Date Initiated by Firm |
February 27, 2017 |
Date Posted |
March 15, 2017 |
Recall Status1 |
Terminated 3 on March 01, 2018 |
Recall Number |
Z-1543-2017 |
Recall Event ID |
76593 |
PMA Number |
P100021 |
Product Classification |
System, endovascular graft, aortic aneurysm treatment - Product Code MIH
|
Product |
Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166E ETBF2313C124E ETBF2513C145E ETBF2516C124E ETBF2316C124E ETBF2513C124E.
Cardiology: The Endurant family of stent graft systems is Intended to treat infra-renal abdominal aortic or aortoiliac aneurysms using an endovascular approach. When placed within the aneurysm, the stent graft provides an alternative conduit for blood flow within the patient's vasculature. |
Code Information |
Product UPN/GTIN Serial number 00613994992048 V05884896 00643169215849 V05981026 00643169215849 V05981027 00643169215849 V05981030 00643169215849 V05981031 00643169215849 V05981032 00643169215849 V05981033 00643169215849 V05981034 00643169215849 V05981035 00613994992024 V05981039 00613994992024 V05981040 00613994992024 V05981044 00613994992024 V05981045 00643169215856 V05981415 00643169215856 V05981416 00643169215856 V05981417 00643169215856 V05981418 00643169215856 V05981419 00643169215856 V05981420 00643169215856 V05981421 00643169215856 V05981422 00643169215856 V05981423 00643169215856 V05981424 00613994991263 V05981721 00613994991263 V05981722 00613994991263 V05981723 00613994991263 V05981724 00613994991263 V05981725 00613994991263 V05981726 00613994991263 V05981728 00643169215795 V05982579 00643169215795 V05982581 00643169215795 V05982584 00643169215795 V05982586 00613994992048 V05983552 00613994992048 V05983556 00613994992048 V05983558 00613994991980 V05984180 00613994991980 V05984181 00613994991980 V05984182 00613994991980 V05984183 00613994991980 V05984184 00613994991980 V05984185 00613994991980 V05984186 00613994991980 V05984187 00613994991980 V05984188 00613994991980 V05984189 00613994992031 V05985031 00613994992048 V05985511 00613994992048 V05985512 00613994992048 V05985513 00613994992048 V05985514 00613994992048 V05985515 00613994992048 V05985516 00613994992048 V05985517 00613994992048 V05985518 00613994992048 V05985519 00613994992048 V05985520 00613994992048 V05985521 00613994992048 V05985522 00613994992048 V05985523 00613994992048 V05985524 00613994992048 V05985525 00613994992048 V05985526 00613994992048 V05985527 00613994992048 V05985528 00613994992048 V05985529 00613994992048 V05985530 00613994644268 V05986025 00613994644268 V05986033 00643169215856 V05986324 00643169215856 V05986325 00643169215856 V05986326 00643169215856 V05986327 00643169215856 V05986328 00643169215856 V05986329 00643169215856 V05986330 00643169215856 V05986331 00643169215856 V05986332 00643169215856 V05986333 00643169215856 V05986334 00643169215856 V05986335 00643169215856 V05986336 00643169215856 V05986338 00643169215856 V05986339 00643169215856 V05986340 00643169215856 V05986341 00643169215856 V05986342 00643169215856 V05986343 00613994991980 V05986382 00613994991980 V05986384 00613994991980 V05986385 00613994991980 V05986386 00613994991980 V05986387 00613994991980 V05986388 00643169215832 V05986404 00643169215832 V05986406 00643169215832 V05986407 00643169215832 V05986408 00643169215832 V05986409 00643169215832 V05986410 00643169215832 V05986411 00643169215832 V05986412 00643169215832 V05986413 00613994644251 V05986594 00613994644251 V05986595 00613994644251 V05986596 00613994644251 V05986597 00613994644251 V05986600 00613994644251 V05986601 00613994644251 V05986603 00613994644251 V05986604 00613994644251 V05986605 00613994644251 V05986606 00613994644251 V05986607 00613994644251 V05986609 00613994644251 V05986610 00643169215856 V05988370 00643169215856 V05988371 00643169215856 V05988372 00643169215856 V05988373 00643169215856 V05988374 00643169215856 V05988375 00643169215856 V05988376 00643169215856 V05988377 00643169215856 V05988378 00613994644251 V05989424 00613994644251 V05989425 00613994644251 V05989426 00613994644251 V05989427 00643169215856 V05990730 00613994991263 V05994424 00613994991263 V05994425 00613994991263 V05994434 00613994991263 V05994435 00613994991263 V05994436 00613994991263 V05994437 00613994991263 V05994438 00613994991263 V05994439 00613994991263 V05994440 00613994991263 V05994441 00613994991263 V05994442 00613994992031 V05996522 00613994992031 V05996523 00613994992031 V05996524 00613994992031 V05996525 00613994992031 V05996526 00613994992031 V05996527 00613994992031 V05996528 00613994992031 V05996529 00613994992031 V05996530 00613994992031 V05996531 00613994992031 V05996532 00613994992031 V05996534 00613994992031 V05996537 00613994992031 V05996539 00613994992031 V05996541 00643169215832 V05997276 00643169215832 V05997277 00643169215832 V05997278 00643169215832 V05997279 00643169215832 V05997280 00613994644251 V05998082 00613994992024 V05998735 00613994992024 V05998736 00613994992024 V05998737 00613994992024 V05998738 00613994992024 V05998739 00613994992024 V05998740 00613994992024 V05998741 00613994992024 V05998742 00613994992024 V05998743 00613994992024 V05998744 00613994644251 V06001373 00613994644251 V06001375 00613994992024 V06006166 00613994992024 V06006167 00613994992024 V06006168 00613994992024 V06006169 00613994992024 V06006170 00613994992024 V06006171 00613994992024 V06006173 00613994992024 V06006174 00613994992048 V06015225 00613994992048 V06015230 00613994992048 V06015231 00613994992048 V06015235 00613994992048 V06015236 00613994992048 V06015237 00613994992048 V06015238 00613994992048 V06015239 00613994992048 V06015240 00613994992048 V06015241 00613994992048 V06015242 00613994992048 V06015243 00613994992048 V06015244 00613994644251 V06016733 00613994644251 V06016735 00613994644251 V06016738 00643169215788 V06017287 00643169215788 V06017290 00613994644251 V06017881 00613994644251 V06017884 00613994644251 V06017885 00613994644251 V06017886 00613994644251 V06017887 00613994991966 V06018652 00613994991966 V06018654 00613994991966 V06018658 00613994991966 V06018661 00613994992048 V06018960 00613994992048 V06018961 00613994992048 V06018962 00613994992048 V06018963 00613994992048 V06018964 00613994992048 V06018965 00613994992048 V06018966 00613994992048 V06018967 00613994992048 V06018968 00613994992048 V06018969 00613994644268 V06019724 00613994644268 V06019725 00613994644268 V06019727 00613994644268 V06019728 00613994644268 V06019730 00613994644268 V06019732 00613994644268 V06019733 00613994644336 V06019734 00613994644336 V06019735 00613994644336 V06019736 00613994644336 V06019740 00613994644336 V06019743 00613994991973 V06020155 00613994991973 V06020156 00613994991973 V06020157 00613994991973 V06020158 00613994991973 V06020159 00613994991980 V06020519 00613994991980 V06020520 00613994991980 V06020521 00613994991980 V06020523 00613994991980 V06020524 00613994991980 V06020525 00613994991980 V06020527 00613994991980 V06020528 00613994992048 V06020529 00613994992048 V06020530 00613994992048 V06020531 00613994992048 V06020532 00613994992048 V06020533 00613994992048 V06020534 00613994992048 V06020535 00613994992048 V06020536 00613994992048 V06020537 00613994992048 V06020538 00613994644282 V06022983 00613994644350 V06023672 00613994644350 V06023673 00613994644350 V06023674 00613994644350 V06023675 00613994644350 V06023677 00613994644350 V06023678 00613994644350 V06023679 00613994644350 V06023680 00613994644343 V06023682 00613994644343 V06023684 00613994644343 V06023685 00613994644343 V06023688 00613994644343 V06023689 00613994644343 V06023690 00613994644343 V06023691 00643169215856 V06025014 00643169215856 V06025015 00643169215856 V06025016 00643169215856 V06025017 00643169215856 V06025018 00643169215856 V06025019 00643169215856 V06025020 00643169215856 V06025021 00643169215856 V06025022 00643169215856 V06025023 00613994644282 V06025292 00613994644282 V06025295 00613994644268 V06025979 00613994644268 V06025980 00613994644268 V06025981 00613994644268 V06025982 00613994644268 V06025983 00613994644329 V06027045 00613994644329 V06027046 00613994644329 V06027047 00613994644329 V06027048 00613994991270 V06027250 00613994991270 V06027251 00613994991270 V06027252 00613994991270 V06027253 00613994991270 V06027254 00613994991270 V06027255 00613994991270 V06027256 00613994991270 V06027257 00613994991270 V06027258 00613994991270 V06027259 00613994992031 V06028387 00613994992031 V06028391 00613994992031 V06028392 00613994991980 V06028395 00613994991980 V06028396 00613994991980 V06028397 00613994991980 V06028398 00613994991980 V06028399 00613994991980 V06028400 00613994991980 V06028401 00613994991980 V06028402 00613994991980 V06028403 00613994991980 V06028404 00643169215856 V06028531 00643169215856 V06028532 00643169215856 V06028533 00643169215856 V06028534 00643169215856 V06028535 00613994992048 V06030367 00613994992048 V06030368 00613994992048 V06030369 00613994992048 V06030370 00613994992048 V06030371 00613994992048 V06030372 00613994992048 V06030373 00613994992048 V06030374 00613994992048 V06030375 00613994992048 V06030376 00613994991201 V06030559 00613994991201 V06030560 00613994991201 V06030561 00613994991201 V06030563 00613994992048 V06030816 00613994992048 V06030817 00613994992048 V06030818 00613994992048 V06030819 00613994992048 V06030820 00613994992048 V06030821 00613994992048 V06030822 00613994992048 V06030823 00613994992048 V06030824 00613994992031 V06030956 00613994992031 V06030957 00613994992031 V06030958 00613994992031 V06030959 00613994992031 V06030960 00613994992031 V06030961 00613994992031 V06030962 00613994992031 V06030963 00613994992031 V06030964 00613994992031 V06030965 00643169215856 V06033761 00643169215856 V06033762 00643169215856 V06033763 00643169215856 V06033764 00643169215856 V06033765 00643169215856 V06033766 00643169215856 V06033767 00643169215856 V06033768 00643169215856 V06033769 00643169215856 V06033770 00613994992048 V06033881 00613994992048 V06033882 00613994992048 V06033883 00613994992048 V06033884 00613994992048 V06033885 00613994992048 V06033886 00613994992048 V06033887 00613994992048 V06033888 00613994992048 V06033889 00613994992048 V06033890 00613994991980 V06034608 00613994991980 V06034609 00613994991980 V06034610 00613994991980 V06034611 00613994991980 V06034612 00613994991980 V06034613 00613994991980 V06034617 00613994992031 V06034618 00613994992031 V06034621 00613994992031 V06034622 00613994992031 V06034624 00613994992031 V06034625 00613994992031 V06034626 00613994992048 V06034692 00613994992048 V06034693 00613994992048 V06034694 00613994992048 V06034695 00613994992048 V06034696 00613994992048 V06034697 00613994992048 V06034698 00613994992048 V06034699 00613994992048 V06034700 00613994991942 V06034738 00613994991942 V06034739 00613994991942 V06034741 00613994991942 V06034743 00613994991942 V06034744 00613994991942 V06034745 00613994991942 V06034746 00613994991942 V06034747 00613994991980 V06034981 00613994991980 V06034982 00613994991980 V06034983 00613994991980 V06034984 00613994991980 V06034985 00613994991942 V06034993 00613994992048 V06035001 00613994991966 V06035306 00613994991966 V06035310 00643169215795 V06035376 00643169215795 V06035377 00643169215795 V06035378 00643169215795 V06035379 00643169215795 V06035380 00643169215849 V06035967 00643169215849 V06035968 00643169215849 V06035969 00643169215849 V06035970 00613994992031 V06036366 00643169215856 V06036388 00643169215856 V06036389 00643169215856 V06036392 00643169215856 V06036395 00613994992048 V06037086 00613994991959 V06038191 00613994991959 V06038195 00613994992017 V06038257 00613994992017 V06038259 00613994644367 V06038696 00613994644367 V06038699 00613994644367 V06038701 00643169215788 V06038843 00643169215788 V06038846 00643169215788 V06038847 00613994644299 V06039631 00613994644299 V06039633 00613994991942 V06039698 00613994991942 V06039699 00613994991942 V06039700 00613994991942 V06039701 00613994991942 V06039702 00613994991942 V06039703 00613994991171 V06039728 00613994991171 V06039729 00613994991171 V06039730 00613994991171 V06039731 00613994991171 V06039732 00613994991171 V06039733 00613994991171 V06039734 00613994991171 V06039736 00613994991171 V06039737 00613994992000 V06039749 00613994992000 V06039757 00613994644282 V06039779 00613994644282 V06039780 00613994644282 V06039782 00613994644282 V06039783 00613994644329 V06039799 00613994644329 V06039803 00643169215757 V06039838 00643169215757 V06039839 00643169215757 V06039840 00643169215757 V06039841 00643169215757 V06039842 00643169215757 V06039843 00643169215757 V06039845 00643169215757 V06039847 00643169215825 V06039848 00643169215825 V06039849 00643169215825 V06039850 00643169215825 V06039851 00643169215825 V06039852 00643169215764 V06039889 00643169215764 V06039890 00643169215764 V06039891 00643169215764 V06039893 00643169215764 V06039894 00643169215764 V06039896 00643169215764 V06039897 00613994991980 V06039903 00613994991980 V06039906 00613994644312 V06041136 00613994644312 V06041142 00613994644312 V06041143 00613994644282 V06042353 00613994644282 V06042354 00613994644282 V06042355 00613994644282 V06042358 00613994644282 V06042359 00613994644329 V06042841 00613994644329 V06042842 00643169215795 V06042996 00643169215795 V06042997 00643169215795 V06042998 00643169215795 V06042999 00643169215795 V06043000 00643169215795 V06043152 00643169215795 V06043153 00643169215795 V06043154 00643169215788 V06043379 00643169215788 V06043380 00643169215788 V06043381 00643169215788 V06043382 00643169215788 V06043383 00643169215788 V06043384 00643169215788 V06043386 00643169215788 V06043388 00613994644299 V06044040 00613994644299 V06044042 00613994991249 V06044239 00613994991249 V06044240 00613994991249 V06044241 00613994991249 V06044242 00613994991249 V06044243 00613994991249 V06044244 00613994991249 V06044245 00613994991249 V06044246 00613994991249 V06044247 00613994644343 V06044471 00613994992000 V06044691 00613994992000 V06044692 00613994992000 V06044693 00613994992000 V06044695 00613994992000 V06044696 00613994992000 V06044699 00613994992000 V06044700 00613994991249 V06045388 00613994991249 V06045389 00613994991249 V06045390 00613994991249 V06045391 00613994991249 V06045392 00613994644367 V06045822 00613994991980 V06048434 00613994991980 V06048435 00613994991980 V06048436 00613994991980 V06048437 00613994991980 V06048438 00613994992048 V06048754 00613994992048 V06048755 00613994992048 V06048757 00613994992048 V06048760 00613994992048 V06048761 00613994992048 V06048764 00613994992048 V06048765 00613994992048 V06048766 00613994992048 V06048767 00613994992048 V06048768 00613994992048 V06048769 00613994992048 V06048770 00613994992048 V06048771 00613994992048 V06048772 00613994992048 V06048773 00613994991188 V06050434 00613994991188 V06050435 00613994991188 V06050436 00613994991188 V06050437 00613994991188 V06050438 00643169215825 V06053882 00643169215825 V06053884 00643169215825 V06053885 00643169215825 V06053886 00643169215825 V06053887 00643169215825 V06053888 00643169215825 V06053889 00643169215825 V06053890 00613994991270 V06053902 00613994991270 V06053903 00613994991270 V06053904 00613994991270 V06053905 00613994991270 V06053906 00613994992000 V06053953 00613994992000 V06053954 00613994992000 V06053959 00613994992000 V06053960 00613994992000 V06053961 00613994992048 V06054042 00613994992048 V06054045 00613994992048 V06054047 00613994992048 V06054048 00613994992048 V06054049 00613994992048 V06054050 00613994992048 V06054051 00613994992000 V06056169 00613994992000 V06056170 00613994992000 V06056171 00613994992000 V06056172 00613994992000 V06056173 00613994992031 V06056279 00613994992031 V06056286 00643169215795 V06056303 00643169215832 V06056304 00643169215832 V06056305 00643169215832 V06056307 00613994992017 V06056399 00613994992017 V06056400 00613994992017 V06056401 00613994992017 V06056402 00613994992017 V06056403 00613994992017 V06056404 00613994992017 V06056406 00613994992017 V06056407 00613994992017 V06056408 00613994991980 V06056409 00613994991980 V06056411 00613994991980 V06056412 00613994991980 V06056413 00613994991973 V06056580 00613994991973 V06056581 00613994991973 V06056582 00613994991973 V06056583 00613994991973 V06056584 00643169215849 V06057075 00643169215849 V06057076 00643169215849 V06057077 00643169215849 V06057078 00643169215849 V06057079 00613994992031 V06057633 00613994992031 V06057635 00613994992024 V06057706 00613994992024 V06057707 00613994992024 V06057708 00613994992024 V06057709 00613994992024 V06057710 00613994991218 V06061265 00613994991218 V06061266 00613994991218 V06061267 00613994991218 V06061268 00613994991218 V06061269 00643169215825 V06061282 00613994992031 V06062573 00613994992048 V06062578 00613994991997 V06062767 00613994991997 V06062769 00613994991997 V06062770 00643169215740 V06063480 00643169215740 V06063483 00643169215740 V06063485 00643169215740 V06063486 00643169215740 V06063487 00643169215740 V06063488 00643169215740 V06063489 00613994991959 V06064144 00613994991959 V06064145 00613994991959 V06064147 00613994991959 V06064148 00613994991959 V06064149 00613994991959 V06064150 00613994991959 V06064151 00613994991959 V06064152 00613994991959 V06064153 00613994991997 V06065138 00613994991997 V06065141 00613994991973 V06065647 00613994991973 V06065648 00613994991973 V06065649 00613994991973 V06065650 00613994991973 V06065652 00613994991973 V06065653 00613994991164 V06066524 00613994991164 V06066525 00613994991164 V06066526 00613994991164 V06066529 00613994991980 V06066544 00613994991980 V06066546 00613994991980 V06066547 00613994991980 V06066548 00613994991980 V06066552 00613994991980 V06066554 00613994991980 V06066555 00613994991980 V06066556 00613994991980 V06066557 00613994991980 V06066559 00613994992024 V06071685 00643169215771 V06072137 00643169215771 V06072139 00643169215771 V06072140 00613994991959 V06075507 00613994991959 V06075508 00613994991959 V06075509 00613994991959 V06075510 00613994991959 V06075511 00613994991959 V06075512 00613994644312 V06079017 00613994644312 V06079022 00613994644312 V06079023 00613994644312 V06079024 00613994991188 V06079611 00613994991188 V06079613 00613994991188 V06079614 00613994991188 V06079615 00613994644329 V06081002 00613994644329 V06081003 00613994644329 V06081004 00613994644329 V06081005 00613994644329 V06081006 00613994644329 V06081009 00613994644329 V06081010 00613994644329 V06081011 00613994992000 V06081896 00613994992000 V06081901 00613994992000 V06081904 00613994992017 V06082298 00613994992017 V06082299 00613994992017 V06082300 00613994992017 V06083701 00613994992017 V06083703 00613994992017 V06083704 00613994992017 V06083705 00613994992017 V06083706 00613994992017 V06083707 00613994992017 V06083708 00613994992017 V06083709 00613994992017 V06083710 00613994992017 V06083711 00613994992017 V06083712 00613994992017 V06083713 00613994992017 V06083714 00613994992017 V06083715 00613994992017 V06083716 00613994992017 V06083717 00643169215818 V06084755 00643169215818 V06084756 00643169215818 V06084757 00643169215818 V06084758 00643169215818 V06084759 00643169215818 V06084760 00643169215818 V06084761 00643169215818 V06084763 00643169215818 V06084764 00613994991973 V06085186 00613994991973 V06085187 00613994991973 V06085188 00613994991973 V06085191 00613994991973 V06085192 00613994991973 V06087777 00613994991973 V06087778 00613994991973 V06087779 00613994991973 V06087780 00613994991973 V06087781 00613994991973 V06087782 00613994991973 V06087783 00613994991973 V06087784 00613994991973 V06087785 00613994991973 V06087786 00613994991997 V06089726 00613994991997 V06089727 00613994991997 V06089728 00613994991997 V06089729 00613994991997 V06089733 00613994991997 V06089735 00643169215764 V06090211 00643169215764 V06090212 00643169215764 V06090217 00643169215764 V06090222 00643169215764 V06090225 00643169215764 V06090226 00643169215757 V06090618 00613994991188 V06092188 00613994991188 V06092189 00613994991188 V06092191 00613994991188 V06092192 00643169215795 V06093535 00643169215795 V06093536 00643169215795 V06093537 00643169215795 V06093538 00613994991942 V06094424 00613994991942 V06094425 00613994991942 V06094426 00613994991942 V06094427 00613994991942 V06094428 00613994991942 V06094429 00613994991942 V06094430 00613994991942 V06094431 00613994991942 V06094432 00613994991942 V06094433 00613994992024 V06095028 00613994992024 V06095029 00613994992024 V06095030 00613994992024 V06095032 00613994992024 V06095033 00613994992024 V06095034 00613994992024 V06095035 00613994991980 V06095036 00613994991980 V06095037 00613994991980 V06095038 00613994991980 V06095039 00613994991980 V06095040 00613994991980 V06095041 00613994991980 V06095042 00613994991980 V06095043 00613994991980 V06095044 00613994991980 V06095045 00643169215764 V06095659 00643169215764 V06095660 00643169215764 V06095661 00643169215764 V06095662 00643169215764 V06095665 00643169215764 V06095666 00643169215764 V06095667 00643169215764 V06095668 00613994991973 V06095959 00643169215795 V06095974 00643169215795 V06095975 00643169215795 V06095976 00643169215795 V06095977 00643169215795 V06095978 00643169215795 V06095979 00643169215795 V06095980 00643169215795 V06095981 00643169215795 V06095982 00643169215795 V06095983 00643169215788 V06096579 00643169215788 V06096580 00643169215788 V06096581 00643169215788 V06096582 00643169215788 V06096583 00643169215788 V06096584 00643169215788 V06096585 00643169215788 V06096586 00643169215788 V06096587 00643169215788 V06096588 00613994991959 V06096599 00613994991959 V06096600 00613994991959 V06096601 00613994991959 V06096602 00613994991959 V06096603 00613994991959 V06096604 00613994991959 V06096605 00613994991959 V06096606 00613994991959 V06096607 00613994991959 V06096608 00613994992031 V06097517 00613994992031 V06097519 00613994992031 V06097520 00613994992031 V06097521 00613994992031 V06097525 00643169215849 V06097557 00643169215849 V06097558 00643169215849 V06097559 00643169215849 V06097563 00643169215849 V06097564 00643169215849 V06097566 00643169215832 V06098105 00643169215832 V06098106 00643169215832 V06098107 00643169215832 V06098108 00643169215832 V06098109 00643169215832 V06098110 00643169215832 V06098111 00643169215832 V06098112 00643169215832 V06098113 00643169215832 V06098114 00613994991188 V06099088 00613994991188 V06099089 00613994991188 V06099090 00613994991188 V06099091 00613994991188 V06099092 00613994992000 V06099103 00613994992000 V06099104 00613994992000 V06099105 00613994992000 V06099106 00613994992000 V06099107 00613994992000 V06099108 00613994992000 V06099109 00613994992000 V06099110 00613994992000 V06099111 00613994992000 V06099112 00613994991270 V06099118 00613994991270 V06099121 00613994991270 V06099122 00613994991270 V06099123 00613994991270 V06099126 00613994991270 V06099127 00613994992048 V06099541 00613994992048 V06099542 00613994992048 V06099543 00613994992048 V06099544 00613994992048 V06099545 00613994992048 V06099546 00613994992048 V06099547 00613994992048 V06099548 00613994992048 V06099549 00613994992048 V06099550 00613994991232 V06099891 00613994991232 V06099892 00613994991232 V06099893 00613994991232 V06099894 00613994991232 V06099895 00613994991232 V06099896 00613994991232 V06099897 00613994991232 V06099898 00613994991232 V06099899 00613994991232 V06099900 00613994991263 V06099911 00613994991263 V06099912 00613994991263 V06099913 00613994991263 V06099915 00613994991263 V06099916 00613994991263 V06099917 00613994991263 V06099918 00613994991263 V06099919 00613994991263 V06099920 00613994991270 V06099921 00613994991270 V06099922 00613994991270 V06099923 00613994991270 V06099924 00613994991270 V06099925 00613994991270 V06099926 00613994991270 V06099927 00613994991270 V06099928 00613994991270 V06099929 00613994991270 V06099930 00613994991959 V06100271 00613994991959 V06100272 00613994991959 V06100273 00613994991959 V06100274 00613994991959 V06100275 00613994991959 V06100276 00613994991959 V06100277 00613994991959 V06100278 00613994991959 V06100279 00613994991959 V06100280 00613994991171 V06101646 00613994991171 V06101647 00613994991171 V06101648 00613994991171 V06101649 00613994991171 V06101650 00613994991973 V06101711 00613994991973 V06101712 00613994991973 V06101713 00613994991973 V06101714 00613994991973 V06101715 00613994991973 V06101716 00613994991973 V06101717 00613994991973 V06101718 00613994991973 V06101719 00613994991973 V06101720 00613994992048 V06102285 00613994992048 V06102286 00613994992048 V06102287 00613994992048 V06102288 00613994992048 V06102289 00613994992048 V06102290 00613994992048 V06102291 00613994992048 V06102292 00613994992048 V06102293 00613994992048 V06102294 00613994992000 V06102295 00613994992000 V06102296 00613994992000 V06102297 00613994992000 V06102298 00613994992000 V06102299 00613994992000 V06102300 00613994992000 V06102351 00613994992000 V06102352 00613994992000 V06102353 00613994991942 V06102989 00613994991942 V06102990 00613994991942 V06102991 00613994991942 V06102992 00613994991942 V06102993 00613994991942 V06102994 00613994991942 V06102995 00613994991942 V06102996 00613994991942 V06102997 00613994991942 V06102998 00613994991164 V06103426 00613994991164 V06103428 00613994991164 V06103431 00613994991164 V06103432 00613994991164 V06103433 00613994991188 V06103434 00613994991188 V06103435 00613994991188 V06103436 00613994991188 V06103437 00613994991188 V06103438 00613994991188 V06103439 00613994991188 V06103440 00613994991188 V06103441 00613994991188 V06103442 00613994644343 V06104066 00613994644343 V06104068 00613994992048 V06104266 00613994992048 V06104271 00613994992048 V06104275 00613994991980 V06104688 00613994991980 V06104689 00613994991980 V06104690 00613994991980 V06104691 00613994991980 V06104692 00613994991980 V06104693 00613994991980 V06104694 00613994991980 V06104695 00613994991980 V06104696 00613994991980 V06104697 00613994992024 V06104699 00613994992024 V06104752 00613994644299 V06105480 00613994991973 V06105678 00613994991973 V06105679 00613994991973 V06105680 00613994991973 V06105681 00613994991973 V06105682 00613994991973 V06105683 00613994991973 V06105684 00613994991973 V06105685 00613994991973 V06105686 00613994991973 V06105687 00613994644282 V06106020 00613994992031 V06106201 00613994992031 V06106203 00613994992031 V06106204 00613994992031 V06106205 00613994992031 V06106206 00613994992031 V06106207 00613994992031 V06106208 00613994992031 V06106209 00613994992031 V06106210 00613994992017 V06106982 00613994992017 V06106983 00613994992017 V06106984 00613994992017 V06106985 00613994992017 V06106986 00613994992017 V06106987 00613994992017 V06106988 00613994992017 V06106989 00613994992017 V06106991 00613994992000 V06107021 00613994992000 V06107022 00613994992000 V06107023 00613994992000 V06107024 00613994992000 V06107025 00613994992000 V06107026 00613994992000 V06107027 00613994992000 V06107028 00613994992000 V06107029 00613994992000 V06107030 00643169215818 V06108110 00643169215818 V06108111 00643169215818 V06108112 00643169215818 V06108113 00643169215818 V06108114 00643169215818 V06108115 00643169215818 V06108116 00643169215818 V06108117 00643169215818 V06108118 00643169215818 V06108119 00613994644299 V06108833 00613994991997 V06109356 00613994991997 V06109357 00613994991997 V06109359 00613994991997 V06109360 00613994991997 V06109361 00613994991997 V06109362 00613994991997 V06109363 00613994991997 V06109364 00613994991997 V06109365 00613994992048 V06109543 00613994992048 V06109544 00613994992048 V06109545 00613994992048 V06109546 00613994992048 V06109547 00613994992048 V06109548 00613994992048 V06109549 00613994992048 V06109550 00613994992048 V06110051 00613994992048 V06110052 00643169215825 V06110093 00643169215825 V06110094 00643169215825 V06110095 00643169215825 V06110096 00643169215825 V06110097 00643169215825 V06110098 00643169215825 V06110099 00643169215825 V06110100 00643169215825 V06110101 00643169215825 V06110102 00613994991942 V06110320 00613994991942 V06110321 00613994991942 V06110322 00613994991942 V06110323 00613994991942 V06110324 00613994991942 V06110325 00613994991942 V06110326 00613994991942 V06110327 00613994991942 V06110328 00613994991942 V06110329 00613994991959 V06110330 00613994991959 V06110331 00613994991959 V06110332 00613994991959 V06110333 00613994991959 V06110334 00613994992031 V06111109 00613994992031 V06111110 00613994992031 V06111111 00613994992031 V06111112 00613994992031 V06111113 00613994992031 V06111114 00613994992031 V06111115 00613994992031 V06111116 00613994991980 V06111149 00613994991980 V06111150 00613994991980 V06111151 00613994991980 V06111152 00613994991980 V06111153 00613994991980 V06111154 00613994991980 V06111155 00613994991980 V06111156 00613994991980 V06111157 00613994991980 V06111158 00613994992048 V06111722 00613994992048 V06111724 00613994992048 V06111725 00613994992048 V06111726 00613994992048 V06111727 00613994992048 V06111728 00613994992048 V06111729 00613994992048 V06111730 00613994992048 V06111731 00613994992017 V06111742 00613994992017 V06111743 00613994992017 V06111744 00613994992017 V06111745 00613994992017 V06111746 00613994992017 V06111747 00613994992017 V06111748 00613994992017 V06111749 00613994992017 V06111750 00613994992017 V06111751 00643169475892 V06111833 00643169475892 V06111834 00643169475892 V06111835 00643169475892 V06111836 00643169475892 V06111837 00613994991249 V06111838 00613994991249 V06111839 00613994991249 V06111840 00613994991249 V06111841 00613994991249 V06111842 00643169475922 V06112572 00643169475922 V06112573 00643169475922 V06112574 00643169475922 V06112575 00643169475922 V06112576 00643169475922 V06112577 00643169475922 V06112578 00643169475922 V06112579 00643169475922 V06112580 00643169475922 V06112581 00643169475915 V06112582 00643169475915 V06112583 00643169475915 V06112584 00643169475915 V06112585 00643169475915 V06112586 00643169475915 V06112587 00643169475915 V06112588 00643169475915 V06112589 00643169475915 V06112590 00643169475915 V06112591 00613994991935 V06113044 00613994991935 V06113045 00613994991935 V06113048 00613994991935 V06113049 00613994991935 V06113050 00613994991935 V06113052 00613994991935 V06113053 00643169475885 V06113065 00643169475885 V06113066 00643169475885 V06113067 00643169475885 V06113068 00613994991959 V06113549 00613994991959 V06113550 00613994991959 V06113557 00613994991959 V06113558 00613994991270 V06113579 00613994991270 V06113580 00613994991270 V06113581 00613994991270 V06113582 00613994991270 V06113583 00613994991270 V06113584 00613994991270 V06113585 00613994991270 V06113586 00613994991270 V06113587 00613994991270 V06113588 00613994992024 V06114855 00613994992024 V06114856 00613994992024 V06114857 00613994992024 V06114858 00613994992024 V06114860 00613994992024 V06114861 00613994992024 V06114862 00613994992024 V06114863 00613994992048 V06115281 00613994992048 V06115282 00613994992048 V06115283 00613994992048 V06115284 00613994992048 V06115285 00613994992048 V06115286 00613994992048 V06115287 00613994992048 V06115288 00613994992048 V06115289 00613994992048 V06115290 00613994992031 V06115311 00613994992031 V06115312 00613994992031 V06115313 00613994992031 V06115314 00613994992031 V06115315 00613994992031 V06115316 00613994992031 V06115317 00613994992031 V06115318 00613994992031 V06115319 00613994992031 V06115320 00613994991218 V06116028 00613994991218 V06116029 00613994991218 V06116030 00613994991218 V06116031 00613994991218 V06116032 00613994991218 V06116033 00613994991218 V06116034 00613994991218 V06116035 00613994991218 V06116036 00613994991218 V06116037 00643169215856 V06116058 00643169215856 V06116059 00643169215856 V06116060 00643169215856 V06116061 00643169215856 V06116062 00643169215856 V06116063 00643169215856 V06116064 00643169215856 V06116065 00643169215856 V06116066 00613994991973 V06116069 00613994991973 V06116070 00613994991973 V06116072 00613994991973 V06116073 00613994991973 V06116074 00613994991973 V06116075 00613994992031 V06116757 00613994992031 V06116758 00613994992031 V06116759 00613994992031 V06116760 00613994992031 V06116761 00613994992031 V06116762 00613994992031 V06116763 00613994992031 V06116764 00613994992031 V06116765 00613994992031 V06116766 00613994992048 V06116767 00613994992048 V06116768 00613994992048 V06116769 00613994992048 V06116770 00613994992048 V06116771 00613994992048 V06116772 00613994992048 V06116773 00613994992048 V06116774 00613994992048 V06116775 00613994991942 V06117487 00613994991942 V06117488 00613994991942 V06117489 00613994991942 V06117490 00613994991942 V06117491 00613994991942 V06117492 00613994991942 V06117493 00613994991942 V06117494 00613994991942 V06117495 00613994992048 V06117497 00613994992048 V06117498 00613994992048 V06117499 00613994992048 V06117500 00613994992017 V06117521 00613994992017 V06117522 00613994992017 V06117525 00613994992017 V06117526 00613994992017 V06117527 00613994992017 V06117528 00643169215849 V06117569 00643169215849 V06117570 00643169215849 V06117571 00643169215849 V06117572 00643169215849 V06117573 00643169215849 V06117574 00643169215849 V06117575 00643169215849 V06117576 00643169215849 V06117577 00643169215849 V06117578 00613994991966 V06117958 00613994991966 V06117961 00613994991966 V06117963 00613994991966 V06117964 00613994991997 V06117968 00613994991997 V06117969 00613994991997 V06117971 00613994992048 V06117973 00613994992048 V06117974 00613994992048 V06117975 00613994992048 V06117976 00613994992048 V06117977 00613994992048 V06117978 00613994992048 V06117979 00613994992048 V06117980 00613994992048 V06117981 00613994644343 V06118210 00613994644343 V06118214 00613994644343 V06118217 00613994644343 V06118218 00643169215795 V06118330 00643169215795 V06118331 00643169215795 V06118332 00643169215795 V06118333 00643169215795 V06118334 00643169215795 V06118335 00643169215795 V06118336 00643169215795 V06118337 00643169215795 V06118338 00643169215795 V06118339 00613994992031 V06118973 00613994992031 V06118974 00613994992031 V06118975 00613994992031 V06118976 00613994992031 V06118977 00613994992031 V06118978 00613994992031 V06118979 00613994992031 V06118980 00613994992031 V06118981 00613994992048 V06119301 00613994992048 V06119302 00613994992048 V06119303 00613994992048 V06119304 00613994992048 V06119305 00613994992048 V06119306 00643169215788 V06119580 00643169215788 V06119581 00643169215788 V06119582 00643169215788 V06119583 00643169215788 V06119584 00643169215856 V06119585 00643169215856 V06119586 00643169215856 V06119587 00643169215856 V06119588 00643169215856 V06119589 00643169215856 V06119590 00643169215856 V06119591 00643169215856 V06119592 00643169215856 V06119593 00643169215856 V06119594 00613994991171 V06119600 00613994991171 V06119601 00613994991171 V06119602 00613994991171 V06119603 00613994991171 V06119604 00613994991171 V06119605 00613994991171 V06119606 00613994991171 V06119607 00613994991171 V06119608 00613994991270 V06119610 00613994991270 V06119611 00613994991270 V06119612 00613994991270 V06119613 00613994991270 V06119614 00613994991270 V06119615 00613994991270 V06119616 00613994991270 V06119617 00613994991270 V06119618 00613994991270 V06119619 00613994991973 V06120173 00613994991973 V06120174 00613994991973 V06120175 00613994991973 V06120176 00613994991973 V06120177 00613994992017 V06120425 00613994991980 V06120431 00613994991980 V06120432 00613994991980 V06120433 00613994991980 V06120434 00613994991980 V06120435 00613994991980 V06120436 00613994991980 V06120437 00613994991980 V06120438 00613994991980 V06120439 00613994992000 V06120445 00613994992024 V06121835 00613994992024 V06121836 00613994992024 V06121837 00613994992024 V06121838 00613994992024 V06121839 00613994992024 V06121840 00613994992024 V06121841 00613994992024 V06121842 00613994992024 V06121843 00613994992024 V06121844 00613994992031 V06122132 00613994992031 V06122133 00613994992031 V06122134 00613994992031 V06122135 00613994992031 V06122136 00613994992031 V06122137 00613994992031 V06122138 00613994992031 V06122139 00613994992031 V06122140 00613994992031 V06122141 00613994992048 V06122152 00613994992048 V06122153 00613994992048 V06122154 00613994992048 V06122155 00613994992048 V06122156 00613994992048 V06122158 00613994991966 V06122585 00613994991966 V06122586 00613994991966 V06122587 00613994991966 V06122588 00613994991966 V06122589 00613994991966 V06122590 00613994991966 V06122591 00613994991966 V06122592 00613994991966 V06122593 00613994991966 V06122594 00613994992031 V06122605 00613994992031 V06122606 00613994992031 V06122607 00613994992031 V06122608 00613994992031 V06122609 00613994992031 V06122610 00613994992031 V06122611 00613994992031 V06122612 00613994992031 V06122613 00613994992031 V06122614 00613994991980 V06124033 00613994991980 V06124035 00613994991980 V06124036 00613994991980 V06124037 00613994991980 V06124042 00613994992017 V06124553 00613994992017 V06124554 00613994992017 V06124555 00613994992017 V06124556 00613994992017 V06124557 00613994992017 V06124558 00613994992017 V06124559 00613994992017 V06124560 00613994992017 V06124561 00613994992017 V06124562 00613994991942 V06124880 00613994991942 V06124881 00613994991942 V06124885 00613994992031 V06124888 00613994992031 V06124889 00613994992031 V06124891 00613994992031 V06124892 00613994992031 V06124895 00613994992024 V06124897 00613994992024 V06124898 00613994992024 V06125001 00613994992024 V06125004 00613994992024 V06125005 00613994991973 V06125012 00613994992000 V06126054 00613994992000 V06126055 00613994992000 V06126058 00613994992000 V06126059 00613994992000 V06126062 00613994992048 V06127105 00643169215856 V06127530 00643169215856 V06127531 00643169215856 V06127532 00643169215856 V06127533 00643169215856 V06127534 00643169215856 V06127535 00643169215856 V06127536 00643169215856 V06127537 00643169215856 V06127538 00643169475922 V06127572 00643169475922 V06127574 00643169215825 V06127600 00643169215825 V06127651 00643169215825 V06127653 00643169215825 V06127654 00643169215825 V06127655 00643169215825 V06127658 00643169215849 V06128219 00643169215849 V06128221 00643169215856 V06128750 00643169215818 V06129335 00643169215818 V06129336 00643169215818 V06129339 00643169215818 V06129340 00643169215818 V06129342 00643169215818 V06129343 00643169215818 V06129344 00613994644268 V06129946 00613994644268 V06129948 00643169215795 V06130283 00613994644251 V06130556 00613994644251 V06130557 00613994644251 V06130558 00613994644251 V06130559 00613994644251 V06130560 00613994644251 V06130562 00613994644251 V06130563 00613994644251 V06130564 00613994644282 V06130566 00613994644282 V06130568 00613994644282 V06130570 00613994644282 V06130572 00613994644282 V06130573 00613994644282 V06130574 00643169215825 V06130824 00643169215825 V06130825 00643169215825 V06130828 00643169215825 V06130829 00643169215825 V06130831 00613994991263 V06131392 00613994991263 V06131393 00613994991263 V06131394 00613994991263 V06131397 00613994991263 V06131401 00613994991249 V06131938 00613994991249 V06131939 00613994991249 V06131940 00613994991249 V06131941 00613994991249 V06131942 00613994991249 V06131943 00613994991249 V06131944 00613994991249 V06131945 00613994991249 V06131946 00613994991249 V06131947 00613994991249 V06132307 00613994991249 V06132309 00613994991249 V06132310 00613994991249 V06132311 00613994991249 V06132312 00613994991249 V06132313 00613994992017 V06132415 00613994992017 V06132416 00613994992017 V06132417 00613994992017 V06132418 00613994992017 V06132419 00613994992017 V06132420 00613994992017 V06132421 00613994992017 V06132422 00613994992017 V06132423 00613994992017 V06132424 00613994992000 V06132505 00613994992000 V06132507 00613994992000 V06132508 00613994992000 V06132510 00613994992000 V06132511 00613994992000 V06132512 00613994992000 V06132513 00613994992000 V06132514 00613994992048 V06132782 00613994991270 V06132784 00613994991270 V06132785 00613994991270 V06132792 00613994992031 V06132813 00613994992031 V06132814 00613994992031 V06132815 00613994992031 V06132816 00613994992031 V06132817 00613994992031 V06132818 00613994992031 V06132819 00613994992031 V06132820 00613994992031 V06132821 00613994992031 V06132822 00613994991249 V06133122 00613994991249 V06133123 00613994991249 V06133124 00613994991249 V06133125 00613994991249 V06133126 00613994991249 V06133127 00613994991249 V06133128 00613994991249 V06133129 00613994991249 V06133130 00613994991249 V06133131 00613994991171 V06133192 00613994992031 V06133216 00613994992031 V06133219 00613994992031 V06133220 00613994992031 V06133222 00613994992048 V06133962 00613994992048 V06133964 00613994992048 V06133966 00613994992048 V06133967 00613994992048 V06133968 00613994992048 V06133970 00613994992048 V06133971 00613994991997 V06133982 00613994991997 V06133984 00613994992017 V06134374 00613994992017 V06134375 00613994992017 V06134376 00613994992017 V06134377 00613994992017 V06134378 00613994992017 V06134379 00613994992017 V06134380 00613994992017 V06134381 00613994992017 V06134382 00613994992017 V06134383 00643169215849 V06135548 00643169215849 V06135550 00643169215849 V06135551 00643169215849 V06135552 00643169215849 V06135554 00643169215849 V06135555 00643169215849 V06135556 00613994992031 V06136801 00613994992031 V06136803 00613994992031 V06136805 00613994992031 V06136806 00613994992031 V06136807 00613994991966 V06137136 00613994992017 V06137172 00613994992017 V06137173 00613994992017 V06137174 00613994992017 V06137175 00613994992017 V06137176 00613994992017 V06137177 00613994992017 V06137178 00613994992017 V06137180 00643169215818 V06138225 00643169215818 V06138226 00643169215818 V06138227 00643169215818 V06138228 00613994992000 V06139183 00613994992000 V06139184 00613994992000 V06139185 00613994992000 V06139186 00613994992000 V06139187 00613994992000 V06139188 00613994992000 V06139189 00613994992000 V06139190 00613994992000 V06139191 00613994992000 V06139192 00643169475878 V06139860 00643169475878 V06139861 00643169475878 V06139864 00643169475878 V06139865 00643169475878 V06139866 00613994991263 V06140870 00613994991263 V06140871 00613994991263 V06140872 00613994991263 V06140873 00613994991263 V06140874 00613994991263 V06140875 00613994991263 V06140876 00613994991263 V06140877 00613994991263 V06140878 00613994991263 V06140879 00613994992031 V06140881 00613994992031 V06140882 00613994992031 V06140883 00643169215849 V06141030 00643169215849 V06141031 00643169215849 V06141032 00643169215849 V06141034 00643169215849 V06141036 00643169215849 V06141038 00643169215849 V06141039 00643169215825 V06142174 00613994991973 V06142276 00613994991973 V06142277 00613994991973 V06142283 00643169215818 V06142684 00613994992031 V06142815 00613994992031 V06142816 00613994992031 V06142817 00613994992031 V06142818 00613994992031 V06142819 00613994992031 V06142820 00613994992031 V06142821 00613994992031 V06142822 00613994992031 V06142823 00613994992031 V06142824 00613994644268 V06143162 00613994644268 V06143163 00613994644268 V06143167 00613994644282 V06143415 00613994644282 V06143416 00613994644282 V06143417 00613994644282 V06143418 00613994644282 V06143419 00613994644282 V06143420 00613994644282 V06143421 00613994644282 V06143422 00613994644282 V06143423 00613994644299 V06143631 00613994644299 V06143633 00613994644299 V06143636 00613994644299 V06143639 00613994644251 V06144165 00613994644251 V06144168 00613994644251 V06144171 00613994644251 V06144172 00613994644251 V06144173 00613994992024 V06144336 00613994992024 V06144337 00613994992024 V06144340 00613994992024 V06144341 00613994992024 V06144342 00613994992024 V06144345 00643169215788 V06144595 00643169215788 V06144596 00643169215788 V06144597 00643169215788 V06144598 00643169215788 V06144600 00643169215788 V06144601 00643169215788 V06144602 00643169215788 V06144603 00613994644282 V06144740 00613994644282 V06144742 00613994644282 V06144743 00613994644282 V06144744 00613994644268 V06145767 00613994644268 V06145773 00613994644268 V06145774 00613994992031 V06146171 00613994992031 V06146172 00613994992031 V06146173 00613994992031 V06146174 00613994992031 V06146175 00613994992031 V06146176 00613994992031 V06146177 00613994992031 V06146178 00613994992031 V06146179 00613994992031 V06146180 00613994644367 V06146518 00613994644367 V06146521 00643169215849 V06149958 00643169215849 V06149959 00643169215832 V06154796 00613994644329 V06157279 00613994644329 V06157280 00613994644329 V06157282 00613994644329 V06157284 00613994644329 V06157285 00613994644329 V06157288 00613994644299 V06157495 00613994644299 V06157496 00613994644299 V06157497 00613994644299 V06157498 00613994644299 V06157499 00613994992024 V06157822 00613994992024 V06157824 00613994992024 V06157825 00613994992048 V06158363 00613994644282 V06158849 00613994644343 V06158865 00613994644251 V06158897 00613994992000 V06158975 00613994644336 V06160079 00613994644336 V06160086 00613994992024 V06160099 00613994992024 V06160100 00613994992024 V06160101 00613994992024 V06160103 00613994992024 V06160104 00613994992024 V06160105 00613994992024 V06160106 00613994992024 V06160107 00643169475885 V06160734 00643169475885 V06160736 00643169475885 V06160740 00643169475885 V06160741 00643169475885 V06160742 00643169215856 V06160764 00643169215856 V06160768 00643169215856 V06160771 00613994992048 V06161045 00613994992048 V06161047 00613994992048 V06161050 00613994644312 V06164460 00613994644312 V06164461 00613994644312 V06164464 00613994644312 V06164465 00613994644312 V06164467 00613994644312 V06164469 00613994644299 V06164470 00613994644299 V06164471 00613994644299 V06164472 00613994644299 V06164473 00613994644299 V06164474 00613994644299 V06164475 00613994644299 V06164477 00613994991997 V06165681 00613994991997 V06165682 00613994992048 V06166717 00613994992048 V06166718 00613994992048 V06166719 00613994992048 V06166720 00613994992048 V06166721 00613994992048 V06166722 00613994992048 V06166723 00613994992048 V06166724 00613994992048 V06166725 00613994992048 V06166726 00613994992048 V06167342 00613994992048 V06167343 00613994992048 V06167344 00613994992048 V06167345 00613994992048 V06167346 00613994992048 V06167347 00613994992048 V06167348 00613994992048 V06167349 00613994992048 V06167350 00613994992048 V06167401 00613994992048 V06167402 00613994992048 V06167403 00613994992048 V06167404 00613994992048 V06167405 00613994992048 V06167406 00613994992048 V06167407 00613994992048 V06167408 00613994992048 V06167409 00613994992048 V06167410 00613994992048 V06167411 00613994992000 V06168975 00613994992000 V06168977 00613994992000 V06168978 00613994992000 V06168979 00613994992000 V06168980 00613994992000 V06168981 00613994992000 V06168982 00613994992000 V06168983 00613994992000 V06168984 00613994991935 V06169755 00613994991935 V06169758 00613994991935 V06169760 00613994991935 V06169762 00613994991935 V06169763 00613994992024 V06169765 00613994992024 V06169766 00613994992024 V06169767 00613994992024 V06169768 00613994992024 V06169769 00613994992024 V06169770 00613994992024 V06169771 00613994992024 V06169772 00613994992024 V06169773 00613994992024 V06169774 00613994992048 V06170840 00613994992048 V06170841 00613994992048 V06170842 00613994992048 V06170843 00613994992048 V06170844 00613994992048 V06170845 00613994992048 V06170846 00613994992048 V06170847 00613994992048 V06170848 00613994992048 V06170849 00613994992031 V06172524 00613994992031 V06172525 00613994992031 V06172526 00613994992031 V06172527 00613994992031 V06172528 00613994992031 V06172529 00613994992031 V06172530 00613994992031 V06172531 00613994992031 V06172532 00613994992031 V06172533 00613994644312 V06173628 00613994644312 V06173629 00613994644312 V06173630 00613994644312 V06173631 00613994644312 V06173632 00613994644312 V06173633 00613994644312 V06173634 00613994644312 V06173635 00613994644312 V06173636 00613994644312 V06173637 00613994992048 V06173821 00613994992048 V06173822 00613994992048 V06173823 00613994992048 V06173824 00613994992048 V06173825 00613994992048 V06173826 00613994992048 V06173827 00613994992048 V06173828 00613994992048 V06173829 00613994992048 V06173830 00613994992048 V06173936 00613994992048 V06173937 00613994992048 V06173938 00613994992048 V06173939 00613994992048 V06173940 00613994992048 V06173941 00613994992048 V06173942 00613994992048 V06173943 00613994992048 V06173944 00613994992048 V06173945 00613994991263 V06174894 00613994991263 V06174895 00613994991263 V06174896 00613994991263 V06174897 00613994992017 V06175000 00613994992017 V06175001 00613994992017 V06175002 00613994992017 V06175003 00613994992017 V06175004 00613994992017 V06175005 00613994644268 V06175279 00613994644268 V06175280 00613994644268 V06175281 00613994644268 V06175282 00613994644282 V06175675 00613994644282 V06175676 00613994644282 V06175679 00613994992031 V06175805 00613994992031 V06175806 00613994992031 V06175807 00613994992031 V06175808 00613994992031 V06175809 00613994992031 V06175810 00613994992031 V06175811 00613994992031 V06175812 00613994992031 V06175813 00613994992031 V06175814 00613994992031 V06175815 00613994992031 V06175816 00613994992031 V06175817 00613994992031 V06175818 00613994992031 V06175819 00613994992031 V06175820 00613994992031 V06175821 00613994992031 V06175822 00613994992031 V06175823 00613994644268 V06176010 00613994644268 V06176011 00613994644268 V06176012 00613994644268 V06176013 00613994644268 V06176014 00613994644299 V06176031 00613994644299 V06176032 00613994644299 V06176034 00613994991249 V06176309 00613994991249 V06176310 00613994991249 V06176311 00613994991249 V06176312 00613994991249 V06176313 00613994991249 V06176314 00613994991249 V06176315 00613994991249 V06176316 00613994991249 V06176317 00613994991249 V06176318 00613994644367 V06176485 00613994644367 V06176487 00613994644367 V06176489 00613994991270 V06176871 00613994991270 V06176872 00613994991270 V06176873 00613994991270 V06176874 00613994644329 V06177038 00613994644329 V06177039 00613994644329 V06177040 00613994644329 V06177041 00613994644329 V06177042 00613994644299 V06177079 00613994644299 V06177084 00613994644299 V06177085 00613994644251 V06177320 00613994992017 V06180216 00613994991188 V06180575 00613994991188 V06180576 00613994991263 V06180587 00613994991263 V06180588 00613994991263 V06180589 00613994991263 V06180590 00613994991263 V06180591 00613994991263 V06180592 00613994991263 V06180593 00613994991263 V06180594 00613994991263 V06180595 00613994992048 V06180736 00613994992048 V06180737 00613994992048 V06180738 00613994992048 V06180739 00613994992048 V06180740 00613994992048 V06180741 00613994992048 V06180742 00613994992048 V06180743 00613994992048 V06180744 00613994992048 V06180745 00613994992048 V06180746 00613994992048 V06180747 00613994992048 V06180748 00613994992048 V06180749 00613994992048 V06180750 00613994992024 V06181268 00613994992024 V06181269 00613994992024 V06181270 00613994992024 V06181271 00613994992024 V06181272 00613994992024 V06181273 00613994992024 V06181274 00613994992024 V06181275 00613994992024 V06181276 00613994992024 V06181277 00613994991935 V06181501 00613994991935 V06181502 00613994991935 V06181503 00613994991935 V06181504 00613994991935 V06181505 00613994991935 V06181508 00613994991935 V06181510 00613994991997 V06181511 00613994991997 V06181512 00613994991997 V06181515 00613994991997 V06181516 00613994991997 V06181517 00613994991997 V06181519 00613994992048 V06181601 00613994992048 V06181602 00613994992048 V06181603 00613994992048 V06181604 00613994992048 V06181605 00613994991263 V06181671 00613994991263 V06181672 00613994991263 V06181673 00613994991263 V06181674 00613994991263 V06181675 00613994991263 V06181676 00613994991263 V06181677 00613994991263 V06181678 00613994991263 V06181679 00613994992031 V06182173 00613994992031 V06182174 00613994992031 V06182175 00613994992031 V06182176 00613994992031 V06182177 00613994992031 V06182178 00613994992031 V06182179 00613994992031 V06182180 00613994992031 V06182181 00613994992031 V06182183 00613994992031 V06182184 00613994992031 V06182185 00613994992031 V06182186 00613994992031 V06182187 00613994992031 V06182188 00613994992031 V06182189 00613994992031 V06182190 00613994992031 V06182191 00613994991270 V06183638 00613994991270 V06183639 00613994991270 V06183640 00613994991270 V06183641 00613994991270 V06183642 00613994991270 V06183643 00613994991270 V06183644 00613994991270 V06183645 00613994991270 V06183646 00613994991270 V06183647 00613994991959 V06184244 00613994991959 V06184245 00613994991959 V06184246 00613994991959 V06184247 00613994991959 V06184248 00613994991959 V06184249 00613994991959 V06184250 00613994991959 V06184251 00613994991959 V06184252 00613994991959 V06184253 00613994992017 V06184426 00613994992017 V06184427 00613994992017 V06184428 00613994992017 V06184432 00643169215856 V06184449 00643169215856 V06184450 00613994991270 V06184648 00613994991270 V06184649 00613994991270 V06184650 00613994991270 V06184651 00613994991270 V06184652 00613994991270 V06184653 00613994991270 V06184654 00613994991270 V06184655 00613994991270 V06184656 00613994992048 V06184698 00613994992048 V06184699 00613994992048 V06184700 00613994992048 V06184701 00613994992048 V06184702 00613994992048 V06184703 00613994992048 V06184704 00613994992048 V06184705 00613994992048 V06184706 00613994992048 V06184707 00643169215856 V06186001 00643169215856 V06186002 00643169215856 V06186003 00643169215856 V06186004 00643169215856 V06186005 00643169215856 V06186006 00643169215856 V06186007 00643169215856 V06186008 00613994991188 V06186723 00613994991188 V06186724 00613994991188 V06186725 00613994991188 V06186726 00613994991188 V06186727 00613994991249 V06186999 00613994991249 V06187000 00613994991249 V06187001 00613994991249 V06187002 00613994991249 V06187003 00643169475892 V06187004 00643169475892 V06187005 00643169475892 V06187006 00643169475892 V06187007 00643169475892 V06187008 00643169475915 V06187014 00643169475915 V06187015 00643169475915 V06187016 00643169475915 V06187017 00643169475915 V06187018 00643169475922 V06187019 00643169475922 V06187020 00643169475922 V06187021 00643169475922 V06187022 00643169475922 V06187023 00613994992048 V06187066 00613994992048 V06187067 00613994992048 V06187073 00643169215849 V06187158 00643169215849 V06187159 00643169215849 V06187160 00643169215849 V06187161 00643169215849 V06187162 00613994991959 V06187685 00613994991959 V06187686 00613994991959 V06187689 00613994991959 V06187691 00643169215849 V06187797 00643169215849 V06187798 00643169215849 V06187799 00643169215849 V06187800 00643169475878 V06188011 00643169475878 V06188012 00643169475878 V06188014 00643169475878 V06188015 00613994992031 V06188016 00613994992031 V06188017 00613994992031 V06188018 00613994992031 V06188019 00613994992031 V06188020 00613994992031 V06188021 00613994992031 V06188022 00613994992031 V06188023 00613994992031 V06188024 00613994992031 V06188025 00643169215849 V06188451 00643169215849 V06188452 00643169215849 V06188453 00643169215849 V06188454 00643169215849 V06188455 00613994992024 V06189133 00613994992024 V06189134 00613994992024 V06189135 00613994992024 V06189136 00613994992024 V06189137 00613994992024 V06189138 00613994992024 V06189139 00613994992024 V06189140 00613994992024 V06189141 00613994992024 V06189142 00643169215801 V06189158 00643169215801 V06189159 00643169215801 V06189160 00643169215801 V06189161 00643169215801 V06189162 00613994992031 V06189343 00613994992031 V06189344 00613994992031 V06189345 00613994992031 V06189346 00613994992031 V06189347 00613994992031 V06189348 00613994992031 V06189349 00613994992031 V06189350 00613994992031 V06189351 00613994992031 V06189352 00643169215764 V06189552 00643169215849 V06190683 00643169215849 V06190684 00643169215849 V06190685 00643169215849 V06190686 00643169215849 V06190687 00643169215849 V06190688 00643169215849 V06190689 00643169215849 V06190690 00643169215849 V06190691 00643169215849 V06190692 00613994991935 V06191237 00613994991935 V06191240 00613994991935 V06191242 00613994991935 V06191243 00613994991935 V06191244 00613994991959 V06191266 00613994991959 V06191267 00613994991959 V06191268 00613994991959 V06191269 00613994991959 V06191270 00613994991959 V06191273 00613994991959 V06191274 00643169475878 V06191376 00643169475878 V06191377 00643169475878 V06191378 00643169475878 V06191379 00643169475878 V06191380 00613994992031 V06191661 00613994992031 V06191662 00613994992031 V06191663 00613994992031 V06191664 00613994992031 V06191665 00613994992031 V06191666 00613994992031 V06191667 00613994992031 V06191668 00613994992031 V06191669 00613994992031 V06191670 00643169215832 V06191686 00643169215832 V06191687 00643169215832 V06191688 00643169215832 V06191689 00643169215832 V06191690 00613994992048 V06192496 00613994992017 V06192506 00613994992017 V06192507 00613994992017 V06192508 00613994992017 V06192510 00613994992017 V06192511 00613994992017 V06192512 00613994992017 V06192513 00613994992017 V06192515 00613994644268 V06192901 00613994644268 V06192902 00613994644268 V06192903 00613994644268 V06192904 00613994644268 V06192905 00613994644268 V06192906 00613994644268 V06192907 00613994644268 V06192911 00613994644268 V06192912 00613994644268 V06192914 00613994991997 V06193936 00613994991997 V06193937 00613994991997 V06193938 00613994991997 V06193939 00613994991997 V06193940 00613994644329 V06193956 00613994644251 V06194461 00613994644251 V06194465 00613994644251 V06194466 00613994644251 V06194467 00613994644251 V06194470 00613994992031 V06195380 00613994992031 V06195381 00613994992031 V06195382 00613994992031 V06195383 00613994992031 V06195384 00613994992031 V06195385 00613994992031 V06195386 00613994992031 V06195387 00613994992031 V06195388 00613994992031 V06195389 00613994644350 V06195541 00613994644350 V06195544 00613994991980 V06195750 00613994991980 V06195751 00613994991980 V06195752 00613994991980 V06195753 00613994991980 V06195754 00613994991980 V06195755 00613994991980 V06195756 00613994991980 V06195757 00613994991980 V06195758 00613994991980 V06195759 00613994991270 V06195830 00613994991270 V06195831 00613994991270 V06195832 00613994991270 V06195833 00613994991270 V06195834 00613994991270 V06195835 00613994991270 V06195836 00613994991270 V06195837 00613994991270 V06195838 00613994991270 V06195839 00613994644268 V06195936 00613994644268 V06195937 00613994644268 V06195939 00613994644312 V06195941 00613994644251 V06195955 00613994644251 V06195956 00613994644251 V06195957 00613994644251 V06195958 00613994644251 V06195959 00613994644251 V06195960 00613994644251 V06195961 00613994644251 V06195962 00613994644251 V06195963 00613994644251 V06195964 00613994991959 V06196042 00613994991959 V06196043 00613994991959 V06196048 00643169215856 V06196086 00643169215856 V06196089 00643169215856 V06196090 00643169215856 V06196091 00643169215856 V06196092 00643169215856 V06196094 00643169215856 V06196096 00643169215856 V06196097 00643169215856 V06196098 00643169215856 V06196099 00643169215856 V06196100 00613994992024 V06196407 00613994992024 V06196408 00613994992024 V06196411 00613994992024 V06196412 00613994992024 V06196415 00613994992024 V06196416 00643169215856 V06196951 00643169215856 V06196952 00643169215856 V06196953 00643169215856 V06196954 00643169215856 V06196955 00613994992048 V06197006 00613994992048 V06197007 00613994992048 V06197008 00613994992048 V06197009 00613994992048 V06197010 00613994992048 V06197011 00613994992048 V06197012 00613994992048 V06197013 00613994992048 V06197014 00613994992048 V06197015 00643169215825 V06197164 00643169215825 V06197165 00643169215825 V06197166 00643169215825 V06197167 00643169215825 V06197168 00613994992048 V06197901 00613994992048 V06197902 00613994992048 V06197903 00613994992048 V06197904 00613994992048 V06197905 00613994992048 V06197906 00613994992048 V06197907 00613994992048 V06197908 00613994992048 V06197909 00613994992048 V06197910 00643169215801 V06198213 00613994992048 V06198246 00613994992048 V06198247 00613994992048 V06198248 00613994992048 V06198249 00613994992048 V06198250 00613994992048 V06198251 00613994992048 V06198252 00613994992048 V06198253 00613994992048 V06198254 00613994992048 V06198255 00643169215825 V06198492 00643169215825 V06198493 00643169215825 V06198494 00643169215825 V06198496 00643169215825 V06198497 00643169215825 V06198498 00643169215825 V06198500 00613994991164 V06198646 00613994991164 V06198647 00613994992031 V06198791 00613994992031 V06198792 00613994992031 V06198793 00613994992031 V06198794 00613994992031 V06198795 00613994992031 V06198796 00613994992031 V06198797 00613994992031 V06198798 00613994992031 V06198799 00613994992031 V06198800 00613994992031 V06199560 00613994992031 V06199562 00613994992031 V06199564 00613994992031 V06199566 00613994991218 V06199761 00613994991218 V06199762 00613994991218 V06199763 00613994991218 V06199764 00613994991942 V06200757 00613994991942 V06200758 00613994991942 V06200759 00613994991942 V06200760 00613994991942 V06200761 00613994991942 V06200764 00613994991942 V06200765 00613994992048 V06201033 00613994992048 V06201034 00613994992048 V06201035 00613994992048 V06201036 00613994992048 V06201037 00613994992048 V06201038 00613994992048 V06201039 00613994992048 V06201040 00613994992048 V06201041 00613994992048 V06201042 00613994992048 V06201043 00613994992048 V06201044 00613994992048 V06201045 00613994992048 V06201046 00613994992048 V06201047 00613994992048 V06201048 00613994992048 V06201049 00613994992048 V06201050 00613994992048 V06201051 00613994992048 V06201052 00613994991249 V06201649 00613994991249 V06201650 00613994991249 V06201651 00613994991249 V06201652 00613994991249 V06201653 00613994991232 V06201744 00613994991232 V06201745 00613994991232 V06201746 00613994991232 V06201747 00613994991232 V06201748 00613994992031 V06202061 00613994992031 V06202062 00613994992031 V06202063 00613994992031 V06202064 00613994992031 V06202065 00613994992031 V06202066 00613994992031 V06202067 00613994992031 V06202070 00613994991935 V06202182 00613994991935 V06202183 00613994991935 V06202185 00613994991935 V06202188 00613994991935 V06202189 00613994992000 V06202901 00613994992000 V06202902 00613994992000 V06202903 00613994992000 V06202904 00613994992000 V06202905 00613994992000 V06202906 00613994992000 V06202907 00613994992000 V06202908 00613994992000 V06202909 00613994992000 V06202910 00613994991270 V06203121 00613994991270 V06203122 00613994991270 V06203123 00613994991270 V06203124 00613994991270 V06203125 00613994991270 V06203126 00613994991270 V06203127 00613994991270 V06203128 00613994991270 V06203129 00613994991270 V06203130 00613994991270 V06204179 00613994991270 V06204180 00613994991270 V06204181 00613994991270 V06204182 00613994991270 V06204183 00613994991270 V06204184 00613994991270 V06204185 00613994991270 V06204186 00613994991270 V06204187 00613994991270 V06204188 00613994991256 V06204287 00613994644329 V06204603 00613994644343 V06204610 00613994644268 V06204623 00613994644343 V06204920 00613994644312 V06205001 00613994644251 V06205232 00613994644251 V06205233 00613994644251 V06205236 00613994644251 V06205239 00613994991980 V06205379 00613994991980 V06205380 00613994991980 V06205383 00613994991980 V06205384 00613994991980 V06205385 00613994991980 V06205387 00613994992048 V06205438 00613994992048 V06205439 00613994992048 V06205440 00613994992048 V06205441 00613994992048 V06205442 00613994992048 V06205443 00613994992048 V06205444 00613994992048 V06205445 00613994992048 V06205446 00613994992048 V06205447 00613994991263 V06206222 00613994991263 V06206223 00613994991263 V06206224 00613994991263 V06206225 00613994991263 V06206226 00613994992048 V06206232 00613994992048 V06206233 00613994992048 V06206234 00613994992048 V06206235 00613994992048 V06206236 00613994992048 V06206237 00613994992048 V06206238 00613994992048 V06206239 00613994991171 V06206628 00613994991171 V06206629 00643169475915 V06206641 00643169475915 V06206642 00643169475915 V06206643 00643169475915 V06206644 00643169475915 V06206645 00613994644312 V06206753 00613994644312 V06206754 00613994992000 V06206984 00613994992000 V06206985 00613994992000 V06206986 00613994992000 V06206987 00613994992000 V06206988 00613994992000 V06206989 00613994992000 V06206990 00613994992000 V06206991 00613994992000 V06206992 00613994992000 V06206993 00613994991263 V06208157 00613994991263 V06208158 00613994991263 V06208159 00613994991263 V06208160 00613994992048 V06208999 00613994992048 V06209000 00613994992048 V06209001 00613994992048 V06209002 00613994992048 V06209003 00613994992048 V06209004 00613994992048 V06209005 00613994992048 V06209006 00613994992048 V06209007 00613994992048 V06209011 00613994992048 V06209014 00613994992048 V06209016 00613994992048 V06209018 00613994991966 V06209528 00613994991966 V06209531 00643169215771 V06209536 00643169215771 V06209537 00643169215771 V06209538 00643169215771 V06209539 00643169215771 V06209540 00613994992024 V06210283 00613994992024 V06210284 00613994992048 V06211740 00613994991959 V06212732 00613994991959 V06212733 00613994991959 V06212735 00613994991959 V06212736 00613994991959 V06212737 00613994991959 V06212738 00643169475885 V06214427 00613994991966 V06214638 00613994991195 V06215971 00613994991188 V06216404 00613994991188 V06216406 00613994991188 V06216409 00613994991980 V06217052 00613994991980 V06217054 00613994644251 V06220459 00613994644251 V06220460 00613994644251 V06220463 00613994644251 V06220465 00613994644251 V06220467 00613994644268 V06220473 00613994644268 V06220477 00613994991218 V06220654 00613994991218 V06220655 00613994991218 V06220656 00613994991218 V06220657 00613994991218 V06220658 00613994991959 V06221627 00613994644251 V06221955 00613994644251 V06221956 00613994644251 V06221958 00613994644251 V06221959 00613994644251 V06221961 00613994644312 V06221992 00613994991942 V06222328 00613994991942 V06222330 00613994991942 V06222333 00613994991942 V06222445 00613994991942 V06222446 00613994991942 V06222447 00613994991942 V06222450 00613994991942 V06222451 00613994991942 V06222454 00613994991980 V06224522 00613994991980 V06224525 00613994991980 V06224527 00613994991980 V06224528 00613994991188 V06225119 00613994991188 V06225120 00613994991188 V06225121 00613994991188 V06225122 00613994991188 V06225123 00613994644251 V06225466 00613994644251 V06225467 00613994644251 V06225468 00613994644251 V06225469 00613994644251 V06225470 00613994991959 V06226280 00613994991959 V06226281 00613994991959 V06226282 00613994991959 V06226283 00613994991959 V06226284 00613994991959 V06226285 00613994991959 V06226286 00613994991959 V06226287 00613994991959 V06226288 00613994991959 V06226289 00613994991201 V06226700 00613994991201 V06226701 00613994991201 V06226704 00613994991973 V06226730 00613994991973 V06226731 00613994991973 V06226732 00613994991973 V06226733 00613994991973 V06226734 00613994991973 V06226735 00613994991973 V06226736 00613994991973 V06226737 00613994991973 V06226738 00613994991973 V06226739 00613994991171 V06229871 00613994991171 V06229872 00613994991171 V06229873 00613994991171 V06229874 00643169215764 V06232205 00643169215764 V06232206 00643169215764 V06232207 00643169215764 V06232208 00643169215764 V06232209 00643169215764 V06232210 00643169215764 V06232211 00643169215764 V06232212 00643169215764 V06232213 00643169215757 V06232235 00643169215757 V06232237 00643169215757 V06232238 00643169215757 V06232240 00643169215757 V06232241 00643169215757 V06232242 00613994991959 V06232608 00613994991959 V06232614 00613994991270 V06234639 00613994991270 V06234640 00613994991270 V06234641 00613994991270 V06234642 00613994991270 V06234643 00613994991270 V06234644 00613994991270 V06234645 00613994991270 V06234646 00613994991270 V06234647 00613994992048 V06234893 00613994992048 V06234896 00613994992048 V06234900 00613994992048 V06234901 00613994992048 V06235979 00613994992048 V06235980 00613994992048 V06235981 00613994992048 V06235982 00613994992048 V06235983 00613994992048 V06235984 00613994992048 V06235985 00613994992048 V06235986 00613994992048 V06235989 00613994992048 V06235990 00613994992048 V06235991 00613994992048 V06235992 00613994992048 V06235993 00613994992048 V06235994 00613994992048 V06235995 00613994992048 V06235996 00613994992048 V06235997 00613994992048 V06235998 00613994644268 V06237149 00613994644268 V06237150 00613994644268 V06237151 00613994644268 V06237152 00613994644268 V06237154 00613994644268 V06237155 00613994644268 V06237156 00613994644268 V06237157 00643169215788 V06237643 00643169215788 V06237644 00643169215788 V06237645 00643169215788 V06237646 00643169215788 V06237647 00643169215788 V06237648 00643169215788 V06237649 00643169215788 V06237650 00643169215788 V06237651 00643169215788 V06237652 00613994992024 V06238326 00613994992024 V06238330 00613994992024 V06238335 00643169215832 V06238925 00643169215832 V06238926 00643169215832 V06238927 00643169215832 V06238928 00643169215832 V06238929 00613994644251 V06238998 00613994644251 V06238999 00613994644251 V06239006 00613994644251 V06239007 00613994644251 V06239008 00613994644251 V06239009 00643169215764 V06239450 00643169215788 V06239539 00643169215788 V06239540 00643169215788 V06239541 00643169215788 V06239542 00643169215788 V06239543 00613994644350 V06239554 00613994644268 V06239560 00613994644251 V06239579 00613994644251 V06239583 00613994644251 V06239584 00613994644251 V06239585 00613994644251 V06239586 00613994644251 V06239587 00643169215856 V06241393 00643169215856 V06241394 00643169215856 V06241395 00643169215856 V06241396 00643169215856 V06241397 00643169215856 V06241398 00643169215856 V06241399 00643169215856 V06241400 00643169215856 V06241401 00643169215856 V06241402 00643169215856 V06242013 00643169215856 V06242014 00643169215856 V06242015 00643169215849 V06243481 00643169215849 V06243482 00643169215849 V06243483 00643169215849 V06243485 00643169215849 V06243486 00643169215849 V06243487 00643169215849 V06243488 00643169215849 V06243489 00643169215788 V06244001 00643169215788 V06244002 00643169215788 V06244005 00643169215788 V06244006 00643169215788 V06244008 00643169215788 V06244010 00643169215832 V06244082 00643169215832 V06244083 00643169215832 V06244084 00643169215832 V06244086 00643169215832 V06244087 00643169215832 V06244089 00643169215832 V06244090 00643169215849 V06244702 00643169215849 V06244703 00643169215849 V06244704 00643169215849 V06244705 00643169215849 V06244706 00643169215849 V06244707 00643169215849 V06244708 00643169215849 V06244709 00643169215849 V06244710 00643169215849 V06244711 00613994992031 V06244732 00613994992031 V06244733 00613994992031 V06244734 00613994992031 V06244735 00613994992031 V06244736 00613994992031 V06244737 00613994992031 V06244738 00613994992031 V06244739 00613994992031 V06244740 00613994992031 V06244741 00643169215849 V06246168 00643169215849 V06246175 00613994644251 V06247512 00613994644251 V06247513 00613994644251 V06247514 00613994644251 V06247515 00613994644251 V06247516 00613994644251 V06247517 00613994644251 V06247518 00613994644251 V06247519 00613994644251 V06247520 00613994644251 V06247521 00613994644251 V06247861 00613994644251 V06247862 00613994644251 V06247863 00613994644251 V06247864 00613994644251 V06247865 00613994644251 V06247866 00613994644251 V06247867 00613994644251 V06247868 00613994644251 V06247869 00613994644251 V06247870 00613994644251 V06248096 00613994644251 V06248098 00613994644251 V06248099 00613994644251 V06248100 00613994644251 V06248103 00613994644251 V06248104 00613994644251 V06248105 00613994992031 V06248837 00613994992031 V06248839 00613994992031 V06248840 00613994992031 V06248846 00613994991270 V06254843 00613994991270 V06254844 00613994991270 V06254845 00613994991270 V06254846 00643169215849 V06255557 00643169215849 V06255558 00643169215849 V06255559 00643169215849 V06255560 00643169215849 V06255561 00643169215849 V06255562 00643169215849 V06255563 00643169215849 V06255564 00643169215849 V06255565 00643169215849 V06255566 00643169475915 V06256045 00643169475915 V06256046 00643169475915 V06256047 00643169475915 V06256048 00643169475915 V06256049 00613994991263 V06256061 00613994992024 V06256767 00613994992024 V06256768 00613994992024 V06256769 00613994992024 V06256770 00613994992024 V06256771 00613994992024 V06256773 00643169475922 V06258146 00643169475922 V06258147 00643169475922 V06258148 00643169475922 V06258149 00643169475922 V06258150 00613994992048 V06258211 00613994992048 V06258212 00613994992048 V06258213 00613994992048 V06258214 00613994992048 V06258215 00613994992048 V06258216 00613994992048 V06258217 00613994992048 V06258218 00613994992048 V06258219 00613994992024 V06258786 00613994992024 V06258787 00613994992024 V06258788 00613994992024 V06258792 00613994992024 V06258794 00613994992048 V06259856 00613994992048 V06259857 00613994992048 V06259858 00613994992048 V06259859 00613994992048 V06259860 00613994992048 V06259861 00613994992048 V06259862 00613994992048 V06259863 00613994992048 V06259864 00613994992048 V06259865 00613994992024 V06261346 00613994992024 V06261347 00613994992024 V06261348 00613994992024 V06261349 00613994992024 V06261351 00613994992024 V06261352 00613994992024 V06261353 00613994992024 V06261354 00613994992024 V06261355 00613994992048 V06261356 00613994992048 V06261357 00613994992048 V06261358 00613994992048 V06261359 00613994992048 V06261360 00613994992048 V06261361 00613994992048 V06261362 00613994992048 V06261363 00613994992048 V06261364 00613994992048 V06261365 00613994992048 V06261366 00613994992048 V06261367 00613994992048 V06261368 00613994992048 V06261369 00613994992048 V06261370 00613994992048 V06261371 00613994992048 V06261372 00613994992048 V06261373 00613994992048 V06261376 00613994992048 V06261377 00613994992048 V06261378 00613994992048 V06261379 00613994992048 V06261380 00613994992048 V06261381 00613994992048 V06261382 00613994992048 V06261383 00613994992048 V06261384 00613994992048 V06261385 00613994644268 V06262134 00613994644268 V06262136 00613994644268 V06262137 00613994644268 V06262138 00613994644268 V06262139 00613994644268 V06262141 00613994644268 V06262267 00613994644268 V06262268 00613994644268 V06262269 00613994644268 V06262270 00613994644268 V06262271 00613994644268 V06262272 00613994644251 V06263603 00613994644251 V06263608 00613994644251 V06263613 00613994644251 V06263614 00613994644251 V06263615 00613994644251 V06263616 00613994644251 V06263617 00613994644251 V06263619 00613994644251 V06263620 00613994644251 V06263622 00613994644251 V06265016 00613994644251 V06265017 00613994644251 V06265019 00643169475922 V06266568 00643169475922 V06266569 00643169475922 V06266570 00643169475922 V06266571 00643169475922 V06266572 00613994991270 V06267698 00613994991270 V06267699 00613994991270 V06267700 00613994991270 V06267702 00613994991270 V06267703 00613994991270 V06267705 00613994991270 V06267707 00613994992031 V06268686 00613994992031 V06268687 00613994992031 V06268688 00613994992031 V06268689 00613994992031 V06268691 00613994992024 V06270106 00613994992024 V06270107 00613994992024 V06270108 00613994992024 V06270109 00613994992024 V06270110 00613994992024 V06270111 00613994992024 V06270112 00613994992024 V06270113 00613994992024 V06270114 00613994992024 V06270115 00613994992031 V06271417 00613994992031 V06271418 00613994992031 V06271419 00613994992031 V06271420 00613994992031 V06271421 00613994992031 V06271422 00613994992031 V06271423 00613994992031 V06271424 00613994992031 V06271425 00613994991263 V06272073 00613994991263 V06272074 00613994991263 V06272075 00613994991263 V06272077 00613994991263 V06272080 00613994992031 V06273435 00613994992031 V06273436 00613994992031 V06273437 00613994992031 V06273438 00613994992031 V06273439 00613994992031 V06273440 00613994992031 V06273441 00613994992031 V06273442 00613994992031 V06273443 00613994992031 V06273444 00613994992048 V06274699 00613994992048 V06274851 00613994992024 V06274875 00613994992024 V06274876 00613994992024 V06274877 00613994992024 V06274879 00613994992024 V06274880 00613994992024 V06274881 00613994992024 V06274882 00613994991270 V06275901 00613994991270 V06275902 00613994991270 V06275903 00613994991270 V06275904 00613994991270 V06275905 00613994991270 V06275906 00613994991270 V06275907 00613994991270 V06275908 00613994991270 V06275909 00613994991270 V06275910 00613994991270 V06276232 00613994991270 V06276233 00613994991270 V06276234 00613994991270 V06276235 00613994991270 V06276236 00613994991270 V06276237 00613994991270 V06276238 00613994991270 V06276239 00613994991270 V06276240 00613994991270 V06276241 00613994992031 V06276641 00613994992048 V06278870 00613994992048 V06278871 00613994992048 V06278872 00613994992048 V06278873 00613994992048 V06278874 00613994992048 V06278875 00613994992048 V06278876 00613994992048 V06278877 00613994992048 V06278878 00613994992048 V06278879 00613994992048 V06279659 00613994992048 V06279661 00613994992048 V06279662 00613994992048 V06279663 00613994992048 V06279664 00613994992048 V06279665 00613994992048 V06279666 00613994992048 V06279667 00613994992024 V06281285 00613994992024 V06281286 00613994992024 V06284990 00613994992024 V06284996 00613994992024 V06284998 00643169475915 V06285959 00643169475915 V06285961 00613994644268 V06286065 00613994644268 V06286068 00613994644268 V06286069 00613994644268 V06286070 00613994644268 V06286071 00613994644268 V06286072 00613994644268 V06286683 00613994644268 V06286685 00613994644268 V06286686 00613994644268 V06286691 00613994644367 V06286699 00613994644367 V06286700 00613994644367 V06286701 00613994992031 V06286966 00613994992031 V06286967 00613994992031 V06286968 00613994992031 V06286970 00613994992031 V06286971 00613994992031 V06286973 00613994992031 V06286974 00613994992031 V06286975 00613994992048 V06287975 00643169215856 V06289211 00643169215856 V06289214 00613994992048 V06290175 00613994992048 V06290176 00613994992048 V06290178 00613994992048 V06290180 00613994992048 V06290181 00613994992048 V06290183 00613994992048 V06290184 00643169475915 V06294532 00643169475915 V06294537 00613994991263 V06294625 00613994991263 V06294626 00613994991263 V06294627 00613994991263 V06294628 00613994991263 V06294629 00613994991263 V06294630 00613994991263 V06294631 00613994991263 V06294632 00613994991263 V06294633 00613994992048 V06297100 00613994992031 V06297518 00613994992031 V06297519 00613994992031 V06297520 00613994992031 V06297521 00613994992031 V06297522 00613994992031 V06297523 00613994992031 V06297524 00613994992031 V06297525 00613994992031 V06297526 00613994992031 V06297527 00613994992048 V06298452 00613994992048 V06298453 00613994992048 V06298454 00613994992048 V06298455 00613994992048 V06298457 00613994992048 V06298458 00613994992048 V06298459 00643169215856 V06299612 00643169215856 V06299613 00643169215856 V06299614 00643169215856 V06299615 00643169215856 V06299616 00643169215856 V06299617 00643169215856 V06299618 00643169215856 V06299619 00643169215856 V06299620 00643169215856 V06299621 00613994992031 V06299877 00613994992031 V06299878 00613994992031 V06299880 00613994992031 V06299881 00613994992031 V06299882 00613994992048 V06299896 00613994992048 V06299900 00613994992048 V06299903 00643169215788 V06300725 00643169215788 V06300726 00643169215788 V06300727 00643169215788 V06300728 00643169215788 V06300729 00643169215788 V06300731 00643169215788 V06300732 00643169215788 V06300734 00643169215849 V06300750 00643169215849 V06300751 00643169215849 V06300752 00643169215849 V06300753 00643169215849 V06300754 00643169215849 V06300755 00643169215849 V06300756 00643169215849 V06300757 00643169215849 V06300758 00643169215849 V06300759 00613994644312 V06301179 00613994644312 V06301184 00613994644367 V06301314 00613994644367 V06301316 00613994991935 V06301566 00613994991935 V06301569 00643169215856 V06302089 00643169215856 V06302090 00643169215856 V06302092 00613994992031 V06302099 00613994992031 V06302100 00613994992031 V06302101 00613994992031 V06302102 00613994992031 V06302103 00613994992031 V06302104 00613994992031 V06302105 00613994992031 V06302106 00613994992031 V06302107 00613994992031 V06302108 00613994991270 V06303281 00613994991270 V06303282 00613994991270 V06303283 00613994991270 V06303284 00613994991270 V06303285 00613994991270 V06303286 00613994991270 V06303287 00613994991270 V06303288 00613994991270 V06303289 00613994991270 V06303290 00613994991973 V06303298 00613994991973 V06303300 00643169475922 V06303331 00643169475922 V06303332 00643169475922 V06303333 00643169475922 V06303334 00643169475922 V06303335 00613994992024 V06303799 00613994992024 V06303800 00613994992024 V06303801 00613994992024 V06303802 00613994992024 V06303803 00613994992024 V06303804 00613994992024 V06303805 00613994992024 V06303806 00613994992031 V06303819 00613994992031 V06303820 00613994992031 V06303822 00613994992031 V06303823 00613994992031 V06303825 00613994992031 V06303826 00613994992031 V06303828 00613994992048 V06303839 00613994992048 V06303840 00613994992048 V06303841 00613994992048 V06303842 00613994992048 V06303843 00613994992048 V06303844 00613994992048 V06303845 00613994992048 V06303846 00613994992048 V06303847 00613994992048 V06303848 00613994992048 V06304438 00613994992048 V06304439 00613994992048 V06304440 00613994992048 V06304441 00613994992048 V06304442 00613994992048 V06304443 00613994992048 V06304444 00613994992048 V06304445 00613994992048 V06304446 00613994992048 V06304447 00613994992031 V06307806 00613994992031 V06307807 00613994992031 V06307808 00613994992031 V06307809 00613994992031 V06307810 00613994992031 V06307813 00613994992031 V06307814 00613994992031 V06307815 00613994992031 V06308596 00613994992031 V06308597 00613994992031 V06308598 00613994992031 V06308599 00613994992031 V06308600 00613994992031 V06308601 00613994992031 V06308602 00613994992031 V06308603 00613994992031 V06308604 00613994992031 V06308605 00613994991270 V06308875 00613994991270 V06308876 00613994991270 V06308877 00613994991270 V06308878 00613994991270 V06308879 00613994991270 V06308880 00613994991270 V06308882 00613994991270 V06308883 00613994991270 V06308884 00613994991935 V06309510 00613994991935 V06309512 00613994644251 V06309841 00613994644251 V06309842 00613994644251 V06309843 00613994644251 V06309845 00613994644251 V06309846 00613994644251 V06309847 00613994644251 V06309848 00613994644251 V06309849 00613994644251 V06309850 00613994644251 V06310930 00613994644251 V06310933 00613994644251 V06310936 00613994644251 V06310937 00613994644268 V06311232 00613994644251 V06311851 00613994644251 V06311852 00643169215832 V06312086 00643169215832 V06312087 00643169215832 V06312088 00643169215832 V06312089 00643169215832 V06312090 00613994644251 V06312781 00613994644251 V06312783 00613994644251 V06312785 00613994992048 V06313703 00613994992048 V06313704 00613994992048 V06313705 00613994992048 V06313706 00613994992048 V06313707 00613994992048 V06313708 00613994992048 V06313709 00613994992048 V06313710 00613994992048 V06313711 00613994992048 V06313712 00613994991256 V06314118 00613994991256 V06314119 00613994991256 V06314120 00613994991256 V06314121 00613994991256 V06314122 00643169215849 V06314598 00643169215849 V06314599 00643169215849 V06314600 00643169215849 V06314601 00643169215849 V06314602 00643169215849 V06314603 00643169215849 V06314604 00643169215849 V06314605 00643169215849 V06314606 00643169215849 V06314607 00613994992048 V06315558 00613994992048 V06315559 00613994992048 V06315560 00613994992048 V06315561 00613994992048 V06315562 00613994992048 V06315563 00613994992048 V06315564 00613994992048 V06315565 00613994992048 V06315566 00613994992048 V06315567 00613994992048 V06315598 00613994992048 V06315599 00613994992048 V06315600 00613994992048 V06315601 00613994992048 V06315602 00613994992048 V06315603 00613994992048 V06315604 00613994992048 V06315605 00613994992048 V06315606 00613994992048 V06315607 00643169215832 V06315654 00643169215832 V06315655 00643169215832 V06315656 00643169215832 V06315657 00643169215832 V06315658 00643169215832 V06315659 00643169215832 V06315660 00643169215832 V06315661 00643169215832 V06315662 00643169215832 V06315663 00613994992048 V06316131 00613994992048 V06316132 00613994992048 V06316133 00613994992048 V06316134 00613994992048 V06316135 00613994992048 V06316136 00613994992048 V06316137 00613994992048 V06316138 00613994992048 V06316139 00613994992048 V06316140 00643169215856 V06316166 00643169215856 V06316169 00613994992031 V06316341 00613994992031 V06316342 00613994992031 V06316343 00613994992031 V06316344 00613994992031 V06316345 00613994992031 V06316346 00613994992031 V06316348 00613994992031 V06316350 00643169215849 V06316935 00643169215849 V06316936 00643169215849 V06316938 00643169215849 V06316940 00643169215849 V06316941 00643169215849 V06316943 00643169215849 V06316944 00613994991270 V06317048 00613994991270 V06317049 00613994991270 V06317050 00613994991270 V06317051 00613994991270 V06317052 00643169215856 V06317363 00643169215856 V06317364 00643169215856 V06317365 00643169215856 V06317366 00643169215856 V06317367 00643169215856 V06317368 00643169215856 V06317369 00643169215856 V06317370 00643169215856 V06317371 00643169215856 V06317372 00613994992048 V06317941 00613994992048 V06317942 00613994992048 V06317943 00613994992048 V06317944 00613994992048 V06317945 00613994992048 V06317946 00613994992048 V06317947 00613994992048 V06317948 00613994992048 V06317949 00613994992048 V06317950 00613994992048 V06317951 00613994992048 V06317952 00613994992048 V06317953 00613994992048 V06317954 00613994992048 V06317955 00613994992048 V06317956 00613994992048 V06317957 00613994992048 V06317958 00613994992048 V06317959 00613994992048 V06317960 00643169475908 V06318451 00643169475908 V06318452 00643169475908 V06318453 00643169475908 V06318454 00613994992048 V06319356 00613994992048 V06319357 00613994992048 V06319358 00613994992048 V06319359 00613994992048 V06319360 00613994992048 V06319361 00613994992048 V06319362 00613994992048 V06319363 00613994992048 V06319364 00613994992048 V06319365 00613994992024 V06319386 00613994992024 V06319387 00613994992024 V06319388 00613994992024 V06319389 00613994992024 V06319390 00613994992048 V06319619 00613994992048 V06319620 00613994992048 V06319621 00613994992048 V06319622 00613994992048 V06319623 00613994992048 V06319624 00613994992048 V06319625 00613994992048 V06319626 00613994992048 V06319627 00613994992048 V06319628 00613994992031 V06319659 00613994992031 V06319660 00613994992031 V06319661 00613994992031 V06319664 00613994992031 V06319665 00613994992031 V06319666 00613994992031 V06319667 00613994992031 V06320271 00613994992031 V06320272 00613994992031 V06320273 00613994992031 V06320274 00613994992031 V06320275 00613994992031 V06320276 00613994992031 V06320277 00613994992031 V06320278 00613994992031 V06320279 00613994992031 V06320280 00613994992024 V06320760 00613994992024 V06320761 00613994992024 V06320762 00613994992024 V06320763 00613994992024 V06320764 00613994992024 V06320765 00613994992024 V06320766 00613994992024 V06320767 00613994992024 V06320768 00613994992024 V06320769 00613994991263 V06320849 00613994991263 V06320850 00613994991263 V06320851 00613994991263 V06320852 00613994991263 V06320853 00613994991263 V06320854 00613994991263 V06320855 00613994991263 V06320856 00613994991263 V06320857 00613994991263 V06320858 00613994992048 V06321451 00613994992048 V06321452 00613994992048 V06321453 00613994992048 V06321454 00613994992048 V06321455 00613994992048 V06321456 00613994992048 V06321457 00613994992048 V06321458 00613994992048 V06321459 00613994992048 V06321460 00613994992048 V06321923 00613994992048 V06321924 00613994992048 V06321925 00613994992048 V06321926 00613994992048 V06321927 00613994992048 V06321928 00613994992048 V06321929 00613994992048 V06321930 00613994992048 V06321931 00613994992048 V06321932 00613994992048 V06322160 00613994992048 V06322161 00613994992048 V06322162 00613994992048 V06322163 00613994992048 V06322164 00613994992048 V06322165 00613994992048 V06322166 00613994992048 V06322167 00613994992048 V06322168 00613994992048 V06322169 00613994992048 V06322579 00613994992048 V06322580 00613994992048 V06322581 00613994992048 V06322582 00613994992048 V06322583 00613994992048 V06322584 00613994992048 V06322585 00613994992048 V06322586 00613994992048 V06322587 00613994992048 V06322588 00613994992048 V06323215 00613994992048 V06323216 00613994992048 V06323217 00613994992048 V06323218 00613994992048 V06323219 00613994992048 V06323220 00613994992048 V06323221 00613994992048 V06323222 00613994992048 V06323223 00613994992048 V06323224 00613994992048 V06323780 00613994992048 V06323781 00613994992048 V06323782 00613994992048 V06323783 00613994992048 V06323784 00613994992048 V06323785 00613994992048 V06323786 00613994992048 V06323787 00613994992048 V06323788 00613994992048 V06323789 00613994992031 V06323820 00613994992031 V06323821 00613994992031 V06323822 00613994992031 V06323823 00613994992031 V06323824 00613994992031 V06323825 00613994992031 V06323826 00613994992031 V06323827 00613994992031 V06323828 00613994992031 V06323829 00613994644268 V06326912 00613994644268 V06326913 00613994644268 V06326914 00613994644268 V06326915 00613994644268 V06326916 00613994644268 V06326917 00613994644268 V06326918 00613994644268 V06326919 00613994644268 V06326920 00613994644268 V06326921 00613994992048 V06327077 00613994992048 V06327078 00613994992048 V06327079 00613994992048 V06327080 00613994992048 V06327081 00613994992048 V06327082 00613994992048 V06327083 00613994992048 V06327084 00613994992048 V06327085 00613994992048 V06327086 00613994991980 V06327139 00613994991980 V06327144 00613994992048 V06327559 00613994992048 V06327560 00613994992048 V06327561 00613994992048 V06327562 00613994992048 V06327563 00613994992048 V06327564 00613994992048 V06327565 00613994992048 V06327566 00613994992048 V06327567 00613994992048 V06327568 00613994992024 V06327589 00613994992024 V06327590 00613994992024 V06327591 00613994992024 V06327592 00613994992024 V06327593 00613994992024 V06327594 00613994992024 V06327595 00613994992024 V06327596 00613994992024 V06327597 00613994992024 V06327598 00613994644329 V06328092 00613994644329 V06328096 00613994644329 V06328097 00613994644329 V06328098 00613994644329 V06328099 00613994644329 V06328100 00613994992031 V06328281 00613994992031 V06328282 00613994992031 V06328283 00613994992031 V06328284 00613994992031 V06328285 00613994992031 V06328286 00613994992031 V06328287 00613994992031 V06328288 00613994992031 V06328289 00613994992031 V06328290 00613994644251 V06328436 00613994644268 V06329036 00613994644268 V06329455 00613994644268 V06329456 00613994644268 V06329457 00613994644268 V06329458 00613994644268 V06329459 00613994644268 V06329460 00613994644268 V06329461 00613994644268 V06329462 00613994644268 V06329463 00613994644268 V06329464 00613994992031 V06329862 00613994992031 V06329863 00613994992031 V06329864 00613994992031 V06329865 00613994992031 V06329866 00613994992031 V06329867 00613994992031 V06329868 00613994992031 V06329869 00613994992031 V06329870 00613994992031 V06329871 00613994644251 V06330426 00613994644251 V06330427 00613994644251 V06330428 00613994644251 V06330429 00613994644251 V06330430 00613994644251 V06330431 00613994644251 V06330432 00613994644251 V06330433 00613994644251 V06330434 00613994644251 V06330435 00643169475878 V06330527 00643169475878 V06330528 00643169475878 V06330529 00613994644343 V06330762 00613994644343 V06330763 00613994644299 V06330792 00613994644299 V06330794 00613994644299 V06330795 00613994644299 V06330797 00613994644299 V06330798 00643169215856 V06331376 00643169215856 V06331377 00643169215856 V06331378 00643169215856 V06331379 00643169215856 V06331380 00643169215856 V06331381 00643169215856 V06331382 00643169215856 V06331383 00643169215856 V06331384 00643169215856 V06331385 00643169215856 V06331386 00643169215856 V06331387 00643169215856 V06331388 00643169215856 V06331389 00643169215856 V06331390 00643169215856 V06331391 00643169215856 V06331392 00643169215856 V06331393 00643169215856 V06331394 00643169215856 V06331395 00613994991270 V06332099 00613994991270 V06332100 00613994991270 V06332101 00613994991270 V06332102 00613994991270 V06332103 00613994991270 V06332104 00613994991270 V06332105 00613994991270 V06332106 00613994991270 V06332107 00613994991270 V06332108 00613994992048 V06332130 00613994992048 V06332131 00613994992048 V06332134 00613994992048 V06332135 00613994992048 V06332136 00613994992048 V06332137 00613994991935 V06332742 00613994991935 V06332743 00613994991225 V06333241 00613994991225 V06333242 00613994991225 V06333243 00613994991225 V06333244 00613994991225 V06333245 00643169215849 V06333536 00643169215849 V06333537 00643169215849 V06333538 00643169215849 V06333539 00643169215849 V06333540 00643169215849 V06333541 00643169215849 V06333542 00643169215849 V06333543 00643169215849 V06333544 00643169215849 V06333545 00643169215795 V06333546 00643169215795 V06333552 00643169215795 V06333553 00643169215795 V06333554 00613994991164 V06334093 00613994991164 V06334094 00613994991164 V06334095 00613994991232 V06334791 00613994991232 V06334792 00613994991232 V06334796 00613994991232 V06334797 00613994991232 V06334798 00613994991249 V06334805 00613994991249 V06334806 00643169215818 V06335505 00643169215818 V06335506 00643169215818 V06335507 00643169215818 V06335508 00643169215818 V06335509 00613994992031 V06502543 00613994992031 V06502544 00613994992031 V06502545 00613994992031 V06502546 00613994992031 V06502547 00613994992031 V06502548 00613994992031 V06502549 00613994992031 V06502550 00643169475915 V06503570 00643169475915 V06503571 00643169475915 V06503572 00643169475915 V06503573 00643169475915 V06503574 00613994992031 V06504151 00613994992031 V06504152 00613994991997 V06504467 00613994991997 V06504468 00613994991997 V06504469 00613994991997 V06504470 00613994991997 V06504471 00613994992017 V06504574 00613994992017 V06504575 00613994992017 V06504576 00613994992017 V06504577 00613994992017 V06504578 00613994992017 V06504579 00613994992017 V06504580 00613994992017 V06504581 00613994992017 V06504582 00613994992017 V06504583 00613994992031 V06504605 00613994992031 V06504606 00613994992031 V06504607 00613994992031 V06504609 00613994992031 V06504610 00613994992031 V06504611 00613994992031 V06504612 00613994992000 V06671257 00613994992000 V06671258 00613994992000 V06671259 00613994992000 V06671260 00613994992000 V06671261 00613994992000 V06671262 00613994992000 V06671263 00613994992000 V06671264 00613994992000 V06671659 00613994992000 V06671660 00613994992000 V06671661 00613994992000 V06671662 00613994992000 V06671663 00613994992000 V06671664 00613994992000 V06671665 00613994992000 V06671666 00613994992000 V06671667 00613994992000 V06671668 00613994991959 V06671789 00613994991959 V06671790 00613994991959 V06671792 00613994991959 V06671793 00643169475878 V06671859 00643169475878 V06671860 00643169475878 V06671861 00643169475878 V06671862 00643169475878 V06671863 00613994992000 V06672292 00613994992000 V06672293 00613994992000 V06672294 00613994992000 V06672295 00613994992000 V06672296 00613994992000 V06672298 00613994992000 V06672299 00613994992000 V06672301 00613994992024 V06672352 00613994992024 V06672353 00613994992024 V06672354 00613994992024 V06672355 00613994992024 V06672356 00613994992024 V06672357 00613994992024 V06672358 00613994992024 V06672359 00613994992024 V06672360 00613994992024 V06672361 00613994992031 V06672448 00613994992031 V06672449 00613994992017 V06672699 00613994992017 V06672700 00613994992017 V06672701 00613994992017 V06672702 00613994992017 V06672703 00613994992017 V06672704 00613994992017 V06672705 00613994992017 V06672706 00613994992017 V06672707 00613994992017 V06672708 00643169475908 V06672715 00613994644343 V06674195 00613994644343 V06674196 00613994644251 V06674200 00613994644251 V06674203 00613994644251 V06674204 00613994644268 V06674901 00613994644268 V06674902 00613994644268 V06674903 00613994644268 V06674904 00613994644268 V06674905 00613994992017 V06675105 00613994992017 V06675106 00613994992017 V06675107 00613994992017 V06675108 00613994992017 V06675109 00613994992017 V06675110 00613994992017 V06675111 00613994992017 V06675112 00613994992017 V06675113 00613994992017 V06675114 00613994991997 V06675868 00613994991997 V06675869 00613994991997 V06675870 00613994991997 V06675871 00613994991997 V06675872 00613994991959 V06677413 00613994991959 V06677416 00613994992048 V06677506 00613994992031 V06679232 00613994992031 V06679233 00613994992031 V06679234 00613994992031 V06679235 00613994992031 V06679236 00613994992031 V06679240 00613994992031 V06679241 00613994992000 V06680113 00613994992000 V06680116 00613994992048 V06680992 00613994992048 V06680993 00613994992048 V06680994 00613994992048 V06680995 00613994992048 V06680996 00613994992048 V06680998 00613994992048 V06680999 00613994992048 V06681000 00613994992048 V06681001 00613994991942 V06681028 00613994991942 V06681029 00613994991942 V06681031 00613994991942 V06681035 00613994991942 V06681036 00613994991942 V06681037 00613994991980 V06681048 00613994991980 V06681049 00613994991980 V06681050 00613994992024 V06681054 00613994992024 V06681055 00613994992048 V06681548 00613994992048 V06681551 00613994992017 V06681553 00613994992017 V06681554 00613994992017 V06681555 00613994992017 V06681556 00613994992017 V06681557 00613994992017 V06681558 00613994992017 V06681559 00613994992017 V06681560 00613994992017 V06681561 00643169215795 V06681603 00643169215795 V06681605 00643169215795 V06681607 00613994991263 V06682259 00613994991263 V06682260 00613994991263 V06682262 00613994991966 V06682263 00613994991966 V06682266 00613994991966 V06682267 00613994991980 V06682501 00613994991980 V06682502 00613994991980 V06682504 00613994991980 V06682506 00613994991980 V06682507 00613994992017 V06682518 00613994992017 V06682519 00613994992017 V06682520 00613994992017 V06682522 00613994992017 V06682524 00613994992017 V06682525 00613994992017 V06682527 00643169215825 V06687682 00613994644350 V06687832 00613994644350 V06687837 00613994644350 V06687838 00613994644350 V06687839 00613994991164 V06690540 00613994991164 V06690541 00613994991164 V06690542 00613994991164 V06690543 00613994991959 V06701200 00613994991959 V06701201 00613994991959 V06701202 00613994991959 V06701203 00613994991959 V06701204 00613994991959 V06701205 00613994991959 V06701206 00613994991959 V06701207 00613994991959 V06701208 00613994991959 V06701209 00613994991959 V06703057 00613994991959 V06703058 00613994991959 V06703059 00613994991959 V06703060 00613994991959 V06703061 00613994991959 V06703062 00613994991959 V06703063 00613994991959 V06703064 00613994991959 V06703065 00613994991959 V06703066 00613994991959 V06703068 00613994991959 V06703069 00613994991959 V06703074 00643169215740 V06705115 00643169215764 V06708139 00613994644367 V06791854 00613994644299 V07309309 00613994644299 V07309310 00613994644299 V07309311 00613994644299 V07309312 00613994644299 V07311127. |
Recalling Firm/ Manufacturer |
Medtronic Vascular, Inc. 3850 Brickway Blvd Santa Rosa CA 95403-8223
|
Manufacturer Reason for Recall |
Medtronic is initiating a voluntary Urgent Medical Device Recall for a subset of Endurant/ Endurant II
Bifurcated Stent Graft Systems of specific models and serial numbers as this specific subset of stent grafts has greater susceptibility to fabric permeability variations that may be associated with endoleaks observed during the initial implant procedure.
|
FDA Determined Cause 2 |
Packaging change control |
Action |
Medtronic sent an Urgent Medical Device Recall letter dated March 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Our records indicate that your facility has received a potentially affected Endurant /Endurant II 23mm and / or 25mm Bifurcated Stent Graft System from the identified subset. As a result, Medtronic is asking that you take the following actions:
Identify and quarantine unused Endurant /Endurant II 23mm and 25mm Bifurcated Stent Graft System from the identified subset of models and serial numbers that are in your inventory .
Return unused products from the attached model and serial number list that are in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570 (#4) and referencing this communication to initiate a return and credit of unused product.
Your Medtronic sales representative can assist you in the return of this product as necessary.
Complete the attached Customer Confirmation Certificate and email it to RS.CFQFCA@Medtronic.com or fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality.
For further questions please call Customer Service at 1 (800) 633-8766. |
Quantity in Commerce |
3331 devices |
Distribution |
Worldwide Distribution - US Distribution including Puerto Rico and to the countries of :
Argentina
Armenia
Australia
Austria
Belgium
Brazil
Bulgaria
Canada
China
Colombia
Costa Rica
Croatia
Cyprus
Czech Republic
Denmark
Dominican Republic
Ecuador
Egypt
El Salvador
Finland
France
French Polynesia
Germany
Greece
Hong Kong
Hungary
Iceland
India
Indonesia
Iran
Ireland
Israel
Italy
Japan
Jordan
Kazakhstan
Korea
Kuwait
Latvia
Lebanon
Lithuania
Luxembourg
Malaysia
Mexico
Montenegro
Netherlands
New Zealand
Norway
Oman
Panama
Paraguay
Peru
Philippines
Poland
Portugal
Reunion
Romania
Russian Federation
Saudi Arabia
Serbia
Singapore
Slovakia
Slovenia
South Africa
Spain
Sri Lanka
Sweden
Switzerland
Taiwan
Thailand
Turkey
United Arab Emirates
United Kingdom
Uruguay
Venezuela
Viet Nam |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = MIH and Original Applicant = Medtronic Vascular
|