| Class 2 Device Recall Glidesheath Slender Hydrophilic Coated Introducer Sheath | |
Date Initiated by Firm | June 07, 2017 |
Create Date | December 08, 2017 |
Recall Status1 |
Terminated 3 on April 25, 2019 |
Recall Number | Z-0226-2018 |
Recall Event ID |
77661 |
510(K)Number | K142183 |
Product Classification |
Introducer, catheter - Product Code DYB
|
Product | Glidesheath Slender Hydrophilic Coated Introducer Sheath, 6FR - 10CM (0.021), REF/Product Code RM*RS6F10PA, STERILE, Rx ONLY |
Code Information |
Lot s 161020, 161021, 161024, 161025, 161027, 161028, 161031, 161101, 161104, 101107, 101108, 161109, 161109, 161110, 161111, 161114, 161121, 161122, 161122, 161123, 161124, 161125, 161128, 161129, 161130, 161206, 161207, 161208, 161209, 161212, 161219, 161220, 161221, 161222, 1612223, 161226, 161227, 161228, 170109, 170112, 170113, 170116, 170117, 170120, 170123, 170124, 170125, 170126, 170127, 170130, 170131, 170201 |
Recalling Firm/ Manufacturer |
Terumo Medical Corp 2101 Cottontail Ln Somerset NJ 08873-1277
|
For Additional Information Contact | 908-302-4916 |
Manufacturer Reason for Recall | An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath. |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | URGENT VOLUNTARY MEDICAL DEVICE RECALL Letters (dated June 7, 2017) and Response Forms were sent to the consignees via UPS on June 7, 2017. |
Distribution | Nationwide and Canada |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DYB
|
|
|
|