Date Initiated by Firm | April 30, 2014 |
Date Posted | December 16, 2014 |
Recall Status1 |
Terminated 3 on March 16, 2015 |
Recall Number | Z-0526-2015 |
Recall Event ID |
69637 |
510(K)Number | K042720 |
Product Classification |
Accelerator, linear, medical - Product Code IYE
|
Product | Varian On-Board Imager Advanced Imaging System.
The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land marks. |
Code Information |
Part Number P1003290001 |
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. 911 Hansen Way Palo Alto CA 94304-1028
|
For Additional Information Contact | K. Jeffery Semone 650-424-6833 |
Manufacturer Reason for Recall | Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc tion assembly cable and the ground stud on the lid of the High Voltage Module. |
FDA Determined Cause 2 | Component change control |
Action | Varian will work with affected customers to establish schedules that minimize impact to their clinic's treatment schedule and send Field Service Engineers with the required hardware to each affected site to make the modification. |
Quantity in Commerce | 32 units |
Distribution | Worldwide Distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYE
|