• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Philips Healthcare MX 16slice CT Scanner System and MX 16slice CT Scanner System Phase 11

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips Healthcare MX 16slice CT Scanner System and MX 16slice CT Scanner System Phase 11see related information
Date Initiated by FirmOctober 09, 2014
Date PostedDecember 16, 2014
Recall Status1 Terminated 3 on July 02, 2015
Recall NumberZ-0190-2015
Recall Event ID 69473
510(K)NumberK083498 K091195 
Product Classification System, x-ray, tomography, computed - Product Code JAK
ProductMX 16-slice CT Scanner System and MX 16-slice CT Scanner System Phase 11 can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles. The system is suitable for all patients. Access Dual CT system can be used as a Head and Whole Body Computed Tomography X-ray System featuring a continuously rotating X-ray tube and detector array with multi-slice capability. The acquired X-ray transmission data is reconstructed by computer into cross-sectional images of the body from the same axial plane taken at different angles.
Code Information Model Number: 728131, Serial Number(s):  EP16E090001,EP16E090002,EP16E090003,EP16E090004,EP16E090006,EP16E090007,EP16E090008, EP16E090009,EP16E090010,EP16E090011,EP16E090012,EP16E090013,EP16E090015,EP16E100001, EP16E100004,EP16E100005,EP16E100006,EP16E100007,EP16E100008,EP16E100010,EP16E100011, EP16E100012,EP16E100013,EP16E100015,EP16E100016,EP16E100017,EP16E100018, EP16E100019,EP16E100020,EP16E100021,EP16E100022,EP16E100023,EP16E100024,EP16E100025, EP16E100026,EP16E100027,EP16E100028,EP16E100029,EP16E100030,EP16E100031,EP16E100032, EP16E100033,EP16E100034,EP16E100035,EP16E100036,EP16E100037,EP16E100038,EP16E100039, EP16E100041,EP16E100043,EP16E100044,EP16E100045,EP16E100046,EP16E100047,EP16E100048, EP16E100049,EP16E100051,EP16E100052,EP16E100055,EP16E100056,EP16E100057,EP16E100058, EP16E100059,EP16E100060,EP16E100061,EP16E100062,EP16E100063,EP16E100067,EP16E100069, EP16E100070,EP16E100071,EP16E100072,EP16E100073,EP16E100074,EP16E100075,EP16E100076, EP16E100077,EP16E100078,EP16E100079,EP16E100080,EP16E100081,EP16E100082,EP16E100083, EP16E100084,EP16E100085,EP16E100086,EP16E100087,EP16E100088,EP16E100090,EP16E100091, EP16E100092,EP16E100093,EP16E100094,EP16E100095,EP16E100096,EP16E100099,EP16E100100, EP16E100101,EP16E100102,EP16E100103,EP16E100104,EP16E100105,EP16E100106,EP16E100107, EP16E100108,EP16E100109,EP16E100110,EP16E100111,EP16E100112,EP16E100113,EP16E110003, EP16E110005,EP16E110006,EP16E110007,EP16E110009,EP16E110011,EP16E110012,EP16E110013, EP16E110014,EP16E110015,EP16E110017,EP16E110018,EP16E110019,EP16E110020,EP16E110021, EP16E110023,EP16E110024,EP16E110025,EP16E110026,EP16E110027,EP16E110028,EP16E110029, EP16E110030,EP16E110031,EP16E110032,EP16E110033,EP16E110034,EP16E110035,EP16E110037, EP16E110038,EP16E110039,EP16E110042,EP16E110043,EP16E110044,EP16E110045,EP16E110046, EP16E110047,EP16E110048,EP16E110049,EP16E110050,EP16E110051,EP16E110052,EP16E110054, EP16E110055,EP16E110056,EP16E110057,EP16E110058,EP16E110060,EP16E110062,EP16E110063, EP16E110064,EP16E110066,EP16E110067,EP16E110069,EP16E110070,EP16E110074,EP16E110075, EP16E110077,EP16E110078,EP16E110083,EP16E110085,EP16E110089,EP16E110090,EP16E110091, EP16E110092,EP16E110093,EP16E110095,EP16E110096,EP16E110097,EP16E110099,EP16E110100, EP16E110101,EP16E110102,EP16E110104,EP16E110105,EP16E110106,EP16E110108,EP16E110109, EP16E110110,EP16E110111,EP16E110115,EP16E110117,EP16E110118,EP16E110119,EP16E120003, EP16E120004,EP16E120005,EP16E120006,EP16E120007,EP16E120008,EP16E120009,EP16E120010, EP16E120011,EP16E120014,EP16E120015,EP16E120016,EP16E120017,EP16E120019,EP16E120021, EP16E120023,EP16E120024,EP16E120025,EP16E120026,EP16E120027,EP16E120029,EP16E120030, EP16E120031,EP16E120032,EP16E120033,EP16E120034,EP16E120035,EP16E120036,EP16E120037, EP16E120039,EP16E120041,EP16E120042,EP16E120043,EP16E120044,EP16E120045,EP16E120046, EP16E120047,EP16E120048,EP16E120050,EP16E120051,EP16E120052,EP16E120054,EP16E120055, EP16E120056,EP16E120058,EP16E120059,EP16E120060,EP16E120061,EP16E120063,EP16E120065, EP16E120067,EP16E120070,EP16E120071,EP16E120072,EP16E120074,EP16E120075,EP16E120076, EP16E120078,EP16E120081,EP16E120083,EP16E120088,EP16E120090,EP16E120091,EP16E120092, EP16E120093,EP16E120095,EP16E120096,EP16E120097,EP16E120098,EP16E120099,EP16E120100, EP16E120101,EP16E120102,EP16E120104,EP16E120105,EP16E120109,EP16E120110,EP16E120111, EP16E120113,EP16E120117,EP16E120118,EP16E120120,EP16E120121,EP16E120122,EP16E120123, EP16E120124,EP16E120125,EP16E120127,EP16E120128,EP16E120129,EP16E120130,EP16E130002, EP16E130003,EP16E130004,EP16E130005,EP16E130006,EP16E130007,EP16E130009,EP16E130010, EP16E130012,EP16E130013,EP16E130014,EP16E130015,EP16E130016,EP16E130017,EP16E130020, EP16E130021,EP16E130022,EP16E130023,EP16E130025,EP16E130026,EP16E130027,EP16E130029, EP16E130030,EP16E130031,EP16E130033,EP16E130034,EP16E130035,EP16E130036,EP16E130037, EP16E130040,EP16E130041,EP16E130042,EP16E130043,EP16E130044,EP16E130045,EP16E130046, EP16E130047,EP16E130048,EP16E130049,EP16E130050,EP16E130051,EP16E130052,EP16E130053, EP16E130055,EP16E130056,EP16E130057,EP16E130058,EP16E130062,EP16E130063,EP16E130064, EP16E130065,EP16E130066,EP16E130067,EP16E130068,EP16E130069,EP16E130070,EP16E130071, EP16E130072,EP16E130073,EP16E130074,EP16E130075,EP16E130076,EP16E130077,EP16E130078, EP16E130079,EP16E130080,EP16E130081,EP16E130082,EP16E130083,EP16E130086,EP16E130087, EP16E130088,EP16E130089,EP16E130091,EP16E130092,EP16E130093,EP16E130094,EP16E130095, EP16E130097,EP16E130098,EP16E130099,EP16E130100,EP16E130101,EP16E130102,EP16E130103, EP16E130105,EP16E130106,EP16E130107,EP16E130109,EP16E130110,EP16E140001,EP16E140002, EP16E140003,EP16E140006,EP16E140007,EP16E140008,EP16E140009,EP16E140010,EP16E140011, EP16E140012,EP16E140013,EP16E140014,EP16E140015,EP16E140016,EP16E140017,EP16E140018, EP16E140019,EP16E140021,EP16E140022,EP16E140023,EP16E140024,EP16E140025,EP16E140026, EP16E140027,EP16E140028,EP16E140029,EP16E140030,EP16E140031,EP16E140032,EP16E140033, EP16E140041,EP16EB110003,EP16EB110004,EP16EB110005,EP16EB110006,EP16EB120006, EP16EB120007, EP16EB120008, EP16EB120009, EP16EB130001, EP16EB130002, EP16EB130003, EP16EB130004, EP16EB130005, EP16EB130006, EP16EB130007, EP16EB130008, EP16EB130009, EP16EB130010, EP16EB130011, EP16EB130012, EP16EB130013, EP16EB130014, EP16EB130015, EP16EB130016, EP16EB130017, EP16EB130018, EP16EB130019, EP16EB130020, EP16EB130021, EP16EB140001, EP16EB140002, P16C080001, P16C080002, P16C100001, P16C100002, P16C100003, P16C100004, P16C100005, P16C100007, P16C100008, P16C100009, P16C100010, P16C100011, P16C100012, P16C100013, P16C100014, P16C100015, P16C100017, P16C100018, P16C110001, P16C110002, P16C110003, P16C110004, P16C110005, P16C110006, P16C110007, P16C110008, P16C110011, P16C110012, P16C110013, P16C110014, P16C110015, P16C110016, P16C110017, P16C110018, P16C110019, P16C110020, P16C110021, P16C110022, P16C110023, P16C110024, P16C110025, P16C110026, P16C110027, P16C110028, P16C110029, P16C110030, P16C110031, P16C110032, P16C110033, P16C110034, P16C110035, P16C110036, P16C110037, P16C120001, P16C120002, P16C120003, P16C120006, P16C120007, P16C120008, P16C120009, P16C120010, P16C120011, P16C120012, P16C120014, P16C120015, P16C120016, P16C120017, P16C120018, P16C120019, P16C120020, P16C120021, P16C120022, P16C120023, P16C120024, P16C120025, P16C120026, P16C120027, P16C120028, P16C120029, P16C120030, P16C120031, P16C120032, P16C120033, P16C120034, P16C120035, P16C120036, P16C120037, P16C120038, P16C120039, P16C120040, P16C120041, P16C120042, P16C120043, P16C120044, P16C120045, P16C120046, P16C120047, P16C120048, P16C120049, P16C120050, P16C120051, P16C130001, P16C130002, P16C130003, P16C130004, P16C130005, P16C130006, P16C130007, P16C130008, P16C130009, P16C130010, P16C130011, P16C130012, P16C130013, P16C130014, P16C130015, P16C130016, P16C130017, P16C130018, P16C130019, P16C130020, P16C130021, P16C130023, P16C130024, P16C130025, P16C130026, P16C130027, P16C130028, P16C130029, P16C130030, P16C130031, P16C130032, P16C130033, P16C130034, P16C130035, P16C130036, P16C130037, P16C130038, P16C130039, P16C130041, P16C130043, P16C130044, P16C130045, P16C130046, P16C130047, P16C130048, P16C130049, P16C130050, P16C130051, P16C130052, P16C130053, P16E080005, P16E080007, P16E080008, P16E090002, P16E090003, P16E090004, P16E090005, P16E090006, P16E090007, P16E090008, P16E090009, P16E090010, P16E090011, P16E090012, P16E090013, P16E090014, P16E090015, P16E090016, P16E090017, P16E090018, P16E090019, P16E090020, P16E090021, P16E090024, P16E090025, P16E090026, P16E090027, P16E090028, P16EF80010, P16EM08003, P16EM09002, P16EM09003, P16EM09004, P16EVO130002, P16EVO130003, P16EVO140001, P16EVO140002, P16EVO140003, P16EVO140005, P16EVO140006, P16EVO140007, P16EVO140008  Model Number: 728132, Serial Number(s):  EP16ER100003,EP16ER100004,EP16ER100005,EP16ER100007,EP16ER100008,EP16ER110003, EP16ER110004,EP16ER110005,EP16ER110006,EP16ER110007,EP16ER110008,EP16ER110009, EP16ER110010,EP16ER110011,EP16ER110012,EP16ER110013,EP16ER110014,EP16ER110015, EP16ER110016,EP16ER110017,EP16ER110018,EP16ER110019,EP16ER110020,EP16ER110021, EP16ER110022,EP16ER110023,EP16ER110024,EP16ER120001,EP16ER120002,EP16ER120003, EP16ER120004,EP16ER120005,EP16ER120006,EP16ER120007,EP16ER120008,EP16ER120009, EP16ER120010,EP16ER120011,EP16ER120012,EP16ER120013,EP16ER120015,EP16ER120016, EP16ER120017,EP16ER120018,EP16ER130001,EP16ER130002,EP16ER130003,EP16ER130004, EP16ER130005,EP16ER130006,EP16ER130007,EP16ER130008,EP16ER130009,EP16ER130010, EP16ER130011,EP16ER130012,EP16ER130013,EP16ER130014,EP16ER130015,EP16ER130016, EP16ER130017,EP16ER130018,EP16ER130019,EP16ER130020,EP16ER130021,EP16ER130022, EP16ER130023,EP16ER130024,EP16ER130025,EP16ER130026,EP16ER130027,EP16ER130028, EP16ER130029,EP16ER130030,EP16ER130031,EP16ER130032,EP16ER130033,EP16ER130034 
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Rd
Cleveland OH 44143-2131
For Additional Information ContactDaniel Brown
440-483-7600
Manufacturer Reason
for Recall
"The orientation of images is displayed incorrectly. When operator chooses gantry on the right side from operation console, and the patient position is HFDL, HFDR, FFDL or FFDR, the patient position picture and text is not consistent. "L" and "R" of scan images are reversed and this is not consistent with the patient actual position. "When a surview which is longe
FDA Determined
Cause 2
Software Design Change
ActionPhilips is informing the consignees of this issue through a Field Safety Notice (FSN) 72800615 on October 09, 2014. The US consignees will receive this notice via certified mail and Philips Key Market representatives will distribute the FSN internationally. Additionally, A Field Service Engineer will contact each site to schedule a time to install the software update.
Quantity in Commerce708
DistributionDomestic Distribution: AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, ND, NJ, NV, NY, OH, PA, PR, SC, TN, and VA. Foreign distribution: Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, China, Colombia, Congo, Ecuador, Egypt, France, Germany, Ghana, Greece, Gutamala, Hungray, India, Indonesia, Ireland, Italy, Japan, Jordan, Latvia, Lybyan Arab JHamahiri, Lithuania, Malawi, Mexico, Myanmar, Namibia, Nepal, Netherlands, Oman, Palestine, Paraguay, Philippines, Poland, Portugal, Romania,Russia, Rwanda, Saudi Arabia, Senegal, South Agrica, South ZKorea, Spain, Sri Lanka, Switzerland, Tunisia, Turkey, Uganda, UAE, Venezuela, Vietnam, Zambia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JAK
-
-