| | Class 2 Device Recall Dimension Vista Intelligent Lab Systems |  |
| Date Initiated by Firm | November 05, 2014 |
| Date Posted | December 20, 2014 |
| Recall Status1 |
Terminated 3 on October 06, 2017 |
| Recall Number | Z-0838-2015 |
| Recall Event ID |
69687 |
| 510(K)Number | K051087 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Dimension Vista 500 Intelligent Lab System or Dimension Vista 1500 Intelligent Lab System:
An in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. |
| Code Information |
Material Nos.: 10284473, 10488224, 10444801, and 10444802 |
| FEI Number |
2517506
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 500 Gbc Dr Ms 514 PO BOX 6101 Newark DE 19702-2466
|
| For Additional Information Contact | Customer Support 800-441-9250 |
Manufacturer Reason for Recall | Customer complaints of discrepant flagged and/or un-flagged patient and QC results. Reagent probe performance may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server. |
FDA Determined Cause 2 | Device Design |
| Action | An Urgent Medical Device Correction, dated November 2014, was issued on November 4th, 2014 to inform customers that Siemens Healthcare Diagnostics has confirmed that reagent probe performance on Dimension Vista systems may decline more quickly than anticipated and have an effect on assays with similar formulations when processed sequentially on the same reagent server. Instructions for replacing an associated reagent probe were provided to resolve the issue. Customers are to retain the letter for their laboratory records and forward it to anyone who has received the affected product. Customers are also to complete and return the Field Correction Effectiveness Check Form. Customer should direct their questions to their Siemens Customer Care Center - Technical Solutions at 1-800-441-9250 or their local Siemens technical support representative. |
| Quantity in Commerce | 1,949 |
| Distribution | Worldwide Distribution -- Austria, Australia, Belgium, Canada, Czech Republic,
Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Saudi Arabia, Slovakia, Slovenia,
South Africa, Spain, Switzerland, United Kingdom, and United States. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|