Class 2 Device Recall Roche cobas p 612 PreAnalytical Instrument calculator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Roche cobas p 612 PreAnalytical Instrument calculatorsee related information
Date Initiated by FirmJuly 15, 2016
Create DateOctober 25, 2016
Recall Status1 Terminated 3 on June 05, 2017
Recall NumberZ-0106-2017
Recall Event ID 75186
Product Classification Calculator/data processing module, for clinical use - Product Code JQP
Productcobas p 612 Pre-Analytical Instrument calculator/data processing module, for clinical use Product Usage: The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module.
Code Information Not applicable
Recalling Firm/
Manufacturer
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis IN 46256-1025
For Additional Information ContactRoche Support Network Customer Support C
800-428-2336
Manufacturer Reason
for Recall
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
FDA Determined
Cause 2
Device Design
ActionOn 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum.
Quantity in Commerce9014 in total
DistributionUS Nationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-