|
Class 3 Device Recall CDl H/S Cuvette Perfusion Pack or Smart Perfusion Pack |
|
Date Initiated by Firm |
January 19, 2016 |
Create Date |
February 25, 2016 |
Recall Status1 |
Terminated 3 on October 24, 2016 |
Recall Number |
Z-1018-2016 |
Recall Event ID |
73126 |
510(K)Number |
K881330
|
Product Classification |
Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
|
Product |
CDI H/S Cuvette, non-sterile, Manufactured by Terumo Cardiovascular Group, Perfusion Pack or Smart Perfusion Pack.
Product Usage: Heart/Lung Perfusion Pack is either a customized tubing pack built to user specifications or a stock tubing pack, designed for use during surgery requiring cardiopulmonary bypass. It is not designed for long-term use (greater than six hours). Perfusion Packs labeled as SMART¿ contain components that have an additive to improve blood compatibility. Non-SMART packs may contain some SMART tubing and/or connectors depending on the customer requirements.
|
Code Information |
Catalog Number 6922, 1/2 inch -lots TE27, TH06 and TK03 Catalog Number 6923, 3/8 inch - lots TE13, TE18, TG15, TH20 and TK17 b. Model Number(s): 020204612, 020208807, 020210805, 020211806, 020493806, 044002300, 044003200, 044004102, 044004700, 044010200, 044016900, 044017601, 044018700, 044019901, 044027700, 044035101, 075104103, 075104404, 020135802, 088511700, 084103204, 047000500, 046001601, 065255600, 066110500, 075203800, 627265002, 627304502, 084118101, 020277801, 075101900, 627309201, 075104800, 627138002, 627241801, 627305101, 627239301, 044006801, 627347601, 075203501, 627232601, 627123006, 627280901, 088523400, 627282602, 044022001, 627252001, 044020900, 627147304, 627147403, 627153904, 046003400, 627308201, 044013600, 088511801, 084515200, 044009000, 627282001, 044022000, 066109900, 627286901, 075201201, 075202000, 084103204, 088504901, 088508000, 088513300, 088513501, 627133104, 627139306, 627165604, 627166101, 627167301, 627167501, 627167601, 627167701, 627168001, 627168101, 627179903, 627190501, 627190801, 627197802, 627198001, 627198102, 627212201, 627297401, 047002300, 044000600, 627251001, 627254501, 627283501, 627283601, 627321201, 044003901, 088504400, 627198903, 627214003, 044028900, 627231801, 627240802, 627243501, 627243601, 627243701, 627244602, 627244701, 627246002, 627252001, 627256901, 627260701, 627262303, 627266301, 627269502, 627286802, 627286901, 627290001, 627303901, 627304901, 627336301, 627336402, 627343801, 627346601, 627346701, 627349401 Catalog Number 6924, 1/4 inch - lots TE20, TF11, TG01 , TG15, TH 13 and TH27 |
Recalling Firm/ Manufacturer |
Sorin Group USA, Inc. 14401 W 65th Way Arvada CO 80004-3503
|
For Additional Information Contact |
Raymond Reid 303-467-6632
|
Manufacturer Reason for Recall |
The CDI System 500 monitor displays the "H/S DISCONNECT AT CUVETTE" error message when the CDI HIS Cuvette does not make a proper connection with the CDI H/S Probe. When this occurs, blood parameter values for HCT, Hgb and S02 do not display.
|
FDA Determined Cause 2 |
Component design/selection |
Action |
A Safety Alert letter dated January 13, 2015 via registered mail to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check the communication of the Cuvette prior to initiating Cardiopulmonary Bypass (CPS).
The Terumo Safety Alert did not request removal or return of the product. The instructions that Sarin has communicated to the customers is identical to the information from Terumo. Acknowledgement of the letter and understanding of
the contents is required for all customer responses. |
Quantity in Commerce |
6859 |
Distribution |
US Nationwide Distribution and Canada. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DWF and Original Applicant = COBE LABORATORIES, INC.
|
|
|
|