Date Initiated by Firm | August 14, 2014 |
Date Posted | October 24, 2014 |
Recall Status1 |
Terminated 3 on August 10, 2015 |
Recall Number | Z-0127-2015 |
Recall Event ID |
69334 |
510(K)Number | K102045 |
Product Classification |
Assay, glycosylated hemoglobin - Product Code LCP
|
Product | Dimension Vista HbA1c |
Code Information |
catalog #K3105A, material #10470481, All in-date lots |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
For Additional Information Contact | Customer Support 800-441-9250 |
Manufacturer Reason for Recall | In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t |
FDA Determined Cause 2 | Device Design |
Action | An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers. |
Quantity in Commerce | 20043 |
Distribution | Worldwide Distribution: US (nationwide) and country of: Canada. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LCP
|