Date Initiated by Firm | October 29, 2015 |
Create Date | November 28, 2017 |
Recall Status1 |
Terminated 3 on June 16, 2020 |
Recall Number | Z-0177-2018 |
Recall Event ID |
78443 |
510(K)Number | K151511 |
Product Classification |
Orthopedic stereotaxic instrument - Product Code OLO
|
Product | Rosa Spine 1.0.2 Stereotaxic Instrument Computer-Assisted Surgical Device
The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by surgeons to guide standard neurosurgical instruments during spine surgery. Guidance is based on an intra-operative plan developed with three-dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The device is indicated for the placement of pedicle screws in lumber vertebrae with a posterior approach. |
Code Information |
Serial #'s; SP14003 and SP14005 |
Recalling Firm/ Manufacturer |
Zimmer Biomet, Inc. 1800 W Center St Warsaw IN 46580-2304
|
For Additional Information Contact | Kevin W. Escapule 574-267-6131 |
Manufacturer Reason for Recall | A design change was initiated to update ROSA Spine 1.0.2 to version ROSA Spine 1.0.2.16 to resolve software bugs and improve usability and stability of the ROSA Spine device. |
FDA Determined Cause 2 | Software design |
Action | Zimmer Biomet issued a retrospective report of a correction taken on 29-Oct-2015. The software issue described was corrected in the new software version 1.0.2.16. There were 2 copies of the affected software in distribution. Field Service Technicians were deployed to the consignee locations to perform the system upgrade. 100% of affected devices were updated with the software solution.
No further action is required. |
Quantity in Commerce | 2 |
Distribution | Worldwide Distribution including France and Germany |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = OLO
|