1247761311 1247771311 1247781311 1247791311 1247801311 1247811311 1247821311 1247831311 1247841311 1247851311 1247861311 1247871311 1247881311 1247891311 1247901311 1247911311 1247921311 1247931311 1247941311 1247951311 1247961311 1247971311 1247981311 1247991311 1248001311 1248011311 1248021311 1248031311 1248041311 1248051311 1248061311 1248071311 1248081311 1248091311 1248101311 1248111311 1248121311 1248131311 1248141311 1248151311 1248161311 1248171311 1248181311 1248191311 1248201311 1248211311 1248221311 1248231311 1248241311 1248251311 1248261311 1248271311 1248281311 1248291311 1248301311 1248311311 1248321311 1248331311 1248341311 1248351311 1248361311 1248371311 1248381311 1248391311 1248401311 1248411311 1248421311 1248431311 1248441311 1248451311 1248461311 1248471311 1248481311 1248491311 1248501311 1248511311 1248521311 1248531311 1248541311 1248551311 1248561311 1248571311 1248581311 1248591311 1251731309 1251751309 1251771309 1251791309 1251811309 1251831309 1251851309 1251891309 1251911309 1251931309 1251951309 1252211309 1252231309 1252251309 1252271309 1252291309 1252561309 1252581309 1252621309 1252641309 1252821309 1252841309 1253991309 1254031309 1254071309 1254111309 1254151309 1254191309 1254231309 1254271309 1254311309 1254321309 1254351309 1254361309 1254381309 1254391309 1254401309 1254421309 1254431309 1254461309 1254471309 1254501309 1254511309 1254541309 1254551309 1254581309 1254591309 1254621309 1254631309 1254661309 1254671309 1254701309 1254711309 1254741309 1254781309 1254821309 1254861309 1254901309 1254941309 1254981309 1255021309 1255681309 1255701309 1255721309 1255741309 1255761309 1255781309 1255801309 1255821309 1255841309 1255861309 1255881309 1255901309 1256041309 1256051309 1256051310 1256061310 1256071310 1256081310 1256091309 1256091310 1256101310 1256111309 1256111310 1256121310 1256131310 1256141310 1256151309 1256151310 1256161310 1256171309 1256171310 1256181310 1256191310 1256201309 1256201310 1256211310 1256221310 1256231310 1256241310 1256251310 1256261310 1256271310 1256281310 1256291310 1256301310 1256311310 1256321310 1256331310 1256341310 1256351310 1256361310 1256371310 1256381310 1256391310 1256401310 1256411310 1256421310 1256431310 1256441310 1256451310 1256461310 1256471310 1256481310 1256491310 1256501310 1256511310 1256521310 1256531310 1256541310 1256551310 1256561310 1256571310 1256581310 1256591310 1256601310 1256611310 1256621310 1256631310 1256641310 1256651310 1256661310 1256671310 1256681310 1256691310 1256711310 1256721310 1256731310 1256741310 1256751310 1256771310 1256791310 1256811310 1256831310 1256851310 1256871310 1256881310 1256891310 1256901310 1256911310 1256921310 1256931310 1256941310 1256951310 1279061310 1279071310 1279081310 1279091310 1279101310 1279111310 1279121310 1279131310 1279141310 1279151310 1279161310 1279171310 1279181310 1279191310 1279201310 1279211310 1279221310 1279231310 1279241310 1279251310 1279261310 1279271310 1279281310 1279291310 1279301310 1279311310 1279321310 1279331310 1279341310 1279351310 1279361310 1279371310 1279381310 1279391310 1279401310 1279411310 1279431310 1279441310 1279451310 1279461310 1279471310 1279491310 1279511310 1279551310 1279571310 1279591310 1279611310 1279631310 1279651310 1279661310 1279671310 1279691310 1279711310 1279731310 1279751310 1279771310 1279791310 1279811310 1279831310 1279851310 1279871310 1279881310 1279891310 1279901310 1279911310 1279921310 1279931310 1279941310 1279951310 1287121308 1287141308 1287161308 1287181308 1287201308 1287221308 1287231308 1287241308 1287251308 1287261308 1287271308 1287281308 1287291308 1287301308 1287311308 1287321308 1287331308 1287341308 1287351308 1287361308 1287371308 1287381308 1287391308 1287401308 1287411308 1287421308 1287431308 1287441308 1287451308 1287461308 1287471308 1287481308 1287491308 1287501308 1287511308 1287521308 1287531308 1287541308 1287551308 1287561308 1287571308 1287581308 1287591308 1287601308 1287611308 1287621308 1287631308 1287641308 1287651308 1287661308 1287671308 1287681308 1287691308 1287701308 1287711308 1287721308 1287731308 1287741308 1287751308 1287761308 1287771308 1287781308 1287791308 1287811308 1287831308 1287851308 1287871308 1287891308 1287901308 1287911308 1287921308 1287931308 1287941308 1287951308 1287961308 1287971308 1287981308 1287991308 1288001308 1288011308 1288021308 1288031308 1288041308 1288051308 1288061308 1288071308 1288081308 1288091308 1288101308 1288111308 1288121308 1288131308 1288141308 1288151308 1288161308 1288171308 1288181308 1288191308 1288201308 1288211308 1288231308 1288251308 1292511308 1292521308 1292531308 1292541308 1292551308 1292561308 1292571308 1292581308 1292591308 1292601308 1292611308 1292621308 1292631308 1292641308 1292651308 1292661308 1292671308 1292681308 1292691308 1292701308 1292711308 1292721308 1292731308 1292741308 1292751308 1292761308 1292771308 1292781308 1292801308 1292821308 1292841308 1292851308 1292861308 1292871308 1292881308 1292891308 1292901308 1292911308 1292921308 1292931308 1292941308 1292951308 1292961308 1292971308 1292981308 1292991308 1293001308 1293011308 1293021308 1293041308 1293061308 1293081308 1293101308 1293111308 1293121308 1293141308 1293161308 1293171308 1293181308 1293191308 1293201308 1293211308 1293221308 1293241308 1293261308 1293281308 1293301308 1293321308 1293361308 1293381308 1293401308 1293421308 1293441308 1293461308 1293481308 1318291308 1318301308 1318311308 1318321308 1318331308 1318341308 1318351308 1318361308 1318371308 1318381308 1318391308 1318401308 1318411308 1318421308 1318431308 1318441308 1318451308 1318461308 1318471308 1318481308 1318491308 1318511308 1318531308 1318551308 1318571308 1318591308 1318611308 1318631308 1318641308 1318651308 1318661308 1318671308 1318681308 1318691308 1318701308 1318711308 1318721308 1318731308 1318741308 1318751308 1318761308 1318771308 1318781308 1318791308 1318801308 1318811308 1318821308 1318831308 1318841308 1318851308 1318861308 1318871308 1318881308 1318891308 1318901308 1318911308 1318921308 1318931308 1318941308 1318951308 1318961308 1318981308 1319001308 1319021308 1319041308 1319051308 1319061308 1319081308 1319091308 1319101308 1319111308 1319121308 1319131308 1319141308 1319151308 1319161308 1319171308 1319181308 1319191308 1319201308 1319211308 1319221308 1319231308 1319241308 1319251308 1319261308 1319271308 1319281308 1319291308 1025221311 1025231311 1025241311 1025251311 1025271311 1025291311 1025301311 1025311311 1025321311 1025331311 1025341311 1025351311 1025361311 1025371311 1025381311 1025391311 1025401311 1025411311 1025421311 1025431311 1025441311 1025451311 1025461311 1025471311 1025481311 1025491311 1025501311 1025511311 1025521311 1025531311 1025541311 1025551311 1025561311 1025571311 1025581311 1025591311 1025601311 1025611311 1025621311 1025631311 1025641311 1025651311 1025661311 1025671311 1025681311 1025691311 1025701311 1025711311 1025721311 1025731311 1025741311 1025751311 1025761311 1025771311 1025781311 1025791311 1025801311 1025811311 1025821311 1025831311 1025841311 1025851311 1025861311 1025871311 1025881311 1025891311 1025901311 1025911311 1025921311 1025931311 1025941311 1025951311 1025961311 1025971311 1025981311 1025991311 1026001311 1026011311 1026021311 1026031311 1026041311 1026051311 1026061311 1026071311 1026081311 1026091311 1026101311 1026111311 1026121311 1026131311 1026141311 1026151311 1026161311 1026171311 1026181311 1028511311 1028521311 1028531311 1028541311 1028551311 1028561311 1028571311 1028581311 1028591311 1028601311 1028611311 1028621311 1028631311 1028641311 1028651311 1028661311 1028671311 1028681311 1028691311 1028701311 1028711311 1028721311 1028731311 1028741311 1028751311 1028761311 1028771311 1028791311 1028801311 1028811311 1028821311 1028831311 1028841311 1028851311 1028861311 1028871311 1028881311 1028891311 1028901311 1028911311 1028921311 1028931311 1028941311 1028951311 1028961311 1028971311 1028981311 1028991311 1029001311 1029011311 1029021311 1029031311 1029041311 1029051311 1029061311 1029071311 1029081311 1029091311 1029101311 1029111311 1029121311 1029131311 1029141311 1029151311 1029161311 1029171311 1029181311 1029191311 1029201311 1029211311 1029221311 1029231311 1029241311 1029251311 1029261311 1029271311 1029281311 1029291311 1029301311 1029311311 1029321311 1029331311 1029341311 1029351311 1029361311 1029371311 1029381311 1029391311 1029401311 1029411311 1029421311 1029431311 1029441311 1029451311 1029461311 1029471311 1029481311 1029491311 1029501311 1029521311 1032841311 1032851311 1032861311 1032871311 1032881311 1032891311 1032901311 1032911311 1032921311 1032931311 1032941311 1032951311 1032961311 1032971311 1032981311 1032991311 1033001311 1033011311 1033021311 1033031311 1033041311 1033061311 1033081311 1033101311 1033121311 1033141311 1033161311 1033181311 1033191311 1033201311 1033211311 1033221311 1033231311 1033241311 1033251311 1033261311 1033271311 1033281311 1033291311 1033301311 1033311311 1033321311 1033331311 1033341311 1033351311 1033361311 1033371311 1033381311 1033391311 1033401311 1033411311 1033421311 1033431311 1033441311 1033451311 1033461311 1033471311 1033481311 1033491311 1033501311 1033511311 1033521311 1033531311 1033541311 1033551311 1033561311 1033571311 1033581311 1033591311 1033601311 1033611311 1033621311 1033631311 1033641311 1033651311 1033661311 1033671311 1033681311 1033691311 1033701311 1033711311 1033721311 1033731311 1033741311 1033751311 1033771311 1033791311 1033811311 1033831311 1033851311 1033861311 1033871311 1033881311 1033891311 1033901311 1033911311 1033921311 1033931311 1033941311 1033951311 1033961311 1033971311 1033981311 1033991311 1034001311 1034011311 1034021311 1034031311 1034041311 1034051311 1034061311 1034071311 1034081311 1034091311 1034101311 1034111311 1034121311 1034131311 1034141311 1034151311 1034161311 1034171311 1034181311 1034191311 1034201311 1034211311 1034231311 1034241311 1034251311 1034271311 1034281311 1034291311 1034301311 1034311311 1034321311 1034331311 1034341311 1034351311 1034361311 1034371311 1034381311 1034391311 1034401311 1034411311 1034421311 1034431311 1034451311 1034471311 1034491311 1034511311 1034531311 1034551311 1034571311 1034591311 1034611311 1034631311 1034651311 1034671311 1034691311 1034711311 1034731311 1034741311 1034751311 1034761311 1034771311 1034781311 1034791311 1034801311 1034811311 1034831311 1034851311 1034861311 1034871311 1034881311 1034891311 1034901311 1034911311 1034921311 1034931311 1034941311 1042331312 1042351312 1042371312 1042771310 1042791310 1042811310 1042831310 1042851310 1042871310 1042881310 1042891310 1042901310 1042911310 1042921310 1042931310 1042941310 1042951310 1042961310 1042971310 1042981310 1042991310 1043001310 1043011310 1043021310 1043031310 1043041310 1043121312 1043141312 1043161312 1043291310 1043291312 1043311310 1043321312 1043331310 1043341312 1043351310 1043371310 1043381312 1043391310 1043401312 1043421312 1043441312 1043501312 1043511312 1043521312 1043591312 1043601312 1043611312 1043641312 1043661312 1043741312 1043801312 1043861312 1043881312 1043901312 1043921312 1043941312 1043961312 1043981312 1044001312 1044021312 1044031312 1044041312 1044051312 1044071312 1044081312 1044091312 1044101312 1044121312 1044141312 1044151312 1044171312 1044191312 1044211312 1044231312 1044251312 1044271312 1044471312 1044491312 1044511312 1044531312 1044541311 1044551312 1044561311 1044571312 1044581311 1044601311 1044621311 1044641311 1044661311 1044681311 1044691312 1044701311 1044711311 1044721311 1044731311 1044751311 1044771311 1044781311 1044791311 1044801311 1044811311 1044821311 1044831311 1044841311 1044851311 1044861311 1044871311 1044881311 1044891311 1044901311 1044911311 1044921311 1044961311 1044981311 1044991311 1045001311 1045021311 1045041311 1045061311 1045081311 1045101311 1045121311 1045131311 1045141311 1045161311 1045171311 1045181311 1045201311 1045221311 1045241311 1045261311 1045281311 1045301311 1045381311 1045401311 1045421311 1045441311 1045461311 1045481311 1045501311 1046961312 1046981312 1047001312 1047021312 1047041312 1047061312 1047081312 1047091312 1047101312 1047111312 1047121312 1047131312 1047141312 1047151312 1047161312 1047181312 1047201312 1047221312 1047241312 1047261312 1047281312 1047301312 1047321312 1047361312 1047381312 1047401312 1047421312 1047441309 1047441312 1047451309 1047461309 1047461312 1047471309 1047481309 1047481312 1047491309 1047501309 1047501312 1047511309 1047521309 1047521312 1047531309 1047541309 1047541312 1047551309 1047561309 1047561312 1047571309 1047581309 1047581312 1047591309 1047591312 1047601309 1047601312 1047611309 1047621309 1047621312 1047631309 1047641309 1047641312 1047661309 1047661312 1047671309 1047681309 1047681312 1047691309 1047691312 1047701309 1047701312 1047711309 1047711312 1047721309 1047721312 1047731309 1047731312 1047741309 1047751309 1047751312 1047761309 1047771309 1047771312 1047781309 1047791309 1047791312 1047801309 1047811309 1047811312 1047821309 1047831309 1047831312 1047841309 1047851309 1047851312 1047861309 1047871309 1047871312 1047881309 1047881312 1047891312 1047901309 1047901312 1047911309 1047911312 1047921309 1047921312 1047931309 1047931312 1047941309 1047941312 1047951309 1047951312 1047961309 1047961312 1047971309 1047971312 1047981309 1047981312 1047991309 1047991312 1048001309 1048001312 1048011309 1048011312 1048021309 1048021312 1048031309 1048031312 1048041309 1048041312 1048051309 1048051312 1048061309 1048061312 1048071309 1048071312 1048081309 1048081312 1048091309 1048091312 1048101309 1048101312 1048111309 1048111312 1048121309 1048121312 1048131309 1048131312 1048141309 1048141312 1048151309 1048151312 1048161309 1048161312 1048171309 1048171312 1048181309 1048181312 1048191309 1048191312 1048201309 1048201312 1048211309 1048211312 1048221309 1048221312 1048231309 1048231312 1048241309 1048241312 1048251309 1048251312 1048261309 1048261312 1048271309 1048271312 1048281309 1048281312 1048291309 1048291312 1048301309 1048301312 1048311309 1048311312 1048321309 1048321312 1048331309 1048331312 1048341309 1048341312 1048351309 1048351312 1048361309 1048361312 1048371309 1048371312 1048381309 1048381312 1048391309 1048391312 1048401309 1048401312 1048411309 1048411312 1048421309 1048421312 1048431309 1048431312 1048441309 1048441312 1048451309 1048451312 1048461309 1048461312 1048471309 1048471312 1048481309 1048481312 1048491309 1048491312 1048501309 1048501312 1048511309 1048511312 1048521309 1048521312 1048531312 1048541312 1048551312 1048561312 1048571312 1048581312 1048591312 1048601312 1048611312 1048621312 1048631312 1048641312 1048651312 1048661312 1048671312 1048681312 1054481310 1054491310 1054501310 1054511310 1054521310 1054531310 1054541310 1054551310 1054561310 1054571310 1054581310 1054591310 1054601310 1054611310 1054621310 1054631310 1054641310 1054651310 1054661310 1054671310 1054681310 1054691310 1054701310 1054711310 1054721310 1054731310 1054741310 1054751310 1054761310 1054771310 1054781310 1054791310 1054801310 1054811310 1054821310 1054831310 1054841310 1054851310 1054861310 1054871310 1054881310 1054891310 1054901309 1054901310 1054911310 1054921309 1054921310 1054931310 1054941309 1054941310 1054951310 1054961309 1054961310 1054971310 1054981309 1054981310 1054991309 1054991310 1055001309 1055001310 1055011309 1055011310 1055021309 1055021310 1055031309 1055031310 1055041309 1055041310 1055051309 1055051310 1055061309 1055061310 1055071309 1055071310 1055081309 1055081310 1055091309 1055091310 1055101309 1055101310 1055111309 1055111310 1055121309 1055121310 1055131309 1055131310 1055141309 1055141310 1055151309 1055151310 1055161309 1055161310 1055171309 1055171310 1055181309 1055181310 1055191309 1055191310 1055201309 1055201310 1055211309 1055221309 1055221310 1055231309 1055241309 1055241310 1055251309 1055261309 1055261310 1055271309 1055281309 1055281310 1055291309 1055291310 1055301309 1055301310 1055311309 1055311310 1055321309 1055321310 1055331309 1055331310 1055341309 1055341310 1055351309 1055351310 1055361309 1055361310 1055371309 1055371310 1055381309 1055381310 1055391309 1055391310 1055401309 1055401310 1055411309 1055411310 1055421309 1055421310 1055431309 1055431310 1055441309 1055441310 1055451309 1055451310 1055461309 1055461310 1055471309 1055471310 1055481309 1055491309 1055501309 1055511309 1055521309 1055531309 1055541309 1055551309 1055561309 1055571309 1055581309 1055591309 1055601309 1055611309 1055621309 1055631309 1055641309 1055651309 1055671309 1055691309 1055711309 1055731309 1055751309 1055771309 1055791309 1055811309 1055831309 1055851309 1055871309 1055891309 1055911309 1055931309 1055951309 1055971309 1055981309 1055991309 1056001309 1056011309 1056031309 1075141310 1075161310 1075181310 1075201310 1075211310 1075231310 1075251310 1075261310 1075271310 1075281310 1075291310 1075301310 1075311310 1075321310 1075331310 1075341310 1075361310 1075381310 1075401310 1075421310 1075441310 1075461310 1075481310 1075501310 1075521310 1075541310 1075561310 1075571310 1075581310 1075601310 1075621310 1075641310 1075661310 1075681310 1075701310 1075721310 1075741310 1075781310 1075821310 1075841310 1075861310 1075881310 1075901310 1075921310 1075991310 1076011310 1076021310 1076031310 1076041310 1076051310 1076071310 1076091310 1076111310 1076131310 1076151310 1101621312 1101641312 1101661312 1101681312 1101741312 1101761312 1101781312 1101881312 1102071312 1102091312 1102191312 1102211312 1102231312 1102251312 1102261312 1102271312 1102281312 1102291312 1102301312 1102311312 1102321312 1102331312 1102341312 1102361312 1102381312 1102401312 1108231311 1108241311 1108251311 1108261311 1108271311 1108281311 1108291311 1108301311 1108311311 1108321311 1108331311 1108341311 1108351311 1108361311 1108371311 1108381311 1108391311 1108401311 1108411311 1108421311 1108431311 1108441311 1108451311 1108461311 1108471311 1108481311 1108491311 1108501311 1108511311 1108521311 1108531311 1108541311 1108551311 1108561311 1108571311 1108581311 1108591311 1108601311 1108611311 1108621311 1108631311 1108641311 1108651311 1108661311 1108671311 1108681311 1108691311 1108701311 1108711311 1108721311 1108731311 1108741311 1108751311 1108761311 1108771311 1108781311 1108791311 1108801311 1108811311 1108821311 1108831311 1108841311 1108851311 1108861311 1108871311 1108881311 1108891311 1108901311 1108911311 1108921311 1108931311 1108941311 1108951311 1108961311 1108971311 1108981311 1108991311 1109001311 1109011311 1109021311 1109031311 1109041311 1109061311 1109081311 1109101311 1109121311 1109141311 1109161311 1109171311 1109181311 1109191311 1109201311 1109211311 1109221311 1109231311 1109241311 1109251311 1109261311 1109271311 1109281311 1109291311 1109301311 1112691311 1112701311 1112711311 1112721311 1112731311 1112741311 1112751311 1112761311 1112771311 1112781311 1112791311 1112801311 1112811311 1112821311 1112831311 1112841311 1112851311 1112861311 1112871311 1112881311 1112891311 1112901311 1112911311 1112921311 1112931311 1112941311 1112951311 1112961311 1112971311 1112981311 1112991311 1113001311 1113011311 1113021311 1113031311 1113041311 1113051311 1113061311 1113071311 1113081311 1113091311 1113101311 1113111311 1113121311 1113131311 1113141311 1113151311 1113161311 1113171311 1113181311 1113191311 1113201311 1113211311 1113221311 1113231311 1113241311 1113251311 1113261311 1113271311 1113281311 1113291311 1113301311 1113311311 1113321311 1113331311 1113341311 1113351311 1113361311 1113371311 1113381311 1113391311 1113401311 1113411311 1113421311 1113431311 1113441311 1113451311 1113461311 1113471311 1113481311 1113491311 1113501311 1113511311 1113521311 1113531311 1113541311 1113551311 1113561311 1113571311 1113581311 1113591311 1113601311 1113611311 1113621311 1113631311 1113641311 1113651311 1113661311 1113671311 1113681311 1113691311 1113701311 1113711311 1113721311 1113731311 1113741311 1113751311 1113761311 1113771311 1113781311 1113791311 1114521311 1114531311 1114541311 1114551311 1114561311 1114571311 1114581311 1114591311 1114601311 1114611311 1114621311 1114631311 1114641311 1114651311 1114661311 1114671311 1114681311 1114691311 1114701311 1114711311 1114721311 1114731311 1114741311 1114751311 1114761311 1114771311 1114781311 1114791311 1114801311 1114811311 1114821311 1114831311 1114841311 1114851311 1114861311 1114871311 1114881311 1114891311 1114901311 1114911311 1114921311 1114931311 1114941311 1114951311 1114961311 1114971311 1114981311 1114991311 1115001311 1115011311 1115021311 1115031311 1115041311 1115051311 1115061311 1115071311 1115081311 1115091311 1115101311 1115111311 1115121311 1115131311 1115141311 1115151311 1115161311 1115171311 1115181311 1115191311 1115201311 1115211311 1115221311 1115231311 1115241311 1115251311 1115261311 1115271311 1115281311 1115291311 1115301311 1115311311 1115321311 1115331311 1115341311 1115351311 1115361311 1115371311 1115381311 1115391311 1115401311 1115411311 1115421311 1115431311 1115441311 1115451311 1115461311 1115471311 1115481311 1115491311 1115501311 1115511311 1115521311 1115531311 1115541311 1116101309 1116111309 1116121309 1116131309 1116141309 1116151309 1116161309 1116171309 1116181309 1116191309 1116201309 1116211309 1116221309 1116231309 1116241309 1116251309 1116261309 1116271309 1116281309 1116291309 1116301309 1116311309 1116321309 1116331309 1116341309 1116351309 1116361309 1116381309 1116391309 1116401309 1116411309 1116421309 1116431309 1116441309 1116451309 1116461309 1116471309 1116481309 1116491309 1116501309 1116511309 1116521309 1116541309 1116551309 1116561309 1116571309 1116581309 1116611309 1116621309 1116641309 1116661309 1116671309 1116681309 1116691309 1116701309 1116711309 1116721309 1116741309 1116761309 1116771309 1116781309 1116791309 1116801309 1116821309 1116841309 1116861309 1116881309 1116901309 1116941309 1116981309 1117001309 1117021309 1117041309 1117061309 1117081309 1117101309 1117121309 1117141309 1117161309 1117181309 1117201309 1127091309 1127101309 1127111309 1127121309 1127131309 1127141309 1127151309 1127161309 1127171309 1127181309 1127191309 1127201309 1127211309 1127221309 1127231309 1127241309 1127251309 1127261309 1127271309 1127281309 1127291309 1127301309 1127311309 1127321309 1127331309 1127341309 1127351309 1127361309 1127371309 1127381309 1127391309 1127401309 1127411309 1127421309 1127431309 1127441309 1127451309 1127461309 1127471309 1127481309 1127491309 1127501309 1127511309 1127521309 1127531309 1127541309 1127551309 1127561309 1127571309 1127581309 1127591309 1127601309 1127611309 1127621309 1127641309 1127661309 1127681309 1127701309 1127721309 1127741309 1127761309 1127781309 1127801309 1127821309 1127841309 1127861309 1127871309 1127881309 1127891309 1127901309 1127911309 1127921309 1127931309 1127941309 1127961309 1127981309 1128001309 1128021309 1128031309 1128041309 1128051309 1128061309 1128071309 1128081309 1128091309 1128101309 1128111309 1128121309 1128131309 1128141309 1128151309 1134731309 1134731310 1134741309 1134741310 1134751309 1134751310 1134761309 1134761310 1134771309 1134771310 1134781309 1134781310 1134791309 1134791310 1134801309 1134801310 1134811309 1134811310 1134821309 1134821310 1134831309 1134831310 1134841309 1134841310 1134851309 1134851310 1134861309 1134861310 1134871309 1134871310 1134881309 1134881310 1134891309 1134891310 1134901309 1134901310 1134911309 1134911310 1134921309 1134921310 1134931309 1134931310 1134941309 1134941310 1134951309 1134951310 1134961309 1134961310 1134971309 1134971310 1134981309 1134981310 1134991309 1134991310 1135001309 1135001310 1135011309 1135011310 1135021309 1135021310 1135031309 1135031310 1135041309 1135041310 1135051309 1135051310 1135061309 1135061310 1135071309 1135071310 1135081309 1135081310 1135091309 1135091310 1135101309 1135101310 1135111309 1135111310 1135121309 1135121310 1135131309 1135131310 1135141309 1135141310 1135151309 1135151310 1135161309 1135161310 1135171309 1135171310 1135181309 1135191309 1135191310 1135201309 1135201310 1135211309 1135211310 1135221309 1135231309 1135231310 1135241309 1135241310 1135251309 1135251310 1135261309 1135271309 1135271310 1135281309 1135281310 1135291309 1135291310 1135301309 1135311309 1135311310 1135321309 1135321310 1135331309 1135331310 1135341309 1135351309 1135351310 1135361309 1135361310 1135371309 1135371310 1135381309 1135391309 1135391310 1135401309 1135401310 1135411309 1135411310 1135421309 1135431309 1135431310 1135441309 1135441310 1135451309 1135451310 1135461309 1135471309 1135471310 1135481309 1135481310 1135491309 1135501309 1135501310 1135511309 1135511310 1135521309 1135531309 1135541309 1135551309 1135551310 1135561309 1135561310 1135571309 1135571310 1135581309 1135591309 1135601309 1135601310 1135611309 1135611310 1135621309 1135621310 1135631309 1135641309 1135641310 1135651309 1135651310 1135661309 1135671309 1135681309 1135681310 1135691309 1135691310 1135701309 1135701310 1135711309 1135721309 1135721310 1135731309 1135731310 1135741309 1135741310 1135751309 1135761309 1135771309 1135781309 1144301310 1144311310 1144321310 1144331310 1144341310 1144351310 1144361310 1144371310 1144381310 1144391310 1144401310 1144411310 1144421310 1144431310 1144441310 1144451310 1144461310 1144471310 1144481310 1144491310 1144501310 1144511310 1144521310 1144531310 1144541310 1144551310 1144561310 1144571310 1144581310 1144591310 1144601310 1144611310 1144621310 1144631310 1144641310 1144651310 1144661310 1144671310 1144681310 1144691310 1144701310 1144711310 1144721310 1144731310 1144741310 1144751310 1144761310 1144771310 1144781310 1144791310 1144801310 1144811310 1144821310 1144831310 1144841310 1144851310 1144861310 1144871310 1144881310 1144891310 1144901310 1144911310 1144921310 1144931310 1144941310 1144951310 1144961310 1144971310 1144981310 1144991310 1145001310 1145011310 1145021310 1145031310 1145041310 1145051310 1145061310 1145071310 1145081310 1145091310 1145101310 1145111310 1145121310 1145131310 1145141310 1145151310 1145161310 1145171310 1145181310 1145191310 1145201310 1145211310 1145221310 1145231310 1145241310 1145251310 1145261310 1161691310 1161701310 1161711310 1161721310 1161741310 1161751310 1161761310 1161781310 1161791310 1161801310 1161811310 1161821310 1161831310 1161841310 1161851310 1161861310 1161871310 1161881310 1161901310 1161911310 1161921310 1161931310 1161941310 1161951310 1161961310 1161971310 1161981310 1161991310 1162001310 1162011310 1162021310 1162031310 1162041310 1162051310 1162061310 1162071310 1162081310 1162091310 1162101310 1162111310 1162121310 1162131310 1162141310 1162151310 1162161310 1162171310 1162181310 1162191310 1162201310 1162211310 1162221310 1162231310 1162241310 1162251310 1162261310 1162271310 1162281310 1162291310 1162301310 1162311310 1162321310 1162331310 1162341310 1162351310 1162361310 1162371310 1162381310 1162391310 1162401310 1162411310 1162421310 1162431310 1162441310 1162451310 1162461310 1162471310 1162481310 1162491310 1162501310 1162511310 1162521310 1162531310 1162541310 1162551310 1162561310 1162571310 1162581310 1185691310 1185701310 1185711310 1185721310 1185731310 1185741310 1185751310 1185761310 1185771310 1185781310 1185791310 1185801310 1185811310 1185821310 1185831310 1185841310 1185851310 1185861310 1185871310 1185881310 1185891310 1185901310 1185911310 1185921310 1185931310 1185941310 1185951310 1185961310 1185971310 1185981310 1185991310 1186001310 1186011310 1186031310 1186051310 1186071310 1186091310 1186111310 1186121310 1186131310 1186141310 1186151310 1186161310 1186171310 1186181310 1186191310 1186201310 1186211310 1186221310 1186231310 1186241310 1186251310 1186261310 1186271310 1186281310 1186291310 1186301310 1186321310 1186341310 1186361310 1186381310 1186401310 1186421310 1186441310 1186461310 1186481310 1186501310 1186521310 1186561310 1186581310 1186601310 1186621310 1186641310 1186661310 1186681310 1186701310 1186721310 1186741310 1186761310 1186781310 1186801310 1191251310 1191261310 1191271310 1191281310 1191291310 1191301310 1191311310 1191321310 1191331310 1191341310 1191351310 1191361310 1191371310 1191381310 1191391310 1191401310 1191431310 1191441310 1191451310 1191471310 1191481310 1191491310 1191511310 1191521310 1191531310 1191551310 1191561310 1191571310 1191581310 1191591310 1191601310 1191611310 1191621310 1191631310 1191641310 1191651310 1191661310 1191671310 1191681310 1191691310 1191701310 1191711310 1191721310 1191731310 1191741310 1191751310 1191761310 1191771310 1191781310 1191791310 1191801310 1191811310 1191821310 1191831310 1191841310 1191851310 1191861310 1191871310 1191881310 1191891310 1191901310 1191911310 1191921310 1191931310 1191941310 1191951310 1191961310 1191971310 1191981310 1191991310 1192001310 1192011310 1192021310 1192031310 1192041310 1192051310 1192061310 1192071310 1192081310 1192091310 1192101310 1192111310 1192121310 1192131310 1192141310 1192151310 1192161310 1192171310 1192181310 1192191310 1192201310 1192211310 1192221310 1192231310 1192241310 1192251310 1192261310 1192271310 1192281310 1192291310 1192301310 1192311310 1192321310 1192331310 1192341310 1192351310 1192361310 1198391311 1198401311 1198411311 1198421311 1198441311 1198461311 1198471311 1198481311 1198501311 1198521311 1198541311 1198561311 1198581311 1198601311 1198621311 1198641311 1198661311 1198681311 1198701311 1198721311 1198741311 1198761311 1198781311 1198991311 1199201311 1199221311 1199241311 1199261311 1199281311 1199301311 1199321311 1199341311 1199361311 1199381311 1199401311 1199411311 1199421311 1201441309 1201461309 1201481309 1201501309 1201511309 1201521309 1201541309 1201561309 1201581309 1201601309 1201621309 1201641309 1201651309 1201661309 1201671309 1201681309 1201691309 1201701309 1201711309 1201721309 1201731309 1201741309 1201751309 1201761309 1201771309 1201781309 1201791309 1201801309 1201811309 1201881309 1201901309 1201921309 1201941309 1201961309 1201981309 1202001309 1202041309 1202061309 1202071309 1202091309 1202101309 1202111309 1202121309 1202131309 1202141309 1202151309 1202161309 1202171309 1202181309 1202191309 1202201309 1202211309 1202221309 1202231309 1202241309 1202251309 1202261309 1202271309 1202281309 1202291309 1202301309 1202311309 1202321309 1202331309 1202341309 1202351309 1202361309 1202371309 1202381309 1202391309 1202401309 1202411309 1202421309 1202431309 1202441309 1202451309 1202461309 1202471309 1202491309 1202511309 1217561309 1217571309 1217581309 1217591309 1217601309 1217611309 1217621309 1217631309 1217641309 1217651309 1217671309 1217691309 1217701309 1217711309 1217721309 1217731309 1217741309 1217851309 1217871309 1217891309 1217931309 1217941309 1217951309 1217961309 1217971309 1217981309 1217991309 1218001309 1218011309 1218031309 1218041309 1218051309 1218071309 1218091309 1218101309 1218111309 1218131309 1218151309 1218161309 1218171309 1218181309 1218191309 1218201309 1218211309 1218231309 1218251309 1218271309 1218291309 1218311309 1218321309 1218341309 1218361309 1218381309 1218391309 1218401309 1218411309 1218421309 1218431309 1218441309 1218451309 1218471309 1218481309 1218491309 1218501309 1218511309 1218521309 1218531309 1218541309 1218551309 1218561309 1218571309 1218581309 1218591309 1218601309 1218611309 1218621309 1218631309 1218641309 1218651309 1218661309 1218671309 1218681309 1218701309 1218711309 1219771311 1219791311 1219811311 1219831311 1219841311 1219851311 1219861311 1219871311 1219881311 1219891311 1219901311 1219911311 1219921311 1219931311 1219941311 1219951311 1219961311 1219971311 1219981311 1219991311 1220001311 1220011311 1220021311 1220031311 1220041311 1220051311 1220071311 1220091311 1220111311 1220131311 1220151311 1220171311 1220191311 1220211311 1220231311 1220251311 1220271311 1220291311 1220311311 1220331311 1220351311 1220371311 1220391311 1220411311 1220421311 1220441311 1220461311 1220491311 1220511311 1220591311 1220611311 1220631311 1220651311 1220671311 1220691311 1220711311 1220731311 1223751311 1223761311 1223771311 1223781311 1223791311 1223801311 1223811311 1223821311 1223831311 1223841311 1223851311 1223871311 1223881311 1223891311 1223901311 1223911311 1223921311 1223941311 1223961311 1223981311 1224001311 1224021311 1224041311 1224061311 1224081311 1224101311 1224121311 1224141311 1224251311 1224371311 1224491311 1224511311 1224531311 1224551311 1224561311 1224571311 1224581311 1224591311 1224601311 1224611311 1224621311 1224631311 1224641311 1224651311 1224661311 1224671311 1224681311 1224691311 1224711311 1224731311 1224751311 1224771311 1224781311 1224791311 1224801311 1224811311 1224821311 1224831311 1224841311 1224851311 1228201311 1228211311 1228221311 1228231311 1228241311 1228251311 1228261311 1228271311 1228281311 1228291311 1228301311 1228321311 1228341311 1228361311 1228381311 1228401311 1228421311 1228441311 1228461311 1228481311 1228501311 1228521311 1228541311 1228561311 1228581311 1228601311 1228621311 1228641311 1228661311 1228681311 1228701311 1228721311 1228731311 1228741311 1228751311 1228761311 1228771311 1228781311 1228791311 1228801311 1228811311 1228821311 1228831311 1228841311 1228851311 1228861311 1228871311 1228881311 1228891311 1228901311 1228911311 1228921311 1228931311 1228941311 1228951311 1228961311 1228971311 1228991311 1229011311 1229021311 1229031311 1229051311 1229071311 1229091311 1229111311 1229131311 1229151311 1229251311 1229271311 1229291311 1229311311 1233921309 1233931309 1233941309 1233951309 1233961309 1233971309 1233981309 1234001309 1234021309 1234041309 1234051309 1234061309 1234071309 1234081309 1234091309 1234101309 1234111309 1234121309 1234131309 1234141309 1234151309 1234161309 1234171309 1234181309 1234191309 1234201309 1234211309 1234221309 1234231309 1234241309 1234251309 1234261309 1234271309 1234281309 1234291309 1234301309 1234311309 1234321309 1234331309 1234341309 1234351309 1234361309 1234371309 1234381309 1234391309 1234401309 1234411309 1234421309 1234431309 1234441309 1234451309 1234461309 1234471309 1234481309 1234491309 1234501309 1234511309 1234521309 1234531309 1234541309 1234561309 1234571309 1234581309 1234601309 1234611309 1234621309 1234631309 1234641309 1234651309 1234661309 1234671309 1234681309 1234691309 1234701309 1234711309 1234721309 1234731309 1234741309 1234751309 1234761309 1234771309 1234781309 1234791309 1234801309 1234811309 1234821309 1234831309 1234841309 1234851309 1234861309 1234871309 1234881309 1234891309 1234901309 1234911309 1234921309 1237061309 1237071309 1237081309 1237091309 1237101309 1237111309 1237121309 1237131309 1237141309 1237151309 1237161309 1237171309 1237181309 1237191309 1237201309 1237211309 1237221309 1237231309 1237241309 1237251309 1237261309 1237271309 1237281309 1237291309 1237301309 1237311309 1237321309 1237331309 1237341309 1237351309 1237361309 1237371309 1237381309 1237391309 1237411309 1237421309 1237431309 1237441309 1237451309 1237461309 1237471309 1237481309 1237501309 1237511309 1237521309 1237541309 1237551309 1237561309 1237581309 1237591309 1237601309 1237621309 1237631309 1237641309 1237661309 1237671309 1237681309 1237701309 1237711309 1237721309 1237741309 1237751309 1237761309 1237781309 1237791309 1237801309 1237821309 1237831309 1237841309 1237861309 1237871309 1237881309 1237891309 1237901309 1237911309 1237921309 1237931309 1237941309 1237951309 1237961309 1237971309 1237981309 1237991309 1238001309 1238011309 1238031309 1238041309 1238051309 1238061309 1238071309 1238081309 1238091309 1238101309 1238111309 1238121309 1238131309 1238141309 1238151309 1238161309 1238171309 1238181309 1238191309 1238211309 1238221309 1238231309 1247141309 1247151309 1247161309 1247181309 1247201309 1247221309 1247241309 1247261309 1247281309 1247301309 1247321309 1247341309 1247361309 1247381309 1247401309 1247421309 1247441309 1247461309 1247481309 1247491309 1247501309 1247511309 1247521309 1247531309 1247541309 1247551309 1247571309 1247581309 1247591309 1247601309 1247611309 1247621309 1247631309 1247641309 1247651309 1247661309 1247671309 1247681309 1247691309 1247701309 1247711309 1247721309 1247741309 1247761309 1247781309 1247801309 1247821309 1247831309 1247841309 1247851309 1247861309 1247871309 1247881309 1247891309 1247901309 1247921309 1247941309 1247951309 1247971309 1247981309 1247991309 1248001309 1248011309 1248021309 1248031309 1248041309 1248051309 1248061309 1248071309 1248081309 1248091309 1248101309 1248111309 1248131309 1248151309 1248171309 1248191309 1248201309 1248221309 1248241309 1251561310 1251581310 1251601310 1251621310 1251661310 1251721310 1251741310 1251751310 1251761310 1251771310 1251781310 1251801310 1251811310 1251821310 1251831310 1251841310 1251851310 1251861310 1251871310 1251881310 1251891310 1251901310 1251921310 1251941310 1251961310 1252001310 1252021310 1252041310 1252061310 1252081310 1252101310 1252121310 1252141310 1252161310 1252181310 1252201310 1252211310 1252221310 1252231310 1252241310 1252251310 1252261310 1252271310 1252281310 1252291310 1252301310 1252311310 1252321310 1252331310 1252341310 1252351310 1252361310 1252371310 1252381310 1252391310 1252401310 1252411310 1252421310 1252431310 1252451310 1252471310 1252491310 1252511310 1252531310 1252551310 1252571310 1252591310 1252611310 1252631310 1252651310 1264271308 1264281308 1264291308 1264301308 1264311308 1264321308 1264331308 1264341308 1264351308 1264361308 1264371308 1264381308 1264391308 1264401308 1264411308 1264421308 1264431308 1264441308 1264451308 1264461308 1264471308 1264481308 1264491308 1264501308 1264511308 1264521308 1264531308 1264551308 1264571308 1264591308 1264611308 1264621308 1264631308 1264641308 1264651308 1264661308 1264671308 1264681308 1264691308 1264701308 1264711308 1264721308 1264731308 1264741308 1264751308 1264761308 1264771308 1264781308 1264791308 1264801308 1264811308 1264821308 1264831308 1264841308 1264851308 1264861308 1264871308 1264881308 1264891308 1264901308 1264911308 1264921308 1264931308 1264941308 1264951308 1264961308 1264971308 1264981308 1264991308 1265001308 1265011308 1265021308 1265031308 1265041308 1265051308 1265061308 1265081308 1265091308 1265111308 1265121308 1265131308 1265151308 1265171308 1265191308 1265211308 1265221308 1265231308 1265241308 1265251308 1265261308 1265281308 1265301308 1265311308 1265331308 1265341308 1265351308 1265361308 1273741308 1273751308 1273761308 1273771308 1273781308 1273791308 1273801308 1273811308 1273821308 1273831308 1273841308 1273851308 1273861308 1273871308 1273881308 1273891308 1273901308 1273911308 1273921308 1273931308 1273941308 1273951308 1273961308 1273971308 1273981308 1273991308 1274001308 1274011308 1274021308 1274031308 1274041308 1274051308 1274061308 1274071308 1274081308 1274091308 1274101308 1274111308 1274121308 1274131308 1274141308 1274151308 1274161308 1274171308 1274181308 1274191308 1274201308 1274211308 1274221308 1274231308 1274241308 1274251308 1274261308 1274271308 1274281308 1274291308 1274301308 1274311308 1274321308 1274331308 1274341308 1274351308 1274361308 1274371308 1274381308 1274391308 1274401308 1274411308 1274421308 1274431308 1274441308 1274451308 1274461308 1274471308 1274481308 1274491308 1274501308 1274511308 1274521308 1274531308 1274541308 1274551308 1274561308 1274571308 1274581308 1274591308 1274601308 1274611308 1274621308 1274631308 1274651308 1274671308 1274681308 1274691308 1274701308 1274711308 1274721308 1274731308 1274741308 1274751308 1274761308 1274771308 1274781308 1274791308 1274801308 1274811308 1274821308 1274831308 1274841308 1274851308 1274861308 1274871308 1274881308 1274891308 1274901308 1274911308 1274921308 1274931308 1274941308 1274951308 1274961308 1274971308 1274981308 1274991308 1275001308 1275011308 1275021308 1275031308 1275041308 1275051308 1275061308 1275071308 1275081308 1275091308 1275101308 1275111308 1275121308 1275131308 1275141308 1275151308 1275161308 1275171308 1275181308 1275191308 1275201308 1275211308 1275221308 1275231308 1275241308 1275251308 1275261308 1275271308 1275281308 1275291308 1275301308 1275311308 1275321308 1275331308 1275341308 1275351308 1275361308 1275371308 1275381308 1275391308 1275401308 1275411308 1275421308 1275431308 1275441308 1275451308 1275461308 1275471308 1275481308 1275491308 1275501308 1275511308 1275521308 1275531308 1275541308 1275551308 1275561308 1275571308 1275581308 1275591308 1275601308 1275611308 1275621308 1275631308 1275641308 1275651308 1275661308 1275671308 1275681308 1275691308 1275701308 1275711308 1275721308 1275731308 1275741308 1275751308 1275761308 1275771308 1275781308 1275791308 1275801308 1275811308 1277911310 1277921310 1277931310 1277941310 1277951310 1277961310 1277971310 1277981310 1277991310 1278001310 1278011310 1278021310 1278031310 1278041310 1278051310 1278061310 1278071310 1278081310 1278091310 1278101310 1278111310 1278121310 1278131310 1278141310 1278151310 1278161310 1278171310 1278181310 1278191310 1278201310 1278211310 1278221310 1278231310 1278241310 1278251310 1278261310 1278271310 1278281310 1278291310 1278301310 1278311310 1278321310 1278331310 1278341310 1278351310 1278361310 1278371310 1278381310 1278391310 1278401310 1278411310 1278421310 1278431310 1278441310 1278451310 1278461310 1278471310 1278481310 1278491310 1278501310 1278511310 1278521310 1278531310 1278541310 1278551310 1278561310 1278571310 1278581310 1278591310 1278601310 1278611310 1278621310 1278631310 1278641310 1278651310 1278661310 1278671310 1278681310 1278691310 1278701310 1278711310 1278721310 1278731310 1278741310 1278751310 1278761310 1278771310 1278781310 1278791310 1278801310 1278811310 1278821310 1278831310 1278841310 1278851310 1278861310 1278871310 1278881310 1278891310 1278901310 1278911310 1278921310 1278931310 1278941310 1278951310 1278961310 1278971310 1278981310 1278991310 1279001310 1279011310 1279021310 1279031310 1279041310 1279051310 1304361308 1304371308 1304381308 1304391308 1304401308 1304411308 1304421308 1304431308 1304441308 1304451308 1304461308 1304471308 1304481308 1304491308 1304501308 1304511308 1304521308 1304531308 1304541308 1304551308 1304561308 1304571308 1304581308 1304591308 1304601308 1304611308 1304621308 1304631308 1304641308 1304651308 1304661308 1304671308 1304681308 1304691308 1304711308 1304721308 1304731308 1304741308 1304801308 1304821308 1304841308 1304861308 1304881308 1304901308 1304921308 1304941308 1304961308 1304981308 1305001308 1305011308 1305021308 1305041308 1305061308 1305081308 1305101308 1305121308 1305141308 1305161308 1305181308 1305201308 1305221308 1305241308 1305251308 1305261308 1305271308 1305281308 1305291308 1305301308 1305311308 1305321308 1305331308 1305341308 1305351308 1305361308 1305371308 1305381308 1305401308 1305411308 1305421308 1316361308 1316371308 1316381308 1316391308 1316401308 1316411308 1316421308 1316431308 1316441308 1316451308 1316471308 1316491308 1316511308 1316521308 1316531308 1316541308 1316551308 1316561308 1316571308 1316581308 1316591308 1316601308 1316611308 1316621308 1316631308 1316641308 1316651308 1316661308 1316671308 1316681308 1316691308 1316701308 1316711308 1316721308 1316731308 1316741308 1316751308 1316761308 1316771308 1316781308 1316791308 1316811308 1316821308 1316831308 1316841308 1316851308 1316861308 1316871308 1316881308 1316891308 1316901308 1316911308 1316921308 1316931308 1316941308 1316951308 1316961308 1316971308 1316981308 1316991308 1317001308 1317011308 1317021308 1317031308 1317041308 1317051308 1317061308 1317071308 1317081308 1317091308 1317101308 1317111308 1317121308 1317131308 1317141308 1317151308 1317161308 1317171308 1317181308 1317191308 1317201308 1317211308 1317221308 1317231308 1317241308 1317251308 1317261308 1319351308 1319381308 1319401308 1319421308 1319431308 1319441308 1319451308 1319471308 1319481308 1319491308 1319501308 1319611308 1319631308 1319651308 1319671308 1319691308 1319711308 1319741308 1319811308 1319881308 1319981308 1320001308 1320021308 1320041308 1320061308 1320081308 1320101308 1320121308 1320141308 1320161308 1320241308 1320261308 1320281308 1320301308 1320351308 |
Recalling Firm/ Manufacturer |
AMO Puerto Rico Manufacturing, Inc. Rd # 402, Km 4.2 Anasco PR 00610
|
Manufacturer Reason for Recall |
Units may be labeled with the incorrect diopter power.
|
FDA Determined Cause 2 |
Error in labeling |
Action |
AMO sent an Urgent Medical Device Recall dated May 13, 2014 to the affected customers. The letter identified the affected product, problem and actions to be taken. This recall only affects the specific Silicone Tecnis IOLs Model Z9002. Implanted IOLs with appropriate post-op refractive outcomes are not affected by this recall. Please underatake the following actions:
1. Stop using and remove from your inventory the affected product.
2.Complete and return the included Facsimile Form as AMO.
If you have inventory to return, please complete the Facsimile Form, and returned to AMO. For questions contact AMO Customer Service Representative 877-266-4543. |
Quantity in Commerce |
77,055 IOLs in total |
Distribution |
Worldwide Distribution - Nationwide (USA and Puerto Rico), and foreign: Latin America, Asia Pacific, Canada, Europe, Africa and Middle East, and Japan. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = HQL and Original Applicant = Johnson & Johnson Surgical Vision, Inc.
|