Date Initiated by Firm |
May 13, 2015 |
Date Posted |
June 16, 2015 |
Recall Status1 |
Terminated 3 on May 20, 2016 |
Recall Number |
Z-1811-2015 |
Recall Event ID |
71081 |
510(K)Number |
K051641
|
Product Classification |
Pump, infusion - Product Code FRN
|
Product |
Alaris PC unit model 8015, Infusion pump. software version 9.17
Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The software is embedded in the Alaris PC unit.
|
Code Information |
2000863 2000879 2000946 12473445 12473494 12473519 12473530 12473531 12473535 12473629 12473840 12473846 12494992 12495525 12496140 12496473 12496474 12496493 12496497 12496514 12496528 12496529 12496551 12496552 12496554 12496568 12496570 12496571 12496572 12496591 12496594 12496595 12505909 12519370 12540760 12577165 12578322 12578596 12578604 12578636 12578645 12578662 12578687 12578750 12578762 12578857 12578899 12579095 12579194 12579499 12579566 12579643 12579689 12581148 12584449 12625758 12625765 12625901 12625916 12625930 12625933 12628604 12630285 12630778 12634157 12634382 12634783 12635363 12635391 12635539 12635928 12636509 12636901 12636967 12636979 12636980 12636984 12636985 12637133 12637353 12637658 12637799 12637853 12637881 12637907 12639373 12639381 12639382 12639406 12639410 12639414 12639495 12639536 12639552 12639555 12639558 12639583 12639590 12639617 12639636 12639637 12639646 12639647 12639650 12639683 12639688 12639699 12639701 12639709 12639710 12639712 12639793 12639795 12639797 12639828 12639839 12639840 12639848 12639849 12639862 12639865 12639881 12639905 12639908 12639940 12639958 12639971 12639990 12640010 12640014 12640059 12640060 12640062 12640063 12640064 12640107 12640115 12640116 12640142 12640143 12640145 12640148 12640169 12640171 12640177 12640201 12640204 12640205 12640236 12640239 12640240 12640241 12640267 12640269 12640273 12640275 12640295 12640296 12640300 12640302 12640317 12640319 12640321 12640332 12640333 12640350 12640352 12640353 12640365 12640366 12640375 12640399 12640400 12640419 12640439 12640440 12640461 12640466 12640470 12640521 12640530 12640533 12640535 12640536 12640546 12640547 12640548 12640567 12640568 12640583 12640584 12640587 12640588 12640590 12640603 12640626 12640627 12640628 12640658 12640671 12640672 12640677 12640694 12640697 12640709 12640737 12640774 12640775 12640777 12640778 12640801 12640809 12640822 12640849 12640856 12640859 12640866 12640867 12640870 12640872 12640875 12640876 12640927 12640931 12640934 12640935 12640949 12640959 12640987 12640988 12640995 12640996 12640997 12641002 12641007 12641009 12641011 12641014 12641015 12641018 12641054 12641058 12641060 12641065 12641067 12641083 12641092 12641100 12641104 12641107 12641110 12641114 12641119 12641120 12641122 12641124 12641127 12641134 12641136 12641140 12641142 12641143 12641146 12641155 12641156 12641158 12641166 12641172 12641176 12641178 12641183 12641185 12641186 12641188 12641194 12641199 12641211 12641228 12641236 12641247 12641295 12643075 12644125 12644172 12644291 12644311 12644320 12644323 12644333 12644337 12644339 12644566 12644615 12644617 12644654 12644675 12644677 12644682 12644683 12644701 12644714 12644716 12644740 12644746 12644791 12645320 12645323 12645333 12645334 12645335 12645336 12645387 12645388 12645389 12645390 12645391 12645410 12645411 12645412 12645413 12645414 12645428 12645513 12645515 12645552 12645564 12645638 12645643 12645757 12645759 12645788 12645791 12645818 12645820 12645837 12645849 12645857 12647652 12651929 12651956 12651979 12651982 12655112 12655116 12655427 12655429 12655440 12655507 12656147 12656172 12656446 12657543 12657760 12666994 12670649 12671576 12671581 12671915 12671928 12682772 12682860 12682974 12687089 12690302 12690303 12690316 12690317 12690318 12690319 12690320 12690326 12690358 12690359 12690362 12690363 12690372 12690373 12690374 12690375 12690382 12690383 12690384 12690385 12690386 12690387 12690388 12690389 12690393 12690394 12690418 12690419 12690444 12690445 12690446 12690449 12690451 12690452 12690453 12690454 12690455 12690456 12690457 12690465 12690466 12690468 12690469 12690470 12690486 12690487 12690495 12690496 12690497 12690498 12690499 12690500 12690501 12690502 12690503 12690504 12690513 12690514 12690515 12690516 12690517 12690518 12690522 12690523 12690525 12690568 12690570 12690573 12690574 12690636 12690637 12690638 12690641 12690642 12690643 12690644 12690645 12690646 12690647 12690648 12690649 12690653 12690654 12690656 12690657 12690658 12690659 12690660 12690661 12690662 12690663 12690664 12690669 12690670 12690672 12690673 12690674 12690677 12690678 12690679 12690680 12690681 12690686 12690687 12690688 12690689 12690703 12690705 12690706 12690707 12690708 12690709 12690711 12690713 12690726 12690727 12690728 12690729 12690730 12690731 12690732 12690751 12690752 12690774 12690775 12690776 12690782 12690783 12690784 12690785 12690786 12690787 12690788 12690789 12690801 12690802 12690803 12690804 12690816 12690817 12690820 12690821 12690825 12690833 12690834 12690835 12690836 12690837 12690838 12690839 12690845 12690846 12690863 12690864 12690872 12690875 12690876 12690976 12691011 12691059 12691060 12691074 12691076 12691077 12691079 12691084 12691085 12691087 12691088 12691107 12691119 12691120 12691122 12691141 12691148 12691152 12691168 12691181 12691182 12691184 12691185 12691186 12691187 12691217 12691218 12691219 12691221 12691222 12691223 12691224 12691225 12691248 12691249 12691252 12691278 12691279 12691280 12691281 12691282 12691283 12691298 12691304 12691305 12691306 12691307 12691308 12691310 12691312 12691313 12691315 12691316 12691355 12691356 12691357 12691375 12691382 12691383 12691393 12691394 12691412 12691413 12691414 12691416 12691421 12691435 12691448 12691471 12691475 12691483 12691487 12691490 12691493 12691504 12691509 12691537 12691547 12691556 12691557 12691562 12691563 12691567 12691573 12691575 12691580 12691585 12691587 12691589 12691591 12691592 12691596 12691603 12691604 12691608 12691610 12691613 12691614 12691615 12691638 12691724 12691726 12691728 12691729 12691732 12691736 12691749 12691753 12691760 12691776 12691779 12691801 12691805 12691809 12691813 12691819 12691820 12691821 12691822 12691823 12691824 12691825 12691826 12691827 12691864 12691865 12691866 12691867 12691869 12691911 12691913 12691914 12691916 12691921 12691922 12691934 12691939 12691940 12691945 12691946 12691947 12691949 12691950 12691951 12691952 12691956 12691957 12691958 12691959 12691981 12691982 12691983 12691984 12691985 12691988 12691989 12691990 12691991 12691999 12692000 12692010 12692011 12692012 12692013 12692016 12692033 12692034 12692035 12692040 12692041 12692042 12692043 12692044 12692046 12692047 12692048 12692051 12692059 12692060 12692061 12692063 12692065 12692066 12692086 12692087 12692088 12692089 12692090 12692093 12692094 12692095 12692096 12692097 12692112 12692113 12692114 12692115 12692116 12692119 12692120 12692121 12692122 12692123 12692146 12692147 12692148 12692149 12692150 12692157 12692158 12692159 12692166 12692167 12692168 12692169 12692171 12692172 12692173 12692174 12692175 12692179 12692180 12692181 12692182 12692183 12692185 12692186 12692187 12692212 12692221 12692222 12692223 12692224 12692227 12692228 12692230 12692232 12692233 12692235 12692236 12692237 12692264 12692265 12692266 12692269 12692284 12692285 12692286 12692287 12692290 12692298 12692305 12692306 12692311 12692312 12692314 12692318 12692326 12692328 12692332 12692333 12692347 12692355 12692356 12692358 12692367 12692374 12692386 12692391 12692392 12692395 12692402 12692407 12692411 12692414 12692415 12692424 12692426 12692433 12692442 12692445 12692454 12692469 12692470 12692471 12692472 12692480 12692557 12692587 12692588 12692597 12692611 12692664 12692732 12709082 12709261 12709344 12709373 12709375 12709383 12756128 12823310 12823530 12823697 12823784 12832846 12833537 12833632 12834543 12834546 12834575 12834591 12834750 12834757 12834766 12834782 12834788 12834796 12834798 12834801 12834810 12834815 12834816 12834821 12834822 12834856 12834882 12834886 12834895 12834917 12834970 12834988 12834997 12835000 12835021 12835022 12835023 12835024 12835037 12835038 12835039 12835056 12835059 12835060 12835132 12887457 12920660 12923109 12924068 12924832 12924913 12924999 12925045 12925050 12925099 12925168 12925368 12925463 12925536 12925682 12925722 12925729 12926178 12957794 12957808 12958518 12977976 12979317 12983528 12983555 12983575 12985603 12990540 12992278 12992435 12993038 12993365 12993366 12994243 12994526 12994645 12994683 12994684 12994685 12994688 12994692 12994693 12994739 12994744 12994745 12994760 12994777 12994783 12994795 12994798 12994804 12994808 12994810 12994840 12994845 12994848 12994876 12994883 12994890 12994891 12994897 12994900 12994908 12994909 12994910 12994913 12994916 12994917 12994918 12994920 12994926 12994932 12994933 12994934 12994937 12994940 12994941 12994943 12994944 12994945 12994948 12994949 12994950 12994955 12994956 12994957 12994958 12994959 12994960 12994961 12994963 12994966 12994967 12994969 12994971 12994973 12994975 12994978 12994981 12994982 12994985 12994994 12994996 12994998 12995014 12995017 12995021 12995024 12995048 12995050 12995051 12995052 12995055 12995074 12995075 12995099 12995100 12995103 12995106 12995107 12995108 12995123 12995124 12995127 12995128 12995138 12995139 12995140 12995142 12995143 12995147 12995150 12995151 12995169 12995170 12995171 12995172 12995183 12995185 12995186 12995187 12995194 12995195 12995196 12995214 12995259 12995260 12995273 12995291 12995293 12995333 12995334 12995338 12995355 12995366 12995378 12995390 12995391 12995433 12995434 12995440 12995441 12995442 12995447 12995476 12995478 12995479 12995481 12995495 12995496 12995497 12995500 12995504 12995508 12995512 12995513 12995520 12995521 12995536 12995538 12995548 12995554 12995555 12995557 12995558 12995590 12995596 12995615 12995625 12995627 12995634 12995636 12995639 12995658 12995659 12995668 12995669 12995673 12995711 12995719 12995720 12995721 12995735 12995763 12995768 12995770 12995773 12995774 12995778 12995785 12995791 12995794 12995837 12995844 12995845 12995849 12995880 12995886 12995913 12995933 12995934 12995938 12995949 12996034 12996054 12996071 12996090 12996140 12996153 12996156 12996157 12996177 12996178 12996190 12996192 12996208 12996211 12996212 12996213 12996239 12996240 12996243 12996245 12996246 12996247 12996251 12996252 12996253 12996264 12996265 12996266 12996275 12996276 12996277 12996281 12996291 12996292 12996293 12996295 12996296 12996297 12996299 12996301 12996302 12996303 12996319 12996320 12996330 12996331 12996332 12996334 12996335 12996336 12996338 12996339 12996340 12996341 12996359 12996375 12996376 12996377 12996379 12996419 12996420 12996421 12996499 12996501 12996502 12996504 12996505 12996511 12996512 12996513 12996517 12996518 12996519 12996529 12996530 12996531 12996532 12996534 12996535 12996536 12996546 12996547 12996549 12996552 12996562 12996564 12996566 12996576 12996577 12996580 12996581 12996600 12996601 12996602 12996603 12996607 12996608 12996612 12996613 12996625 12996628 12996635 12996637 12996638 12996652 12996660 12996661 12996662 12996669 12996670 12996673 12996691 12996692 12996695 12996696 12996697 12996730 12996731 12996734 12996751 12996753 12996754 12996755 12996763 12996764 12996765 12996766 12996782 12996788 12996808 12996872 12996899 12996902 12996904 12996905 12996921 12996922 12996923 12996925 12996926 12996934 12996935 12996936 12996942 12996944 12996956 12996957 12996960 12996961 12996967 12996968 12996982 12996992 12996993 12996994 12996997 12997106 12997119 12997122 12997137 12997140 12997141 12997147 12997185 12997193 12997198 12997240 12997307 12997395 12997416 12997421 13104896 13105472 13135813 13170902 13173483 13173495 13173764 13173876 13173914 13174220 13174257 13174266 13174277 13174445 13174463 13174792 13175148 13175562 13175570 13175939 13175952 13175958 13175960 13175988 13175999 13176046 13176047 13176212 13176335 13176376 13176377 13176378 13177213 13177223 13177272 13177295 13177317 13177375 13177389 13177390 13177391 13177399 13177400 13177401 13177402 13177403 13177437 13177438 13177439 13177464 13177467 13177480 13177482 13177483 13177484 13177521 13177522 13177538 13177540 13177555 13177559 13177638 13177645 13177646 13177649 13177659 13177660 13177662 13177665 13177666 13177668 13177680 13197886 13198270 13198289 13198301 13198334 13198351 13198419 13198424 13198425 13198436 13198439 13198448 13199008 13199143 13199153 13345286 13345401 13345751 13345797 13345858 13345884 13345967 13346779 13346828 13346843 13346929 13346968 13346973 13347137 13347159 13347187 13347281 13347391 13347417 13347440 13347458 13347498 13347520 13347551 13347555 13347563 13347596 13347602 13347620 13347639 13347650 13347652 13347678 13347691 13347788 13347800 13347853 13347866 13347870 13347882 13347883 13347936 13347944 13347960 13347977 13347981 13347986 13347996 13348007 13348054 13348070 13348082 13348087 13348115 13348134 13348140 13348183 13348221 13348231 13348232 13348249 13348270 13348353 13348366 13348412 13348427 13348431 13348501 13348508 13348509 13348523 13348550 13348560 13348592 13348593 13348617 13502105 13502172 13502193 13502297 13502700 13502993 13503168 13545823 13546069 13559911 13559973 13560036 13560325 13560326 13560880 13560899 13561226 13562177 13571361 13571379 13571407 13571422 13571430 13571485 13571486 13571488 13580815 13581289 13582035 13582070 13582266 13582379 13582383 13582391 13582396 13582427 13585759 13585782 13585898 13585935 13585947 13585960 13585998 13586036 13586051 13586119 13586122 13613192 13614382 13635380 13635497 13635620 13636530 13639751 13639778 13639851 13639881 13639907 13639956 13640022 13640023 13640025 13640032 13640041 13640056 13640074 13640080 13640088 13644802 13645869 13646195 13646684 13647589 13647677 13648658 13649249 13649579 13649874 13650049 13650224 13650332 13650360 13650391 13650413 13650415 13650419 13650450 13650488 13650518 13650740 13650747 13651040 13651123 13651139 13659176 13660643 13661272 13661273 13661276 13661340 13661492 13661507 13661508 13661557 13662082 13662099 13662125 13662187 13662380 13662443 13662505 13662534 13663018 13663340 13663522 13719652 13775421 13775475 13775951 13790865 13792476 13792481 13792484 13792508 13792510 13792511 13792512 13792532 13792533 13792534 13792587 13792639 13793461 13793517 13793527 13793535 13793565 13793567 13793587 13793591 13793647 13793746 13793757 13793758 13793814 13793816 13796338 13797792 13799536 13800160 13800873 13801017 13801027 13801066 13801124 13801134 13801219 13801220 13801221 13801229 13801244 13801247 13801294 13828741 13828798 13828800 13828811 13828833 13828872 13831258 13831824 13832774 13833797 13834146 13834479 13834581 13835226 13835235 13835239 13835298 13835300 13835335 13835342 13835348 13835351 13835362 13835363 13835366 13835367 13835380 13835387 13835389 13835395 13835397 13835399 13835402 13835416 13835417 13835421 13835422 13835424 13835428 13835429 13835434 13835435 13835436 13835443 13835444 13835445 13835446 13835459 13835463 13835471 13835479 13835536 13835539 13835542 13835553 13835566 13835567 13835568 13835569 13835584 13835591 13835594 13835611 13835615 13835633 13835635 13835648 13835664 13835688 13835689 13835690 13835691 13835701 13835707 13835708 13835711 13835712 13835713 13835714 13835782 13835787 13835788 13835790 13835804 13835809 13835811 13835813 13835814 13835815 13835817 13835822 13835845 13835847 13835848 13835849 13835850 13835853 13835874 13835891 13835892 13835906 13835908 13835918 13835937 13835938 13835940 13835986 13835987 13835990 13835991 13835992 13836050 13836052 13836056 13836085 13836086 13836089 13836092 13836093 13836108 13836110 13836111 13836114 13836115 13836116 13836117 13836122 13836124 13836133 13836134 13836138 13836143 13836151 13836157 13836158 13836370 13836377 13836380 13836392 13836393 13836394 13836395 13836435 13836437 13836438 13836442 13836443 13836444 13836448 13836453 13836454 13836455 13836456 13836457 13836458 13836479 13836480 13836485 13836501 13836503 13836509 13836511 13836512 13836513 13836514 13836520 13836526 13836527 13836528 13836529 13836532 13836578 13836579 13836588 13836591 13836592 13836609 13836611 13836613 13836614 13836625 13836628 13836639 13836678 13836764 13836783 13836812 13836828 13836832 13836869 13836886 13839219 13839938 13839966 13861250 13862493 13862534 13863147 13863725 13864130 13864192 13864630 13864812 13864928 13864934 13864961 13865019 13865066 13865115 13865164 13865246 13865288 13865289 13865314 13865318 13865326 13865327 13865333 13865337 13865338 13865344 13865345 13865347 13865348 13865350 13865416 13865420 13865437 13865439 13865441 13865451 13865452 13865460 13865463 13865469 13865472 13865478 13865525 13865529 13865531 13865533 13865536 13865546 13865551 13865610 13865631 13865632 13865634 13865666 13865667 13865669 13865675 13865679 13865705 13865719 13865739 13865749 13865825 13865835 13865914 13865929 13865936 13865975 13866029 13866030 13866035 13866045 13866061 13866065 13866072 13866074 13866076 13866082 13866086 13866096 13866152 13866185 13866187 13866192 13866195 13866199 13866214 13866225 13866240 13866276 13866325 13866336 13866339 13866341 13866363 13866367 13866384 13866387 13866390 13866393 13866418 13866422 13866423 13866432 13866435 13866439 13866442 13866446 13866459 13866472 13866489 13866499 13866609 13866643 13866645 13866646 13866649 13866700 13866711 13866712 13866713 13866751 13866767 13866772 13866773 13866783 13866803 13866806 13866811 13866812 13866814 13866821 13866822 13866855 13866856 13866857 13866869 13866877 13866882 13866885 13866937 13866938 13866949 13866950 13866951 13866962 13866971 13866991 13867007 13867010 13867107 13867126 13867127 13867132 13867134 13867153 13867183 13867187 13867190 13867193 13867208 13867210 13867222 13867240 13867317 13867319 13867323 13867350 13867375 13867376 13867381 13867516 13867528 13867533 13867592 13867595 13867596 13867663 13867665 13867703 13867705 13867707 13867712 13867717 13867718 13867720 13867733 13867734 13867739 13867744 13867748 13867749 13867750 13867752 13867753 13867754 13867757 13867758 13867760 13867776 13867781 13867782 13867783 13867796 13867800 13867801 13867830 13867832 13867857 13867859 13867867 13867877 13867882 13867883 13867892 13867905 13867906 13867910 13867911 13867916 13867940 13867944 13867954 13867955 13867959 13867979 13867981 13867983 13868105 13868109 13868112 13868130 13868132 13868183 13868222 13868230 13868303 13868348 13868354 13868357 13868395 13868424 13868461 13868463 13868467 13868478 13868482 13868499 13868513 13868517 13868519 13868525 13868532 13868535 13868537 13868546 13868554 13868564 13868569 13868571 13868577 13868580 13868585 13868603 13868613 13868627 13868629 13868630 13868631 13868633 13868634 13868636 13868736 13868767 13868768 13868779 13868842 13868859 13868889 13868908 13869070 13869077 13869079 13869119 13869120 13869124 13869138 13869169 13869170 13869186 13869189 13869205 13869225 13869262 13869263 13869265 13869272 13869274 13869297 13869316 13869437 13869439 13869441 13869443 13869450 13869486 13869498 13869501 13869520 13869521 13869541 13869568 13869571 13869639 13869640 13869649 13869675 13869686 13869760 13869761 13869763 13869783 13869784 13869787 13869802 13869852 13869853 13869854 13869857 13869859 13869861 13869862 13869892 13869956 13869977 13869992 13870048 13870054 13870061 13870082 13870085 13870119 13870124 13870162 13870164 13870169 13870173 13870174 13870180 13870182 13870183 13870184 13870185 13870186 13870187 13870188 13870189 13870193 13870202 13870207 13870215 13870230 13870236 13870238 13870242 13870244 13870261 13870268 13870274 13870294 13870328 13870334 13870338 13870405 13870407 13870454 13870455 13870477 13870503 13870507 13870508 13870599 13870600 13870620 13870627 13870638 13870644 13870682 13870697 13870700 13870727 13870733 13870745 13870749 13870754 13870757 13870779 13870781 13870801 13870806 13870825 13870839 13870841 13870847 13870880 13870889 13870892 13870917 13870949 13871016 13871302 13871415 13871443 13871566 13871586 13871675 13871694 13871696 13871707 13871733 13871790 13871903 13871907 13871919 13871921 13871944 13871949 13871956 13871969 13871972 13871978 13871982 13872025 13872087 13872090 13872109 13872111 13872156 13872243 13872280 13872327 13872329 13872330 13872331 13872346 13872402 13872407 13872409 13872410 13872487 13872488 13872489 13872490 13872491 13872492 13872494 13872495 13872496 13872502 13872503 13872504 13872505 13872506 13872509 13872510 13872511 13872513 13872548 13872549 13872660 13872661 13872662 13872663 13872664 13872665 13872693 13872696 13872706 13872707 13872709 13872710 13872745 13872746 13872747 13872748 13872749 13872750 13872784 13872786 13872787 13872788 13872789 13872790 13872817 13872821 13872866 13872867 13872868 13872869 13872870 13872883 13872884 13872897 13872907 13873090 13873096 13873097 13873098 13873099 13873100 13873132 13873133 13873237 13873756 13876159 13876873 13876899 13877123 13877362 13877376 13877388 13877440 13877523 13878124 13878603 13880654 13880819 13881595 13882097 13883443 13883454 13883474 13883475 13883476 13883478 13883481 13883488 13883491 13883492 13883498 13883499 13883501 13883507 13883517 13883518 13883519 13883521 13883527 13883528 13883550 13883554 13883557 13883558 13883559 13883589 13883593 13883605 13883606 13883607 13883609 13883611 13883613 13883614 13883615 13883617 13883619 13883621 13883622 13883625 13883628 13883633 13883635 13883636 13883637 13883638 13883640 13883642 13883648 13883674 13883678 13883682 13883685 13883686 13883689 13883695 13883699 13883704 13883705 13883708 13883715 13883716 13883718 13883724 13883725 13883740 13883755 13883762 13883764 13883766 13883767 13883774 13883781 13883786 13883793 13883795 13883814 13883829 13883862 13883887 13883890 13883892 13883974 13884588 13885927 13886445 13886555 13886740 13886998 13887172 13887241 13888019 13888121 13888347 13888383 13888384 13888385 13888423 13888430 13888443 13888447 13888476 13888495 13888686 13888696 13888698 13888702 13888723 13888760 13888840 13888850 13888882 13888886 13889070 13889130 13889216 13889240 13889309 13889350 13889357 13889365 13889368 13889400 13889420 13889421 13889422 13889479 13889518 13889523 13889527 13889528 13889530 13889532 13889539 13889540 13889565 13889569 13889596 13889599 13889603 13889605 13889630 13889633 13889644 13889656 13889661 13889724 13889760 13889765 13889794 13889805 13889878 13889879 13889898 13889903 13889904 13889909 13889927 13889933 13889934 13889935 13889945 13889948 13889949 13889951 13889957 13889962 13889976 13889977 13889978 13889979 13889980 13889988 13889989 13889999 13890000 13890001 13890002 13890003 13890021 13890023 13890025 13890026 13890029 13890040 13890041 13890042 13890058 13890073 13890074 13890114 13890116 13890117 13890119 13890120 13890124 13890126 13890127 13890158 13890159 13890177 13890179 13890180 13890181 13890198 13890206 13890217 13890247 13890262 13890264 13890265 13890266 13890267 13890268 13890270 13890277 13890293 13890302 13890303 13890335 13890338 13890343 13890347 13890353 13890366 13890367 13890368 13890369 13890370 13890371 13890372 13890389 13890390 13890391 13890393 13890394 13890395 13890397 13890398 13890430 13890431 13890433 13890434 13890435 13890436 13890440 13890442 13890443 13890445 13890446 13890447 13890448 13890459 13890461 13890476 13890482 13890483 13890487 13890488 13890492 13890494 13890496 13890497 13890498 13890509 13890510 13890511 13890512 13890550 13890552 13890554 13890556 13890558 13890559 13890560 13890566 13890569 13890575 13890576 13890579 13890582 13890711 13890712 13890713 13890714 13890715 13890717 13890719 13890722 13890724 13890725 13890727 13890728 13890730 13890731 13890739 13890740 13890741 13890742 13890743 13890744 13890745 13890746 13890751 13890752 13890753 13890755 13890765 13890768 13890770 13890771 13890772 13890773 13890775 13890776 13890777 13890779 13890788 13890789 13890794 13890795 13890797 13890798 13890865 13890866 13890869 13890870 13890871 13890880 13890886 13890887 13890889 13890892 13890893 13890895 13890897 13890900 13890901 13890910 13890913 13890921 13890930 13890954 13890955 13890983 13890987 13891019 13891021 13891022 13891078 13891079 13891080 13891095 13891096 13891100 13891103 13891114 13891115 13891117 13891123 13891130 13891133 13891135 13891136 13891144 13891254 13891264 13891267 13891268 13891289 13891312 13891314 13891323 13891340 13891342 13891343 13891374 13891382 13891390 13891416 13891418 13891419 13891420 13891423 13891424 13891440 13891443 13891444 13891445 13891446 13891447 13891448 13891450 13891451 13891452 13891457 13891458 13891459 13891460 13891461 13891462 13891464 13891465 13891469 13891491 13891492 13891493 13891496 13891497 13891499 13891509 13891510 13891511 13891512 13891514 13891515 13891516 13891518 13891519 13891521 13891523 13891525 13891528 13891529 13891531 13891535 13891536 13891537 13891547 13891550 13891551 13891555 13891559 13891575 13891576 13891577 13891579 13891581 13891587 13891607 13891637 13891638 13891639 13891640 13891643 13891645 13891650 13891652 13891655 13891656 13891657 13891660 13891661 13891663 13891666 13891667 13891668 13891671 13891672 13891674 13891675 13891677 13891678 13891679 13891680 13891697 13891699 13891703 13891704 13891705 13891707 13891708 13891709 13891712 13891714 13891715 13891717 13891721 13891723 13891745 13891746 13891748 13891749 13891750 13891767 13891779 13891780 13891783 13891799 13891835 13891836 13891837 13891839 13891840 13891846 13891850 13891853 13891934 13891957 13891960 13891961 13891965 13891968 13891975 13891976 13891981 13891983 13891984 13891994 13891999 13892000 13892008 13892028 13892029 13892030 13892031 13892034 13892043 13892054 13892058 13892059 13892078 13892085 13892086 13892087 13892092 13892106 13892110 13892117 13892124 13892128 13892141 13892148 13892154 13892155 13892157 13892184 13892193 13892194 13892198 13892199 13892202 13892359 13892360 13892367 13892382 13892386 13892390 13892391 13892392 13892437 13892438 13892440 13892441 13892453 13892465 13892466 13892506 13892550 13892551 13892552 13892553 13892584 13892607 13892697 13892707 13892732 13892733 13892735 13892753 13892760 13892764 13892766 13892773 13892789 13892790 13892793 13892797 13892800 13892801 13892831 13892832 13892833 13892835 13892836 13892853 13892855 13892857 13892860 13892861 13892866 13892868 13892870 13892872 13892885 13892913 13892952 13892958 13892960 13892978 13893015 13893027 13893028 13893029 13893047 13893052 13893059 13893068 13893069 13893072 13893075 13893079 13893080 13893098 13893104 13893105 13893107 13893109 13893144 13893146 13893147 13893148 13893149 13893150 13893154 13893156 13893158 13893160 13893169 13893173 13893244 13893245 13893247 13893265 13893267 13893268 13893274 13893278 13893279 13893289 13893317 13893324 13893327 13893344 13893355 13893373 13893403 13893407 13893432 13893453 13893490 13893499 13893539 13893547 13893548 13893549 13893580 13895886 13918347 13921569 13921570 13922042 13922049 13922092 13922107 13922164 13922261 13922280 13922284 13922291 13922301 13922302 13922303 13922312 13922330 13922331 13922366 13922369 13922370 13922372 13922375 13922402 13922405 13922418 13922423 13922429 13922442 13922446 13922447 13922449 13922451 13922484 13922487 13922496 13922498 13922500 13922503 13922537 13922641 13922646 13922655 13922679 13922680 13922713 13922738 13922782 13923285 13937202 13937378 13937415 13937417 13937971 13937973 13937984 13938043 13939408 13939580 13940296 13940604 13940724 13940770 13940939 13940980 13940990 13941041 13941057 13941165 13941173 13941207 13941279 13941316 13941341 13941443 13941486 13941524 13941617 13941976 13941985 13942034 13942060 13942067 13942097 13942148 13942150 13942234 13942274 13942295 13942304 13942308 13942325 13942333 13942386 13942388 13942389 13942390 13942392 13942394 13942395 13942401 13942404 13942441 13942450 13942455 13942457 13942458 13942459 13942478 13942510 13942553 13942571 13942572 13942893 13942894 13943037 13943053 13943055 13943066 13943070 13943072 13943079 13943080 13943082 13943087 13943089 13943090 13943091 13943092 13943171 13943172 13943173 13943174 13943175 13943176 13943177 13943179 13943180 13943181 13943182 13943183 13943185 13943188 13943190 13943218 13943219 13943235 13943236 13943260 13943261 13943262 13943263 13943264 13943271 13943358 13943359 13943369 13943370 13943376 13978282 13979409 13979683 13979688 13979703 13979741 13979747 13979782 13979912 13980573 14008401 14016999 14017252 14018511 14019550 14021306 14022147 14022578 14022649 14023116 14023146 14023244 14023471 14023534 14023540 14023572 14023589 14023596 14023604 14023628 14023630 14023666 14023678 14023691 14023706 14023709 14023715 14023721 14023722 14023796 14023799 14023802 14023806 14023808 14023826 14023828 14023829 14023830 14023831 14023832 14023833 14023834 14023836 14023867 14023903 14023904 14023906 14023908 14023909 14023910 14023911 14023913 14023914 14023915 14023918 14023919 14023920 14023921 14023922 14023928 14023940 14023958 14023970 14023971 14023973 14023974 14023977 14023978 14023979 14023981 14023982 14023983 14023984 14023985 14023989 14023990 14023991 14023994 14023995 14023996 14023997 14023998 14023999 14024001 14024008 14024013 14024014 14024015 14024117 14024118 14024120 14024121 14024122 14024123 14024124 14024125 14024126 14024127 14024128 14024129 14024130 14024131 14024133 14024134 14024135 14024136 14024137 14024138 14024139 14024140 14024141 14024142 14024143 14024144 14024145 14024167 14024168 14024169 14024170 14024171 14024172 14024174 14024175 14024177 14024178 14024179 14024184 14024186 14024193 14024202 14024205 14024211 14024213 14024240 14024241 14024254 14024257 14024263 14024266 14024268 14024269 14024271 14024273 14024274 14024276 14024374 14024375 14024376 14024377 14024378 14024382 14024386 14024388 14024389 14024390 14024391 14024392 14024393 14024395 14024396 14024397 14024398 14024399 14024400 14024402 14024403 14024404 14024405 14024406 14024408 14024409 14024412 14024421 14024423 14024448 14024449 14024452 14024454 14024455 14024456 14024485 14024493 14024495 14024518 14024520 14024521 14024522 14024550 14024559 14024563 14024575 14024599 14024601 14024603 14024651 14024654 14024655 14024656 14024657 14024661 14024662 14024666 14024717 14024749 14024768 14024783 14024802 14024817 14024818 14024819 14024820 14024822 14024823 14024824 14024825 14024826 14024827 14024828 14024829 14024830 14024832 14024833 14024834 14024836 14024837 14024838 14024841 14024842 14024843 14024848 14024849 14024851 14024852 14024854 14024855 14024858 14024859 14024860 14024861 14024862 14024863 14024864 14024870 14024871 14024872 14024873 14024874 14024875 14024877 14024878 14024879 14024881 14024882 14024883 14024884 14024886 14024887 14024888 14024889 14024890 14024891 14024892 14024893 14024894 14024895 14024896 14024897 14024898 14024899 14024915 14024918 14024927 14024984 14024997 14025465 14025483 14025501 14025539 14025548 14025627 14025633 14025634 14025878 14025888 14025909 14025970 14026000 14026005 14026006 14026007 14026012 14026026 14026028 14026029 14026034 14026038 14026080 14026082 14026083 14026087 14026089 14026090 14026091 14026092 14026093 14026117 14026118 14026120 14026123 14026131 14026134 14026136 14026137 14026155 14026156 14026159 14029332 14031150 14032647 14032899 14033622 14033685 14035892 14035997 14038362 14038400 14038423 14039134 14039587 14039591 14039801 14040154 14040177 14040451 14040839 14040840 14040848 14041149 14041412 14041499 14041609 14041611 14041853 14041892 14051594 14079423 14079531 14080100 14080602 14082529 14083543 14084007 14084038 14084212 14084536 14084710 14084861 14084891 14084989 14085007 14085009 14085035 14085079 14085083 14085100 14085119 14085202 14085213 14085219 14085220 14085222 14085226 14085243 14085324 14085327 14085333 14085545 14085546 14085583 14085592 14085926 14085933 14086042 14086122 14086169 14086201 14086446 14086463 14086492 14086583 14086678 14086696 14086697 14086713 14086744 14086758 14086772 14086784 14086788 14086801 14086816 14086817 14086825 14086834 14086873 14086926 14086955 14086957 14086960 14086983 14086987 14087003 14087029 14087031 14087051 14087091 14087140 14087142 14087143 14087144 14087148 14087151 14087204 14087232 14087233 14087352 14087359 14087369 14087384 14087412 14087446 14087494 14087534 14087726 14087730 14087737 14087742 14087836 14087865 14087866 14088011 14088013 14088318 14088321 14088361 14088362 14088392 14088396 14088407 14088491 14088494 14088495 14088496 14088498 14088501 14088503 14088512 14088514 14088516 14088517 14088518 14088520 14088522 14088523 14088527 14088528 14088530 14088531 14088532 14088534 14088535 14088537 14088602 14088604 14088630 14088643 14088645 14088653 14088654 14088656 14088657 14088669 14088671 14088683 14088684 14088685 14088692 14088693 14088703 14088704 14088705 14088706 14088707 14088719 14088720 14088721 14088729 14088730 14088731 14088733 14088735 14088772 14088773 14088774 14088775 14088805 14088806 14088807 14088808 14088810 14088811 14088812 14088813 14088814 14088829 14088830 14088831 14088832 14088833 14088844 14088848 14088849 14088850 14088855 14088860 14088863 14088864 14088865 14088866 14088867 14088868 14088869 14088870 14088874 14088879 14088880 14088881 14088882 14088883 14088919 14088922 14088924 14088925 14088927 14088928 14088929 14088930 14088931 14088932 14088933 14088934 14088935 14088936 14088963 14088966 14089003 14089007 14089009 14089021 14089022 14089023 14089041 14089043 14089044 14089057 14089058 14089059 14089060 14089061 14089068 14089070 14089073 14089074 14089075 14089078 14089081 14089083 14089090 14089091 14089092 14089112 14089130 14089133 14089157 14089174 14089182 14089186 14089187 14089188 14089190 14089201 14089202 14089208 14089209 14089210 14089231 14089232 14089238 14089239 14089245 14089255 14089256 14089257 14089258 14089262 14089263 14089264 14089265 14089289 14089292 14089294 14089295 14089296 14089297 14089300 14089301 14089302 14089305 14089306 14089308 14089333 14089334 14089337 14089338 14089339 14089340 14089341 14089344 14089346 14089350 14089351 14089357 14089358 14089359 14089360 14089361 14089362 14089363 14089364 14089389 14089390 14089392 14089394 14089401 14089402 14089403 14089404 14089419 14089420 14089422 14089437 14089438 14089440 14089441 14089442 14089443 14089444 14089463 14089466 14089476 14089482 14089483 14089485 14089489 14089490 14089493 14089516 14089517 14089519 14089520 14089525 14089542 14089544 14089547 14089549 14089566 14089567 14089568 14089569 14089571 14089572 14089573 14089574 14089780 14089781 14089786 14089787 14089789 14089790 14089791 14089792 14089825 14089828 14089829 14089855 14089857 14089858 14089859 14089860 14089868 14089869 14089871 14089873 14089875 14089876 14089877 14089879 14089880 14089882 14089883 14089884 14089957 14089960 14089961 14089962 14089963 14089975 14089977 14089996 14089998 14090000 14090001 14090002 14090003 14090004 14090014 14090016 14090017 14090018 14090035 14090037 14090038 14090039 14090040 14090041 14090042 14090043 14090045 14090063 14090064 14090065 14090066 14090067 14090068 14090069 14090070 14090071 14090072 14090073 14090075 14090076 14090077 14090078 14090079 14090087 14090089 14090090 14090091 14090092 14090095 14090096 14090097 14090098 14090099 14090106 14090107 14090120 14090121 14090122 14090123 14090124 14090125 14090126 14090127 14090128 14090129 14090135 14090136 14090137 14090138 14090139 14090140 14090141 14090142 14090143 14090144 14090145 14090146 14090147 14090148 14090149 14090150 14090151 14090152 14090153 14090154 14090175 14090176 14090177 14090178 14090179 14090180 14090181 14090182 14090183 14090184 14090185 14090186 14090187 14090202 14090203 14090219 14090221 14090222 14090223 14090224 14090230 14090231 14090232 14090233 14090234 14090237 14090239 14090240 14090241 14090242 14090244 14090245 14090253 14090256 14090260 14090263 14090264 14090265 14090266 14090267 14090268 14090270 14090273 14090307 14090308 14090309 14090310 14090311 14090312 14090313 14090314 14090315 14090320 14090321 14090322 14090323 14090324 14090325 14090326 14090327 14090346 14090347 14090358 14090359 14090365 14090366 14090367 14090368 14090369 14090370 14090371 14090372 14090373 14090380 14090382 14090386 14090388 14090389 14090390 14090391 14090418 14090419 14090420 14090421 14090425 14090466 14090467 14090468 14090543 14090544 14090546 14090547 14090549 14090551 14090552 14090553 14090554 14090555 14090556 14090557 14090558 14090559 14090560 14090561 14090562 14090563 14090564 14090565 14090566 14090567 14090568 14090569 14090570 14090576 14090577 14090586 14090605 14090614 14090615 14090617 14090619 14090620 14090621 14090622 14090623 14090624 14090625 14090626 14090627 14090628 14090640 14090641 14090642 14090643 14090644 14090645 14090646 14090647 14090648 14090659 14090660 14090661 14090662 14090692 14090693 14090696 14090697 14090701 14090716 14090717 14090718 14090719 14090720 14090777 14090780 14090783 14090784 14090785 14090786 14090787 14090788 14090789 14090790 14090791 14090792 14090793 14090794 14090795 14090796 14090797 14090798 14090799 14090817 14090818 14090819 14090820 14090821 14090887 14090888 14090889 14090890 14090891 14090903 14090904 14090922 14090923 14090925 14090927 14090928 14090930 14090931 14090933 14090934 14090935 14090937 14090939 14090941 14090943 14090945 14090946 14090957 14090958 14090959 14090960 14090980 14090994 14090995 14090996 14090997 14090998 14090999 14091000 14091001 14091004 14091041 14091042 14091043 14091044 14091045 14091046 14091047 14091048 14091049 14091054 14091055 14091056 14091063 14091064 14091065 14091066 14091067 14091068 14091069 14091070 14091071 14091072 14091073 14091074 14091075 14091076 14091097 14091098 14091104 14091105 14091106 14091122 14091123 14091124 14091125 14091229 14091230 14091231 14091232 14091233 14091234 14091235 14091236 14091237 14091238 14091239 14091240 14091241 14091242 14091243 14091259 14091260 14091262 14091263 14091265 14091293 14091294 14091295 14091296 14091297 14091298 14091299 14091300 14091301 14091302 14091303 14091304 14091305 14091307 14091309 14091311 14091312 14091314 14091327 14091328 14091329 14091330 14091331 14091373 14091374 14091375 14091376 14091377 14091378 14091379 14091380 14091381 14091382 14091383 14091384 14091385 14091386 14091387 14091388 14091394 14091395 14091396 14091397 14091398 14091404 14091405 14091406 14091407 14091408 14091409 14091410 14091411 14091412 14091413 14091433 14091434 14091435 14091436 14091437 14091500 14091501 14091502 14091503 14091504 14091505 14091506 14091507 14091508 14091509 14091510 14091511 14091512 14091513 14091514 14091515 14091516 14091517 14091518 14091519 14091523 14091524 14091525 14091526 14091527 14091538 14091539 14091540 14091541 14091542 14091548 14091549 14091550 14091551 14091552 14091558 14091559 14091560 14091561 14091562 14091565 14091568 14091570 14091571 14091572 14091573 14091574 14091575 14091576 14091577 14091578 14091579 14091580 14091581 14091582 14091588 14091589 14091590 14091591 14091592 14091593 14091594 14091596 14091597 14091608 14091609 14091610 14091611 14091612 14091613 14091614 14091615 14091616 14091617 14091619 14091620 14091621 14091622 14091656 14091660 14091661 14091662 14091663 14091664 14091665 14091666 14091682 14091683 14091684 14091685 14091686 14091687 14091688 14091689 14091690 14091691 14091692 14091693 14091706 14091707 14091708 14091709 14091710 14091711 14091712 14091713 14091714 14091716 14091725 14091728 14091733 14091734 14091736 14091737 14091738 14091739 14091740 14091741 14091742 14091743 14091745 14091746 14091747 14091748 14091749 14091750 14091751 14091752 14091753 14091754 14091755 14091767 14091768 14091769 14091770 14091771 14091772 14091773 14091774 14091775 14091776 14091777 14091778 14091779 14091780 14091781 14091789 14091790 14091791 14091792 14091793 14091794 14091795 14091796 14091797 14091798 14091799 14091800 14091825 14091828 14091830 14091837 14091839 14091843 14091853 14091854 14091858 14091859 14091860 14091861 14091862 14091868 14091869 14091870 14091871 14091872 14091874 14091875 14091876 14091877 14091878 14091879 14091880 14091881 14091884 14091885 14091886 14091887 14091888 14091890 14091891 14091892 14091893 14091894 14091895 14091896 14091897 14091898 14091904 14091908 14091910 14091911 14091912 14091913 14091914 14091915 14091916 14091917 14091918 14091919 14091920 14091921 14091922 14091923 14091924 14091925 14091926 14091927 14091928 14091929 14091930 14091931 14091932 14091933 14091934 14091935 14091944 14091945 14091946 14091947 14091948 14091997 14091999 14092004 14092005 14092006 14092007 14092008 14092009 14092011 14092012 14092024 14092025 14092026 14092027 14092028 14092029 14092030 14092031 14092032 14092033 14092034 14092035 14092036 14092037 14092048 14092049 14092050 14092051 14092052 14092053 14092054 14092055 14092056 14092057 14092058 14092059 14092060 14092061 14092062 14092063 14092064 14092065 14092066 14092070 14092071 14092073 14092074 14092075 14092081 14092082 14092083 14092084 14092085 14092092 14092093 14092094 14092095 14092096 14092097 14092098 14092099 14092100 14092101 14092102 14092103 14092104 14092105 14092106 14092107 14092108 14092114 14092115 14092116 14092117 14092118 14092145 14092146 14092147 14092149 14092160 14092161 14092162 14092163 14092164 14092165 14092166 14092167 14092168 14092169 14092224 14092225 14092226 14092227 14092228 14092229 14092230 14092231 14092232 14092233 14092234 14092235 14092236 14092237 14092238 14092239 14092240 14092241 14092242 14092255 14092256 14092257 14092258 14092259 14092260 14092261 14092262 14092263 14092264 14092265 14092266 14092267 14092268 14092269 14092284 14092285 14092286 14092287 14092288 14092289 14092290 14092291 14092292 14092293 14092294 14092296 14092297 14092298 14092299 14092300 14092301 14092302 14092303 14092304 14092305 14092306 14092307 14092308 14092309 14092310 14092311 14092312 14092313 14092314 14092315 14092316 14092317 14092318 14092319 14092320 14092321 14092322 14092323 14092324 14092325 14092326 14092327 14092328 14092329 14092330 14092331 14092332 14092333 14092334 14092335 14092336 14092337 14092338 14092339 14092340 14092341 14092342 14092343 14092344 14092345 14092346 14092347 14092348 14092349 14092350 14092351 14092352 14092353 14092354 14092355 14092356 14092357 14092358 14092359 14092360 14092361 14092362 14092363 14092364 14092365 14092366 14092367 14092368 14092369 14092370 14092371 14092372 14092373 14092374 14092375 14092376 14092377 14092378 14092380 14092381 14092382 14092383 14092384 14092385 14092386 14092387 14092388 14092389 14092390 14092391 14092392 14092393 14092394 14092396 14092397 14092398 14092399 14092400 14092401 14092402 14092403 14092404 14092405 14092406 14092407 14092408 14092409 14092434 14092435 14092436 14092437 14092438 14092439 14092445 14092446 14092447 14092448 14092449 14092450 14092451 14092452 14092453 14092454 14092455 14092456 14092463 14092464 14092465 14092466 14092467 14092468 14092469 14092471 14092472 14092473 14092474 14092475 14092476 14092477 14092478 14092479 14092480 14092481 14092482 14092483 14092484 14092485 14092486 14092487 14092488 14092489 14092490 14092516 14092518 14092520 14092521 14092522 14092532 14092533 14092534 14092535 14092536 14092537 14092538 14092539 14092540 14092541 14092545 14092546 14092547 14092548 14092549 14092550 14092551 14092559 14092563 14092564 14092565 14092566 14092567 14092568 14092632 14092634 14092635 14092636 14092638 14092640 14092641 14092642 14092644 14092645 14092646 14092647 14092648 14092649 14092650 14092651 14092652 14092653 14092654 14092655 14092657 14092666 14092667 14092668 14092669 14092670 14092671 14092672 14092673 14092674 14092675 14092676 14092677 14092678 14092679 14092680 14092682 14092683 14092684 14092686 14092689 14092690 14092691 14092692 14092693 14092694 14092695 14092696 14092697 14092698 14092699 14092700 14092701 14092703 14092704 14092706 14092708 14092709 14092710 14092711 14092712 14092713 14092714 14092727 14092728 14092729 14092730 14092752 14092753 14092754 14092755 14092756 14092758 14092759 14092760 14092761 14092764 14092765 14092766 14092773 14092774 14092775 14092776 14092778 14092779 14092780 14092781 14092782 14092783 14092784 14092785 14092786 14092787 14092788 14092789 14092791 14092794 14092795 14092796 14092797 14092799 14092800 14092801 14092802 14092803 14092804 14092807 14092808 14092809 14092810 14092811 14092812 14092813 14092814 14092816 14092820 14092821 14092822 14092823 14092824 14092825 14092831 14092832 14092833 14092890 14092891 14092892 14092893 14092894 14092895 14092896 14092897 14092898 14092902 14092904 14092906 14092907 14092908 14092909 14092910 14092915 14092921 14092922 14092923 14092924 14092925 14092972 14092973 14092974 14092975 14092976 14092982 14092983 14092984 14092985 14092986 14092987 14092988 14092991 14092993 14092994 14092995 14093011 14093017 14093019 14093020 14093026 14093027 14093028 14093029 14093030 14093036 14093037 14093038 14093039 14093040 14093043 14093059 14093061 14093062 14093068 14093069 14093070 14093071 14093072 14093078 14093079 14093080 14093081 14093082 14093093 14093094 14093095 14093099 14093101 14093104 14093111 14093113 14093114 14093115 14093116 14093117 14093118 14093119 14093120 14093122 14093123 14093124 14093126 14093127 14093198 14093199 14093200 14093201 14093202 14093203 14093204 14093205 14093206 14093207 14093210 14093212 14093214 14093216 14093223 14093224 14093225 14093226 14093229 14093230 14093248 14093255 14093261 14093263 14093264 14093273 14093274 14093281 14093284 14093286 14093312 14093333 14093354 14093364 14093412 14093429 14093456 14093460 14093504 14093505 14093507 14093508 14093510 14093511 14093512 14093524 14093525 14093527 14093575 14093576 14093577 14093578 14093579 14093580 14093581 14093582 14093583 14093590 14093592 14093593 14093595 14093596 14093597 14093615 14093616 14093617 14093618 14093619 14093626 14093645 14093648 14093649 14093650 14093657 14093658 14093660 14093661 14093663 14093665 14093666 14093668 14093700 14093701 14093710 14093740 14093749 14093752 14093758 14093790 14093829 14093832 14093834 14093835 14093860 14093872 14093894 14093905 14093907 14093914 14102634 14102676 14105795 14109278 14110105 14113111 14113219 14118989 14118993 14119477 14120003 14120257 14120293 14120305 14120307 14120326 14120328 14120378 14120381 14120465 14120466 14120472 14120501 14120551 14120556 14120581 14120600 14120742 14120769 14120959 14120960 14120975 14120976 14120977 14121219 14121222 14121408 14121479 14121519 14121520 14121523 14121525 14121543 14121544 14121545 14121548 14121549 14121555 14121571 14121575 14121583 14121594 14121599 14121633 14121634 14121640 14121642 14121643 14121752 14121755 14121756 14121790 14121791 14121807 14121813 14121815 14121820 14121851 14121861 14121879 14121881 14121882 14121897 14121912 14121983 14122002 14122006 14122013 14122014 14122015 14122018 14122020 14122021 14122022 14122023 14122024 14122041 14122042 14122044 14122046 14122047 14122048 14122049 14122050 14122051 14122052 14122053 14122059 14122060 14122065 14122069 14122071 14122079 14122082 14122084 14122090 14122091 14122092 14122093 14122094 14122131 14122135 14122137 14122140 14122141 14122149 14122151 14122152 14122159 14122160 14122210 14122211 14122214 14122243 14122246 14122249 14122252 14122254 14122285 14122693 14122749 14122789 14122797 14122840 14122973 14123047 14123143 14123220 14123222 14123224 14123226 14123228 14123232 14123233 14123245 14123261 14123272 14123273 14123274 14123275 14123302 14123305 14123306 14123333 14123334 14123335 14123336 14123337 14123338 14123339 14123340 14123343 14123344 14123345 14123346 14123347 14123348 14123354 14123356 14123357 14123373 14123374 14123411 14123432 14124294 14125928 14126498 14126530 14127395 14127540 14127547 14127555 14130964 14131012 14131072 14131074 14131084 14131087 14131255 14131265 14131455 14131604 14131606 14131662 14131663 14131684 14132000 14133563 14134325 14134710 14135157 14135857 14136070 14136184 14136350 14137156 14137160 14137240 14137250 14137319 14137343 14137398 14137399 14137403 14137457 14137922 14138020 14138122 14138201 14138220 14138272 14138315 14138355 14138361 14138399 14138440 14138441 14138442 14138468 14138600 14138655 14138713 14138779 14138796 14138797 14138845 14138848 14138853 14138857 14138873 14138874 14138878 14138931 14138944 14138947 14138972 14138985 14138986 14138988 14139001 14139006 14139034 14139059 14139061 14139084 14139086 14139099 14139102 14139135 14139145 14139147 14139151 14139165 14139247 14139261 14139268 14139275 14139333 14139414 14139415 14139495 14139500 14139544 14139551 14139552 14139561 14139566 14139601 14139602 14139605 14139608 14139609 14139613 14139619 14139629 14139676 14139679 14139685 14139686 14139689 14139690 14139692 14139694 14139695 14139702 14139707 14139708 14139710 14139723 14139725 14139733 14139734 14139748 14139749 14139753 14139755 14139796 14139802 14139807 14139812 14139817 14139818 14139824 14139825 14139826 14139828 14139830 14139831 14139832 14139834 14139835 14139836 14139837 14139838 14139840 14139841 14139842 14139876 14139877 14139882 14139890 14139892 14139894 14139897 14139898 14139899 14139900 14139901 14139908 14139910 14139914 14139915 14139916 14139917 14139923 14139946 14139956 14139959 14139960 14139961 14139987 14139988 14139991 14140002 14140003 14140004 14140005 14140007 14140024 14140025 14140026 14140042 14140043 14140045 14140046 14140126 14140646 14140647 14140648 14140649 14140650 14140669 14140670 14140674 14140675 14140676 14140679 14140681 14140682 14140690 14140691 14140692 14140693 14140697 14140698 14140699 14140702 14140703 14140704 14140705 14140706 14140713 14140715 14140720 14140721 14140722 14140723 14140725 14140726 14140729 14140730 14140731 14140733 14140747 14140748 14140749 14140750 14140751 14140752 14140753 14140760 14140764 14140766 14140768 14140771 14140772 14140773 14140774 14140775 14140776 14140778 14140779 14140780 14140782 14140783 14140784 14140785 14140786 14140790 14140791 14140792 14140793 14140794 14140795 14140796 14140797 14140798 14140802 14140804 14140805 14140814 14140944 14140977 14140992 14140993 14141053 14141102 14141109 14141112 14141114 14141133 14141324 14141390 14141466 14141497 14141645 14141866 14147990 14148386 14148728 14148794 14149112 14149115 14149136 14149144 14149155 14149166 14149198 14149201 14149205 14149218 14149377 14149384 14149386 14149409 14149465 14149471 14149513 14149519 14149650 14149721 14149733 14149737 14149758 14149761 14149781 14149823 14149889 14149890 14149906 14149984 14149988 14150042 14150045 14150049 14150080 14150094 14150112 14150115 14150160 14150165 14150179 14150212 14150252 14150253 14150261 14150263 14150287 14150370 14150404 14150442 14150445 14150453 14150455 14150488 14150535 14150582 14150595 14151084 14151122 14152862 14153285 14153288 14153289 14153295 14153330 14153332 14153333 14153630 14153682 14153727 14153771 14153947 14153955 14154047 14154088 14154089 14154139 14154149 14154167 14154177 14154180 14154187 14154215 14154260 14154285 14154295 14154353 14154363 14154367 14154370 14154372 14154375 14154384 14154386 14154391 14154404 14154405 14154406 14154407 14154411 14154414 14154416 14154419 14154423 14154430 14154434 14154435 14154439 14154441 14154443 14154444 14154447 14154458 14154470 14154475 14154478 14154481 14154486 14154487 14154488 14154499 14154508 14154512 14154518 14154520 14154526 14154533 14154534 14154539 14154543 14154549 14154553 14154568 14154570 14154571 14154597 14154605 14154611 14154612 14154616 14154619 14154620 14154623 14154635 14154676 14154717 14154770 14154779 14154819 14154857 14154883 14154900 14154921 14154946 14154948 14154949 14154970 14154979 14154989 14155010 14155027 14155061 14155077 14155105 14155156 14155198 14155911 14156144 14156181 14156183 14156197 14156208 14156211 14156238 14156257 14156284 14156381 14156393 14156584 14156596 14156608 14156609 14156636 14156665 14156679 14156683 14156712 14156963 14156971 14157023 14157025 14157070 14157270 14157291 14157342 14157657 14157766 14157805 14157821 14157838 14157862 14157900 14157910 14158217 14158230 14158429 14158511 14164228 14164302 14164329 14164335 14164336 14165064 14165165 14173348 14173373 14183635 14183642 14183643 14183644 14183645 14183646 14183647 14183648 14183649 14183650 14183664 14183665 14183666 14183667 14183668 14183679 14183681 14183682 14183724 14183725 14183726 14183727 14183728 14183734 14183735 14183736 14183737 14183738 14183739 14183741 14183750 14183751 14183752 14183753 14183754 14183760 14183761 14183762 14183763 14183764 14183766 14183767 14183768 14183769 14183770 14183781 14183782 14183783 14183784 14183795 14183796 14183797 14183798 14183799 14183800 14183801 14183802 14183803 14183804 14183805 14183816 14183817 14183818 14183819 14183820 14183865 14183866 14183867 14183868 14183869 14183870 14183871 14183872 14183873 14183874 14183878 14183879 14183880 14183881 14183904 14183905 14183906 14183907 14183908 14183912 14183914 14183915 14183916 14183917 14183920 14183921 14183922 14183923 14183924 14183925 14183926 14183927 14183928 14183929 14183930 14183931 14183932 14183933 14183934 14183938 14183939 14183940 14183941 14183942 14183943 14183944 14183945 14183946 14183947 14183983 14183984 14183985 14183986 14183987 14183994 14183995 14183996 14183997 14183998 14183999 14184000 14184001 14184002 14184003 14184007 14184008 14184010 14184012 14184015 14184016 14184017 14184018 14184019 14184027 14184046 14184053 14184057 14184058 14184059 14184060 14184073 14184203 14184205 14184230 14184346 14184407 14184453 14184492 14184529 14184549 14184556 14184558 14184559 14184560 14184561 14184564 14184565 14184567 14184570 14184643 14184644 14184647 14184697 14184723 14184724 14184804 14184838 14184845 14184848 14184854 14184902 14184905 14184914 14184915 14184924 14184949 14184950 14184956 14184960 14184961 14184962 14184965 14187060 14187154 14187392 14187776 14187788 14188119 14188190 14201892 14202509 14202670 14202878 14202950 14216113 14216723 14218058 14218148 14219058 14219093 14219547 14220105 14220205 14220225 14221068 14221722 14222062 14236572 14236953 14237053 14237171 14237482 14237504 14237763 14238030 14238258 14238494 14238573 14238621 14238761 14239087 14239198 14239650 14239724 14239730 14239731 14239831 14239930 14240445 14240473 14240735 14241227 14241228 14241230 14241250 14241290 14241291 14241292 14241293 14241294 14241295 14241323 14241329 14241341 14241360 14241408 14241411 14241417 14241420 14241429 14241474 14241476 14241478 14241486 14241506 14241517 14241521 14241522 14241527 14241530 14241543 14241545 14241546 14241557 14241561 14241580 14241581 14241583 14241584 14241607 14241613 14241614 14241629 14241631 14241634 14241635 14241638 14241639 14241640 14241648 14241651 14241659 14241690 14241729 14241766 14241774 14241776 14241780 14241782 14241798 14241799 14241886 14241889 14241891 14241903 14241924 14241975 14241996 14241999 14242012 14242029 14242034 14242036 14242040 14242042 14242050 14242051 14242053 14242088 14242107 14242108 14242110 14242111 14242115 14242125 14242127 14242138 14242142 14242152 14242153 14242156 14242158 14242160 14242188 14242210 14242219 14242223 14242225 14242239 14242260 14242271 14242273 14242302 14242305 14242313 14242314 14242328 14242347 14242350 14242351 14242353 14242354 14242355 14242357 14242358 14242360 14242361 14242470 14242473 14242475 14242476 14242480 14242483 14242501 14242525 14242526 14242527 14242556 14242557 14242558 14242559 14242568 14242569 14242575 14242596 14242602 14242603 14242618 14242619 14242623 14242624 14242635 14242640 14242661 14242665 14242666 14242764 14242768 14242769 14242771 14242790 14242791 14242793 14242872 14242873 14242898 14242903 14242907 14242910 14242915 14242916 14242924 14242943 14242956 14242967 14242973 14242980 14242990 14242997 14242998 14242999 14243007 14243033 14243037 14243041 14243043 14243044 14243045 14243047 14243083 14243084 14243085 14243099 14243100 14243101 14243102 14243103 14243104 14243115 14243116 14243117 14243118 14243119 14243122 14243127 14243128 14243166 14243210 14243211 14243215 14243216 14243217 14243229 14243230 14243231 14243244 14243245 14243285 14243298 14243299 14243301 14243305 14243326 14243339 14243345 14243388 14243399 14243469 14244577 14244668 14244673 14245019 14245535 14246137 14246171 14246337 14246354 14246356 14247634 14247767 14248126 14249177 14249863 14249865 14249866 14249879 14249906 14249908 14250546 14250548 14250555 14250641 14250973 14254447 14272449 14272451 14272513 14272798 14272902 14272924 14272971 14273861 14276894 14277054 14278605 14278826 14280556 14281093 14281535 14281692 14282029 14282911 14283026 14283340 14283526 14283655 14283724 14283787 14284214 14284523 14284552 14285034 14285053 14285061 14285093 14285095 14285144 14285146 14285166 14285180 14285207 14285210 14285266 14285270 14285272 14285285 14285288 14285289 14285290 14285291 14285292 14285295 14285296 14285298 14285302 14285305 14285306 14285317 14285319 14285320 14285322 14285327 14285328 14285368 14285603 14285605 14285706 14285709 14285721 14285722 14285723 14285770 14285771 14285772 14285774 14285817 14285818 14285819 14285821 14285834 14285835 14285836 14285837 14285838 14285842 14285844 14285845 14285846 14285936 14285937 14285938 14285939 14285940 14285967 14286025 14286125 14286127 14286141 14286160 14286172 14286174 14286175 14286187 14286204 14286210 14286217 14286229 14286232 14286233 14286234 14286268 14286280 14286297 14286298 14286300 14286301 14286342 14286346 14286352 14286356 14286358 14286425 14286426 14286427 14286430 14286436 14286441 14286453 14286456 14286485 14286486 14286487 14286498 14286516 14286519 14286527 14286530 14286531 14286532 14286533 14286534 14286535 14286541 14286542 14286543 14286544 14286545 14286546 14286547 14286549 14286550 14286578 14286585 14286603 14286690 14286691 14286723 14286801 14286802 14286804 14286805 14286827 14286831 14286843 14286845 14286846 14286852 14286853 14286920 14286934 14286936 14286937 14286938 14286961 14286962 14286963 14286964 14286965 14287084 14287085 14287092 14287096 14287098 14287102 14287103 14287104 14287105 14287106 14287117 14287118 14287119 14287122 14287123 14287124 14287126 14287155 14287157 14287158 14287160 14287161 14287162 14287163 14287164 14287178 14287182 14287183 14287184 14287370 14287371 14287378 14287379 14287380 14287381 14287382 14287384 14287385 14287386 14287394 14287395 14287397 14287399 14287428 14287429 14287430 14287431 14287432 14287465 14287471 14287485 14287487 14287488 14287489 14287498 14287499 14287500 14287504 14287506 14287534 14287535 14287536 14287537 14287538 14287539 14287540 14287541 14287542 14287543 14287544 14287545 14287546 14287550 14287556 14287610 14287612 14287613 14287614 14287616 14287618 14287621 14287622 14287623 14287624 14287666 14287674 14287675 14287676 14287677 14287704 14287705 14287706 14287707 14287708 14287763 14287764 14287765 14287766 14287767 14287802 14287803 14287805 14287806 14287823 14287825 14287830 14287836 14287837 14287924 14287925 14287926 14287927 14287928 14287989 14287991 14287992 14287993 14287994 14287996 14288040 14288041 14288042 14288043 14288047 14288072 14288096 14288097 14288098 14288099 14288125 14288220 14288221 14288222 14288223 14288224 14288267 14288268 14288269 14288270 14288271 14288299 14288300 14288302 14288338 14288341 14288342 14288343 14288344 14288374 14288375 14288422 14288423 14288425 14288426 14288435 14288451 14288452 14288453 14288454 14288455 14288468 14288513 14288514 14288515 14288516 14288517 14288531 14288547 14288549 14288550 14288551 14288552 14288553 14288554 14288582 14288583 14288584 14288585 14288586 14288669 14288670 14288671 14288672 14288673 14288674 14288675 14288676 14288677 14288689 14288690 14288691 14288692 14288693 14288694 14288695 14288702 14288713 14288714 14288715 14288716 14288717 14288718 14288719 14288721 14288722 14288723 14288724 14288725 14288726 14288727 14288728 14288729 14288730 14288735 14288737 14288747 14288748 14288749 14288750 14288751 14288752 14288753 14288758 14288772 14288773 14288782 14288783 14288784 14288785 14288787 14288795 14288799 14288801 14288809 14288810 14288811 14288812 14288813 14288814 14288815 14288816 14288817 14288818 14288819 14288820 14288821 14288822 14288823 14288824 14288825 14288828 14288829 14288830 14288831 14288836 14288837 14288838 14288843 14288844 14288845 14288846 14288847 14288848 14288849 14288850 14288851 14288852 14288853 14288877 14288878 14288893 14288895 14288896 14288915 14288916 14288917 14288918 14288919 14288935 14288936 14288937 14288940 14288946 14288957 14288958 14288962 14288963 14288965 14288966 14288967 14288968 14288969 14288970 14289052 14289053 14289063 14289065 14289066 14289067 14289068 14289070 14289071 14289072 14289073 14289074 14289075 14289076 14289093 14289094 14289095 14289096 14289097 14289125 14289126 14289127 14289128 14289135 14289136 14289137 14289138 14289139 14289167 14289168 14289169 14289171 14289175 14289176 14289235 14289236 14289237 14289238 14289239 14289250 14289251 14289252 14289253 14289254 14289275 14289276 14289277 14289278 14289279 14289297 14289298 14289299 14289320 14289321 14289322 14289323 14289324 14289343 14289344 14289345 14289346 14289347 14289348 14289350 14289351 14289352 14289380 14289382 14289383 14289384 14289421 14289422 14289424 14289425 14289477 14289478 14289479 14289480 14289482 14289483 14289484 14289485 14289505 14289507 14289508 14289509 14289510 14289551 14289553 14289554 14289555 14289580 14289613 14289618 14289624 14289627 14289636 14289637 14289643 14289659 14289660 14289661 14289662 14289663 14289664 14289676 14289702 14289706 14289708 14289710 14289711 14289713 14289718 14289720 14289731 14289733 14289751 14289753 14289755 14289756 14289758 14289761 14289762 14289767 14289769 14289790 14289791 14289792 14289793 14289794 14289795 14289801 14289807 14289808 14289820 14289821 14289822 14289823 14289824 14289826 14289833 14289870 14289873 14289874 14289891 14289906 14289908 14289911 14289914 14289927 14289928 14289932 14289934 14289966 14289973 14289974 14289975 14289976 14289978 14289988 14289993 14289996 14289998 14289999 14290000 14290058 14290059 14290060 14290061 14290062 14290064 14290080 14290082 14290086 14290088 14290090 14290093 14290094 14290095 14290096 14290097 14290098 14290099 14290104 14290117 14290129 14290145 14290147 14290148 14290153 14290155 14290156 14290162 14290164 14290165 14290166 14290167 14290168 14290173 14290175 14290176 14290181 14290184 14290185 14290186 14290187 14290188 14290190 14290289 14290425 14290551 14290580 14290585 14290593 14290600 14290632 14290636 14290643 14290648 14290665 14290683 14290686 14290710 14290795 14290824 14290829 14290865 14290866 14290893 14290905 14290920 14290939 14290960 14290962 14290963 14290964 14290980 14291007 14291008 14291011 14291059 14291066 14291128 14291129 14291136 14291145 14291165 14291168 14291170 14291189 14291190 14291191 14291192 14291195 14291216 14291217 14291218 14291231 14291232 14291233 14291234 14291236 14291244 14291247 14291248 14291271 14291273 14291292 14291294 14291312 14291322 14291325 14291326 14291343 14291349 14291353 14291389 14291391 14291417 14291423 14291458 14291478 14291480 14291486 14291496 14291498 14291499 14291515 14291516 14291519 14291535 14291557 14291558 14291559 14291565 14291566 14291567 14291574 14291575 14291583 14291584 14291585 14291605 14291606 14291609 14291612 14291615 14291744 14291755 14291790 14291799 14292033 14292116 14295035 14296116 14297560 14298116 14298336 14298589 14298973 14298976 14298993 14299077 14299165 14299262 14299298 14299960 14299978 14300018 14300026 14300054 14302945 14303390 14303591 14303685 14303877 14303901 14303912 14304294 14305163 14305223 14305378 14305385 14305437 14305553 14305600 14305603 14305608 14305612 14305621 14305631 14305632 14305687 14305696 14305700 14305737 14305739 14305741 14305742 14305745 14305746 14305749 14305750 14305752 14305757 14305758 14305763 14305765 14305773 14305774 14305780 14305781 14305783 14305784 14305789 14305805 14305812 14305814 14305815 14305819 14305833 14305837 14305838 14305841 14305842 14305843 14305844 14305846 14305851 14305852 14305866 14305869 14305872 14305873 14305874 14305883 14305886 14305967 14306281 14307787 14307814 14307880 14307977 14307980 14307986 14307990 14308148 14308223 14308227 14308228 14308346 14308814 14308845 14308889 14308921 14308947 14308948 14308949 14308950 14308990 14308991 14309015 14309083 14309107 14309110 14309112 14309115 14309153 14309162 14309185 14309186 14309226 14309232 14309235 14309327 14309329 14309330 14309331 14309338 14309339 14309340 14309341 14309342 14309343 14309344 14309345 14309360 14309362 14309363 14309364 14309366 14309368 14309388 14309396 14309397 14309398 14309418 14309423 14309469 14309471 14309491 14309492 14310521 14310670 14311010 14311094 14311703 14311832 14311911 14311953 14312032 14312075 14312111 14312200 14312255 14312316 14312328 14312329 14312330 14312353 14312354 14312385 14312388 14312389 14312402 14312403 14312404 14312405 14312406 14312409 14312410 14312412 14312417 14312420 14312422 14312429 14312430 14312431 14312436 14312437 14312438 14312439 14312446 14312449 14312450 14312465 14312466 14312467 14312468 14312479 14312482 14312483 14312496 14312502 14312504 14312506 14312511 14312518 14312538 14312539 14312541 14312544 14312546 14312549 14312566 14312568 14312573 14312575 14312589 14312591 14312593 14312595 14312596 14312633 14312634 14312640 14312642 14312643 14312646 14312674 14312675 14312676 14312678 14312680 14312687 14312691 14312695 14312696 14312697 14312706 14312707 14312709 14312725 14312751 14312775 14312776 14312778 14312787 14312795 14312810 14312811 14312822 14312823 14312824 14312825 14312830 14312831 14312832 14312834 14312883 14312884 14312911 14312913 14312914 14312915 14312916 14312917 14312918 14312946 14312954 14312959 14312980 14313002 14313011 14313017 14313049 14313078 14313079 14313120 14313149 14313220 14313325 14313364 14313383 14313391 14313427 14313488 14313611 14313644 14313713 14313794 14313852 14313859 14313910 14313985 14313994 14313995 14313998 14314032 14314046 14314047 14314057 14314060 14314085 14314087 14314090 14314147 14314158 14314160 14314164 14314193 14314244 14314270 14314273 14314283 14314360 14314367 14314389 14314398 14314414 14314416 14314440 14314468 14314476 14314477 14314484 14314485 14314490 14314595 14314638 14314640 14314642 14314644 14314645 14314656 14314657 14314662 14314666 14314668 14314723 14314730 14314765 14314767 14314772 14314773 14314782 14314784 14314787 14314817 14314837 14314840 14314852 14314853 14314904 14314922 14314929 14314942 14314958 14314962 14314964 14315030 14315033 14315034 14315036 14315037 14315039 14315055 14315112 14315114 14315116 14315145 14315147 14315148 14315152 14315162 14315225 14315281 14315282 14315325 14315327 14315329 14315331 14315358 14315362 14315363 14315366 14315374 14315381 14315397 14315401 14315406 14315419 14315420 14315421 14315422 14315429 14315430 14315436 14315437 14315447 14315449 14315450 14315459 14315461 14315462 14315463 14315481 14315482 14315485 14315488 14315490 14315491 14315492 14315495 14315511 14315554 14315566 14315567 14315569 14315570 14315571 14315573 14315574 14315575 14315576 14315577 14315578 14315579 14315580 14315582 14315588 14315590 14315599 14315612 14315621 14315623 14315627 14315628 14315629 14315631 14315639 14315640 14315653 14315655 14315657 14315664 14315668 14315669 14315670 14315671 14315674 14315675 14315677 14315678 14315681 14315684 14315685 14315686 14315687 14315689 14315690 14315691 14315692 14315695 14315702 14315707 14315713 14315714 14315715 14315721 14315727 14315756 14315758 14315759 14315771 14315773 14315776 14315777 14315784 14315785 14315788 14315790 14315792 14315797 14315799 14315811 14315812 14315833 14315873 14315875 14315886 14315901 14315905 14315908 14315916 14315918 14315919 14315923 14315924 14315952 14315953 14315957 14315958 14315959 14315961 14315962 14315966 14315968 14315976 14315977 14315979 14315980 14315983 14315984 14315986 14315987 14315988 14315995 14315998 14316005 14316006 14316007 14316013 14316014 14316016 14316017 14316020 14316023 14316026 14316044 14316074 14316083 14316084 14316085 14316086 14316087 14316103 14316106 14316107 14316108 14316109 14316115 14316116 14316117 14316118 14316125 14316126 14316127 14316128 14316130 14316131 14316132 14316133 14316134 14316135 14316136 14316137 14316138 14316139 14316140 14316155 14316157 14316158 14316168 14316169 14316170 14316172 14316173 14316174 14316175 14316191 14316192 14316193 14316194 14316195 14316196 14316197 14316198 14316226 14316227 14316228 14316247 14316248 14316249 14316251 14316252 14316253 14316255 14316256 14316257 14316258 14316259 14316273 14316274 14316276 14316277 14316278 14316290 14316291 14316292 14316299 14316300 14316301 14316303 14316304 14316305 14316306 14316308 14316309 14316310 14316326 14316327 14316329 14316330 14316331 14316332 14316339 14316341 14316342 14316349 14316356 14316364 14316405 14316406 14316407 14316408 14316409 14316410 14316412 14316415 14316420 14316421 14316436 14316438 14316445 14316469 14316475 14316481 14316482 14316484 14316486 14316494 14316495 14316496 14316499 14316502 14316503 14316511 14316512 14316513 14316520 14316556 14316558 14316559 14316567 14316573 14316575 14316576 14316593 14316594 14316598 14316599 14316602 14316603 14316604 14316605 14316611 14316612 14316636 14316642 14316643 14316644 14316645 14316646 14316647 14316649 14316655 14316663 14316664 14316666 14316667 14316673 14316681 14316683 14316684 14316686 14316688 14316694 14316731 14316733 14316749 14316750 14316757 14316759 14316760 14316771 14316773 14316790 14316791 14316793 14316886 14316901 14316923 14316926 14316936 14316953 14316990 14316993 14316994 14317001 14317011 14317012 14317015 14317019 14317032 14317063 14317068 14317075 14317084 14317085 14317090 14317094 14317098 14317102 14317178 14317180 14317186 14317187 14317191 14317205 14317212 14317232 14317240 14317243 14317262 14317269 14317271 14317273 14317277 14317278 14317281 14317284 14317301 14317308 14317309 14317310 14317348 14317349 14317357 14317367 14317369 14317433 14317443 14317481 14317485 14317488 14317512 14317515 14317521 14317524 14317527 14317529 14317534 14317561 14317578 14317582 14317585 14317609 14317613 14317615 14317616 14317617 14317622 14317623 14317647 14317652 14317656 14317658 14317670 14317671 14317674 14317763 14317765 14317768 14317771 14317783 14317790 14317792 14317793 14317794 14317796 14317800 14317801 14317802 14317803 14317804 14317805 14317821 14317866 14317868 14317869 14317870 14317871 14317892 14317893 14317894 14317896 14317897 14317898 14317905 14317907 14317908 14317909 14317910 14317912 14317913 14317914 14317917 14317919 14317936 14317938 14317939 14317955 14317956 14317957 14317959 14317960 14317965 14317966 14317968 14317971 14317975 14317992 14317993 14317994 14318009 14318011 14318014 14318025 14318026 14318028 14318038 14318039 14318048 14318049 14318068 14318069 14318070 14318071 14318074 14318084 14318086 14318094 14318095 14318097 14318108 14318117 14318119 14318120 14318129 14318131 14318132 14318133 14318135 14318139 14318140 14318143 14318144 14318155 14318156 14318161 14318169 14318177 14318179 14318180 14318181 14318190 14318191 14318192 14318199 14318200 14318202 14318203 14318204 14318206 14318227 14318228 14318229 14318230 14318231 14318238 14318239 14318240 14318241 14318321 14318322 14318323 14318324 14318325 14318328 14318385 14318386 14318390 14318391 14318392 14318394 14318398 14318407 14318408 14318410 14318411 14318417 14318418 14318420 14318421 14318423 14318425 14318446 14318447 14318455 14318477 14318478 14318479 14318480 14318481 14318482 14318484 14318486 14318487 14318489 14318500 14318501 14318503 14318505 14318506 14318511 14318528 14318529 14318535 14318536 14318538 14318541 14318542 14318548 14318549 14318556 14318557 14318558 14318566 14318567 14318569 14318575 14318578 14318580 14318581 14318583 14318593 14318594 14318595 14318596 14318597 14318602 14318604 14318657 14318661 14318667 14318677 14318678 14318679 14318682 14318683 14318684 14318685 14318688 14318689 14318691 14318692 14318698 14318700 14318703 14318704 14318705 14318730 14318735 14318736 14318737 14318740 14318741 14318743 14318744 14318749 14318750 14318751 14318758 14318762 14318763 14318765 14318766 14318769 14318770 14318771 14318772 14318773 14318774 14318775 14318777 14318778 14318781 14318782 14318783 14318784 14318785 14318787 14318797 14318799 14318800 14318801 14318804 14318805 14318806 14318816 14318817 14318827 14318828 14318829 14318830 14318831 14318840 14318841 14318842 14318843 14320389 14320390 14320396 14320397 14320424 14320427 14320428 14320429 14320430 14320431 14320452 14320459 14320460 14320461 14320478 14320493 14320494 14320507 14320508 14320509 14320510 14320511 14320525 14320542 14320546 14320580 14320599 14320628 14320691 14320716 14320848 14320851 14321109 14321110 14321114 14321115 14321179 14321185 14321464 14321532 14321541 14321542 14321562 14321567 14321568 14321624 14321625 14321626 14321628 14321706 14321708 14321718 14321721 14321722 14321747 14321748 14321808 14321815 14321832 14321842 14321843 14321844 14321845 14321857 14321867 14321869 14321889 14321890 14321891 14321980 14321981 14322003 14322014 14322015 14322016 14322017 14322034 14322035 14322036 14322047 14322049 14322074 14322075 14322076 14322077 14322078 14322152 14322153 14322154 14322155 14322174 14322176 14322177 14322178 14322181 14322198 14322215 14322404 14323574 14324102 14324148 14324152 14324436 14324499 14324513 14324525 14324568 14324602 14324656 14324784 14324786 14324788 14324799 14324803 14324827 14324838 14324880 14324885 14324902 14324916 14324918 14324933 14325047 14325051 14325086 14325101 14325167 14325220 14325235 14325263 14325272 14325274 14325310 14325321 14325409 14325500 14325501 14325506 14325538 14325541 14325555 14325647 14325720 14325722 14325747 14325749 14325779 14325790 14325812 14325882 14326004 14326036 14326038 14326056 14326058 14326079 14326203 14326206 14326240 14326247 14326256 14326298 14326335 14326366 14327740 14328365 14328366 14328374 14328377 14328414 14328466 14328467 14328468 14328472 14328473 14328482 14328504 14328505 14328512 14328527 14328528 14328548 14328565 14328566 14328585 14328599 14328604 14328625 14328639 14328646 14328663 14328669 14328672 14328681 14328690 14328691 14328696 14328724 14328728 14328729 14328733 14328736 14328741 14328755 14328757 14328764 14328769 14328777 14328833 14328835 14328837 14328845 14328859 14328875 14328876 14328896 14328900 14328907 14328922 14328924 14328936 14328937 14328943 14328959 14328971 14328974 14328977 14328978 14328994 14328999 14329022 14329030 14329036 14329067 14329077 14329086 14329090 14329111 14329118 14329131 14329134 14329162 14329166 14329173 14329175 14329185 14329187 14329192 14329198 14329199 14329208 14329211 14329218 14329238 14329239 14329240 14329276 14329281 14329282 14329283 14329290 14329302 14329322 14329326 14329328 14329329 14329330 14329332 14329338 14329345 14329348 14329366 14329367 14329399 14329410 14329475 14329476 14329481 14329482 14329483 14329495 14329500 14329531 14329539 14329543 14329550 14329591 14329599 14329600 14329606 14329607 14329608 14329610 14329613 14329649 14329654 14329656 14329657 14329701 14329723 14329728 14329731 14329743 14329782 14329783 14329788 14329806 14329815 14329823 14329832 14329870 14329873 14329876 14329879 14329883 14329899 14329919 14329920 14329922 14329923 14329927 14329928 14329939 14329942 14329966 14329968 14329970 14329975 14329978 14330000 14330001 14330004 14330006 14330014 14330020 14330023 14330024 14330068 14330069 14330071 14330073 14330074 14330079 14330080 14330090 14330093 14330124 14330125 14330128 14330140 14330145 14330146 14330162 14330163 14330176 14330189 14330192 14330200 14330213 14330214 14330231 14330232 14330255 14330259 14330285 14330301 14330303 14330304 14330319 14330335 14330337 14330345 14330367 14330369 14330382 14330389 14330392 14330399 14330416 14330417 14330423 14330426 14330434 14330435 14330448 14330460 14330462 14330463 14330519 14330520 14330525 14330529 14330546 14330547 14330554 14330578 14330593 14330601 14330603 14330617 14330618 14330649 14330663 14330725 14330750 14330754 14330764 14330801 14330865 14330871 14330972 14330999 14331044 14331045 14331047 14331082 14331085 14331110 14331179 14331201 14331202 14331203 14331229 14331270 14331286 14331287 14331293 14331295 14331297 14331325 14331326 14331343 14331350 14331353 14331354 14331385 14331390 14331428 14331448 14331479 14331481 14331483 14331484 14331486 14331489 14331501 14331516 14331517 14331519 14331538 14331551 14331553 14331577 14331626 14331627 14331628 14331633 14331643 14331647 14331681 14331687 14331715 14331750 14331761 14331772 14331780 14331791 14331793 14331832 14331838 14331909 14331914 14331922 14331951 14331957 14332006 14332009 14332059 14332064 14332140 14332141 14332144 14332195 14332324 14332386 14332390 14332402 14332429 14332431 14332463 14332506 14332508 14332512 14332568 14332583 14332587 14332595 14332875 14332976 14333063 14333070 14333078 14333081 14333160 14333209 14333302 14333325 14333379 14333498 14333523 14333548 14333564 14333624 14333626 14333643 14333651 14333689 14333722 14333723 14333785 14334106 14334134 14334188 14334189 14334225 14334313 14334359 14334481 14334483 14334520 14334540 14334541 14334725 14334749 14334949 14334952 14334964 14334981 14335088 14335141 14335144 14335166 14335253 14335436 14335437 14335559 14335640 14335819 14335973 14336153 14336168 14336231 14336261 14336268 14336313 14336317 14336361 14336897 14337248 14337415 14340150 14340275 14340343 14340346 14340519 14340597 14340687 14340736 14340937 14341012 14341470 14341836 14342202 14342282 14342730 14342768 14343048 14343304 14343312 14343325 14343435 14343504 14343518 14343521 14343522 14343685 14343694 14343784 14343872 14343926 14344533 14344735 14345088 14345185 14345576 14345676 14346843 14346930 14346992 14347399 14347509 14352902 14352963 14352971 14352976 14352987 14353093 14353094 14353120 2000474 12418940 12419048 12419285 12419492 12419570 12419680 12419733 12419761 12419804 12419814 12419824 12419903 12419924 12430202 12430221 12430224 12430251 12430271 12430285 12430287 12430288 12430304 12430339 12430383 12430389 12430419 12430420 12434321 12434346 12434636 12434638 12434932 12484226 12484323 12490168 12491254 12491430 12491470 12491493 12491627 12492192 12492398 12492521 12504759 12507821 12508026 12508059 12508142 12508213 12508511 12508537 12508710 12508764 12508824 12508826 12508935 12509012 12509108 12509115 12509124 12509244 12509327 12509359 12510364 12510390 12510425 12510426 12510617 12510781 12510981 12511247 12511298 12511352 12511460 12511481 12511484 12511494 12511505 12511506 12511530 12511539 12511576 12511578 12511580 12511582 12511616 12511617 12511618 12511619 12511649 12511654 12511657 12511673 12511681 12511700 12511702 12511703 12511730 12511840 12511874 12511878 12511974 12511980 12512012 12512157 12512158 12512176 12512213 12512220 12512228 12512249 12512255 12512367 12512421 12512477 12512510 12512543 12512554 12512661 12512662 12512700 12512898 12512906 12513106 12513260 12513263 12513265 12513318 12513319 12513321 12513336 12513354 12513355 12513364 12513369 12513370 12513371 12513384 12513385 12513407 12513432 12513438 12513511 12513520 12513592 12513594 12513596 12513623 12513641 12513642 12513669 12513684 12513693 12513714 12513732 12513784 12513794 12513797 12513798 12513808 12513833 12513840 12514235 12514328 12514329 12514461 12514611 12514692 12514723 12514966 12515011 12515021 12515042 12515173 12515177 12515361 12515386 12515387 12515445 12515476 12515483 12515486 12515553 12515554 12515557 12515579 12515583 12515586 12515587 12515588 12515702 12515705 12515741 12515743 12515744 12515776 12515777 12515779 12515781 12515782 12515784 12515816 12515817 12515820 12515854 12515857 12515892 12515941 12515942 12515998 12516029 12516030 12516121 12516168 12516189 12516295 12516306 12516307 12516308 12516366 12516368 12516409 12516428 12516429 12516460 12516579 12516581 12516640 12516654 12516690 12516691 12516693 12516815 12516818 12516840 12516849 12516905 12516906 12516963 12516978 12516980 12516983 12517112 12517130 12517148 12517160 12517162 12517163 12517278 12517341 12517360 12517361 12522749 12523027 12523237 12523390 12523517 12523581 12523613 12523653 12523915 12524006 12524049 12524330 12524579 12525040 12525109 12525151 12525189 12525258 12525299 12525308 12525312 12525330 12525369 12525470 12525686 12525931 12525946 12525957 12526067 12526113 12526125 12526140 12526148 12526149 12526176 12526177 12526211 12526220 12526224 12526227 12526273 12526337 12526342 12526408 12526533 12526579 12526612 12526613 12526711 12526719 12526744 12526746 12526759 12526799 12526826 12526832 12526833 12526847 12526865 12526892 12526901 12526903 12526987 12527003 12527005 12527024 12527025 12527047 12527052 12527073 12527094 12527134 12531577 12543593 12561986 12564545 12567805 12567959 12567987 12567992 12569587 12583249 12583276 12584855 12585566 12588910 12597115 12597243 12601712 12602806 12602850 12602994 12603182 12603201 12603285 12603354 12603363 12603410 12603412 12603414 12603419 12629910 12630434 12630572 12631359 12631613 12631629 12631634 12631709 12631778 12632006 12634498 12634499 12634627 12634921 12635219 12635465 12635727 12636064 12636115 12636521 12637083 12641117 12650638 12650690 12650696 12650698 12650706 12650719 12650735 12650770 12650801 12650929 12651005 12651018 12651043 12651044 12651061 12651416 12651844 12651849 12651872 12654518 12654554 12655083 12655106 12655474 12656009 12656195 12656201 12656222 12656421 12658619 12662027 12663761 12663799 12663813 12663875 12663993 12664019 12664025 12664095 12664138 12664179 12664232 12664233 12664279 12664288 12664330 12664331 12664395 12664405 12664431 12664446 12664449 12664482 12664493 12664499 12664517 12664581 12664582 12664587 12664600 12664683 12664685 12664705 12664740 12664789 12664888 12664895 12664902 12664905 12664967 12665050 12665118 12665282 12665289 12665334 12665335 12665336 12665350 12665515 12665519 12665520 12665554 12665782 12665811 12665827 12667883 12669840 12670042 12670197 12670379 12670512 12670600 12670647 12671132 12671477 12671555 12671595 12671859 12672066 12674031 12674425 12674427 12674440 12674500 12674502 12674520 12674529 12674544 12674700 12674847 12674849 12674886 12674889 12674896 12674899 12674902 12675041 12675061 12675073 12675074 12675087 12675088 12675098 12675106 12675125 12675141 12675145 12675149 12675153 12675156 12675172 12675173 12675184 12675186 12675188 12675190 12675198 12675215 12675236 12675285 12675302 12675317 12675336 12675359 12675361 12675362 12675370 12675380 12675386 12675392 12675407 12675409 12675419 12675426 12675428 12675431 12675432 12675435 12675442 12675445 12675449 12675450 12675506 12675556 12675557 12675561 12675592 12675595 12675647 12675648 12675649 12675651 12675698 12675699 12675700 12675742 12675744 12675885 12675887 12675889 12675909 12675970 12675998 12676008 12676013 12676019 12676020 12676032 12676034 12676087 12676093 12676110 12676111 12676127 12676142 12676143 12676152 12676153 12676154 12676155 12676161 12676170 12676177 12676181 12676182 12676185 12676207 12676209 12676210 12676213 12676220 12676221 12676233 12676240 12676241 12676246 12676252 12676253 12676258 12676259 12676261 12676263 12676264 12676276 12676277 12676279 12676284 12676291 12676304 12676308 12676321 12676330 12676333 12676338 12676342 12676357 12676358 12676360 12676363 12676364 12676379 12676380 12676385 12676386 12676391 12676392 12676394 12676397 12676398 12676400 12676403 12676404 12676406 12676407 12676418 12676430 12676440 12676442 12676553 12677433 12677454 12677509 12677517 12677518 12677519 12677520 12678086 12678103 12678136 12678137 12678138 12678140 12678141 12678143 12678144 12678146 12678149 12678150 12678165 12678166 12678171 12678174 12678176 12678181 12678182 12678195 12678196 12678201 12678202 12678213 12678216 12678217 12678218 12678219 12678220 12678236 12678307 12678498 12678539 12678571 12678588 12678595 12678598 12678599 12678600 12678606 12678607 12678608 12678625 12678627 12678630 12678633 12678634 12678637 12678638 12678911 12679903 12679942 12682622 12683056 12683929 12683942 12683994 12685155 12688809 12708986 12708993 12709003 12709006 12709018 12709019 12709020 12709027 12709045 12709816 12709820 12709821 12709822 12709839 12709859 12709860 12709864 12709865 12709866 12709867 12709870 12709872 12709890 12709892 12709982 12709986 12709987 12709988 12709991 12709995 12710001 12710003 12710004 12710005 12710006 12710007 12710017 12710019 12710027 12710031 12710061 12710087 12710088 12710089 12710090 12710091 12710092 12710095 12710097 12710098 12710099 12710100 12710101 12710102 12710121 12710123 12710128 12710162 12710163 12710164 12710165 12710166 12710168 12710169 12710170 12710171 12710181 12710182 12710184 12710185 12710186 12710193 12710201 12710203 12710204 12710205 12710206 12710221 12710282 12710284 12710285 12710286 12710287 12710288 12710289 12710290 12710312 12710313 12710314 12710317 12710318 12710319 12710320 12710360 12710367 12713284 12713317 12713343 12713423 12713439 12713467 12713483 12713537 12713567 12713569 12713627 12713812 12724648 12725546 12725755 12726712 12726774 12728668 12728920 12729104 12729787 12735548 12735549 12735572 12735607 12735702 12735709 12735753 12736177 12736269 12736511 12736591 12736607 12736674 12736814 12745784 12750722 12751167 12751470 12752270 12752357 12752385 12760364 12765692 12767554 12797593 12840629 12841116 12841366 12842137 12842262 12842360 12842585 12843045 12843075 12843630 12843650 12843743 12844211 12844256 12844391 12844434 12844515 12844603 12844629 12844695 12844759 12844802 12844811 12844822 12844828 12844884 12844898 12844921 12844922 12845018 12845038 12845061 12845063 12845114 12845115 12845131 12845132 12845133 12845156 12845160 12845192 12845195 12845255 12845256 12845325 12845364 12845424 12845457 12845469 12845515 12845519 12845602 12845740 12845750 12845760 12845761 12845762 12845797 12845802 12845815 12845829 12845844 12845859 12845872 12845876 12845879 12845885 12845888 12845889 12845891 12845897 12845900 12845924 12845960 12845972 12845973 12845978 12846002 12846062 12846070 12846081 12846133 12846145 12846147 12846155 12846187 12846204 12846214 12846261 12846279 12846300 12846307 12846317 12846338 12846340 12846349 12846375 12846470 12846479 12846504 12846537 12846557 12846575 12846576 12846596 12846617 12846662 12846670 12846672 12846673 12846689 12846694 12846709 12846742 12846744 12846756 12846757 12846758 12846759 12846765 12847617 12848094 12848104 12848123 12848174 12848188 12848239 12848265 12848268 12848283 12848305 12848320 12848343 12848357 12848366 12848369 12848398 12848399 12848409 12848420 12848421 12848423 12853633 12854030 12855145 12856815 12856880 12856946 12856965 12857098 12857166 12857176 12857202 12857225 12857286 12857489 12857504 12857665 12857745 12857783 12857873 12858278 12860424 12860477 12860515 12860548 12860581 12860729 12860759 12861049 12861058 12861204 12861860 12862113 12862531 12867422 12867557 12867566 12867577 12867690 12875577 12882583 12888009 12889244 12921261 12921657 12921709 12935232 12936208 12946267 12946385 12946472 12946538 12946769 12946805 12946924 12949366 12952356 12952368 12953424 12958836 12958870 12958941 12958960 12958964 12958967 12958982 12959012 12959016 12959044 12959134 12961854 12962024 12985966 12985974 12986219 12986320 12986616 12987548 12987959 12987966 12987983 12988025 12988112 12988138 12988150 12988298 12988324 12988348 12988454 12988590 12988620 12988652 12989025 12989031 12989057 12989268 12989789 12989848 12994088 12994205 12994303 12994422 12994534 12994577 12994643 12994806 12994819 12994842 12997668 12997994 12998563 13005557 13005585 13005600 13005614 13005630 13005640 13005686 13005691 13005710 13006030 13006061 13020971 13021047 13021586 13021618 13021629 13021802 13030431 13038487 13040204 13040299 13040362 13040514 13040538 13040559 13040572 13040668 13040754 13040776 13040920 13040939 13041089 13041323 13046425 13046737 13066421 13067161 13085045 13089796 13090166 13090191 13094950 13094965 13095223 13095301 13095322 13095353 13095414 13095423 13095607 13096005 13119113 13119615 13131323 13143818 13144033 13144214 13146148 13170695 13171336 13208165 13376909 13391392 13403459 13404464 13405216 13410449 13410717 13503626 13505041 13530247 13532335 13568485 13611820 13701339 13720507 13724686 13724970 13729417 13729456 13729507 13729682 13729741 13729769 13729834 13729835 13729848 13729854 13729915 13729930 13729977 13729978 13730031 13730085 13730090 13730151 13730211 13730240 13730267 13730389 13730398 13730480 13730481 13730734 13730736 13730739 13730764 13730786 13730797 13730870 13730950 13730951 13730952 13730970 13731006 13731009 13746009 13749280 13752144 13756742 13759060 13759119 13761311 13761372 13761443 13761452 13761507 13761509 13761548 13761559 13761581 13761591 13761609 13761639 13761649 13761651 13761667 13761674 13761675 13761678 13761683 13761686 13761687 13761696 13761705 13761714 13761718 13761725 13761803 13761831 13761873 13761877 13761880 13761881 13761907 13761908 13761945 13761947 13761948 13761955 13761985 13767634 13768254 13768311 13768426 13768429 13768728 13768737 13768752 13768774 13768838 13768875 13768915 13768979 13769074 13769148 13769197 13796919 13797017 13797038 13797112 13797271 13797496 13797675 13797707 13797777 13797789 13798454 13798633 13798764 13798928 13798989 13798995 13798996 13798998 13799018 13799063 13799134 13799135 13799136 13799141 13799178 13799222 13799551 13799570 13799657 13799857 13800185 13800292 13800440 13800737 13801609 13801952 13842293 13842356 13842387 13842750 13842807 13842821 13842941 13842953 13842954 13842958 13842963 13859538 13866463 13878350 13890902 13891727 13891728 13892107 13892116 13892188 13892549 13892554 13892572 13892586 13892606 13892632 13892643 13892649 13892699 13892734 13892757 13892798 13893243 13893321 13893394 13893544 13893560 13893647 13893675 13893677 13893701 13893703 13893705 13893747 13893751 13893821 13893822 13893823 13893882 13893937 13894235 13894329 13894394 13894457 13894851 13894862 13894863 13894872 13895045 13895169 13895242 13895296 13895302 13895315 13895392 13895393 13895394 13895419 13895421 13895429 13895441 13895445 13895446 13895450 13895474 13895477 13895480 13895498 13895503 13895506 13895507 13895522 13895559 13895565 13895581 13895612 13895616 13895617 13895618 13895672 13895732 13895734 13895740 13895745 13895747 13895749 13895758 13895759 13895760 13895761 13895762 13895763 13895764 13895765 13895780 13895783 13895784 13895810 13895824 13895830 13895834 13895837 13895841 13895846 13895848 13895849 13895863 13895867 13895871 13895879 13895885 13895887 13895888 13895890 13895892 13895893 13895894 13895895 13895899 13895901 13895902 13895903 13895904 13895905 13895906 13895914 13895916 13895917 13895918 13895919 13895920 13895922 13895927 13895935 13895938 13895940 13895941 13895942 13895949 13895952 13895953 13895955 13895956 13895957 13895958 13895959 13895960 13895961 13895964 13895965 13895974 13895975 13895976 13895977 13895978 13895979 13895980 13895981 13895982 13895983 13895991 13895994 13895995 13895996 13895997 13895999 13896000 13896002 13896003 13896006 13896007 13896018 13896019 13896020 13896021 13896024 13896025 13896027 13896028 13896029 13896030 13896031 13896032 13896033 13896034 13896040 13896045 13896046 13896048 13896050 13896051 13896052 13896055 13896056 13896060 13896061 13896062 13896064 13896075 13896098 13896140 13896141 13896144 13896146 13896147 13896148 13896150 13896153 13896154 13896159 13896163 13896164 13896165 13896166 13896167 13896168 13896170 13896172 13896173 13896174 13896175 13896177 13896178 13896182 13896186 13896187 13896188 13896189 13896191 13896194 13896195 13896196 13896197 13896198 13896202 13896203 13896204 13896205 13896206 13896207 13896208 13896215 13896216 13896218 13896219 13896220 13896221 13896222 13896227 13896232 13896233 13896234 13896235 13896236 13896237 13896238 13896239 13896246 13896247 13896248 13896250 13896251 13896252 13896260 13896261 13896263 13896264 13896265 13896270 13896272 13896273 13896274 13896275 13896276 13896277 13896281 13896282 13896284 13896285 13896287 13896288 13896289 13896290 13896292 13896293 13896295 13896296 13896297 13896298 13896299 13896301 13896302 13896303 13896328 13896329 13896330 13896331 13896332 13896333 13896336 13896337 13896342 13896343 13896344 13896349 13896350 13896353 13896354 13896355 13896356 13896357 13896358 13896359 13896363 13896364 13896368 13896369 13896370 13896375 13896376 13896377 13896378 13896380 13896381 13896382 13896384 13896385 13896386 13896391 13896393 13896397 13896399 13896400 13896402 13896403 13896404 13896405 13896406 13896408 13896410 13896458 13896459 13896460 13896464 13896465 13896466 13896467 13896471 13896473 13896478 13896514 13896522 13896525 13896527 13896536 13896584 13896628 13896629 13896635 13896636 13896651 13896657 13896676 13896677 13896679 13896794 13900398 13900401 13900432 13900433 13900434 13900435 13900442 13900443 13900448 13900450 13900451 13900452 13900453 13900456 13900465 13900466 13900467 13900468 13900487 13900489 13900491 13900492 13900531 13900542 13900559 13900560 13900561 13900572 13900575 13900582 13900583 13900597 13900598 13900600 13900614 13900615 13900616 13900617 13900624 13900643 13900644 13900645 13900646 13900647 13900648 13900656 13900686 13900688 13900689 13900690 13900691 13900693 13900710 13900711 13900727 13900967 13901409 13901894 13902626 13903684 13903707 13903712 13903714 13903772 13903773 13903786 13903795 13903848 13903850 13903910 13903936 13904307 13904310 13904327 13904426 13904440 13904442 13904443 13904444 13904445 13904446 13904466 13904472 13904473 13904477 13904480 13904492 13904493 13904510 13904520 13904521 13904522 13904523 13904538 13904539 13904540 13904542 13904543 13904544 13904545 13904546 13904547 13904576 13904585 13904587 13904589 13904591 13904593 13904609 13904610 13904611 13904618 13904625 13904626 13904640 13904641 13904642 13904655 13904657 13904659 13904661 13904663 13904684 13904685 13904686 13904687 13904689 13904691 13904692 13904694 13904715 13904716 13904719 13904721 13904722 13904723 13904725 13904726 13904734 13904735 13904736 13904745 13904748 13904756 13904757 13904760 13904761 13904762 13904763 13904764 13904767 13904787 13904789 13904798 13904802 13904829 13904842 13904843 13904856 13904858 13904859 13904861 13904862 13904872 13904875 13904883 13904888 13904889 13904895 13904896 13904899 13904906 13904907 13904908 13904918 13904919 13904920 13904921 13904922 13904950 13904951 13904953 13904962 13904963 13904964 13904965 13904980 13904981 13905000 13905003 13905006 13905008 13905028 13905036 13905039 13905040 13905041 13905050 13905051 13905052 13905066 13905074 13905101 13905102 13905104 13905109 13905115 13905198 13905209 13905210 13905211 13905215 13905218 13905219 13905234 13905254 13905280 13905305 13905313 13905331 13905332 13905339 13905352 13905375 13905378 13905380 13905381 13905384 13905400 13905407 13905409 13905410 13905412 13905413 13905415 13905416 13905417 13905419 13905431 13905435 13905441 13905458 13905459 13905473 13905475 13905482 13905487 13905506 13905512 13905514 13905516 13905531 13905563 13905565 13905566 13905569 13905570 13905572 13905585 13905588 13905589 13905599 13905601 13905631 13905632 13905633 13905635 13905636 13905652 13905653 13905656 13905657 13905668 13905669 13905674 13905682 13905698 13905700 13905702 13905703 13905705 13905736 13905742 13905771 13905772 13905774 13905775 13905776 13905782 13905783 13905810 13905834 13905853 13905862 13905863 13905877 13905878 13905883 13905886 13905894 13905939 13905982 13905983 13905995 13905997 13905998 13906001 13906003 13906021 13906028 13906034 13906077 13906134 13906140 13906154 13906170 13906175 13906177 13906191 13906195 13906200 13906227 13906232 13906234 13906249 13906272 13906274 13906286 13906296 13906299 13906310 13906323 13906408 13906410 13906417 13906421 13906448 13906456 13906471 13906473 13906490 13906491 13906507 13906530 13906547 13906568 13906579 13906592 13906613 13906633 13906637 13906638 13906646 13906664 13906680 13906704 13906705 13906717 13906718 13906719 13906724 13906726 13906727 13906728 13906736 13906737 13906743 13906745 13906747 13906754 13906755 13906756 13906757 13906762 13906766 13906777 13906790 13906793 13906816 13906825 13906827 13906828 13906832 13906835 13906836 13906837 13906862 13906897 13906996 13907220 13907299 13907351 13907352 13907479 13907482 13907607 13907694 13907698 13907763 13907814 13907845 13907885 13907941 13907942 13908322 13908327 13908385 13908422 13908458 13908468 13908479 13908547 13908619 13908694 13908697 13909509 13909825 13909991 13910683 13910700 13910711 13910721 13910730 13910736 13910742 13910749 13910751 13910766 13910767 13910782 13910787 13910788 13910806 13910823 13910863 13910865 13910879 13910881 13910882 13910883 13910892 13910961 13910963 13910965 13910967 13910968 13910970 13910972 13910991 13910998 13911013 13911015 13911026 13911033 13911059 13911082 13911085 13911103 13911189 13911206 13911231 13911232 13911238 13911264 13911318 13911320 13911337 13911349 13911352 13911353 13911380 13911381 13911400 13911421 13911423 13911426 13911439 13911446 13911447 13911449 13911456 13911458 13911465 13911467 13911468 13911473 13911477 13911483 13911484 13911490 13911499 13911514 13911523 13911545 13911556 13911558 13911584 13911587 13911646 13911655 13911666 13911702 13911747 13911765 13911800 13911805 13911822 13911826 13911831 13911875 13911914 13911926 13911937 13911948 13911950 13912022 13912127 13912144 13912159 13912160 13912279 13912407 13912590 13912630 13912640 13912690 13912744 13912795 13912865 13912868 13912886 13912894 13912934 13912935 13912937 13912938 13912982 13912997 13913023 13913027 13913033 13913053 13913057 13913058 13913061 13913088 13913096 13913098 13913119 13913131 13913132 13913147 13913153 13913157 13913164 13913191 13913192 13913195 13913208 13913222 13913223 13913232 13913240 13913247 13913248 13913250 13913251 13913252 13913258 13913265 13913267 13913268 13913271 13913275 13913287 13913288 13913289 13913297 13913299 13913304 13913308 13913311 13913312 13913313 13913315 13913316 13913321 13913322 13913337 13913342 13913365 13913383 13913384 13913389 13913391 13913396 13913398 13913417 13913418 13913438 13913440 13913445 13913456 13913462 13913474 13913476 13913478 13913482 13913571 13913573 13913579 13913580 13913585 13913611 13913623 13913624 13913635 13913638 13913639 13913655 13913674 13913675 13913679 13913699 13913703 13913704 13913706 13913712 13913713 13913728 13913741 13913759 13913766 13913785 13913788 13913793 13913794 13913796 13913822 13913828 13913829 13913837 13913897 13913911 13913961 13913967 13913974 13914016 13914051 13914120 13914123 13914169 13914170 13914182 13914183 13914200 13914209 13914217 13914250 13914349 13914381 13914388 13914391 13914393 13914404 13914464 13914466 13914473 13914588 13914589 13914618 13914619 13914624 13914634 13914637 13914644 13914645 13914646 13914650 13914651 13914666 13914707 13914729 13914739 13914740 13914742 13914751 13914752 13914795 13914796 13914801 13914830 13914842 13914847 13914949 13914952 13914987 13915143 13915190 13924199 13924477 13924994 13925642 13925791 13925804 13925934 13926897 13926905 13926969 13926972 13926996 14056942 14059737 14068338 14068515 14068792 14069341 14069376 14069561 14069893 14070605 14093968 14094235 14146202 14146249 14147908 14147923 14148316 14148333 14148335 14148354 14148361 14148399 14148413 14148414 14148421 14148423 14148435 14148457 14148460 14148469 14148560 14148568 14148601 14148602 14148608 14148612 14148624 14148678 14148689 14148694 14148792 14148835 14148846 14148848 14148922 14148947 14148953 14148985 14148989 14148993 14149048 14149075 14149083 14149103 14149106 14149131 14149169 14149185 14149193 14149199 14149221 14149289 14149400 14149741 14149765 14149810 14149876 14149888 14095852 14096233 14350323 14350346 14350759 14352737 14352764 14352789 14352795 14352800 14352831 14352853 14352899 14352914 14352920 14352927 14352964 14352981 14352988 14352989 14352990 14353044 14353068 14353095 14353123 14353145 14353151 14353152 14353157 14354152 |
Recalling Firm/ Manufacturer |
CareFusion 303, Inc. 10020 Pacific Mesa Blvd San Diego CA 92121-4386
|
For Additional Information Contact |
858-617-2000
|
Manufacturer Reason for Recall |
An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.
|
FDA Determined Cause 2 |
Software design |
Action |
CareFusion sent an Urgent Medical Device Recall Notification letter dated May 13, 2015 to customers. The letter identified the affected product, problem and the recommended actions to be taken. Customers are informed that CareFusion has released a software upgrade to address this issue. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the software upgrade installation. Customers with recall related questions are instructed to contact CareFusion Support Center at (888) 562-6018, 7am-4pm PT, supportcenter@carefusion.com. Customers with adverse event report questions are instructed to contact customer advocacy at (888) 812-3266, 24 hrs/day, 7 days a week, customerfeedback@carefusion.com. Customers with technical questions regarding the Alaris System are instructed to contact Technical Support at (888) 812-3229, 6am-5pm PT, Mon-Fri, dl-us-inf-techsupport@carefusion.com. |
Quantity in Commerce |
63 units |
Distribution |
US Nationwide Distribution |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = CARDINAL HEALTH, ALARIS PRODUCTS
|