• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Zimmer

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Zimmer see related information
Date Initiated by Firm January 11, 2016
Date Posted March 12, 2016
Recall Status1 Terminated 3 on April 13, 2017
Recall Number Z-1113-2016
Recall Event ID 72900
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
Product Product 15 consists of all product code JDI
, and same usage:
Item no:
805002802 ASTEL FEM HEAD +0X28MM DI
805002805 ASTEL FEM HEAD +10.5X28MM
805002803 ASTEL FEM HEAD +3.5X28MM
805002804 ASTEL FEM HEAD +7X28MM DI
805002801 ASTEL FEM HEAD -3.5X28MM

Usage:
For use in total or hemi hip arthroplasty





























































































Code Information lot no.: 62154349 62263555 62270747 62315652 62389960 62212034 62270748 62389957 37107513 62154348 62212021 62270746 62299145 62389956 62299145N 
Recalling Firm/
Manufacturer
Zimmer Manufacturing B.V.
Turpeaux Industrial Park Rd 1 km 123.4
Mercedita PR 00715
For Additional Information Contact Consumer Relations Call Center
800-447-5633
Manufacturer Reason
for Recall
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
FDA Determined
Cause 2
Packaging
Action Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to 1.Review the notification and ensure that relevant personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product. 3. Your Zimmer Biomet sales representative will remove the recalled product from your facility. 4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com. 5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative.
Quantity in Commerce 0
Distribution US Nationwide Distribution
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-