| Class 2 Device Recall Sysmex |  |
Date Initiated by Firm | July 01, 2016 |
Date Posted | September 09, 2016 |
Recall Status1 |
Terminated 3 on February 10, 2017 |
Recall Number | Z-2770-2016 |
Recall Event ID |
74654 |
Product Classification |
Stains, hematology - Product Code KQC
|
Product | Fluorocell PLT reagent, Catalog #CD994563
Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers |
Code Information |
Lot number A5053, Expiration date 12/21/2016 Lot number A5054, Expiration date 12/21/2016 Lot number A5041, Expiration date 9/2/2016 Lot number A5042, Expiration date 9/2/2016 Lot number A5044, Expiration date 10/2/2016 Lot number A5046, Expiration date 11/9/2016 Lot number A5047, Expiration date 11/17/2016 Lot number A5048, Expiration date 11/17/2016 Lot number A5049, Expiration date 11/26/2016 Lot number A5050, Expiration date 11/26/2016 Lot number A5055, Expiration date 12/24/2016 |
Recalling Firm/ Manufacturer |
Sysmex America Inc 577 Aptakisic Rd Lincolnshire IL 60069-4325
|
For Additional Information Contact | Sysmex Technical Assistance Center 888-879-7639 |
Manufacturer Reason for Recall | There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the Impedance platelet (PLT-I) value and a false low PLT-F value. In most cases, the falsely decreased PLT-F results displayed a "PLT Abn Scattergram" message with asterisks beside the results indicating the data is unreliable. In rare instances, platelet flags may be absent. The PLT-I counts are unaffected.
The PLT-F is a reflex test used to confirm PLT-I results due to flagging or low values. As such, the likely impact to patient results is low. Affected results display a discrepancy between the PLT-I and PLT-F, with lower than expected PLT-F results that do not match the patient's clinical picture. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | Sysmex sent a Product Notification Letter dated August 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. . Customers were instructed to cease use and discard the affected lots, and reply to Sysmex America, Inc. using the provided fax-back form.
An updated Product Notification (Attachment 1) was mailed to all XN-Series customers on August 12, 2016 informing them of the additional recalled lots. This notification included the phone number for the Technical Assistance Center at 1-888-879-7639 in the U.S., and 1-888-679-7639 in Canada. |
Quantity in Commerce | 3868 units |
Distribution | Worldwide Distribution - US (Nationwide, Washington DC) Canada; Cuba; Chile; and Dominican Republic. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|