Date Initiated by Firm |
January 15, 2016 |
Date Posted |
February 16, 2016 |
Recall Status1 |
Terminated 3 on May 08, 2017 |
Recall Number |
Z-0814-2016 |
Recall Event ID |
73093 |
Product Classification |
Cabinet, table and tray, anesthesia - Product Code BRY
|
Product |
Jadak Barcode Scanner, Part No. 350587-32, 350587-41, 351377-32.
Product Usage: An accessory for the Pyxis Anesthesia ES system and Pyxis Anesthesia system 4000 stations. The Jadak Barcode Scanner is used to scan medication labels during refill of the Pyxis Anesthesia ES system and Pyxis Anesthesia system 4000 stations and dispensing medication for patients. |
Code Information |
14488129 14488133 14488134 14463683 14463684 14463687 14463688 14463689 14463690 14463691 14463692 14463693 14463694 14463695 14463696 14463697 14463698 14463700 14463701 14463702 14463703 14463704 14463707 14463708 14463709 14463710 14463711 14463712 14463713 14463714 14463715 14463716 14463717 14467646 14467647 14513579 13941887 13941888 13941889 13941890 13941891 13941892 13941893 13941894 13941895 13941896 13941897 13941898 13941899 13943819 13943820 13972753 14474968 14510555 14510556 14510165 14510164 14510494 14510495 14510557 14510559 14510561 14510563 14510564 14510565 14510569 14510570 14510571 14510572 14454894 14477691 14477692 14477693 14477694 14477695 14477696 14477697 14477698 14477699 14477700 14477701 14477702 14477703 14477704 14477705 14495537 14387081 14387087 14387088 14387089 14387090 14387091 14387093 14387094 14387095 14387098 14387099 14387100 14387101 14387102 14387103 14387104 14387105 14387106 14387107 14405257 14405258 13966765 13966766 13966767 13966768 13966769 13966770 13966771 13966772 13966773 13966774 14398441 14398442 14398443 14398446 14398447 14398448 14398449 14398450 14398451 14398453 14398455 14398456 14398457 14398458 14398460 14398461 14398462 14398463 14398465 14398466 14398467 14398468 14469978 14475143 14440312 14440314 14440315 14440317 14232880 14232881 14232884 14232890 14232892 14250969 14043833 14043835 14043837 14043839 14463731 14463732 14463734 14463735 14463736 14398404 14423374 14423375 14423376 14423377 14423379 14423380 14423381 14423382 14423383 14423384 14423385 14423386 14423388 14423389 14423390 14423391 14423392 14423393 14423394 14423395 14423396 14423397 14423400 14423401 14423402 14423403 14423404 14232847 14232850 14232856 14232858 14232861 14232862 14232864 14232865 14232866 14232867 14232869 14232870 14232871 14232873 14232874 14232916 14248133 14248134 14248135 14248136 14248137 14248138 14248139 14248140 14248141 14248142 14248143 14248144 14248145 14248146 14248147 14390571 14390573 14390574 14390576 14390577 14189659 14189660 14189661 14189662 14189663 14189664 14189665 14189666 14189667 14189668 14189669 14189670 14189674 14189677 14189678 14067380 14067381 14067383 14067385 14067386 14067387 14067388 14192820 14192821 14192822 14192823 14197970 14197971 14197972 14448398 14448399 14163235 14163236 14163237 14163239 14163240 14163241 14163242 14163243 14163244 14163245 14163246 14163248 14184864 14184865 14184868 14184871 14184873 14472945 14472946 14472948 14512641 14160600 14160601 14160602 14160603 14160604 14160605 14160606 14160607 14160608 14160609 14160610 14160611 14160612 14160613 14470667 14055456 14055457 14055459 14055462 14440342 14440344 14440345 14440346 14440347 14440348 14440349 14440351 14441520 14029004 14029005 14029006 14029007 14029008 14489747 14489748 14489749 14489751 14489752 14489753 14489754 14489755 14489757 14489758 14489759 14489760 14113578 14127073 14127074 14127075 14127076 14127077 14127078 14127079 14127080 14127081 14127082 14127083 14127084 14127085 14127086 14423434 14423435 14423436 14423437 14423438 14423439 14423440 14423441 14423442 14423443 14423444 14262702 14262703 14262704 14262707 14262709 14262712 14262713 14155679 14155680 14155681 14155682 14493386 14493387 14493388 14493389 14493390 14493391 14513562 14513564 13943167 13943168 13943169 13943170 13956952 13961865 13961868 13961869 13961873 13961875 13961877 13961878 13961879 13961880 13961881 13961882 13995344 14282544 14398487 14388207 14388208 14388217 14388218 14388219 14388221 14388222 14388223 14388224 14388226 14388227 14388229 14475146 14475147 14475148 14475149 14475150 14475151 14475152 14475153 14475154 14475156 14475157 14475158 14475159 14475160 14475161 14475163 14475164 14475165 14503110 14503111 14503112 14503114 14503117 14180020 14194891 14194892 14194894 14194899 14451616 14451617 14451618 14451619 14151753 14151755 14151756 14151757 14151758 14151759 14451640 14451641 14479837 14479839 14479840 14479841 14479842 14479843 14479844 14479845 14479847 14479848 14479849 14479850 14479851 14479852 14479853 14479854 14479855 14479856 14479857 14479858 14479859 14479860 14479861 14479862 14479863 14479864 14479865 14479866 14493734 14490614 14490615 14490616 14490617 14490618 14490620 14490621 14490626 14490628 14490630 14493732 14493733 14354746 14354747 14354748 14354749 14354750 14354751 14354752 14354753 14354754 14354755 14428899 14428900 14428901 14428902 14428903 14221472 14221475 14221477 14221478 14221481 14221485 14221488 14221489 14221490 14221491 14221492 14221493 14221494 14221495 14221496 14232918 14262819 14262820 14262821 14262822 14262823 14262824 14262825 14262826 14262827 14262828 14262829 14502179 14480512 14480514 14480515 14480516 14480517 14480519 14444172 14444174 14463718 14463719 14463720 14463721 14463722 14463723 14463725 14463726 14463727 14463728 14463729 14463730 14071093 14071095 14472885 14472886 14472887 14472888 14472889 14472890 14472891 14472892 14472893 14472894 14472896 14472897 14493392 14227454 14227455 14262900 14262901 14262902 14262903 14262904 14262905 14404249 14404250 14502181 14448391 14448392 14448393 14448394 14448397 14387345 14387346 14387349 14387351 14387352 14262863 14262864 14262865 14262866 14262867 14262868 14262869 14262870 14262874 14262875 14262877 14262881 14262885 14262886 14262887 14262888 14262889 14298638 14298640 14119039 14119040 14119041 14119043 14119044 14119045 14119046 14384456 14384457 14063052 14063053 14063054 14063055 14063056 14063057 14063058 14063059 14063060 14063061 14063062 14063063 14063064 14063065 14063066 14063067 14063068 14063069 14063070 14063071 14063072 14063073 14063074 14063075 14063076 14063077 14063078 14063079 14043796 14043800 14043803 14043806 14043807 14043808 14043809 14043810 14043812 14043813 14043814 14043815 14043816 14146607 14146608 14160637 14160639 14160640 14160641 14160642 14160643 14160644 14160645 14160646 14160647 14160648 14160649 14160650 14160651 14160652 14160653 14160654 14160655 14160656 14160658 14160659 14160660 14160661 14160662 14160663 14160664 14160665 14160666 14160667 14183130 14183131 14183132 14183133 14183134 14183135 14183136 14183137 14183138 14183139 14183140 14183141 14183142 14183143 14183144 14183145 14183146 14183147 14183148 14183149 14183150 14183151 14191750 14201967 14173165 14173166 14173167 14173168 14173169 14173170 14175292 14156454 14156456 14209762 14438053 14454499 14428861 14428862 14428863 14160614 14160616 14160617 14160619 14160620 14160621 14279798 14279799 14279800 14279801 14279802 14279805 14279809 13995379 13995380 13995381 13995383 14282507 14045085 14504286 14504287 14504288 14504290 14504291 14504292 14504293 14504294 14504295 14504296 14504297 14504299 14504300 14504301 14504302 14504303 14504304 14384251 14140114 14145005 14145007 14145008 14145009 14145010 14145011 14145012 14145013 14145014 14145015 14145017 14145019 14145021 14145022 14145023 14318062 14197945 14236729 14419480 14419481 14419482 14442346 14442347 14480271 14480272 14480273 14480274 14480275 14405260 14487111 14487112 14487113 14487114 14487115 14487117 14487118 14486068 14486069 14486070 14486071 14486072 14486073 14486074 14486075 14486076 14487085 14487088 14487091 14487092 14487095 14487096 14487097 14487098 14487099 14487101 14487102 14487103 14487104 14487105 14487106 14487108 14419243 14419244 14419247 14419249 14419250 14419251 14480290 14412333 14412334 14214232 14214233 14214234 14144330 14144331 14144332 14144333 14144335 14144336 14144337 14144338 14144339 14144340 14144341 14144342 14144343 14478956 14478957 14289050 14444151 14444152 14444153 14281886 14066674 14066675 14066676 14066677 14066678 14066679 14066680 14221512 14221515 14221516 14221517 14221518 14221519 14232876 14232877 14232878 14232879 14448862 14448863 14448864 14448865 14398397 14398398 14398399 14398400 14017168 14017169 14017170 14017171 14017172 14017173 14017174 14017175 14017176 14017177 14017178 14017179 14053839 14053840 14053841 14053842 14487109 14487110 14487119 14053815 14053816 14053817 14053819 14053821 14053823 14053824 14053825 14053826 14053827 14053829 14053831 14053832 14053833 14053834 14053835 14053836 14067389 14067391 14067392 14493385 14493382 14493383 14493384 14213805 14383646 14383647 14383648 14383651 14383652 14383653 14480267 14146548 14146550 14146553 14146556 14388158 14388159 14388160 14388161 14388163 14319316 14319317 14319318 14319319 14319320 14383633 14383634 14383635 14383636 14383637 14383638 14398388 14398389 14146611 14170037 14170040 14170043 14170044 14170045 14170046 14170047 14170048 14200535 14200537 14200538 14200539 14215656 14215657 14215658 14215659 14493735 14493736 14384346 14384347 14384348 14384349 14384350 14384351 14384352 14423445 14423446 14423447 14423448 14423449 14423450 14423451 14423452 14423453 14423454 14423456 14423457 14500222 14500223 14500224 14500225 14428865 14428866 13950921 13950922 13950923 13950924 13950925 13950926 13950927 13950928 13950929 13950930 13950931 13950932 13950933 13950934 13950935 13995632 13995633 13995634 13995635 13995636 13995637 13995638 13995639 13995640 14045006 14045009 14045010 14045013 14045017 14045020 14045021 14045022 14045023 14045024 14045025 14518418 14518421 14518422 14518423 14518424 14384373 14384374 14384375 14384376 14384377 14384378 14384379 14384380 14384381 14384382 14384383 14384384 14384385 14384386 14384387 14496639 14394669 14394670 14394671 14394672 14394673 14394674 14394675 14394676 14394677 14394678 14394679 14394680 14394681 14394683 14394684 14394686 14394687 14394689 14394691 14394694 14394697 14394699 14463737 14463738 14463739 14463740 14463741 14463742 14463743 14463744 14463745 14262997 14262996 14428879 14428880 14428881 14428882 14428883 14428884 14428885 14428886 14432201 14432203 14320315 14398372 14456799 14456800 14514803 14285646 14285647 14285648 14285649 14451631 14451632 14451633 14451634 14451635 14451637 14451638 14461269 14423410 14423411 14423412 14423413 14423414 14423415 14423416 14423417 14423418 14423419 14423420 14423421 14423422 14423423 14423424 14423425 14423426 14423427 14423428 14423429 14423430 14423431 14423432 14423433 14298674 14298676 14298678 14298679 14298681 14298684 14298685 14165523 14281471 14281472 14281473 14281474 14281475 14281476 14281477 14281478 14281479 14281480 14256151 14256152 14256153 14256154 14256155 14282545 14282546 14251014 14251017 14251020 14251025 14251028 14471972 14471973 14436125 14436127 14294555 14294556 14294557 14294558 14294559 14294560 14294561 14294562 14433870 14433871 14433873 14433874 14433875 14433876 14433878 14001913 14001914 14001916 14001918 14001920 14001923 14001925 14001928 14001930 14001931 14001932 14001933 14001934 14001935 13964453 13964454 13964455 13964457 13964460 13964462 14157985 14157986 14282557 14282561 14282565 14475124 14475125 14475126 14475127 14475128 14475129 14475130 14475131 14475132 14475133 14475134 14475135 14475136 14475137 14475138 14475139 14475140 14475141 14475142 14394705 14394706 14394707 14398429 14398430 14398432 14398433 14398434 14398435 14398436 14398437 14398438 14191742 14191743 14191744 14191745 14191746 14175830 14175832 14175833 14175834 14209668 13993467 13993468 13993469 13993470 13993471 13993472 13993473 13993474 13993475 13993477 13993478 13993482 13993485 13993486 13993487 13993488 13993489 13993490 13993491 13993492 13993494 13993496 13993497 13993499 13993501 14495578 14495579 14495580 14165506 14165507 14165508 14165509 14165510 14165511 14165512 14165513 14165514 14165515 14165517 14165519 14165521 14165522 14160622 14160623 14160624 14160625 14160626 14160627 14160628 14160629 14160630 14160631 14160632 14160633 14160634 14160636 14160669 14160670 14160671 14160672 14160673 14160674 14160675 14160676 14160677 14160678 14160679 14429781 14430679 14430680 14435787 14170332 14175803 14175804 14175805 14175806 14175807 14175808 14175809 14175810 14175811 14175812 14175813 14175814 14175815 14175816 14175817 14175819 14175820 14175821 14175822 14175823 14441531 14227429 14227430 14227431 14227433 14227437 14227439 14227441 14280312 14280313 14280314 14382706 14382707 14382708 14382709 14382710 14382712 14382713 14382715 14382716 14382717 14382718 14382719 14382720 14382722 14382723 14382724 14382725 14382726 14382727 14382728 14382729 14382730 14382731 14382732 14382733 14382734 14382735 14382736 14382737 14382738 14382739 14382740 14382741 14382742 14154339 14188702 14188703 14188704 14188705 14374149 14374150 14483523 14483524 14483525 14483526 14483527 14483528 14483530 14483532 14483533 14519580 14519581 14519582 14519584 14504306 14504307 14504308 14504309 14504310 14504312 14504313 14504314 14504315 14504316 14504317 14504318 14504319 14504320 14504322 14504323 14504324 14504325 14504326 14504327 14504328 14504329 14504330 14480282 14345819 14353650 14353651 14353652 14353653 14353654 14353655 14353656 14353657 14353658 14353659 14353660 14353661 14353662 14353663 14353664 14353665 14353666 14414487 14414488 14478958 14478959 14457532 14468058 14468059 14468060 14468061 14468062 14468063 14468064 14468065 14468066 14468067 14468068 14468069 14468070 14468071 14468072 14468073 14468074 14468075 14468076 14468077 14468078 14468079 14468080 14468081 14468082 14468083 14468084 14452854 14452856 14452857 14452858 14340514 14340515 14340516 14340517 14340544 14180018 14180019 14180021 14180022 14197946 14197947 14197949 14197950 14197951 14197952 14197953 14197954 14197955 14197956 14197957 14197958 14197959 14197960 14197961 14197962 14197963 14197964 14197965 14197966 14197967 14197968 14197969 14360090 14360091 14360094 14360096 14360097 14360098 14360099 14360100 14360101 14360102 14372160 14372161 14372162 14372163 14372164 14372165 14372166 14372168 14372169 14372170 14372171 14372172 14372173 14372174 14372175 14372176 14372177 14372178 14372179 14372180 14372181 14372182 14372183 14372184 14372185 14372186 14394665 14405235 14405236 14405237 14405238 14405239 14405240 14405241 14405242 14405243 14405244 14405245 14405246 14405247 14428868 14428869 14160332 14175775 14175776 14175777 14175779 14175781 14175784 14175786 14175787 14175788 14175789 14175790 14175791 14175792 14175793 14175794 14175795 14175796 14175799 14175800 14175801 14389201 14389203 14389205 14389206 14389207 14156036 14156037 14156038 14156039 14156040 14156041 14240563 14318532 14318533 14442348 14014855 14014856 14014857 14014858 14014859 14014860 14014861 14014862 14014863 14014864 14014865 14014866 14014867 14014868 14014869 14014870 14014871 14014872 14014874 14014875 14014876 14014878 14014879 14037482 14125279 14140063 14140064 14140065 14140066 14140067 14140068 14140069 14140070 14168197 14168198 14168199 14168200 14168201 14168202 14168203 14168204 14168205 14168206 14168207 14168208 14168209 14168210 14168211 14168212 14168213 14168214 14168215 14168216 14202252 14474274 14474275 14474276 14474277 14474278 14292959 14346437 14346440 14346443 14346444 14346445 14346446 14346447 14346448 14346449 14346450 14346452 14346454 14346455 14346456 14346457 14346458 14346459 14346461 14346463 14361272 14361277 14451623 14451624 14451625 14451626 14451627 14451628 14488098 14488099 14488100 14488101 14488102 14488103 14488104 14488105 14488106 14488107 14488108 14488109 14488110 14488111 14488112 14488113 14488114 14488115 14488116 14488117 14428190 14428852 14428853 14428854 14428855 14428856 14279915 14279916 14279917 14125315 14125318 14125320 14125322 14125323 14125324 14125325 14282548 14352701 14354605 14354606 14354607 14354608 14354609 14354610 14354611 14354612 14354613 14354614 14354615 14354616 14354617 14354618 14354619 14354621 14388666 14388667 14388668 14388669 14388670 14388671 14388672 14474271 14474272 14474273 13945998 13945999 13946000 13946002 13946003 13946005 13946006 13946007 14062773 14055441 14055442 14055443 14055444 14055445 14055446 14055448 14055451 14055452 14055453 14055454 14055455 14380030 14380031 14380032 14380033 14380034 14380035 14380036 14380037 14380038 13947221 13947223 13947233 13973152 13973153 13973154 13973155 13973158 13973162 13973163 13973164 13973165 13973166 13973167 13973168 13973169 13973170 14440328 14381253 14381254 14381252 14381255 14381256 14381257 14381258 14381260 14381261 14381262 14381263 14381264 14381265 14381266 14381267 14381268 14381269 14381270 14469249 14491253 14491255 14514804 14071247 14071248 14071249 14071250 14160599 14194968 14412497 14412498 14412499 14412500 14412501 14412502 14071252 14071253 14071254 14071255 14071256 14071257 14438036 14438037 14438038 14438040 14438041 14438042 14438043 14438044 14438045 14438046 14438047 14438048 14438049 14438050 14438051 14438052 14352697 14352698 14448866 14448883 14448884 14448885 14448887 14319321 14319322 14319323 14319324 14319325 14452859 14452860 14452861 14452862 14452863 14452864 14452865 14452866 14452867 14452868 14452869 14452870 14452871 14452872 14452873 14452874 14452875 14125327 14125328 14125329 14125332 14125334 14125336 14125337 14125338 14125339 14125340 14154350 14389214 14389215 14389216 14389217 14389218 14389219 14389220 14389221 14462590 14462591 14462592 14462594 14462595 14462596 14462597 14462598 14462599 14250278 14250279 14250280 14414627 14414628 14414629 14414630 14414631 14414633 14414634 14414635 14414636 14414637 14414639 14414640 14414641 14414642 14414643 14414644 14414645 14414646 14414647 14414648 14414649 14414650 14414651 14414652 14414653 14414654 14414655 14414656 14414657 14414658 14414659 14414660 14414661 14414662 14414663 14414664 14414665 14414666 14414667 14414668 14414669 14414670 14414671 14414672 14472090 14510500 14510501 14510502 14510503 14510504 14510505 14510506 14510507 14510508 14510509 14510513 14510516 14510518 14510520 14510521 14510523 14510524 14510525 14510526 14510528 14510529 14510531 14510532 14510534 14510536 14510537 14510538 14510539 14510540 14510542 14510543 14510545 14510546 14510547 14510548 14510549 14510550 14508307 14508308 14508310 14508313 14508314 14508315 14508316 14508318 14508319 14398469 14398470 14398471 14398472 14398473 14398474 14398475 14398476 14398477 14398478 14398479 14398481 14398483 14398484 14398485 13950335 13950336 13950337 13970081 14017136 14017137 14017138 14017139 14017140 14017141 14017142 14017143 14017144 14282549 14374091 14374092 14374093 14374094 14374095 14374096 14374097 14374098 14434965 14434966 14434967 14434968 14434969 14434970 14434972 14434973 14434975 14434976 14434977 14434979 14434980 14434981 14443268 14408955 14478937 14478938 14478939 14478941 14478942 14478943 14478945 14478946 14478948 14478949 14478950 14478951 14478953 14478955 14468729 14468724 14468725 14468726 14468727 14468728 14471947 14471948 14471949 14471950 14471951 14471952 14471953 14471954 14471955 14471956 14471957 14471958 14471960 14471961 14471962 14471963 14471964 14471965 14471966 14471967 14471968 14484494 14484496 14412467 14412469 14412470 14412472 14412473 14412475 14412477 14412478 14412479 14412480 14412481 14412482 14412483 14412484 14412485 14074809 14197948 14213807 14213808 14213809 14213810 14213811 14213812 14213813 14213814 14213815 14213821 14213825 14401090 14401092 14419254 14419256 14419258 14419259 14419260 14419261 14387361 14387362 14352692 14352693 14352694 14352695 14352696 14130387 14130388 14130389 14130390 14130391 14130392 14130393 14130394 14130395 14130396 14130397 14130398 14130402 14130406 14130407 14130408 14130410 14130411 14130412 14130413 14130414 14130416 14130418 14130425 14130428 14130429 14130430 14130432 14130433 14130435 14130436 14130437 14130438 14130439 14221520 14221521 14221522 14221523 14221524 14221525 14475170 14475171 14475172 14475173 14475175 13943145 13943146 13943147 14472933 14472934 14472935 14472937 14472938 14472939 14472940 14472941 14472942 14472943 14472944 14421053 14421054 14421055 14421056 14421057 14421058 14421059 14421060 14421062 14421065 14421066 14421067 14421068 14421069 14421070 14421071 14136518 14136519 14149957 14149960 14149962 14149965 14149969 14149971 14149973 14149974 14149975 14149976 14149978 14149980 14149983 14149987 14149992 14149994 14149998 14150000 14150001 14150002 14150004 14150007 14150009 14150011 14150012 14150013 14254678 14254679 14254680 14254681 14254682 14254683 14254684 14254685 14254686 14254687 14254698 14276216 14508200 14348096 14348097 14348098 14348099 14348100 14348104 14281034 14281035 14281036 14281037 14281038 14281039 14281040 14281041 14335857 14298709 14298710 14298711 14490632 14490634 14490635 14490636 14490637 14490638 14490639 14490641 14490644 14490645 14490646 14490648 14490649 14490651 14490652 14490653 14490654 14490655 14490656 14490657 14490659 14490660 14490661 14490662 14490663 14490664 14490665 14490667 14490668 14490670 14490672 14490673 14490674 14490675 14398440 14433879 14484993 14484994 14484995 14484996 14484997 14484998 14485000 14485002 14485003 14485005 14485006 14485008 14485009 14485010 14485011 14485012 14485013 14485014 14485018 14485019 14485021 14485023 14485024 14485025 14485026 14485027 14485028 14485029 14485030 14485031 14485032 14485033 14485034 14485035 14485037 14485039 14485040 14485041 14436096 14436098 14436099 14436100 14436101 14436102 14436103 14436105 14436106 14436107 14436108 14436110 14436111 14436113 14436114 14436115 14436116 14436117 14436118 14436119 14436121 14436122 14436123 14436124 14390588 14390590 14390591 14390592 14390594 14390596 14390597 14390598 14390601 14390605 14390606 14390607 14390609 14390610 14390611 14390612 14390614 14390616 14390617 14390618 14390619 14390621 14390622 14390623 14390624 14390625 14390626 14390628 14390629 14390630 14390632 14390633 14390635 14390637 14390638 14390639 14390641 14390642 14390643 14390644 14390645 14390646 14390648 14390649 14390650 14438080 14453677 14453678 14453680 14453681 14453682 14453684 14453685 14453686 14453687 14453689 14453690 14278984 14278985 14278986 14278987 14397524 14397525 14397527 14397528 14406592 14406593 14406594 14406595 14406596 14406597 14406598 14406599 14406600 14406601 14406602 14406603 14406604 14406605 14406606 14406607 14406608 14406609 14406611 14406614 14406615 14406616 14406617 14475168 14475169 14250269 14250271 14250272 14250273 14250274 14250275 14250276 14250277 14480289 14018045 14018046 14018048 14018049 14018050 14018051 14018052 14018053 14018054 14018055 14018066 14018067 14018068 14018069 14018070 14018071 14018072 14018073 14018075 14018076 14018077 14452876 14452877 14414719 14414720 14414721 14414723 14414724 14414725 14414726 14414727 14414728 14414729 14414730 14414731 14419093 14361264 14384253 14384256 14384257 14384258 14384259 14384260 14384261 14384263 14384265 14448900 14448901 14448902 14448903 14448904 14448905 14448906 14448909 14448910 14448911 14448912 14448913 14448914 14448916 14448917 14448918 14448919 14448920 14448921 14421098 14421099 14441521 14441522 14441523 14441524 14441525 14441526 14441527 14441528 14441529 14441530 14453676 14453694 14146558 14146561 14146563 14146565 14146566 14146567 14146568 14146569 14146570 14146571 14146572 14146573 14228576 14282542 14479868 14506956 14506957 14506958 14115265 14153548 14153549 14153550 14153551 14153552 14153553 14153554 14153555 14153556 14153557 14153558 14153559 14153560 14153561 14153562 14153563 14153564 14153565 14153566 14153567 14153568 14156458 14156460 14156461 14156463 14156465 14156466 14156468 14156470 14156472 14156474 14517150 14517151 14517152 14517153 14517154 14517155 14517156 14517157 14517158 14517159 14517160 14517162 14517163 14517164 14517165 14517166 14517168 14517170 14517171 14517172 14340565 14489746 14488130 14488131 14488132 14508288 14508289 14508290 14508291 14508292 14508295 14508296 14508297 14508298 14508299 14508300 14508301 14508303 14508304 14508306 14448384 14448385 14448386 14448387 14448388 14448389 14260053 14260054 14260055 14260056 14260057 14260058 14260059 14260060 14298708 14472883 14472884 14472918 14472919 14472920 14472926 14472927 14472928 14472929 14472931 14472932 14488899 14488900 14488901 14488902 14113579 14115288 14115291 14140071 14140072 14140073 14140074 14140075 14140076 14140077 14140078 14140079 14140080 14140081 14140082 14140083 14140084 14140085 14140086 14140087 14140088 14140089 14140090 14140091 14140092 14140093 14140094 14170049 14170050 14170051 14170052 14170053 14170054 14175293 14160376 14160377 14170056 14170057 14170058 14170059 14170060 14170061 14170062 14170055 14153348 14153355 14153356 14153357 14153358 14153362 14153363 14153364 14153365 14153366 14153367 14153368 14153369 14153370 14153371 14153372 14153373 14153374 14153375 14412319 14412320 14412321 14412322 14412323 14412324 14412325 14412326 14412327 14412328 14412329 14412330 14412331 14412332 14464878 14464879 14464880 14464881 14464882 14464883 14464884 14464885 14464886 14464887 14464888 14464889 14464890 14464891 14464892 14464893 14464894 14464895 14469257 14469258 14470310 14470311 14470312 14168193 14185266 14185267 14185268 14185269 14185270 14185271 14185272 14185273 14185274 14185275 14185276 14185277 14185278 14185279 14185281 14185283 14185284 14185285 14185286 14185287 14185288 14185289 14185290 14185291 14185292 14185293 14185294 14185295 14185296 14185298 14185299 14185300 14185301 14185302 14237031 14237032 14237033 14237035 14237036 14237037 14237038 14237039 14237040 14237041 14237042 14237043 14237044 14237045 14237046 14237047 14237048 14237049 14237050 14237051 14237055 14237058 14237059 14237060 14237061 14237062 14237063 14237064 14237065 14237066 14237067 14237068 14237069 14237070 14237071 14237072 14148040 14192786 14192787 14192788 14192789 14192790 14192791 14192792 14292986 14353110 14353111 14353112 14353113 14353114 14353115 14353116 14353117 14353118 14353121 14353124 14189636 14189637 14189638 14189639 14189642 14453691 14453692 14499827 14467426 13995641 13995642 13995643 13995644 13995645 13995646 13995647 13995648 13995649 14215638 14215639 14215640 14215641 14215642 14215643 14215644 14215645 14215646 14215647 14215648 14215652 14215654 14215655 14261360 14261361 14261362 14261363 14261364 14261365 14261366 14261367 14261368 14261369 14261370 14261373 14261374 14261375 14267795 14269904 14269905 14269906 14269907 14269908 14269909 14269910 14269911 14269912 14269913 14269914 14269915 14269916 14269917 14464877 14269920 14213806 14269918 14269919 14269921 14269922 14207633 14207634 14207635 14207636 14207637 14207638 14207639 14207640 14245364 14009020 14040977 14040978 14040979 14040980 14040981 14040982 14040983 14040984 14040985 14040986 14040987 14394704 14426165 14426166 14426167 14426168 14426169 14426171 14426172 14426174 14426179 14426180 14426181 14376771 14379203 14379204 14379205 14379207 14379208 14379209 14379210 14379211 14379212 14379213 14379214 14379216 14379219 14379221 14379222 14379224 14379225 14379226 14379227 14379228 14379232 14379233 14379234 14379235 14379236 14456801 14440330 14483535 14483536 14483537 14483538 14483540 14483541 14483542 14483543 14483544 14483545 14483546 14483548 14483549 14483550 14483551 14483553 14483554 14483555 14483556 14483557 14483558 14461271 14461272 14478447 14478448 14478449 14478450 14478451 14478452 14478453 14478454 14478455 14478456 14478457 14478458 14478459 14478460 14478461 14478462 14478463 14478464 14478465 14478466 14478467 14496651 14496652 14496653 14496655 14496656 14496657 14496660 14496661 14496663 14496664 14496666 14496667 14496668 14496669 14213853 14213856 14213857 14230231 14230232 14230233 14230234 14230235 14230236 14230237 14230238 14230257 14230258 14025924 14025927 14025933 14025938 14025939 14025940 14025941 14025942 14025943 14348087 14348089 14348090 14421050 14421051 14421052 14055466 14055469 14055471 14055473 14055474 14055477 14055480 14055482 14055483 14055485 14055488 14055491 14055492 14055493 14055494 14055495 14463748 14490484 14490486 14490488 14490489 14151141 14151142 14151143 14151144 14200852 14200853 14200854 14227456 14227457 14262734 14262736 14262737 14262738 14262739 14262741 14294363 14319335 14319336 14319337 14319338 14480269 14483260 14483261 14451525 14451527 14451528 14398387 14175286 14175287 14175288 14175289 14175290 14484489 14484490 14484492 14484493 14392755 14392756 14392759 14392760 14392761 14392762 14392763 14392764 14392765 14392766 14392767 14392768 14392769 14392770 14392771 14392772 14392773 14392774 14392775 14392776 14392777 14480270 14474268 14474269 14449664 14449665 14449666 14449667 14449668 14449669 14449671 14449672 14449673 14449674 14475166 14475167 14442360 14442361 14443252 14443253 14443254 14443255 14443256 14443257 14443259 14443260 14443261 14443262 14443263 14443264 14443265 14443266 14443267 14451639 14173220 14173221 14173222 14173223 14167182 14167184 14167185 14167186 14167187 14167188 14167189 14167190 14167191 14167192 14167193 14356229 14356230 14356231 14356232 14457513 14457514 14457515 14457516 14444163 14444164 14444165 14444166 14444167 14444168 14444169 14444170 14444171 14136371 14136372 14136373 14146534 14154351 14154343 14154344 14154345 14154346 14154347 14154348 14154349 14408945 14408946 14408947 14408948 14408949 14408950 14408951 14408953 14408954 14157280 14157281 14157282 14157283 14157284 14157285 14157286 14157287 14157600 14157601 14157602 14157603 14157604 14157605 14157606 14157607 14279677 14355667 14355668 14159803 14230229 14230230 14262688 14262689 14262690 14262691 14262692 14262693 14262694 14262695 14262696 14290473 14290474 14290475 14290476 14290477 14290478 14290479 14290480 14290481 14290482 14290483 14290484 14290485 14290487 14290488 14290490 14290491 14290493 14290494 14290495 14290496 14290497 14290498 14291742 14318073 14318075 14318076 14318077 14318078 14318079 14318081 14318083 14318085 14318087 14318090 14318091 14318092 14344110 14344111 14344112 14344113 14344114 14344115 14390244 14390245 14474279 14474280 14474281 14298700 14298701 14298702 14298703 14298704 14298705 14298706 14298707 14374224 14374225 14374226 14374227 14374228 14374229 14374230 14374231 14374232 14398390 14398391 14398392 14398393 14356226 14356227 14356228 14355673 14355674 14355675 14355676 14355677 14202232 14202233 14202234 14202235 14202236 14202237 14202238 14202239 14202240 14458437 14260035 14260036 14260037 14260038 14260039 14260040 14260041 14260042 14260043 14260044 14260045 14260046 14455524 14455525 14455528 14442362 14442363 14442364 14159410 14159411 14159412 14159413 14159414 14159415 14163098 14160680 14160681 14136358 14136359 14343561 14344100 14344101 14344102 14344103 14344104 14344105 14344106 14344107 14344108 14344109 14383625 14383627 14071099 14071101 14071104 14071106 14071108 14071109 14071112 14071114 14389222 14389223 14389224 14389225 14389226 14389227 14389228 14389229 14389230 14390579 14508283 14508284 14508285 14250336 14250338 14250339 14250341 14250343 14421072 14421073 14421075 14421077 14421078 14421079 14421080 14421081 14421082 14421083 14421084 14421085 14421086 13946924 13946925 13946926 13946927 13946928 13946929 13946930 14437890 14437891 14318873 14318874 14318875 14451613 14451614 14451615 14004757 14004760 14004763 14004764 14004765 14004766 14004767 14004768 14004769 14004770 14227458 14227459 14227460 14227461 14227462 14227463 14227464 14227465 14388682 14388683 14388684 14388685 14388686 14388687 14388688 14388689 14388690 14388691 14388692 14388693 14388694 14388695 14388659 14388660 14388661 14388662 14388663 14388664 14420165 14441532 14441533 14441534 14166333 14184874 14184875 14184876 14184877 14184878 14184879 14184880 14184882 14184883 14184884 14452849 14452850 14452851 14452852 14452853 14423405 14423406 14423407 14423408 14163092 14163093 14163094 14146593 14146594 14146595 14146596 14146597 14146598 14146599 14146600 14146601 14146602 14146603 14146604 14146605 14146606 14144344 14144345 14144348 14170401 14170402 14170403 14170404 14170405 14170406 14284496 14284498 13944427 13944428 13944429 13944430 13944431 13944433 14419459 14419460 14419461 14419462 14419463 14419465 14419466 14419467 14419468 14419469 14419470 14419471 14419472 14419473 14419474 14419475 14419476 14419477 14419478 14419479 14412486 14412487 14412488 14412489 14412490 14412491 14412492 14412493 14412494 14412495 14412496 14445591 14438079 14493703 14493706 14493707 14493708 14493709 14493710 14493711 14493712 14493713 14493714 14493715 14493719 14493721 14493723 14493724 14493726 14493727 14493728 14493729 14493730 14493731 14494531 14361279 14361280 14361281 14441518 14441519 14461273 14461274 14461275 14461276 14461277 14461278 14461279 14461280 14461281 14461282 14461283 14461284 14461285 14461287 14173224 14173225 14173226 14173227 14207682 14289036 14289037 14289038 14289039 14289040 14289041 14289042 14289043 14289044 14289045 14289046 14289047 14289048 14289049 14197944 14025944 14025945 14025946 14025947 14025948 14025949 14025950 14025951 14025952 14025953 14025954 14025955 14388216 14454896 14477680 14477681 14477682 14477683 14477684 14477685 14477686 14477687 14370354 14370355 14370356 14370357 14370358 14370359 14370360 14370362 14370363 14370364 14398405 14398406 14398407 14398409 14398410 14398412 14398413 14398416 14398419 14398420 14398421 14398422 14255102 14255103 14255104 14255105 14255106 14044987 14051915 14051917 14051919 14051923 14051924 14051925 14051930 14051931 13944422 13944423 13944424 13944425 13944426 14335226 14335227 14335228 14335229 14335230 14335231 14335232 14335233 14335234 14335240 14335241 14335244 14335246 14335247 14335248 14335249 14335250 14335251 14335255 14335256 14335257 14335259 14335261 14335264 14063098 14063099 14063100 14063101 14063102 14063103 14063104 14063106 14063109 14063111 14063113 14063114 14063115 14063116 14063117 14236731 14236734 14236737 14236741 14236743 14236744 14236745 14236748 14236752 14236757 14236760 14236761 14236762 14236763 14236764 14236765 14236766 14236767 14236768 14236769 14236770 14236771 14236774 14236776 14221770 14221773 14221775 14221778 14221782 14221785 14221788 14221793 14221797 14221800 13930588 13930595 13930596 13930597 13930598 13930599 13930600 13930601 13930602 13930603 13995021 14180023 14180024 14180025 14180026 14180027 14180028 14180029 14180030 14180031 14180032 14180033 14180034 14180036 14352209 14352210 14352211 14352212 14352213 14352214 14352215 14352216 14352217 14434614 14434616 14434621 14434622 14434624 14434627 14434629 14434630 14434631 14262750 14262751 14262752 14262753 14262754 14262755 14262756 14262760 14262763 14262764 14262765 14262766 14262767 14262768 14262769 14262770 14262771 14262772 14262773 14262774 14262775 14262776 14262777 14262778 14262779 14262780 14262781 14262782 14262783 14262784 14262785 14262786 14262787 14262788 14262789 14262790 14262791 14262792 14262793 14262796 14262800 14262804 14262808 14262809 14262810 14455252 14455253 14455254 14455255 14455256 14455257 14455258 14455259 14455260 14455261 14455262 14455263 14455264 14455265 14455266 14455267 14455268 14455269 14455270 14455271 14455272 14455273 14455274 14455275 14455276 14455277 14455278 14455279 14455280 14455281 14455282 14455283 14455284 14455285 14455286 13922773 14454892 14466206 14466207 14466208 14466209 14230221 14230222 14230223 14451629 14514805 14514806 14055562 14055563 14055564 14055565 14055566 14055567 14448867 14448868 14448869 14448871 14448873 14448874 14448876 14448879 14448877 14448878 14404513 14315585 14315586 13882960 13882961 13882962 13882963 13882966 13882969 13882972 13882993 13882994 13882995 13882996 13882997 13882998 13882999 13883000 13883001 13883002 13883003 13883004 14394848 14394849 14394850 14163298 14189648 14189649 14189650 14189651 14189652 14189653 14189654 14189655 14189656 14189657 14189658 14125343 13887725 13887729 13887730 13887733 13887734 13893783 13914542 13914547 13914549 13914550 13914551 13914552 13914553 13949128 14154340 14154341 14163101 14163102 14163103 14163104 14163106 13943805 13943806 13943807 13943808 13943809 14470278 14470279 14470280 14470281 14463661 14463662 14496640 14496643 14496644 14496645 14496646 14496648 14496649 14064406 14064407 14064408 14064409 14064410 14064411 14064412 14064413 14064414 14064415 14064416 14064417 14064418 14064419 14064420 14064421 14064422 14064423 14064424 14064425 14064426 14064427 14064428 14064429 14064430 14064431 14064432 14064433 14064434 14463668 14463669 14463670 14463671 13892232 13914496 13914500 13914503 14007729 14007730 14007731 14007732 14007733 14007734 14007735 14007736 14007737 14007738 14007739 14007740 14007741 14007742 14007743 14007744 14007745 14007746 14007747 14007748 14067390 14037935 14037936 14037937 14037938 14037941 14037942 14037943 14037944 14232919 14385503 13920668 13920669 13920670 13920671 13920672 13920673 13920674 13920675 13920676 13920677 13920678 13920679 13922221 13922222 13922223 13922224 13922225 13922226 13922227 13922228 13922229 14163286 14163287 14426183 14426184 14426185 14426186 14044608 14044609 14044610 14044611 14044612 14044613 14044614 14044615 14044618 14153597 14153598 14153599 14151133 14472898 14418761 14015380 14015381 14015382 14015383 14015384 14015385 14015386 14015387 14015388 14015389 14015390 14015391 14015392 14015394 14015396 14182977 14182978 14182979 14182980 14182981 14182982 14182983 13888972 13888973 13888974 13888975 13888976 13888977 13888978 13888979 14202242 14158291 14180894 14180895 14180896 14180897 14180898 14180899 14180900 14180901 14180902 14180903 14180904 14180905 14180906 14180907 14180909 14180910 14180911 14180913 14180914 14180915 14197935 14197936 13897772 13897773 13897774 13897775 13897776 13897777 13897778 13897779 13897780 13897781 13897782 13897783 13897784 13897785 13897786 13897787 13897788 14156476 14156478 14156479 14156480 14156481 14156482 14156483 14156484 14156485 14156486 14156487 14156488 14156489 14156490 14156491 14156492 14156493 14156494 14156495 14156496 14175291 14298694 13896704 13896705 13896706 13896707 13896709 13896710 13896712 13896714 13896715 13896716 14042354 14042355 14042356 14042357 14042358 14042359 14042360 14042361 14042362 14458438 14458439 14458441 14458442 14458443 14458445 14458446 14458447 14458448 14458451 14458452 14500236 14500237 14500238 14500239 14500240 14500241 13883174 13883175 13883176 13883177 13883178 13883179 13883180 13883181 13883182 13883183 13883184 13883185 13883186 13883187 13883188 13883189 13894653 13899127 13899128 13899129 13899130 13899131 13899132 13899133 13899134 13899135 13899136 13899138 14419089 14419090 14051926 14051927 14051928 14051929 14202253 14202254 14202255 14202256 14202257 14202258 14017180 14017181 14017182 14017183 14017184 14017185 14017187 14017188 14475144 14475145 14463746 13932024 13932025 13932026 13932027 13932028 13932029 13932030 13932031 13932032 13932033 13932034 13932035 13932036 13932037 13932038 13932039 13932040 13932041 13932042 13932043 13932044 13932045 13932046 13932047 13932048 13932049 13932050 13932051 14495547 14495550 14495551 14495552 14495553 14495554 14495556 14495557 14467427 14340537 14340538 14340539 14158003 14158004 14158005 14158006 14158007 14158008 14158009 14392778 14392779 14392780 14392781 14392783 14513580 14384359 14384360 14384361 14384362 14384363 14384358 14449644 14449645 14449646 14449648 14449649 14449650 14449651 14449652 14449653 14449654 14449655 14449656 14449657 14449658 14449659 14449660 14449661 14449662 14449663 14463651 14463652 14465004 14039842 14039843 14039844 14039845 14039846 14039847 14039848 13951840 13951842 13883669 13883670 13883671 13883672 13883673 13883675 13883676 14335885 14335886 14335890 14335894 14335895 14335907 14335910 14335911 13888601 13888602 13888604 14285650 14285651 14285652 14285653 14285654 14285655 14285656 14357334 13946952 13946953 13946954 13946955 13946956 13946957 14315608 14315609 14315610 14263024 14263025 14263026 14263027 14263028 14263029 14148548 14148549 14148550 14148551 14148552 14148559 14148561 14148562 14148565 14148569 14148571 14148574 14148576 14148577 14148578 14148579 14148580 14148581 14151769 14151770 14151771 14151772 14151773 14151774 14151775 14151776 14151777 14151778 14151779 14151780 14151781 14154305 14154306 14154307 14154308 14154309 14154310 14154311 14154312 14154313 14154314 14154315 14154316 14154317 14154318 14154319 14159820 14483270 14157987 14157988 14157989 14157990 14157991 14157992 14157993 14157994 14157995 14157996 14157997 14157998 14157999 14158000 14158001 14158002 13952115 13952116 13952117 13952118 13952119 13952120 13952121 14151129 14151130 14151131 14151132 13921017 13921018 13921019 13921021 13921024 13921025 13921026 13921027 13921028 13921029 13921030 13921031 13921032 13921033 14229998 14263011 14263012 14263013 14263014 14263015 14263016 14263017 14263020 14352097 14352098 14352099 14352100 14352101 14352102 14352103 14352104 14352105 14352106 14352107 14352108 14493700 14230220 13909291 13909294 13909298 13909302 13909303 13909304 14035689 14119052 14119053 14119054 14119055 14119056 14119057 14119058 14119059 14119061 14119064 14119066 14119067 14119068 14160210 14467644 14467645 13921013 13921014 13921015 13921016 14318017 14318018 14160369 14160370 14160371 14160372 14160373 14160374 14160375 14440322 14440323 14440324 14440325 14440326 14440327 14004750 14004751 14004752 14004753 14004754 14004755 14004756 14480283 14480284 14480285 14480286 14480287 13940458 13940459 13940460 13940461 14071258 14071259 14495534 14495535 13888964 13888968 13888971 14278949 14278950 14278951 14278952 14278953 14003473 14003474 14003476 14003477 14003479 14003480 14003481 14003483 14003484 14003485 14003486 14003487 14003488 14003489 14003490 14003493 14003495 14003496 14003497 14003498 14003499 14003500 14003502 14003503 14003504 14438064 14438065 14438066 14438067 14438068 14438069 14438070 14438071 14438072 14438073 14438074 14438075 14438076 14438077 14475176 14475177 14475178 14475179 14475180 14475181 14475182 14475183 14475184 14475187 14475188 14475189 14475190 14475191 14475192 14475193 14475194 14475195 14475196 14475197 14074765 14074766 14074767 14074770 14074772 14074774 14074775 14074776 14074777 14186608 14186610 14186613 14021716 14021717 14021718 14021719 14021720 14021721 14021722 14021723 14021724 14021725 14294366 14294367 14294368 14294369 14294370 14294371 14294372 13995619 14185306 13884357 13884358 13884359 13884360 13884361 13884362 13884363 13884364 13884365 13891933 13891935 13891936 13891939 13891942 13891943 13891944 13891945 13891946 13891947 13891948 13891950 14036832 14055498 14055499 14055500 14055501 14055502 14055503 14055504 14055505 14055506 14062766 14062767 14062768 14062769 14062770 14155685 14155686 14155687 14155688 14155689 13929291 13929293 13929294 13932021 13932022 13932023 14414673 14414674 14414675 14414676 14383632 14263005 14263006 14263007 14263008 14263009 14263010 14374104 14374106 14374107 14374109 14374111 14374112 14318985 14318986 14024070 14024071 14024072 14024073 14024074 14024075 14024076 14024077 14024078 14024079 14024080 14024081 14024082 14024083 14024084 14024085 14024086 14024087 14024088 14024089 14024090 14024091 14024092 14499833 14499834 14151736 14151740 14151745 14151749 14151752 13952030 13952031 13952032 13952033 13952034 13952035 13952036 13952037 13952038 13952039 13952040 14444175 13921000 13921001 13921002 13921003 14154333 14154334 14154335 14154336 14154337 14154338 13880009 13880012 13880014 13880015 13880016 13955368 13955370 13955371 13955372 13955373 13955374 13955375 13955376 14151760 14151761 14151762 14151763 14151764 14151766 14151767 14151768 14490476 14394793 14394794 14394795 14394796 14394797 14394798 14394800 14394802 14394803 14404253 14440030 14189600 14189601 14189602 14189603 14189604 14189605 14189606 14189607 14189611 14189614 14189615 14189616 14189617 14189618 14189620 14189625 14189629 14189630 14189631 14189632 14189633 14189634 14189635 14450988 13892226 13894972 13894973 13894974 13894975 13894976 13894977 13894978 13894979 13894980 13894981 13894982 13894983 13894984 13984498 14201938 14201939 14201940 14201941 14201942 14201943 14201944 14201945 14201946 14201947 14201948 14421092 14421093 14421094 14421095 14421096 14421097 14508195 14508198 14335929 14335930 14335931 14335932 14335933 14379811 14379812 14379813 14379816 14379817 13893755 13893756 13893757 13893758 13893759 13893760 13893761 13893762 13893763 13893764 13893765 13893766 13893770 13893773 13893775 13893777 13893778 13893779 13893780 13893781 13893782 13909358 14071243 14071245 14071237 14071240 14071242 14071244 14071218 14071221 14071224 14071227 14071229 14071232 14071234 14071235 14071246 14160212 14398371 14185303 14185304 14407440 14407441 14477688 14477689 14477690 14149464 14398494 14398495 14398496 14398497 14398499 14398501 14398502 14398503 14398504 14428779 14155716 14155717 14155718 14155719 14155720 14155721 14155722 14155723 14155724 14155725 14155726 14155727 14155728 14155729 14495529 14478962 14478963 14478964 14478966 14478967 14478968 14478969 14478970 14478971 14478972 14245359 14295655 14295656 14295657 14295658 14295659 14295660 14295661 14295662 14295663 14295664 14295665 14295666 14295667 14295668 14295669 14295670 14295671 14295672 14295673 14115263 14115264 14354572 14354573 14354575 14354576 14354577 14354578 14037945 14037946 14037947 14037948 14037949 14037950 14037951 14037952 14037953 14037954 14037955 14037956 14037957 14037958 14037959 14037960 14037961 14037962 14356239 14226784 14262906 14262907 14262908 14262909 14262910 14262911 14262912 14262914 14262915 14262917 14262918 14262919 14262920 14262921 14262922 14262923 14262924 14262925 14262926 14262927 14262928 14262929 14262931 14262932 14262934 14262936 14262938 14262939 14262941 14262942 14262944 14262947 14262950 14262951 14262952 14262953 14262954 14262955 14262956 14262957 14262961 14262962 14262965 14262966 14262967 14262968 14262969 14262970 14262971 14262972 14262973 14262974 14262975 14262976 14262977 14262978 14262979 14262980 14262981 14262982 14262983 14262984 14262985 14284620 14384761 14202241 14048635 14506914 14506915 14506916 14506917 14506918 14506919 14506921 14506923 14506926 14506929 14506931 14506933 14506934 14506935 14506936 14506937 14506938 14506939 14506940 14506941 14506942 14506944 14506946 14506947 14506948 14506949 14506950 14506952 14506954 14506955 14513569 14513570 14513572 14513573 14513574 14513575 14474265 13933909 13933910 13944394 13944395 13944396 13944397 13944398 13944399 13944400 13944401 13944402 13944403 13944404 13944405 13944406 14004738 14004739 14004740 14004741 14004744 14149942 14149943 14149944 14149945 14149947 14149951 13899124 13899125 13899126 13947085 13947086 13947087 13970083 14034347 14004797 14004799 14004801 14004803 14004805 14469239 14480498 14480500 14480502 14480503 14480504 14480505 14480507 14480508 14480510 14480511 14468054 14468055 14468056 14468057 14515733 14515734 14515735 14515736 14515738 14515739 13909213 13909214 13909216 13909217 14426187 14426188 14426189 14426190 14426191 14426192 14426194 14426196 14426197 14426198 14426199 14426200 14426201 14074803 14074804 14074805 14074807 14074808 14115270 14115271 14115273 14115274 14119076 14119077 14119080 14119082 14119086 14119089 14119090 14393799 14393802 14393803 13930850 13930851 13930852 13930853 13930854 14213835 14213840 14213841 14213842 14213843 14213844 14213845 14213846 14213847 14213848 14213849 14213850 14213851 14189680 14189683 14189685 14189687 14189688 14189689 14189691 14189692 14189693 14189696 14189699 14189702 14189704 14189705 14189706 14189707 14189709 14189710 14189712 14045026 14045027 14045028 14045029 14045030 14045031 14045032 14045034 14045035 14045037 14045038 14045039 14045040 14045041 14071208 14071209 14071211 14071214 14380800 14380801 14380803 14380804 14380805 14380808 14380811 14380812 14361326 14361327 13933228 13933229 14035690 14045058 14045059 14045060 14045064 14045067 14045068 14045069 14045070 14045071 14045072 14045073 14045074 14045075 14045076 14045078 14045080 14045082 14045083 14045084 13972834 13972835 13972836 13972837 13972838 13903367 13903371 13903375 13903377 13903378 14040968 14040969 14040970 14040971 14040972 14040973 14040974 14040975 13985426 13985427 13985428 13985429 13985430 13985431 13982558 13982559 13982560 13982561 13982562 13982563 13982564 13982565 13982566 13982567 13982568 13982569 13982570 13982571 13982572 13982573 13982574 13982575 13982576 14370328 14370329 14370330 14029003 14036827 14036828 14036829 14036830 14340548 14340549 14340550 14340551 14340552 14340553 14340554 14340555 14340556 14340562 14340563 14340564 13883157 14379972 14379973 14379974 14379975 13888598 14209847 14209849 14143107 14143108 14143109 14143110 14143112 14143114 14143117 14143118 14143121 14143124 14143126 14143128 14143129 14143130 14143131 14143132 14143133 14143134 14143135 14143136 14143138 14143139 14155683 14160211 13944974 13944975 13944976 14008961 14008962 14008963 14227417 14227419 14227420 14227421 14227422 14227423 14227424 14227425 14227426 14227427 14227428 13893435 13893436 13893437 13893438 13893440 13893441 13893442 13893443 13893444 13893445 14500347 14500358 14500359 14379718 14379719 14379720 14379721 14379722 14379723 14379724 14379725 14379726 14474600 13940435 13940436 13940437 13940438 13940439 13940440 13940441 13943672 13943673 14262862 14065612 14065613 14065614 14065615 14065616 14065617 14065618 14065620 14065621 14405952 14405953 13883159 13883160 13883161 13883162 13883163 13883164 13883165 13883166 13883167 13883168 13883169 14157608 14157609 14157610 14157611 14157612 14157613 14157616 14160121 14167064 14167065 14167066 14167067 14157059 14384224 14227447 14227449 14227451 14227452 14227453 13943136 13943137 13943138 13943139 13943140 13962234 14026933 14448929 14448930 14448931 14448932 14448933 14166340 14166341 14166342 14166343 13924625 13924627 13924628 13924630 13924631 13924632 13924633 14031609 14031610 14031611 14031612 14031613 14031615 14166334 14166335 14166336 14166337 14065623 14065624 14065625 14065626 14488118 14488119 14488120 14488121 14488122 14488123 14488124 14488125 14488126 14488127 14488128 14040989 14040990 14040991 14040992 14040993 14040994 14040995 14040996 14040997 14040998 14040999 14041000 14041001 14471989 14192824 14192825 14192826 14192827 14192828 14192829 14192830 14192831 14192832 14209790 13967257 13967258 13967259 13967260 13967261 13967262 13967263 13967265 13967266 13967267 13967268 14445067 14445068 14445069 14445070 14445071 14445072 14445073 14445074 14445075 14445076 14445077 14445078 14445079 14445080 14445081 14445082 14445083 14445084 14445085 14445086 14445087 14445088 14445089 14445090 14445091 14017159 14017160 14017161 14017162 14017163 14017164 14043831 14043832 14068744 14068745 14068746 14068747 14068749 14068750 14068751 14068752 14068753 14068754 14068755 14068756 14068757 14068758 14068759 14068760 14068761 14068762 14068763 14068764 14068765 14068766 14068767 14068768 14068769 14068770 14068771 14068772 14068773 14068774 14068775 14068776 14068777 14068778 14068779 14055507 14055508 14055509 14055510 14055511 14055512 14055513 14055516 14055518 14055520 14055521 14055524 14055526 14055529 14055530 14055531 14442365 14442366 14442367 14442368 14442369 14442370 14442371 14442372 14442373 14442374 14442375 14442376 14442377 14442378 14442379 14442380 14442381 14442382 14442383 14442384 14442385 14442386 14442387 14442388 14442389 14442390 13964490 13928254 13928255 13928256 13928257 13928258 13940425 13940426 13940427 13940428 13947384 14412710 14388183 14388184 14388186 14388187 14388188 14188707 14188710 14188713 14186941 14186942 14186943 14186944 14186945 14186947 14186948 14186949 14186950 14186951 14407895 14483259 14352091 14352092 14352093 14352094 14352095 14163088 14163089 14163090 14163091 14388175 14388176 14421191 14055555 14055556 14055557 14055558 14055559 14055560 14055561 14146535 14384353 14384354 13998573 13998574 13998575 13998576 13998577 13998578 13998579 13998580 13998581 13998584 13998585 13998586 13998587 13998588 13998590 13998591 13998593 13998596 13998597 13998599 13998602 14489761 14252277 14252278 14175282 14175283 14125274 14125275 14125276 14125277 14125278 14292960 14292961 14292962 14292963 14292964 14292965 14292966 14292967 14292968 14292970 14292973 14292974 14292975 14292976 14292977 14292978 14292979 14292981 14292983 14292985 13914505 13914508 13914510 13914512 13914513 13914514 13914515 13914517 13914537 14000379 14014023 14014024 14014025 14014026 14014027 14014028 14398378 13930855 13930856 13930857 13930858 14180017 14226796 14245369 14245370 14245371 14245372 14004832 13903356 13903357 13903359 13903361 13903363 13991384 13991385 13991386 14000358 14000360 14000363 14000366 14000369 14000373 14000374 14000375 14000376 14000377 14007718 14008964 14008965 14008966 14008967 14008969 14008971 14008973 14008976 14008977 14008978 14008980 13966753 13882923 13882924 13882925 13882926 13874601 13874602 13932078 13932079 13932080 13932081 13877151 13877152 13877153 13877154 13877155 13877156 14318881 14318883 14318884 14318885 14318886 14318887 14318888 14318889 14318890 13941855 13941856 13941857 13941858 13887683 13887685 13887689 13887691 14379194 14379727 14379729 14379730 14379731 14379732 14379733 14379734 14379735 14379736 14379737 14379738 14379740 14379741 13877312 13973150 13998566 13998567 13998568 13998569 13877314 13877317 13877319 13877321 13877322 13877323 13946932 13946933 14255098 14144994 14144995 14144996 14144997 14144998 14144999 14145000 14145002 14440033 13908492 14157051 14157052 14157053 14157054 14157055 14157056 14157057 14157058 13896122 13896123 13896124 13896125 13896126 13896127 13896128 13896129 13896130 13896131 13896132 13896133 13896134 13982837 13982839 13982840 13982841 13982842 13982844 14041002 14041003 14383665 14383667 14383668 14383669 14383670 14383671 14383674 14383675 13908496 13908497 13908498 13908499 13908500 13908501 13908502 13908503 14125342 14180882 14180883 14180884 13930827 13930828 13930829 13930832 13930833 13930834 13930835 13930836 13930837 13930838 13930839 13930840 13930842 13930843 13930844 13930845 13930846 13930847 13930848 13930849 14207667 14398373 14398374 14398375 14398376 14398377 14263004 13949123 13949124 13949125 13949126 13949127 14398490 14398491 14398492 14398493 14356220 14356221 14356222 14361331 14361336 14361340 14483266 14483267 14483268 14483269 14418717 14418718 14418719 14469250 14469251 14469252 14469253 14469254 14469255 13997852 13997855 13997857 13997858 13997859 14355678 14355679 14355680 14355681 14355682 14355683 13882487 13882488 13882489 13882490 13882491 13882492 13882493 13882494 13882495 13882496 13882497 13882498 13882499 13882500 13882501 13920271 13920272 14398488 14401084 14428840 14428841 14428842 14428843 14428844 14428845 14428846 14428847 14428848 14428849 14428850 14434953 14434954 14434956 14434957 14434958 14434959 14434961 14434962 14434963 14434964 14434948 14434949 14434950 14434952 14143140 14143141 14143144 14143146 14143148 14143149 14143150 14143151 14143153 14153593 13956927 13956928 13956929 13956930 13956931 13956932 13956933 13956934 13956935 13956937 13956938 13956940 13956943 13956944 14055414 14151782 14163293 14163294 14163295 14163296 14163297 13921733 13933848 13933849 13933851 13933853 13933856 13933857 13933860 13933861 13933862 13933863 13933864 13933865 13933866 13933867 13933868 13933869 13933870 13933871 13933872 13933874 13933876 13933877 13933878 13933879 13933880 13933881 13941861 13941862 13941863 13941864 13941865 13941866 13941867 13941868 13941869 13877124 13877125 13877126 13877127 13877128 13877129 13877132 13877134 13877135 13888944 13888946 13888947 13888949 13888950 13888951 13888953 13888954 13888957 13888961 13894985 13894987 13894988 13894989 13894990 13894991 13894992 14419483 14461293 14461294 13924634 13924635 13924636 13924637 13924638 14340448 14340449 14340450 14340454 14340455 14340456 14340457 14340458 14340459 14340460 14340530 14136541 14136542 14136543 14136544 14136545 14136546 14136547 14136549 14136550 14136551 14136552 14318984 14158730 14158731 14158732 14158733 14158734 14158735 14158736 14158737 14158738 14158739 14158740 14158741 14158742 14158743 14158744 14158745 14158746 14158747 14158748 14158749 14158750 14180868 14180871 14180872 14180873 14180874 14180875 14180887 14180891 14180892 14445092 14445093 14445094 14445095 14445096 14445097 14490477 14136523 14136524 14136528 14136531 14136535 14136537 14136538 14136539 14136540 13941870 13941871 13941872 13941873 13941874 13941875 14183122 14183123 14183124 14183125 14183126 14183127 14200872 14200874 14406591 14109509 14109512 14214885 14230239 14230240 14230241 14230242 14230243 14230244 14230245 14230246 14230247 14230248 14230249 14230250 14230251 14230252 14230253 14230254 14230255 14230256 14398370 14214228 13933896 13933897 13933898 13933899 13933900 13933901 13933902 13933903 13933904 13933905 13933907 13933908 13944407 13899992 13899993 13899994 13899995 14483262 14483263 14483264 13919305 13919306 13919307 13919308 13919309 13919310 13919311 13919312 13919313 13919314 13919315 13918486 13918488 13918489 13918492 13918493 13918494 13918496 13918497 14024309 14024310 14024311 14024312 14024313 14024314 13888606 13888607 13888608 14062771 14062772 14031614 13978955 13978956 13978957 13995617 13995618 14045086 14045087 14045088 14045089 14045090 14045091 14045093 14045095 14045096 14045097 14045099 14045100 14045101 14055532 14055533 14055534 14055537 14055539 14055542 14055545 14055546 14055548 14055549 14055550 14055552 14226797 14226798 14153582 14153583 14153584 14153585 14153586 14153587 14153588 14153589 14153590 14153591 14153592 14458474 14458475 14458476 14458477 14458478 14458479 14458480 14458481 14458482 14163261 14163262 14163263 14163264 14163265 14163266 14163267 14163268 14163269 14163270 14163271 14163272 13964483 14154352 14053843 14053844 14053845 14053846 14053847 14053848 14053849 14053850 14197937 14074778 14074779 14074780 14074781 14074783 14074786 14074787 14074790 14074791 13995621 14008981 14008983 14008984 14008985 14008986 14008987 14008988 14008989 14008990 14008991 14008992 14008993 14008994 14008996 14008999 14009001 14009002 14009003 14009004 14009005 14009006 14009007 14245373 14245374 14191730 14191731 14191732 14191733 14191734 14191735 14191736 14191737 14191738 14191739 14191740 14191741 13990556 14028317 14028318 14028319 14028320 14028322 14028324 14028326 14028327 14028328 14028329 13909234 13909235 13909236 13909237 14201949 14201950 13877327 13877324 13877325 13877326 13877288 13877290 13877291 13877292 13877293 13877294 13877295 13877296 13877297 13877298 13877299 13877300 13877301 13877302 13877303 13877304 13877306 13877308 13877309 13877310 13877328 13891415 14245379 14245380 14245381 13902673 13902674 13902675 13902676 13902677 13902678 14457517 14457518 14457520 14457521 14457522 14457523 14457524 14457525 14457526 14163099 14163100 13890396 13890399 13890400 13890401 13890402 13890403 13890404 13890406 13896490 13896493 13896496 13896497 13909247 13909248 13909249 13909250 13890409 13890411 14175826 14175827 14175828 14175829 14472414 14472415 14472416 14472418 14167196 14194940 14194941 14194945 14194949 14194951 14194952 14194953 14194954 14194955 14194956 14194957 14194958 14194959 14194960 14194961 14194962 14194963 14194964 14194966 14514807 14514808 14221498 14221499 14221500 14221501 14221502 14221503 14221504 14221506 14250939 14250940 14250941 14250942 14250943 14250944 14250945 14250946 14250947 14352089 14352090 14112789 14112790 14112791 14112792 14112793 14112794 14136560 14136562 14136563 13928248 13928249 13928250 13928251 13928252 13928253 13940429 13940430 13940431 13940432 14158713 14029989 14029992 14029996 14029997 14029998 14029999 13883677 13888643 13888644 13888646 13888647 13888648 13888649 13987442 13987443 13987444 13987445 13987446 13987447 13987448 14007685 14007686 14007687 14007688 14007689 14022348 14034337 14469967 14184863 14513576 14024683 14024684 14024685 14024686 14048634 14166359 14166360 14166361 14166362 14166363 14166364 14166366 14166367 14166370 13887735 13887736 13887737 13887739 13887741 13887743 13887744 13887745 13908533 13909093 13909208 13909210 13909211 13909212 14127117 14127122 14127125 14127130 14127131 14127132 14127133 14127134 14127135 14127136 14127137 14127138 14127139 14136553 14136554 14136555 14136556 14136557 14136558 14136559 14296095 14296096 14298688 14298690 14298691 13956924 13956925 13956926 14151783 14262744 14001912 14317890 14317891 14026931 13970086 13970084 13903364 13924596 13924598 13924600 13924602 13924603 13924606 13924608 13924609 13964464 13964465 13964466 13964467 13964468 13964469 13964470 13964471 13964472 13964473 13964474 13964475 13964476 13964477 13964478 13964480 13964481 14165504 14165505 14420163 14420164 14428803 14428805 14428806 14428807 14428808 14452848 13877329 13877330 13877331 13877332 13877333 13877334 13877335 13879992 13879993 13879994 13879995 13879996 13879997 13879998 13879999 13880000 13880001 13880002 13880003 13880004 13880006 13880007 13880008 13880308 13880309 13880311 13880313 13880314 13880315 14194967 13907752 13907753 13907754 13907755 13907757 13907759 13907760 13907761 13907762 13907765 13907768 13907769 13907770 13975574 14202265 14202266 14202267 14202268 14202269 14202270 14202271 14136360 14136361 14148041 14148042 14148043 14148044 14148045 14148046 14148047 14148048 14148049 14160445 14160446 14160447 14160448 14160449 14160450 14160451 14160452 14160453 14160454 14160455 14160456 14160457 14160458 14160459 14160460 14160461 14160462 14160463 14160464 14160465 14160466 14160467 14160468 14160469 14160470 14160471 14160472 14160473 14160474 14160475 14160476 14160477 14160478 14160479 14160480 14160481 14510554 14001890 14001893 14001894 14001895 14001896 14001897 14001898 14001899 14001903 14001906 14001908 14001911 13950338 13950339 13950340 13950341 13950342 13950343 13950344 13952052 13952053 13952054 13952055 13952056 13952057 13952058 13952059 13952061 13952062 13952063 13952065 13952066 13952067 13952068 13952069 13952070 13952071 13952073 13952074 13952075 13952076 13952079 13952081 13952083 13952085 13952087 13952090 13952092 13952094 13952096 13952098 13952099 13952100 13952101 13952102 13952103 13952105 13952106 13952107 13952108 13952109 13952110 13952111 13952112 13952113 14153594 14153595 14153596 14188714 14188715 14188716 14188719 14188724 14188725 14188726 14188727 14188728 14188729 14188730 14188731 14188732 14188733 14319370 14319371 14319372 14319373 14319374 14319375 14319376 14319377 14352699 14352700 14175835 14175836 14175837 14175838 14175839 14175840 14175841 14175842 14175843 14175844 14175845 14384234 14384236 14384238 14384240 14384241 14384243 14384244 14384245 14384246 14384247 14384248 14384249 14404514 14512645 14512646 14352096 14024315 14024316 14024317 14024318 14043793 14319963 14319964 14319965 14319966 14319967 14009010 14009011 14009012 14009013 14009014 14009017 14504305 13889481 13889482 13918475 13918476 13918477 13918478 13918479 13918480 13918481 13918482 13918483 13918484 13918485 14282547 14241983 14241985 14241986 14241989 14241990 14241994 14241995 14191747 14163273 14163275 14163276 14163277 14163278 14163279 14163280 14163281 14163283 14163284 14163285 14166380 14166381 14166382 14166384 14166386 14166387 14166388 13908493 13908494 13908495 13891414 13921734 13921735 13921736 13921737 13921738 13921739 13921740 13921741 13921744 13921746 13921747 13921748 13921749 13921750 13921751 13921752 14148260 14148261 14469240 14469241 14469242 14469243 14469244 14469245 14469246 14469247 14469248 14030003 14030004 14030005 14030006 14030007 14030008 14030009 14030010 14030011 14030012 14030013 14030014 14030015 14030016 14030017 14030018 14030019 13909350 13909351 13946907 13946908 13946909 13946910 13946911 13946912 13946913 13946914 13946915 13946916 13946917 13946918 13946919 13946920 13946921 13946922 13946923 14428870 14428871 14428872 14428873 14428874 14428875 14428876 14428877 14428878 14384364 14384365 14384366 14384367 14384368 14384369 14384370 14394804 14394805 14394806 14394807 14394808 14394809 14394810 14394811 14394812 14394813 14394814 14394815 14394816 14394817 14394818 14394819 14394820 14394821 14394822 14394823 14394824 14394827 14394828 14394829 14394830 14394831 14394832 14394833 14394834 14394835 14394836 14394837 14394838 14394840 14394841 14394843 14394845 14394846 14394847 14394851 14394852 14394854 14394855 14394856 14394857 14394858 13941841 13941842 13941843 13941844 13941845 13941846 13941847 13941848 13941849 13941850 13941851 13941852 13941853 13941854 13991387 13991388 13991389 13991390 13991391 13991392 13991395 13991397 13991400 13991403 13991406 13991408 13991410 13991411 13991412 13991413 13991414 13991415 13991416 13991417 13991418 13991419 13991420 13991421 13991422 13991423 13991424 13991425 13991426 13991427 13991428 14024103 14024104 14024105 14024106 14024107 14024108 14024109 14024110 14024111 14024112 14024114 14024115 14024116 14202227 13882480 14397523 13877136 13877137 13877138 13877139 13877140 13877141 13877142 13877143 14226783 14315679 14315682 14053837 14053838 14160200 14160201 14160202 14160203 14160204 14160205 14160206 14160207 14160208 14160209 14044621 14044623 14044624 14044625 14044626 13877311 13903380 13903381 13903382 13888650 13888651 13888654 13888656 13908525 13908526 13908527 13908528 13908529 13908530 13908531 13908532 14039816 14039817 14039818 14039819 14039820 14039822 14039824 14039825 14039826 14039827 14039828 14039829 14039830 14039831 14039832 14039833 14039834 14039835 14039836 14039837 14039838 14039839 14039840 14039841 14380818 14472899 13908513 13908514 13908515 13908516 13908517 13908518 13908519 13908520 13908521 13908522 13908523 13908524 13966328 13966333 13966335 13966336 13966337 13966338 13966339 13972724 13972725 14226770 14226772 14226774 14226777 14226778 14226779 14226780 14226781 14226782 13879509 13879510 14200541 14200542 14200543 14182969 14182970 14182971 14182972 14182973 14182974 14182975 14182976 14066666 14066667 14066668 14066669 14066670 14066671 14066672 14066673 13884355 13884356 13884381 14357251 14357252 14357253 14357254 14357255 14357259 14357262 14357263 14357264 14357265 14357266 14357267 14357268 14357269 14357270 14357271 14357272 14357273 14357274 14357275 14357276 14357277 14357278 14357279 14357280 14357281 14357282 14357283 14357284 14357285 14357286 14357287 14357289 14357291 14357292 14357293 14357294 14357295 14357296 14357297 14357298 14357299 14357300 14357301 14357302 14357303 14357304 14388180 14388182 14462600 14462601 14462602 14462604 14462605 14462606 14462607 14462608 14462610 14462611 14462612 14462613 14462615 14462616 14462617 14462618 14480291 14157288 14443086 14493737 14493738 14493739 14493740 14493741 13940448 13940449 13941886 13946934 13946935 13946936 13946937 13946938 13946939 13946940 13946941 13946942 13946943 13946944 13946945 13946946 13946947 13946948 13946949 13946950 13946951 14154342 14159805 14159806 14159807 14159808 14159809 14159810 14159811 14159812 14159813 14159814 14159815 14159816 14159817 14159818 14159819 14320362 13956953 13956954 13956955 13956956 13956957 14355670 14355671 13987449 14014008 14014009 14014010 14014011 14014012 14014013 14014014 14014015 14014016 14014017 14014018 14014019 14014021 14406618 14406619 14406620 14406621 13947568 14167150 14167151 14167152 14167153 14167155 14163249 14163250 14163251 14163252 14163253 14160183 14160185 14160188 14160190 14160191 14160192 14160193 14160194 14160195 14160196 14160197 14160198 14160199 14163288 14163289 14163290 14163291 14163292 14143446 14143447 14458459 14458460 14458461 14458462 14458463 14458465 14458466 14458467 14458468 14458469 14370411 14370412 14370413 14370414 14370416 14370417 14370418 14370419 14370420 14370421 14370422 14370423 14370424 14370425 14370426 14370427 14370428 14370429 14370431 14370433 14370434 14370435 14370436 14370437 14370438 14370439 14370441 14370442 14370444 14370447 14370449 14370451 14370452 14370453 14370454 14372158 14461288 14461290 14461292 14136364 14136366 14136367 14136368 14136369 14136370 14017165 14017166 14017167 13940442 13940443 13940444 13940445 13940446 13940447 13943670 13943671 14414678 13909359 13909361 13909362 13909363 13909364 13909365 13909366 13909367 13909368 13909369 13909370 13909371 13909372 13909373 13909374 13909375 13909376 13909377 13909378 13909379 13914559 13914560 13914561 13914562 13914563 13914565 13914567 13914568 13914569 13914570 13914571 13914572 13914573 13914574 14470282 14470286 14480520 14480521 14480522 14480524 14480526 14480527 14480528 14480529 14480532 14480533 14480534 14480535 14480536 14480537 14480538 14480539 14480541 14480542 14480543 14480544 14480545 14480547 14480548 14480550 14480551 14480555 14480556 14480557 14480558 13891406 13891407 13891408 13891409 13891410 13891411 13891412 13891413 13919153 13950328 13950329 13950330 13950331 13950332 13950333 13950334 13941885 13943810 13943811 13943812 13943813 13943814 13943815 13943816 13943817 14192793 14192794 14192795 14192796 14192797 14192798 14192799 14192800 14192801 14192802 14192803 14192804 14192805 14192806 14192807 14192808 14192809 14192810 14192811 14192812 14192813 14192814 14192815 14192816 14192817 14192818 14192819 14414679 14414680 14414681 14414682 14414683 14414684 14414685 14414686 14414687 14414688 14414689 14414690 14414691 14414692 14414693 14414694 14414695 14414696 14414697 14414698 14414699 14414700 14414703 14414705 14414706 14414707 14414708 14414709 14414710 14414711 14414712 14414714 14414715 14414716 14414717 14414718 14335853 14335854 14185307 14185308 14185309 14346476 14346477 14346478 14346479 14346480 14346481 14346482 14346483 14346484 14346485 14346486 14346487 14346490 14346492 14346495 14346498 14346500 14346502 14346503 14346504 14346505 14346506 14346507 14346508 14346509 14346510 14346511 14346512 14346516 14346517 14346518 13970088 13972832 13972833 13984454 13984457 13984460 13984464 13984468 13984469 13984470 13984471 13984472 13984473 13984474 13984475 13984476 13984477 14278947 14448923 14448924 14448925 14448927 14448928 14388665 14127087 14127089 14127093 14127094 14127096 14127099 14127101 14127102 14127103 14127104 14127107 14127112 14438078 14469968 14469969 14469971 14469972 14469974 14469976 13920667 13940423 13907771 13907772 13940424 14040976 14398427 14398428 14020734 14020735 14020736 14020737 14020738 14020740 14020742 14020743 14020746 14020747 14020748 14020752 14020753 14020754 14020756 14020757 14020758 14020766 14020767 14020769 14020774 14020775 14020794 13941757 13941758 13941759 13941760 13941761 13941762 13941763 13941764 13941765 13941766 13941767 13941768 13941769 13941770 13941772 13941774 13941777 13941778 13941779 14119042 13993493 14226799 14226800 14226801 14226802 14230211 14230212 14230213 14230214 14230215 14230216 14230217 14230218 14230219 14256142 14256143 14256144 14256149 14256150 14478436 14478437 14478438 14478439 14478440 14478441 14478442 14478443 14478444 14478445 14478446 13877147 13877150 14213829 14213833 14463749 14109514 14045042 14045043 14045044 14045045 14045046 14045047 14045051 14045054 14045055 14045056 14045057 14024319 14024320 14024321 14024322 14024323 14024324 14024325 14024326 14024327 14024328 14024329 14024330 14024331 14404593 14404594 13914483 13914484 13914485 13914486 13914488 13914489 13914490 13914492 14214235 14428887 14428888 14428889 14428890 14428891 14428892 14428893 14428894 14428895 13962228 13987651 13987652 13987653 13987654 13987655 13987656 13987657 13987659 13987663 13987664 13987666 13987667 13987668 13987669 13987670 13987671 13987672 13987673 13987674 13987675 13987676 13987677 13987678 13987679 13987681 13987682 13987683 13987684 13987685 13987686 13899996 13899997 13899998 13899999 13941755 13941756 13947052 13947053 13947054 13947055 13947056 13947057 13947058 13947059 13947060 13947061 13947062 13947063 13947064 13947065 14212135 14212137 14175755 14175756 14175757 14175758 14175759 14175760 14175761 14175762 14175763 14175765 14175766 14175767 14175768 14175770 14175771 14175772 14175773 14407414 14407415 14407416 14407417 14407418 14407419 14407420 14407421 14407422 14407423 14407424 14407425 14407426 14407427 14407428 14407429 14407430 14407431 14407432 14407433 14407434 14407435 14407436 14407437 14064435 14064436 14064437 14188693 14188694 14188695 14188696 14188697 14188698 14188699 14188700 14188701 14384225 14384226 14384227 14428141 14428143 14428144 14428146 14428148 14428150 14428152 14428154 14428156 14428159 14428162 14428164 14428166 14428168 14428171 14428173 14428175 14428177 14428179 14428180 14428181 14428182 14428183 14428184 14428185 14428186 14428187 14428188 14428189 14428191 14428194 14428195 14428196 14428197 14201951 14201952 14201953 14201954 14201955 14201956 14201957 14201958 14201959 14201960 14201961 14201962 14201963 14201964 14201965 14201966 14186912 14186913 14186914 14186915 14186916 14186917 14186918 14186919 14186921 14186922 14186923 14186924 14186925 14186926 14186927 14186928 14186929 14186930 14186931 14186932 14186933 14186935 14186936 14186938 14186940 13887694 13887695 13887697 13887699 13887700 13887701 13887702 13887703 13887704 13887705 13887707 13887709 13887713 13887716 13887717 13887718 13887719 13887720 13887721 13887722 13887723 13887724 14004807 14004809 14004811 14004814 14004817 14004821 14004823 14004825 14004827 14004829 14184891 14254676 14207656 14207657 14207658 14207659 14207660 14207661 14207662 14207663 14207664 14207665 14207666 14463653 14463654 14463655 14463656 14463657 14463658 14463659 14463660 14142080 14142081 14142082 14142083 14142084 14142085 14142086 14142087 14142088 14142089 14142090 14142091 14142092 13940450 13940451 13940452 13940453 13940454 13940455 13940456 13940457 13928239 13928240 13928241 13928242 13928243 13928244 13928245 13928246 13928247 14163254 14163255 14163256 14163257 14163258 14163259 14163260 14168194 14168195 14168196 13946290 13946291 13946293 13946294 13946295 14394708 14394710 14154324 14154325 14154326 14154327 14154328 14154330 14154332 13992666 14428851 13884371 13884372 13884373 13884374 13884375 13884376 13884377 13884378 13884379 13884380 14387363 14387364 14387365 14387366 14387367 14387368 14387369 14387371 14387372 14387373 14387374 14387375 14387376 14387377 14387378 14387379 14470283 14470284 14474572 14474573 14474574 14474575 14474576 14474577 14474578 14474579 14474580 14474581 14474582 14474583 14474584 14474585 14474586 14474587 14474588 14474589 14474590 14474591 14474592 14474593 14474594 14474595 14474596 14474597 14474598 14474599 14483559 14483560 14483561 14483562 14483564 14483566 14483567 14483568 14483569 14483570 14483571 14483572 14483573 14483574 14483575 13894993 13919155 14470277 14146574 14146575 14146576 14146577 14146578 14146579 14146580 14146581 14146582 14146583 14146584 14146585 14146586 14146587 14146588 14146589 14146590 14146591 14146592 14168190 14168191 14168192 14175824 14175825 13951836 13951837 13951838 14457527 14148539 14148540 14148541 14148542 14148543 14148544 14148545 14148546 14148547 14148553 14148554 14148555 14148556 14148557 14148558 14495539 14068780 14068781 14068782 14068783 14068784 14068785 14068786 14068788 14068790 14068791 14068793 14068794 14068796 14068797 13984478 13984479 13984480 13984481 13984482 13984483 13984484 13984490 13984491 13984492 14146536 14146537 14146538 14146539 14146540 14146541 14146542 14146543 14146544 14146545 14356223 14356224 14370350 14384447 14384449 13984485 13984486 13984487 13984488 13984489 13972831 14024300 14024301 14024302 14024303 14024304 14024305 14024306 14024307 14024308 14209850 14209851 14209852 14209853 14209854 14394713 14394714 14394715 14394716 14394717 14394718 14479870 14151127 14340540 14340541 13901768 13901771 13901773 13901775 13901776 13901778 13901780 13901781 13901782 13901783 13901784 13901785 13901786 13901787 13901788 13902658 13902659 13902660 13902661 13902662 13902664 13902666 13909251 14002102 14002103 14002104 14002105 14002107 14002108 14352474 13896498 13919288 13919289 13919290 13919291 13919293 13919294 13919295 13919296 13919297 13919298 13919299 13919301 14513581 14513582 14513583 14513584 14513585 13914493 13914494 13971471 13971472 13971473 13971474 13971475 13971476 13971477 13971478 13971479 13971480 13971481 13971482 13971483 13971484 13971485 13971486 13973146 13973147 13973148 13973149 14500226 14500227 14500228 14500229 14500230 14500231 14500232 14500233 14500234 14500235 13883650 13883651 13883652 13883653 13883654 13883655 13883656 13883657 13883658 13883659 13883660 13883661 13883662 13883663 13883664 13883665 13883666 13883667 13883668 13956917 13956918 13956919 13956921 13956922 13956923 13941860 13943393 13943394 13943395 13943396 13918498 13918500 13922233 13922234 13922236 13973151 13995349 13995350 13995351 13995352 13995353 13995355 13995357 13995358 13995359 13995360 13995361 13995362 13995363 13995364 13995365 13995366 13995367 13995368 13995369 13995370 14004771 14004772 14004773 14004775 14004777 14004779 14004784 14004787 14004790 14004792 14004795 13922237 13922238 13922239 13922240 13922241 14262830 14262831 14262832 14262833 14262834 14262835 14262836 14262837 14262839 14262842 14262845 14262848 14262850 14262851 14262852 14262853 14262856 14262858 14262859 14262860 14397511 14397513 14397514 14397515 14397517 14397518 14397519 14397520 14397521 14397522 14250974 14470621 14470622 14470623 14470626 14470627 14470628 14470629 14470630 14470631 14470632 14470634 14470636 14470637 14470638 14470639 14470640 14470641 14470642 14470643 14470644 14470645 14470646 14470647 14470648 14470649 14470650 14470651 14470652 14470653 14470654 14470655 14470656 14470659 14470662 14470664 14170032 14392606 14477679 14189711 14202259 14202260 14202261 14202262 14202263 14202264 14191751 14148257 14148258 14148259 13995371 13995372 13995373 13995374 13995375 13995376 13995377 13995378 14282575 14282576 14282577 14282578 14282579 14282580 14282581 14282582 14282583 14440029 14007719 14007721 14007722 14007725 14007726 14007727 14007728 14038473 14038474 14038475 14038476 14038477 14062765 14063080 14063082 14063083 14063084 14063087 14063091 14063097 14407438 14209787 14209788 14214236 14458470 14458472 14458473 14207668 14207669 14207671 14207673 14207674 14207675 14207679 14207681 13948933 13948934 13948935 13948936 14477706 14477707 14477708 14477709 14477710 13927165 13927167 13927168 13927169 13927171 13927172 14158714 14158715 14158716 14158720 14158725 14158727 14292987 14292988 14292989 14292990 14292991 14292992 14292993 14292994 14405259 14472900 14472901 14472902 14472903 14472904 14472906 14472907 14472908 14472909 14472910 14472911 14472912 14472913 14472915 14472916 14472917 14488135 14468732 13966327 13966340 13966341 14202272 14226786 14226787 14226788 14226789 14226790 14226791 14226792 14226793 14226794 14226795 14356225 13967256 13967269 13967270 13967271 13967272 13967273 14457528 14457529 14457530 14457531 14428896 14428897 14428898 14158010 14158011 14158012 14158013 14158014 14158015 14158016 14158017 14207456 14483253 14483254 14483255 14483257 14483265 14483258 13941906 13941907 13941908 13941909 13941910 13941911 13941912 13941913 13941914 13941915 13941916 13917775 13917776 14120538 14384355 14384356 14384357 14458453 14458454 14458455 14458456 14458457 14458458 13952041 13952042 13952043 14236727 14236728 14153569 14153570 14153571 14153572 14153573 14153574 14153575 14153576 14153577 14153578 14153579 14153580 14153581 14172806 14167194 14167195 14180893 14344099 14345927 14345928 14345929 14345930 14345931 14345932 14345933 14345934 14345935 14345936 14345937 14345939 14345942 14345947 14419091 13993465 13993466 14480263 14480264 14480265 14480266 14502121 14502122 14502123 14502124 14502145 14502146 14502148 14502149 14502150 14502152 14502153 14502154 14502155 14502156 14502158 14502159 14502160 14502161 14502162 14502165 14502167 14502168 14502169 14502170 14502172 14502173 14502174 14502175 14502176 14502177 14502178 14503177 14232893 14232898 14232904 14232909 14232911 14232913 14262745 14356233 14356234 14356235 14356236 14356237 14356238 14404251 14404252 14448880 14448882 14355452 14355454 14355456 14355458 14040988 14071096 14071098 13899279 13899281 13940462 13940463 14191748 14191749 14278948 14019265 14019266 14019267 14019268 14019269 14019270 14019271 14019272 14019273 14136520 14136521 13899278 13900000 13900001 13900002 13900003 13900004 13909220 13909224 13909228 13909231 13909232 13909352 13909353 13909354 13909355 13909357 13917780 13917781 13917782 13917783 13917784 13924611 13924612 13924615 13924617 13924620 13929567 13929568 13929569 13929570 13929571 13932073 13932074 13932075 13932076 13932077 13940433 13940434 13941834 13941835 13941837 13941839 13941840 13941917 13941918 13941919 13943141 13943142 13943143 13943144 13993010 14160378 13914518 13914519 13914520 13914521 13914522 13914524 13914525 13914526 13914527 13914529 13914531 13914532 13914533 13914534 13914535 13914536 13914538 13914540 13914541 13950936 13964485 13964486 13964487 13964488 13964489 14463663 14463664 14463665 14463666 14463667 13917767 13917768 13917769 13917772 14024093 14024094 14024095 14024096 14024097 14024098 14024099 14024100 14024101 14024102 13940464 13949746 13949747 13966751 13966752 14021726 14021727 14021728 14042351 14042352 14042353 14055416 14055417 14055418 14067376 14067379 13949917 13949918 13949919 14151145 14471969 13882479 13903104 13903354 13903355 14157060 13952044 13952045 13952046 13952048 13952051 14361270 14361271 14000380 13930859 13930860 13930862 14390581 14390583 14390584 14390585 14118695 14160482 13909287 13909288 13909290 14022351 14022352 14022353 14022354 14022355 14022356 14022360 14022363 14022364 14022365 14022367 14428867 13989680 13989681 14154320 14154321 14154322 14154323 14499829 14499830 14499831 14499832 14499835 13966763 13919156 13919157 13919158 13919159 14024803 14024804 14024805 14024806 14024807 14024808 14024809 14024810 14024811 14025979 14025981 14025983 14025987 14025990 14025991 13950917 13950918 13950919 13950920 13989674 13989677 13989678 13989679 14031582 14031583 14031584 14031585 14031586 14031588 14031591 14031592 14031594 14031595 14031596 14031598 14031599 14031600 14031601 13883679 13883680 14283024 14283025 13951844 13951845 13951846 13951847 13888593 13888594 13888595 13888596 13888597 14196539 14428857 14428858 14428859 14428860 13889483 13889485 13889486 13889487 13889488 13921728 13921730 13921731 13921732 14227414 14227415 14151137 14024332 14024333 14024334 14024335 14024336 14024337 14024338 14024339 14024340 14071260 14071261 14071262 14071263 14318497 14374033 14374034 14374035 14374036 14374037 14374038 14374039 14374042 14374043 14374044 14163085 14163086 14163087 14402427 14402428 14402430 14402432 14402434 14402436 14402437 14402438 14402439 14402442 14402445 14402448 14402450 14402451 14402452 14402453 13877240 13877241 13877243 13877244 13877245 13877246 13877247 13877248 13877249 13877250 13877251 13877252 13924621 13924622 13924624 14166373 14015400 14015401 14015405 14015406 14015407 14015408 14015409 14015410 14015411 14015412 14015413 14015414 14015415 14015417 14015418 14015419 14015421 14015422 14015423 14015424 14015425 13877157 13877158 13877159 13877160 13877202 13877203 13877204 13877205 13877206 14130440 13949109 13949110 13949111 13949112 13949113 13949114 13949115 13949116 13949117 13949119 13949120 13949122 14480262 14183497 14183498 14183499 14183510 14183511 14355651 14355652 14355653 14355654 14355655 14355656 14355658 14355659 13909240 13909243 13909244 13909246 14251041 14251045 14156497 14156498 14467721 14474601 14474602 14474603 14480268 14467418 14467419 14467420 14467421 14467422 14467423 14467424 14467425 14373348 14373349 14373351 14373353 14373357 14373359 14373360 14373362 14373363 14373364 14373366 14373367 14373368 14373369 14373371 14373374 14373377 14148050 14163095 14163096 14163097 13971468 13971469 13971470 14207627 14207628 14207629 14207631 14180876 14180878 14180880 14180881 14227445 14250970 14497460 14497461 14497462 14497463 14497464 14497465 14517173 14031603 14031604 14031605 14031606 14031607 14031608 14184885 14184886 14184889 14184890 14184892 14155521 14155522 14155523 14155524 14155525 14155526 14155527 14155528 14155529 13970089 13970091 13970093 13970096 13970097 13970099 13970102 13970106 13970110 13970111 13970112 13970113 13877253 13877254 13877257 13877260 13877261 13877262 13877284 13877285 13877286 13877287 14285640 14421087 14421088 14421089 14421091 14148529 14148530 14148531 14148532 14148533 14148534 14148535 14148536 14148537 14148538 13989665 13989666 13989667 13989668 13989669 13989670 14167068 14167069 14167070 14221497 14423370 14423371 14423372 13875773 14007684 14025956 14025957 14025958 14025959 14025960 14025961 14025962 14025963 14025964 14025965 14025967 14025969 14025971 14025972 14025973 14025974 14025975 14025976 14025977 14145024 13990539 13990540 13990542 14044996 14044997 14044998 14044999 14045000 14045001 14045002 14045004 14045005 13949748 13949749 13949750 13949751 13949752 13987650 13921006 13921007 13921008 13921009 13921010 13921011 13921012 14167073 14185305 14194903 14194907 14194909 14194910 14194911 14194912 14194913 14194914 14194915 14194917 14194918 14194919 14194922 14194923 14194924 14194927 14194930 14194934 14194936 14194938 14384345 14003469 14003470 14003471 14003472 14007691 14007694 14007695 14007696 14007697 14007699 13892227 13892228 13892229 13892230 13870999 13871001 13877144 14217081 14217082 14217083 14217084 14217086 14217087 14217088 14217089 14217090 14217091 14217092 14217093 14217094 14217095 14217096 14217097 14217098 14217099 14217101 14217103 14217105 14217108 14217109 14217111 14217112 14217113 14217114 14217115 14217117 14217120 14217121 14217122 14241815 14241816 14241817 14241819 14241820 14241821 14143558 14143559 14143562 14143563 14143564 14143565 14143566 14143567 14143568 14143569 |
Recalling Firm/ Manufacturer |
CareFusion 303, Inc. 10020 Pacific Mesa Blvd San Diego CA 92121-4386
|
For Additional Information Contact |
858-617-2000
|
Manufacturer Reason for Recall |
Customers reported issues when scanning medications with the scanners. When a user scans a medication using the affected barcode scanner, information on a different medication could be communicated to the Pyxis product.
|
FDA Determined Cause 2 |
Device Design |
Action |
BD sent an Urgent Medical Device Recall Notification letter dated January 15, 2015 to all affected customers The letter identified the affected product, problem and actions to be taken. Customers are instructed to complete the enclosed Customer Response Card by selecting the desired deployment option and return the card to CareFusion. Customers with questions are instructed to contact the CareFusion Support Center at (888) 562-6018. Customers with adverse event reports are instructed to contact CareFusion Customer Advocacy at (800) 836-7842. Customers with technical questions are instructed to contact CareFusion Technical Support at (800) 727-6102. |
Quantity in Commerce |
9,345 units |
Distribution |
Worldwide:Distribution - US Nationwide and the countries of Saudi Arabia, Bahrain, Australia, United Arab Emirates, Qatar, Mexico, Guam and the Bahamas. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|