• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Zimmer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Zimmersee related information
Date Initiated by FirmJanuary 11, 2016
Date PostedMarch 12, 2016
Recall Status1 Terminated 3 on April 13, 2017
Recall NumberZ-1106-2016
Recall Event ID 72900
510(K)NumberK060710 
Product Classification Screw, fixation, bone - Product Code HWC
ProductProduct 8 consists of all products under product code HWC, and same usage: Item no: 47483501201 3.5MM CORT. SCREW 12MM LN 47483501401 3.5MM CORT. SCREW 14MM LN 47483501601 3.5MM CORT. SCREW 16MM LN 47483501801 3.5MM CORT. SCREW 18MM LN 47483503601 3.5MM CORT. SCREW 36MM LN 47483507501 3.5MM CORT. SCREW 75MM LN Product Usage: A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Code Information lot no: 61894346 61892658 61892659 61894347 61888124 61614768 62036752 62036753 62036754 
Recalling Firm/
Manufacturer
Zimmer Manufacturing B.V.
Turpeaux Industrial Park Rd 1 km 123.4
Mercedita PR 00715
For Additional Information ContactConsumer Relations Call Center
800-447-5633
Manufacturer Reason
for Recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
FDA Determined
Cause 2
Packaging
ActionZimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity in Commerce99
DistributionUS Nationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HWC
-
-