Class 2 Device Recall VITEK 2 AST N276 |
|
Date Initiated by Firm |
April 20, 2017 |
Date Posted |
May 22, 2017 |
Recall Status1 |
Terminated 3 on July 22, 2020 |
Recall Number |
Z-2449-2017 |
Recall Event ID |
77026 |
Product Classification |
unknown device name - Product Code N/A
|
Product |
VITEK¿ 2 Gram Negative Susceptibility card (AST-N276), REF 414286, 20 cards per carton. |
Code Information |
6960139203 14/Apr/18 |
Recalling Firm/ Manufacturer |
Biomerieux Inc 595 Anglum Rd Hazelwood MO 63042-2320
|
For Additional Information Contact |
Jeff Scanlan 314-731-8694
|
Manufacturer Reason for Recall |
The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
|
FDA Determined Cause 2 |
Packaging |
Action |
The firm posted their recall notification on their customer website on 4/20/2017. The firm sent the recall notification letter to their customers via FedEx on 4/21/2017 for delivery 4/24/2017. A follow-up recall notice will be issued on 5/08/2017 which identified 3 additional lots impacted by the recall. |
Quantity in Commerce |
477 cartons |
Distribution |
Nationwide distribution in the U.S, including military and government sites. The product was also distributed to Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe. |
|