Date Initiated by Firm |
October 06, 2016 |
Create Date |
May 04, 2017 |
Recall Status1 |
Terminated 3 on July 26, 2019 |
Recall Number |
Z-2030-2017 |
Recall Event ID |
76059 |
510(K)Number |
K011444
|
Product Classification |
Table and attachments, operating-room - Product Code BWN
|
Product |
5.0cm Heavy Duty Short Attachment |
Code Information |
All serial numbers shipped to customers prior to 09/27/2016; DFU 18-0048; Part No. SHORT-HD. |
Recalling Firm/ Manufacturer |
The Anspach Effort, Inc. 4500 Riverside Dr Palm Beach Gardens FL 33410-4235
|
For Additional Information Contact |
Anne Brisson 800-327-6687
|
Manufacturer Reason for Recall |
Supplied Directions for Use (DFU) did not contain a recommended service interval.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
Consignees were sent a recall notice via mail, dated 10/10/2016. In the notice, the firm notifies consignees of the Directions For Use (DFU) correction that will include recommended servicing intervals. All DFUs distributed after 09/28/2016 will have the correct DFU. The recalling firm instructs consignees to review the service intervals provided for product received, forward the notification to necessary parties and/or facilities, and complete the attached Verification Form. Complete Verification Forms are to be sent by fax to 877-546-5069 or by email to Anspach4870@stericycle.com. Consignees are also asked to maintain a copy of the notice for their records. Any questions can be directed to Customer Service at 1-800-327-6887, option 1. |
Quantity in Commerce |
2339 |
Distribution |
Worldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = BWN and Original Applicant = THE ANSPACH EFFORT, INC.
|