Date Initiated by Firm | September 15, 2014 |
Date Posted | January 13, 2015 |
Recall Status1 |
Terminated 3 on May 20, 2015 |
Recall Number | Z-0841-2015 |
Recall Event ID |
69660 |
510(K)Number | K123603 |
Product Classification |
Image-intensified fluoroscopic x-ray system, mobile - Product Code OXO
|
Product | GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product
Used for general surgical applications and musculoskeletal procedures to visualize. |
Code Information |
GE OEC Brivo 865 Plus Serial Numbers: B4S13011 B4S13012 B4S13016 B4S13018 B4S13020 B4S13029 B4S13031 B4S13035 B4S13038 B4S13043 B4S13054 B4S13056 B4S13059- B4S13061 B4S14002- B4S14004 B4S14007 B4S14015 B4S14016 B4S14022 B4S14022 B4S1403 3 B4S1403 4 |
Recalling Firm/ Manufacturer |
GE OEC Medical Systems, Inc 384 N Wright Brothers Dr Salt Lake City UT 84116-2862
|
For Additional Information Contact | 801-536-4516 |
Manufacturer Reason for Recall | mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3). |
FDA Determined Cause 2 | Radiation Control for Health and Safety Act |
Action | Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit.
Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8. |
Quantity in Commerce | 61 units installed in US |
Distribution | Worldwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = OXO
|