| Class 2 Device Recall Natus neoBLUE blanket LED Phototherapy system |  |
Date Initiated by Firm | October 28, 2016 |
Create Date | December 15, 2016 |
Recall Status1 |
Terminated 3 on January 15, 2021 |
Recall Number | Z-0770-2017 |
Recall Event ID |
74587 |
510(K)Number | K103589 |
Product Classification |
Unit, neonatal phototherapy - Product Code LBI
|
Product | Natus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatal Hyperbilirubinemia.
006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad
006254 neoBLUE blanket LED Light Box
006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad
007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR
007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR
007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK
007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK
007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS
007296 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS |
Code Information |
************SERIAL NUMBERS OF UNITS DISTRIBUTED TO US GOVERMENT ACCOUNTS 000089, 000104, 000128, 001342, 001364, 001367, 001371, 001372, 001373, 001374, 001375, 001376, 001377, 001391, 001392, 001393, 001394, 001441, 001442, 001455, 001456, 001463, 001495, 001496, 001497, 001878, 001893, 001912, 001915, 001992, 001993, 002040, 002041, 002165, 002166, 002167, 002168, 002289, 002290, 002291, 002292, 002320, 002462, 002463, 002762, 002763, 002766, 002860, 002870, 002873, 002888, 003594, 003595, 003600, 003604, 003632, 003633, and 003893. ***********SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE US ..000284, ..000287, ..000297, ..000299, ..000310, ..000324, ..000328, ..000365, .000301, .000313, .000321, .000322, .000323, .000352, .000415, .000419, .000420, .000426, .000427, .000435, .000445, .000449, .000467, .000483, .000508, .000518, .000523, .000531, .000541, .000581, .000651, .000670, .000695, .000696, .000697, .000698, 000030, 000031, 000032, 000033, 000035, 000036, 000039, 000041, 000042, 000043, 000044, 000046, 000048, 000050, 000051, 000052, 000053, 000056, 000062, 000070, 000076, 000081, 000082, 000085, 000086, 000087, 000088, 000093, 000094, 000139, 000140, 000143, 000144, 000146, 000162, 000163, 000164, 000166, 000167, 000170, 000180, 000181, 000187, 000189, 000190, 000193, 000195, 000202, 000205, 000208, 000215, 000218, 000221, 000223, 000234, 000235, 000238, 000240, 000242, 000243, 000244, 000245, 000246, 000248, 000249, 000250, 000253, 000254, 000255, 000257, 000263, 000275, 000276, 000281, 000290, 000291, 000292, 000293, 000303, 000317, 000320, 000325, 000332, 000333, 000335, 000336, 000340, 000343, 000344, 000345, 000349, 000350, 000351, 000352, 000353, 000354, 000366, 000411, 000414, 000417, 000418, 000422, 000424, 000425, 000428, 000430, 000432, 000436, 000441, 000442, 000443, 000446, 000452, 000453, 000454, 000456, 000457, 000458, 000459, 000460, 000462, 000469, 000472, 000475, 000477, 000479, 000483, 000489, 000493, 000495, 000499, 000500, 000502, 000511, 000516, 000522, 000528, 000530, 000547, 000551, 000553, 000560, 000561, 000563, 000565, 000570, 000574, 000599, 000604, 000615, 000616, 000619, 000620, 000625, 000628, 000631, 000641, 000643, 000646, 000647, 000648, 000655, 000656, 000662, 000664, 000667, 000671, 000675, 000680, 000681, 000682, 000683, 000684, 000687, 000693, 000694, 000697, 000708, 000709, 000711, 000712, 000715, 000717, 000718, 000719, 000720, 000731, 000732, 000734, 000735, 000736, 000740, 000741, 000743, 000744, 000745, 000749, 000751, 000752, 000754, 000758, 000760, 000762, 000763, 000766, 000767, 000768, 000769, 000770, 000771, 000772, 000774, 000775, 000776, 000781, 000782, 000783, 000784, 000785, 000786, 000787, 000788, 000789, 000790, 000792, 000793, 000794, 000796, 000797, 000798, 000799, 000800, 000801, 000802, 000803, 000804, 000805, 000806, 000807, 000808, 000809, 000810, 000812, 000813, 000814, 000815, 000816, 000817, 000818, 000820, 000821, 000822, 000825, 000826, 000831, 000833, 000836, 000847, 000848, 000849, 000850, 000851, 000852, 000854, 000857, 000858, 000859, 000861, 000862, 000863, 000865, 000866, 000867, 000868, 000869, 000870, 000871, 000873, 000874, 000875, 000876, 000877, 000878, 000879, 000880, 000882, 000884, 000891, 000892, 000893, 000899, 000900, 000901, 000902, 000903, 000904, 000905, 000907, 000908, 000909, 000910, 000912, 000913, 000914, 000915, 000916, 000921, 000922, 000923, 000924, 000925, 000935, 000936, 000937, 000938, 000939, 000940, 000941, 000943, 000944, 000945, 000946, 000947, 000950, 000952, 000953, 000957, 000958, 000959, 000960, 000961, 000962, 000963, 000965, 000966, 000967, 000970, 000972, 000973, 000974, 000976, 000978, 000979, 000981, 000988, 000991, 000992, 000993, 001005, 001006, 001010, 001011, 001012, 001013, 001018, 001019, 001020, 001021, 001022, 001031, 001035, 001036, 001037, 001038, 001039, 001040, 001041, 001042, 001043, 001044, 001046, 001047, 001048, 001050, 001056, 001058, 001059, 001085, 001086, 001087, 001088, 001095, 001103, 001104, 001105, 001106, 001109, 001110, 001111, 001113, 001115, 001117, 001118, 001119, 001121, 001122, 001123, 001124, 001125, 001126, 001127, 001128, 001129, 001130, 001131, 001133, 001134, 001135, 001136, 001137, 001138, 001139, 001140, 001141, 001143, 001144, 001146, 001149, 001150, 001153, 001154, 001155, 001156, 001161, 001162, 001163, 001164, 001165, 001166, 001167, 001190, 001205, 001206, 001207, 001208, 001209, 001211, 001213, 001214, 001215, 001216, 001217, 001218, 001219, 001220, 001221, 001222, 001223, 001224, 001225, 001227, 001228, 001230, 001231, 001233, 001235, 001236, 001239, 001241, 001242, 001249, 001252, 001253, 001254, 001255, 001256, 001257, 001260, 001261, 001262, 001265, 001266, 001268, 001269, 001271, 001272, 001274, 001275, 001277, 001280, 001281, 001283, 001288, 001289, 001290, 001291, 001292, 001293, 001294, 001296, 001297, 001299, 001300, 001301, 001302, 001305, 001307, 001309, 001311, 001312, 001315, 001316, 001319, 001321, 001327, 001329, 001330, 001331, 001334, 001335, 001338, 001340, 001346, 001348, 001349, 001350, 001351, 001353, 001354, 001356, 001357, 001360, 001361, 001362, 001363, 001366, 001368, 001370, 001378, 001379, 001380, 001381, 001382, 001383, 001384, 001385, 001386, 001387, 001388, 001389, 001390, 001397, 001398, 001399, 001400, 001401, 001402, 001403, 001404, 001406, 001407, 001408, 001409, 001411, 001414, 001415, 001416, 001419, 001420, 001425, 001427, 001428, 001429, 001430, 001431, 001432, 001433, 001434, 001435, 001436, 001437, 001439, 001440, 001443, 001444, 001445, 001446, 001447, 001452, 001454, 001457, 001467, 001470, 001472, 001473, 001474, 001475, 001476, 001477, 001478, 001481, 001486, 001492, 001494, 001499, 001500, 001503, 001507, 001528, 001532, 001540, 001541, 001542, 001543, 001544, 001545, 001546, 001547, 001548, 001549, 001550, 001551, 001552, 001553, 001554, 001556, 001557, 001558, 001559, 001560, 001562, 001564, 001565, 001566, 001567, 001568, 001569, 001579, 001580, 001581, 001582, 001583, 001584, 001586, 001587, 001588, 001589, 001590, 001591, 001592, 001593, 001594, 001596, 001597, 001598, 001599, 001600, 001601, 001602, 001603, 001607, 001613, 001614, 001615, 001616, 001617, 001618, 001619, 001620, 001622, 001623, 001624, 001625, 001626, 001627, 001629, 001630, 001631, 001632, 001633, 001634, 001636, 001637, 001638, 001640, 001644, 001645, 001646, 001647, 001648, 001649, 001650, 001651, 001652, 001656, 001657, 001658, 001661, 001662, 001663, 001665, 001666, 001667, 001680, 001681, 001682, 001683, 001685, 001686, 001687, 001689, 001691, 001693, 001695, 001696, 001698, 001701, 001702, 001703, 001709, 001711, 001712, 001714, 001715, 001716, 001717, 001718, 001719, 001720, 001721, 001729, 001733, 001738, 001740, 001741, 001742, 001748, 001752, 001753, 001755, 001757, 001758, 001760, 001761, 001762, 001764, 001765, 001766, 001769, 001771, 001772, 001774, 001777, 001779, 001780, 001781, 001785, 001799, 001813, 001814, 001815, 001817, 001818, 001821, 001822, 001823, 001834, 001836, 001840, 001844, 001845, 001846, 001848, 001850, 001851, 001852, 001855, 001856, 001858, 001860, 001861, 001862, 001863, 001864, 001866, 001867, 001868, 001869, 001872, 001873, 001874, 001875, 001876, 001879, 001880, 001881, 001882, 001883, 001890, 001894, 001895, 001896, 001897, 001898, 001899, 001900, 001902, 001903, 001904, 001905, 001906, 001907, 001908, 001909, 001910, 001916, 001917, 001918, 001919, 001920, 001921, 001922, 001923, 001924, 001925, 001926, 001927, 001928, 001929, 001930, 001932, 001935, 001936, 001937, 001938, 001939, 001940, 001941, 001942, 001943, 001946, 001947, 001948, 001949, 001950, 001951, 001953, 001954, 001955, 001956, 001957, 001958, 001959, 001960, 001961, 001962, 001963, 001964, 001965, 001967, 001969, 001970, 001973, 001974, 001979, 001982, 001983, 001984, 001987, 001988, 001990, 001991, 001994, 002001, 002009, 002010, 002014, 002016, 002017, 002018, 002019, 002020, 002021, 002022, 002023, 002024, 002026, 002027, 002031, 002032, 002034, 002037, 002038, 002043, 002044, 002045, 002046, 002047, 002048, 002049, 002075, 002076, 002077, 002078, 002079, 002080, 002086, 002087, 002088, 002089, 002090, 002091, 002092, 002141, 002142, 002143, 002144, 002145, 002146, 002147, 002148, 002150, 002151, 002152, 002154, 002156, 002157, 002158, 002159, 002160, 002161, 002162, 002163, 002164, 002169, 002170, 002171, 002172, 002173, 002175, 002176, 002179, 002180, 002181, 002182, 002183, 002185, 002186, 002198, 002199, 002200, 002201, 002202, 002203, 002208, 002209, 002212, 002215, 002217, 002218, 002219, 002225, 002226, 002227, 002228, 002233, 002234, 002235, 002236, 002237, 002239, 002240, 002241, 002243, 002247, 002248, 002249, 002250, 002252, 002253, 002254, 002259, 002260, 002261, 002262, 002263, 002264, 002266, 002267, 002268, 002269, 002270, 002271, 002272, 002273, 002274, 002275, 002276, 002277, 002281, 002284, 002285, 002286, 002287, 002288, 002293, 002294, 002295, 002296, 002305, 002306, 002308, 002309, 002310, 002311, 002312, 002313, 002315, 002316, 002318, 002319, 002320, 002321, 002322, 002323, 002324, 002325, 002333, 002334, 002335, 002337, 002338, 002339, 002344, 002345, 002352, 002356, 002357, 002359, 002360, 002361, 002363, 002366, 002367, 002368, 002369, 002370, 002372, 002376, 002378, 002379, 002380, 002381, 002382, 002383, 002386, 002387, 002388, 002389, 002392, 002393, 002394, 002395, 002396, 002397, 002398, 002399, 002400, 002401, 002403, 002404, 002409, 002415, 002416, 002417, 002421, 002422, 002429, 002430, 002431, 002432, 002435, 002436, 002437, 002438, 002439, 002440, 002441, 002442, 002443, 002444, 002445, 002446, 002448, 002449, 002451, 002452, 002453, 002454, 002455, 002456, 002457, 002458, 002459, 002460, 002461, 002469, 002471, 002473, 002476, 002478, 002480, 002482, 002483, 002484, 002485, 002486, 002488, 002489, 002490, 002491, 002492, 002493, 002494, 002495, 002496, 002497, 002498, 002499, 002501, 002502, 002503, 002504, 002505, 002506, 002507, 002508, 002509, 002513, 002514, 002516, 002517, 002519, 002520, 002521, 002527, 002528, 002529, 002530, 002533, 002536, 002539, 002543, 002544, 002545, 002546, 002652, 002653, 002654, 002655, 002657, 002658, 002659, 002660, 002661, 002662, 002663, 002669, 002670, 002671, 002672, 002673, 002675, 002676, 002694, 002695, 002696, 002697, 002698, 002699, 002700, 002701, 002702, 002703, 002704, 002715, 002718, 002719, 002722, 002723, 002724, 002727, 002728, 002729, 002732, 002733, 002735, 002736, 002738, 002739, 002751, 002752, 002753, 002754, 002755, 002757, 002758, 002765, 002767, 002768, 002770, 002771, 002773, 002774, 002775, 002776, 002777, 002778, 002779, 002781, 002782, 002783, 002786, 002787, 002790, 002791, 002792, 002793, 002799, 002800, 002801, 002803, 002807, 002808, 002810, 002811, 002812, 002813, 002814, 002815, 002816, 002817, 002818, 002819, 002820, 002821, 002822, 002823, 002824, 002825, 002827, 002828, 002829, 002830, 002831, 002836, 002837, 002838, 002839, 002840, 002841, 002842, 002858, 002859, 002861, 002863, 002866, 002867, 002868, 002869, 002871, 002872, 002874, 002875, 002876, 002877, 002878, 002879, 002880, 002884, 002886, 002887, 002889, 002890, 002891, 002892, 002893, 002894, 002895, 002899, 002910, 002911, 002913, 002914, 002919, 002926, 002931, 002934, 002946, 002947, 002952, 002959, 002960, 002961, 002962, 002963, 002964, 002965, 002966, 002967, 002968, 002969, 002970, 002971, 002972, 002973, 002974, 002975, 002976, 002977, 002979, 002980, 002981, 002982, 002983, 002984, 002985, 002986, 002987, 002988, 002989, 002990, 002991, 002992, 002993, 002994, 003007, 003009, 003010, 003014, 003015, 003016, 003018, 003019, 003020, 003021, 003022, 003023, 003025, 003026, 003027, 003028, 003029, 003030, 003031, 003032, 003033, 003034, 003035, 003036, 003037, 003039, 003040, 003041, 003042, 003043, 003044, 003045, 003046, 003047, 003053, 003068, 003069, 003070, 003071, 003072, 003074, 003075, 003076, 003077, 003078, 003083, 003095, 003102, 003103, 003104, 003105, 003106, 003107, 003108, 003111, 003112, 003113, 003114, 003115, 003116, 003117, 003118, 003119, 003120, 003121, 003122, 003123, 003124, 003125, 003127, 003128, 003129, 003135, 003136, 003137, 003138, 003139, 003140, 003141, 003142, 003143, 003149, 003150, 003151, 003152, 003153, 003154, 003155, 003156, 003160, 003161, 003583, 003584, 003590, 003592, 003593, 003596, 003597, 003601, 003603, 003605, 003606, 003607, 003608, 003609, 003610, 003611, 003613, 003614, 003615, 003616, 003617, 003618, 003619, 003620, 003621, 003622, 003623, 003624, 003625, 003626, 003627, 003629, 003630, 003631, 003634, 003638, 003639, 003640, 003641, 003642, 003643, 003645, 003646, 003647, 003648, 003649, 003650, 003651, 003657, 003658, 003667, 003668, 003669, 003670, 003671, 003672, 003673, 003674, 003675, 003680, 003681, 003682, 003683, 003684, 003685, 003686, 003687, 003688, 003690, 003691, 003692, 003693, 003694, 003695, 003696, 003697, 003698, 003699, 003700, 003701, 003702, 003708, 003709, 003710, 003711, 003712, 003713, 003714, 003715, 003716, 003717, 003718, 003719, 003720, 003721, 003722, 003723, 003724, 003728, 003729, 003730, 003731, 003732, 003733, 003736, 003737, 003739, 003740, 003743, 003744, 003745, 003746, 003747, 003749, 003750, 003751, 003752, 003753, 003754, 003755, 003756, 003757, 003759, 003760, 003761, 003762, 003763, 003765, 003766, 003767, 003772, 003773, 003774, 003775, 003780, 003786, 003788, 003789, 003791, 003792, 003793, 003794, 003795, 003796, 003797, 003798, 003799, 003800, 003801, 003802, 003803, 003804, 003805, 003806, 003809, 003810, 003811, 003812, 003813, 003814, 003815, 003816, 003817, 003821, 003822, 003823, 003832, 003833, 003834, 003836, 003837, 003838, 003840, 003841, 003845, 003846, 003847, 003853, 003855, 003856, 003858, 003859, 003860, 003861, 003862, 003863, 003866, 003868, 003869, 003870, 003871, 003872, 003873, 003875, 003877, 003878, 003879, 003882, 003885, 003891, 003892, 003896, 003898, 003899, 003900, 003904, 003908, 003910, 003911, 003912, 003913, 003915, 003916, 003918, 003919, 003921, 003924, 003926, 003927, 003930, 003932, 003933, 003934, 003935, 003936, 003937, 003938, 003939, 003940, 003941, 003942, 003943, 003944, 003945, 003946, 003947, 003948, 003949, 003950, 003951, 003952, 003953, 003954, 003955, 003957, 003959, 003961, 003962, 003963, 003964, 003965, 003966, 003967, 003968, 003970, 003971, 003972, 003975, 003976, 003977, 003978, 003979, 003980, 003981, 003982, 003983, 003984, 003988, 003989, 003990, 003991, 003992, 003994, 003995, 003996, 003997, 003998, 004000, 004001, 004002, 004003, 004004, 004005, 004006, 004007, 004008, 004010, 004011, 004012, 004013, 004014, 004015, 004016, 004017, 004018, 004019, 004020, 004021, 004022, 004023, 004024, 004026, 004027, 004028, 004029, 004032, 004033, 004034, 004035, 004041, 004042, 004043, 004072, 004207, 004208, 004209, 004210, 004211, 004212, 004213, 004214, 004216, 004217, 004218, 004219, 004220, 004221, 004222, 004223, 004224, 004225, 004226, 004227, 004228, 004229, 004231, 004232, 004233, 004234, 004236, 004237, 004238, 004240, 004241, 004246, 004247, 004248, 004249, 004250, 004251, 004252, 004253, 004254, 004255, 004256, 004257, 004258, 004259, 004260, 004261, 004263, 004264, 004268, 004269, 004270, 004271, 004273, 1215R002299, 2298, 2300, 2302, 2303, 2304, NR002392, R002300, R002301, R002357, R002359, R002360, R002362, R002364, R002365, R002368, R002369, R002370, R002389, R002413, R002415, R002416, and R002417. ************SERIAL NUMBERS OF UNITS DISTRIBUTED OUTSIDE US ..000283, 000061, 000064, 000066, 000067, 000069, 000078, 000079, 000121, 000123, 000125, 000129, 000130, 000131, 000136, 000137, 000147, 000148, 000149, 000150, 000152, 000153, 000155, 000168, 000169, 000173, 000174, 000182, 000183, 000192, 000194, 000196, 000198, 000200, 000201, 000211, 000212, 000216, 000219, 000227, 000280, 000295, 000296, 000298, 000300, 000302, 000306, 000309, 000315, 000326, 000334, 000356, 000357, 000368, 000369, 000412, 000423, 000429, 000437, 000451, 000471, 000515, 000521, 000536, 000537, 000539, 000542, 000544, 000548, 000549, 000550, 000581, 000586, 000589, 000590, 000591, 000592, 000593, 000607, 000608, 000609, 000610, 000617, 000624, 000626, 000630, 000633, 000635, 000636, 000637, 000638, 000639, 000642, 000650, 000653, 000654, 000657, 000672, 000673, 000685, 000686, 000691, 000692, 000701, 000706, 000707, 000713, 000714, 000716, 000721, 000723, 000724, 000725, 000726, 000728, 000729, 000737, 000738, 000739, 000746, 000747, 000748, 000757, 000764, 000765, 000777, 000778, 000779, 000780, 000840, 000885, 000886, 000887, 000888, 000889, 000890, 000896, 000897, 000918, 000919, 000920, 000931, 000932, 000933, 000934, 000956, 000975, 000977, 000980, 000994, 000995, 000996, 000997, 001001, 001002, 001007, 001008, 001009, 001014, 001015, 001016, 001017, 001049, 001051, 001052, 001053, 001054, 001055, 001057, 001060, 001061, 001062, 001063, 001064, 001065, 001066, 001069, 001070, 001071, 001072, 001073, 001074, 001075, 001076, 001077, 001078, 001079, 001080, 001081, 001082, 001083, 001084, 001089, 001090, 001091, 001092, 001093, 001097, 001098, 001099, 001100, 001101, 001102, 001114, 001172, 001178, 001179, 001187, 001188, 001192, 001193, 001194, 001198, 001244, 001245, 001246, 001247, 001248, 001250, 001251, 001320, 001325, 001326, 001328, 001332, 001333, 001336, 001339, 001341, 001343, 001345, 001352, 001355, 001358, 001359, 001365, 001395, 001396, 001417, 001418, 001423, 001450, 001451, 001453, 001458, 001459, 001461, 001462, 001479, 001490, 001491, 001509, 001511, 001512, 001513, 001514, 001515, 001516, 001519, 001520, 001521, 001524, 001526, 001561, 001570, 001571, 001572, 001573, 001574, 001576, 001577, 001578, 001604, 001605, 001606, 001608, 001642, 001643, 001659, 001660, 001668, 001669, 001671, 001672, 001673, 001674, 001675, 001676, 001677, 001678, 001694, 001726, 001727, 001728, 001730, 001731, 001732, 001734, 001735, 001736, 001743, 001744, 001745, 001746, 001749, 001750, 001751, 001763, 001773, 001789, 001793, 001796, 001798, 001801, 001802, 001804, 001843, 001859, 001884, 001885, 001888, 001889, 001891, 001892, 001933, 001934, 001968, 001976, 001977, 001985, 001995, 001996, 001997, 001998, 001999, 002000, 002002, 002003, 002004, 002005, 002007, 002011, 002012, 002013, 002025, 002028, 002036, 002039, 002051, 002052, 002053, 002054, 002055, 002056, 002057, 002058, 002059, 002061, 002062, 002065, 002066, 002067, 002068, 002069, 002070, 002071, 002072, 002073, 002074, 002081, 002082, 002083, 002084, 002085, 002187, 002188, 002191, 002192, 002193, 002194, 002195, 002197, 002204, 002205, 002206, 002207, 002210, 002211, 002213, 002214, 002216, 002220, 002221, 002222, 002223, 002224, 002229, 002230, 002231, 002232, 002244, 002245, 002255, 002256, 002257, 002282, 002307, 002318, 002342, 002347, 002350, 002351, 002353, 002354, 002364, 002371, 002372, 002373, 002374, 002375, 002376, 002382, 002383, 002384, 002385, 002391, 002394, 002401, 002402, 002404, 002405, 002406, 002407, 002408, 002409, 002410, 002411, 002412, 002413, 002414, 002423, 002424, 002425, 002426, 002428, 002450, 002464, 002465, 002467, 002477, 002479, 002510, 002511, 002531, 002532, 002534, 002535, 002664, 002665, 002666, 002667, 002668, 002677, 002679, 002681, 002682, 002683, 002684, 002687, 002688, 002689, 002691, 002692, 002693, 002706, 002708, 002709, 002710, 002712, 002713, 002714, 002720, 002730, 002731, 002734, 002737, 002740, 002741, 002742, 002743, 002744, 002745, 002746, 002747, 002748, 002749, 002750, 002756, 002759, 002760, 002761, 002764, 002772, 002797, 002804, 002805, 002806, 002809, 002826, 002834, 002835, 002864, 002865, 002881, 002882, 002916, 002920, 002923, 002927, 002928, 002929, 002930, 002932, 002933, 002935, 002936, 002937, 002938, 002939, 002940, 002941, 002942, 002943, 002944, 002945, 002948, 002949, 002950, 002954, 002955, 002956, 002957, 002996, 002997, 002999, 003000, 003001, 003002, 003003, 003005, 003011, 003012, 003017, 003038, 003051, 003052, 003055, 003056, 003057, 003058, 003059, 003060, 003061, 003062, 003063, 003065, 003066, 003067, 003073, 003081, 003084, 003085, 003086, 003087, 003088, 003089, 003090, 003091, 003092, 003093, 003094, 003096, 003097, 003098, 003099, 003100, 003101, 003130, 003131, 003132, 003133, 003134, 003144, 003145, 003146, 003148, 003157, 003158, 003159, 003162, 003585, 003586, 003587, 003588, 003591, 003598, 003599, 003636, 003637, 003644, 003652, 003653, 003654, 003655, 003656, 003676, 003678, 003704, 003705, 003706, 003707, 003725, 003726, 003727, 003734, 003735, 003738, 003741, 003742, 003770, 003771, 003776, 003777, 003778, 003779, 003781, 003782, 003790, 003807, 003808, 003818, 003819, 003820, 003835, 003842, 003843, 003844, 003848, 003849, 003850, 003851, 003852, 003854, 003864, 003865, 003867, 003876, 003880, 003881, 003884, 003886, 003887, 003888, 003890, 003895, 003897, 003902, 003903, 003905, 003906, 003907, 003909, 003917, 003923, 003925, 003931, 003960, 003969, 003993, 004242, 004243, 004267, 004272, 1517, 1911000001, NR002393, R002358, R002366, R002371, R002390, and R002414. |
Recalling Firm/ Manufacturer |
Natus Medical Incorporated 5900 1st Ave S Seattle WA 98108-3248
|
Manufacturer Reason for Recall | neoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding discoloration/ degradation of fiberoptic bundle at connection of pad and box. |
FDA Determined Cause 2 | Device Design |
Action | Natus began sending the URGENT RECALL NOTIFICATION / FIELD SAFETY NOTICE neoBLUE blanket system performance letter dated October 27, 2016 and the Technical Bulletin to US customers on October 28, 2016.
Natus plans to notify their international customers on December 2, 2016.
Customers were informed in March 2015 of a potential issue of discoloration/degradation/melting of the fiber optic bundle at the pad connector which is inserted in the neoBLUE blanket light box. Included with that letter was a Technical Bulletin which described the issue and addressed many of your potential questions.
The purpose of this current letter (dated October 27, 2016) is to communicate Natus' plan for a follow-on field correction for neoBLUE blanket systems shipped between 2011 through March 21, 2016 (Serial Numbers up to and including xxxx004282), and to remind customers of the instructions contained in the neoBLUE blanket Technical Bulletin.
Natus is in the process of confirming an updated neoBLUE blanket configuration which will not be susceptible to the degradation described above. However, the date of project completion is uncertain at this time.
Natus will contact customers when the new configuration is available, at which time Natus intend to initiate a field corrective action to replace or upgrade your device(s).
Please contact Natus Quality_Programs@natus.com if you have any further questions. Please call (800) 426-0353. |
Quantity in Commerce | 2619 units (1925 units in the US and 694 units OUTSIDE the US) |
Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of : including American Samoa and Guam, and in the following countries:
Albania, Argentina, Australia, Austria, Bangladesh, Canada, Chile, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, Indonesia, Iraq, Israel, Italy, Korea, Kuwait, Latvia, Malaysia, Mongolia, Namibia, Netherlands, Nigeria, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia (Slovak Republic), Slovenia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Virgin Islands (British). |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LBI
|
|
|
|