• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Kappa

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Kappasee related information
Date Initiated by FirmMarch 28, 2017
Create DateApril 10, 2017
Recall Status1 Terminated 3 on September 17, 2018
Recall NumberZ-1774-2017
Recall Event ID 76862
510(K)NumberK152407 
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
ProductKappa, Catalog Number: MS18853 in combination with Scio, Scio Four, Scio Four Oxi plus, Scio Four Oxi, Scio Four plus. Catalog Number(s): MS18597/MS18596/MS18853. Product Usage: The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patients medical condition. The Infinity Delta Series (Delta/Delta XL/Kappa) monitors are intended to be used on adult, pediatric, and neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2, which for the neonatal population, is to only be used when the patient is not under gas anesthesia. For combination with Scio gas module: Scio gas module samples breathing gases from adults and pediatrics. The gas module continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the In-finity monitors.
Code Information VF10.0 software
Recalling Firm/
Manufacturer
Draegar Medical Systems, Inc.
6 Tech Dr
Andover MA 01810-2434
For Additional Information Contact
800-437-2437
Manufacturer Reason
for Recall
It was reported that a set low O2 alarm does not go off although the measured O2 level is below the alarm limit.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn March 28, 2017, US consignees were sent an Urgent Medical Recall letter and Customer Reply and Order card. Delta family monitors running software version 10.0 in facilities that also have at least one Scio module will be downgraded to software version VF9.1 as a temporary solution Free of Charge. Draeger Medical Systems, Inc. is developing a new software version to resolve the issue (VF10.1). Once available, all Delta family monitors that were running software version VF10.0 will be upgraded with VF10.1 Free of Charge. For all other countries, the responsible local Draeger Sales and Service will mail the letter and card.
Quantity in Commerce2,156 (total for all devices)
DistributionWorldwide - US Nationwide distribution in the states of: AL, CA, GA, IL, LA, MA NY, OH, PA, TN, TX and VA and the countries of: AR AT BE BG BN BO BY CA CH CL CO CZ DE EG ES FI FR GB GR HR HU ID IE IL IN IT LT MX MY NL NO NP PA PE PK PL PT RO RS RU SA SE SK TH TR TW UG UZ VN ZA
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MHX
-
-