| Class 2 Device Recall Siemens Luminos Agile Max system | |
Date Initiated by Firm | September 02, 2014 |
Date Posted | October 02, 2014 |
Recall Status1 |
Terminated 3 on February 24, 2015 |
Recall Number | Z-0005-2015 |
Recall Event ID |
69258 |
510(K)Number | K111292 |
Product Classification |
System, x-ray, fluoroscopic, image-intensified - Product Code KPR
|
Product | Siemens Luminos Agile Max system
The Luminos Agile is intended to be used as a universal imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable to emergency treatment on an outpatient basis, as well as for bedside examinations. |
Code Information |
Siemens Luminos Agile Max system material #s: 10762472, with serial numbers: 61013 61015 61011 61014 61017 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA, Inc 51 Valley Stream Pkwy Malvern PA 19355
|
For Additional Information Contact | Customer Support 610-219-6300 |
Manufacturer Reason for Recall | Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images |
FDA Determined Cause 2 | Software design |
Action | Siemens sent a Safety Advisory Notice, dated September 2, 2014, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disable the automatic RIS worklist update and perform this procedure manually. Customers should disable the Measurement Field Display and Collimator Display by setting the display time for both setting to 0 sec (if editable). Customers should forward the safety notice to the new owner and inform Siemens of the new owner where possible. Customers with questions should call 610-219-2119 or 610-219-4834.
For questions regarding this recall call 610-219-6300. |
Quantity in Commerce | 5 |
Distribution | Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|