Class 2 Device Recall Double Offset Rasp Handles Version 2 PIN REWORK |
![see related information](../images/related-info.gif) |
Date Initiated by Firm |
July 23, 2014 |
Date Posted |
August 14, 2014 |
Recall Status1 |
Terminated 3 on September 15, 2015 |
Recall Number |
Z-2229-2014 |
Recall Event ID |
68656 |
Product Classification |
Orthopedic manual surgical instrument - Product Code LXH
|
Product |
VER2 HANDLE, PIN REWORK, RT Nonsterile
|
Code Information |
Item Number 82-0167-613-32 Lot Number 97010151 |
Recalling Firm/ Manufacturer |
Zimmer, Inc. 1800 W Center St Warsaw IN 46580-2304
|
For Additional Information Contact |
Consumer Relations Call Center 800-447-5633
|
Manufacturer Reason for Recall |
The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring.
|
FDA Determined Cause 2 |
Device Design |
Action |
On July 23, 2014, Zimmer issued recall notification to their consignees. Distributors were notified via electronic mail and courier. Recall notification included a description of the issue, affected product, possible risks to patient, and instructions for responding to the recall.
Questions or concerns regarding the recall are directed to Zimmer customer call center @ 1-800-348-2759 M-F, 8am-8pm ET. |
Quantity in Commerce |
412 units |
Distribution |
Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, IN, KS, KY, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, and VA, and the countries of Australia, France, Germany, Italy, Switzerland, and Canada. |