| | Class 2 Device Recall cobas 6000 analyzer series, cobas 8000 modular analyzer series, and MODULAR Analytics system |  |
| Date Initiated by Firm | June 29, 2017 |
| Date Posted | January 16, 2018 |
| Recall Status1 |
Terminated 3 on January 24, 2020 |
| Recall Number | Z-0378-2018 |
| Recall Event ID |
78760 |
| 510(K)Number | K060373 K100853 K953239 |
| Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
| Product | Brand Name:
cobas c 6000 MODULAR Series System e 601
cobas c 8000 MODULAR Series System - e 602
Roche / Hitachi MODULAR Analytics Combination System
Model/Catalog/Part Number:
MODULAR Analytics E 170
03022109001
03023109973
03739040001
03739040692
05023572001
05023599001
cobas e 601 module
04745922001
04745922692
05036348001
05036348692
05860652001
cobas e 602 module
05990378001
The Roche/Hitachi MODULAR Analytics System (E170) is a fully automated, random-access, software- controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. The cobas 6000 series (e 601) is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of Ion selective electrodes (ISE), a photometric analysis unit an immunoassay analysis module. The cobas 8000 Modular Analyze Series (e602) is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit. |
| Code Information |
Serial Number 2072-26 27P6-20 15S2-20 2717-18 2719-02 1544-11 1686-03 1534-12 1698-04 1549-04 2275-12 2263-16 2263-17 2497-16 1849-01 1570-15 26F0-17 2256-18 2564-30 2850-05 2384-12 2384-18 2399-08 2230-19 2230-20 2379-14 2381-02 27Z8-01 2058-22 1909-13 1909-01 27V5-20 15U5-18 2308-01 2087-07 1984-05 1985-07 26D3-30 2497-01 1861-18 1861-19 26D0-27 26D1-21 1480-14 1480-15 1488-01 1488-02 26C3-12 26F0-19 26F0-27 26D9-24 2107-35 2348-12 2348-13 1423-16 2237-18 2240-09 2240-10 16J8-05 16J8-06 16L7-20 16K2-17 16K2-18 16K2-19 16K2-20 16K3-01 16K3-06 16K3-07 16K3-08 16K3-09 16L4-03 16L4-04 26D1-25 26D1-26 26D1-28 26D1-29 2328-03 2437-13 2072-29 2072-30 2064-12 1931-11 2080-35 1936-08 2049-15 27Z0-09 27Z0-10 1911-11 2581-18 2062-14 2074-05 1430-16 1430-17 1691-17 1691-18 1985-13 1989-19 1739-19 2578-25 13B8-12 1169-06 1169-08 1169-09 1211-16 1211-17 1211-18 1224-10 1458-04 2543-10 27J6-15 2048-05 2406-27 26B3-30 1927-19 1914-13 1914-11 27J7-06 2822-07 2822-08 2535-23 2306-04 1159-19 2867-11 2711-02 27Y6-19 2107-20 2195-34 2106-20 2079-34 27M5-13 15D1-02 15C9-10 15F0-17 15F0-18 2137-15 1872-01 2195-40 2102-30 2428-02 2539-08 1377-14 2495-01 1490-16 15C5-16 15C5-17 15C5-20 15C6-01 2250-02 26H6-29 2423-15 2539-06 1377-12 2307-26 2307-27 2227-11 2080-24 1128-05 1128-06 14B1-02 14B2-13 2467-05 27Z8-09 26E4-05 2413-14 2423-19 2086-27 27Z0-14 27Z0-16 2145-16 2321-26 2321-25 26F1-20 26F1-19 1305-40 26G7-23 26G8-09 27L8-12 27X4-14 2719-07 27S8-20 15U1-20 26A4-28 26D1-22 2368-06 1128-19 2427-29 2539-07 1377-13 1357-08 1357-09 1357-10 1344-05 1344-06 14B1-01 1344-02 1344-03 1344-04 1469-03 1469-04 1469-05 1476-20 1482-11 1483-01 1483-17 1483-18 1483-19 1483-20 1484-16 1484-17 1484-18 1490-18 1490-19 1490-20 14A3-01 14A3-02 14A3-03 15B7-01 15B7-02 2430-27 2430-29 15B7-03 2327-25 2347-04 2369-04 2332-04 2332-09 27S4-02 27S4-03 2230-14 2230-18 1211-15 1211-14 15D2-04 15D2-05 15D3-12 15F0-15 15F0-16 16P4-02 16P4-03 2061-21 14A9-16 14A9-17 14A9-18 15B7-04 2256-19 2865-20 2867-01 2070-11 2070-13 2720-17 16X4-17 2102-29 2192-37 2196-02 2535-18 2842-03 2896-04 16P4-04 16P4-07 16P4-08 16P4-09 2311-12 2311-13 1871-09 1125-07 1128-02 2059-28 15H8-11 15H8-12 1996-01 2396-04 2399-06 2068-24 2283-14 2080-02 2577-04 2047-10 2107-05 26E4-10 27J7-04 2585-07 2581-08 2581-13 2327-11 2396-07 2413-01 2413-04 185731 1867-06 2037-07 2423-14 1549-02 27Z8-02 1860-19 2074-25 2087-09 2103-35 1924-15 2192-02 2437-19 2437-20 27Z0-03 2437-17 1913-03 1913-04 2488-05 2402-05 26C3-11 26C9-12 26E4-08 26F3-21 1545-05 1844-14 2822-09 2822-10 1303-31 1303-32 1357-19 1357-20 2402-30 2195-38 2196-01 2080-11 2080-17 2226-17 2439-12 2107-02 1988-03 2086-40 2195-24 2195-06 2195-08 2072-39 2041-12 2047-07 2192-15 2192-08 1736-04 1991-08 1871-06 2037-14 2055-30 2055-35 2873-02 2891-13 1990-05 192506 2894-12 1936-14 1928-19 2848-17 2867-08 2058-21 2058-25 2073-01 2111-04 130833 130834 2850-06 1334-12 1334-13 1334-14 1476-01 1476-02 1476-12 2049-29 2049-30 28E5-14 2347-11 2367-09 27Z8-12 2842-02 192418 1936-09 28A2-10 28AO-16 2208-17 2227-10 15J5-02 15J5-03 1852-15 2256-15 1461-03 1461-04 1476-11 16M7-01 16M7-16 14B3-06 14B3-07 16K3-15 16K3-16 16K3-17 16K3-18 16K3-19 16K3-20 16K4-07 16K4-11 16K4-20 16K5-01 16K5-02 16K5-03 16K5-04 16K5-05 16K5-09 16K5-10 16K7-04 16K7-05 16K7-06 16L4-05 16L4-06 16L4-07 16L4-08 16L7-19 16L5-03 16L7-17 16P1-13 16P1-14 16P1-15 16P2-19 1417-20 2581-25 2587-14 1431-14 2065-12 1545-03 16M7-13 16M8-01 16M8-02 16P1-11 2275-13 2267-19 2275-09 2275-10 2056-14 2047-08 185715 185817 26C3-13 2137-09 15D9-17 15D9-18 2486-24 2489-30 2054-06 2054-10 26E0-18 2467-15 16J8-12 16J8-13 1871-07 2567-22 2367-13 2304-14 2304-15 2202-07 2288-10 1487-16 1488-03 1480-16 1488-04 2427-27 1857-13 2486-06 27J7-03 27Z0-17 27Z0-18 27Z0-19 2494-30 2071-34 2576-22 26E4-06 1124-05 1124-06 1132-03 1132-04 16K5-06 2493-18 2497-12 2848-14 2848-15 2842-01 2848-12 2896-03 2263-13 2263-14 2050-29 2547-29 2550-19 2058-17 2058-20 2054-07 2054-08 2328-05 2439-13 1535-09 2548-03 184826 2497-13 2039-09 2061-35 1905-11 0753-22 1685-16 1570-14 2135-17 2140-11 2493-28 2493-29 2850-09 2850-10 2581-07 2581-22 1867-14 27P6-19 15S2-19 26G7-24 26G7-25 26H3-28 2108-36 192717 2493-23 2494-21 1225-14 2822-20 2824-01 2585-04 1910-02 2540-23 2540-24 1694-19 2104-01 2047-36 0884-01 1991-18 2389-18 2107-29 2108-27 26D6-03 26D5-22 1174-08 2189-02 2189-03 2086-13 2086-21 2446-13 1557-12 2486-23 2867-03 191007 2137-14 16K5-12 16K5-13 16K5-14 16K5-15 2104-02 1725-20 26E4-04 26A9-09 2869-20 16Z2-20 2581-26 2389-13 1142-15 2075-40 0817-20 1927-15 1156-01 1160-04 1156-02 1156-09 1160-06 1331-10 1331-11 1331-12 2072-34 1548-08 2262-18 2405-26 16M8-03 16R5-08 15B7-12 15D2-01 15D2-02 15D2-03 16P2-20 2720-20 2451-02 2451-06 2451-09 2468-18 1985-12 1860-15 2850-03 2850-04 2067-31 2086-28 0829-17 191305 2107-11 1848-01 1848-03 27Z0-15 2704-19 2047-39 0884-04 1421-11 2423-16 2423-17 1482-12 28E7-02 2413-02 2413-03 2486-05 2074-17 2486-04 2486-16 2327-23 2398-05 2438-15 2438-16 2479-09 2479-11 2311-07 2311-11 2425-11 2427-18 2227-07 2233-16 27X4-18 2427-21 2381-03 2382-04 2822-19 2867-07 2850-01 2382-06 2227-02 26C7-30 14H8-04 1017-06 1021-03 1021-05 2108-31 2108-32 2414-14 2367-11 2367-12 2451-19 2406-28 27M5-12 2368-20 2369-01 2398-14 26G4-01 26G4-09 1686-01 2369-05 16S0-01 2080-19 1874-06 26D8-26 26E4-14 26D5-24 26D4-26 26D4-27 2039-07 2585-27 26A4-26 2570-30 13A9-09 2240-13 2240-15 1914-18 2072-31 2072-18 2304-13 2362-19 1122-18 1924-13 2049-14 16R5-15 27L8-11 1868-08 2446-12 2535-20 28E7-16 2548-24 1387-17 1486-20 1490-17 14A3-04 14A3-05 14A3-06 14A3-07 1118-13 2534-05 2534-06 1872-12 2047-37 0884-02 2413-09 2891-15 26H1-28 2858-02 16J8-16 16J8-17 16K2-01 27P2-10 2704-20 2441-10 2441-11 2422-18 2425-12 2219-18 2234-07 1988-11 26B5-23 2379-13 2406-29 2407-04 2894-20 2896-01 1919-05 2102-31 1872-07 1476-03 1479-07 2576-21 185503 27S2-18 2228-15 1005-29 2413-11 2429-30 1263-20 27P2-16 27P2-17 2108-05 2108-21 2425-14 2039-13 2041-16 2413-06 26F1-17 27S4-04 27S4-05 2585-19 2585-21 2493-05 2493-19 1910-04 2422-29 26D6-02 26E0-16 2842-11 2107-31 2848-18 2193-39 27R3-06 27R3-08 27L1-09 2842-10 1914-20 0763-09 2347-15 2348-16 27Z8-08 1691-15 2427-25 2427-26 2207-08 2226-09 26C7-29 14H8-03 2577-02 2577-01 26D4-29 26D4-30 1927-09 2848-16 1459-12 1459-13 1476-05 1476-06 2073-10 26C3-07 190703 2049-21 2369-03 2587-10 2305-22 2070-21 27J7-09 2066-27 2103-40 26H3-26 2848-19 2061-36 0801-27 2581-23 2584-23 2584-30 2395-29 2396-02 1870-16 2073-38 2080-30 2578-24 13B8-11 2071-20 2071-21 2065-04 27W3-18 1344-01 1344-07 1344-08 2288-12 2288-07 27Z8-18 27Z8-19 28E7-13 28E7-14 2140-06 2140-07 2257-02 2079-30 16P1-10 1479-08 1476-07 1476-08 1476-09 1476-10 1988-02 2137-17 14A9-01 14A9-05 2563-30 27R8-19 15S9-19 26B4-26 27J7-02 26H6-26 27Y6-16 27S0-18 27S0-19 26D9-30 2473-07 2486-21 27M5-14 2368-03 1128-16 2319-06 28A3-02 2363-01 26H5-28 28D3-18 1122-20 14M7-18 1616-18 16X4-19 2711-01 2232-19 2708-15 26D6-04 26D8-24 1872-15 1873-05 2054-11 2193-11 2140-03 1855-33 1927-06 1927-20 2074-15 2107-07 2074-14 2209-20 2078-16 2085-33 2137-11 2347-16 1914-17 2358-18 1914-15 0763-06 1118-12 0763-04 15F5-15 15F5-16 2077-38 2078-20 15H2-05 16K4-05 2074-30 2408-22 2580-28 2580-29 1679-13 2497-02 2497-03 26A9-06 1138-10 1139-01 1127-08 1138-08 16M7-05 1988-06 2148-15 28E5-13 2486-27 2493-04 1912-20 0761-08 2374-11 1684-07 1684-08 1570-17 1479-06 1480-17 15F5-12 15F5-13 15H2-08 15H2-09 15D9-15 15D9-16 1691-19 1548-16 2126-08 15H7-15 15H7-16 15H7-17 15H7-18 2068-33 1857-12 2850-15 2106-17 2858-03 2858-04 27P2-07 27P2-08 26D8-20 1418-03 1418-04 1561-12 1925-04 16P4-11 16P4-12 16P4-16 16P4-17 16P4-18 2187-36 2187-37 2187-38 2331-01 2085-09 2086-17 2086-18 2086-15 2086-16 2086-19 2188-08 2188-09 2188-10 2188-36 2188-37 2188-38 2102-06 2102-07 2102-08 2253-16 2253-17 2253-18 2260-11 2260-12 2260-13 26H6-24 2493-27 1858-04 1681-09 1860-21 1861-24 2065-10 1860-20 1860-17 2050-04 2566-26 27P0-20 2850-16 16K6-11 16K6-12 16K6-13 16K6-14 16K6-15 16K6-16 2061-20 2085-12 2085-14 2283-15 2065-06 2081-07 2081-08 2081-09 2081-10 2081-11 2083-04 2083-05 2084-22 2084-23 2083-06 2083-07 2083-08 2188-31 2188-34 2188-35 2192-21 2192-22 2192-26 2192-27 2260-17 2260-18 2260-19 2260-20 2262-11 2262-12 2260-14 2260-15 2260-16 2071-25 2075-20 2263-15 2283-18 2451-17 2451-18 2347-19 2348-11 28E5-12 1733-15 2052-01 2052-02 16K6-17 16K6-18 16K6-19 2086-25 2086-26 2189-07 2082-39 2084-05 2084-12 2084-13 2084-14 2084-15 2084-06 2084-07 2192-31 2101-39 2257-08 2257-09 2257-10 2257-11 27S2-05 27S2-06 2050-06 2050-07 15D6-18 15D9-12 15D9-13 15D9-14 2071-22 2086-23 2086-24 2252-05 2255-20 2263-12 2084-08 2084-09 2084-10 2084-11 2085-04 2085-10 2188-04 2187-33 2188-39 2188-40 2102-03 2195-23 2146-17 2146-18 2146-19 2253-19 2253-20 2254-01 2257-04 2577-03 16K5-17 1616-17 27Z0-07 1905-05 1307-41 2050-01 1544-13 1536-04 2085-11 2084-19 2084-20 2084-21 2413-08 2891-14 2446-14 2446-16 1537-15 1989-20 2050-08 15H7-11 15H7-12 15H7-13 2082-40 2083-01 2084-16 2084-17 2084-18 2188-32 2189-37 2189-38 2189-39 2251-16 2251-17 2251-18 2452-01 2469-06 2232-18 2234-02 26C3-08 26C3-09 2488-02 2232-20 2039-10 1357-15 1357-16 1357-17 1357-18 26G8-08 27R8-20 15S9-20 2493-13 2493-15 2469-05 2469-07 2234-01 2348-17 2145-17 2850-12 2850-13 2145-11 2145-12 15D3-13 15D5-04 15D5-06 15D6-17 1308-35 1308-36 1419-12 1422-17 16J8-14 16J8-15 2079-18 2081-12 2192-18 2147-05 2147-06 2085-05 2085-07 2085-08 2190-14 2188-01 2119-14 28D7-19 1616-19 1921-01 1921-02 1537-17 1905-01 1905-02 2050-03 2050-05 15E3-04 15E3-05 15E3-06 15E3-01 15E3-02 15E3-03 2085-16 2085-17 2189-01 2189-05 2189-06 2256-01 2278-09 2081-06 2188-05 2188-06 2188-07 2190-17 2190-18 2190-21 2190-03 2190-04 2190-05 2189-40 2190-01 2190-02 2189-36 2192-24 2086-12 2106-28 2106-29 2110-09 2110-10 2188-02 2119-13 2056-21 2147-04 2251-15 2253-12 2253-13 2253-14 2257-05 2257-06 2257-07 2059-05 2140-08 2140-10 2473-02 2473-03 2076-22 1544-10 1733-18 1725-19 1857-27 1556-12 1681-08 1694-20 16M6-20 16M7-02 16M7-03 16M8-04 2076-19 2076-20 2076-21 2083-03 2085-18 2083-02 2085-13 2275-14 2275-15 2275-16 2278-07 2190-13 2190-15 2190-16 2192-28 2192-29 2290-05 2290-06 1711-16 2192-12 2189-04 2086-14 2086-20 2190-10 2190-19 2190-20 26E4-12 26H6-23 27L7-20 2069-01 2087-03 26D8-21 26D8-25 2873-01 2891-18 1921-03 1557-10 1736-07 1855-12 1905-03 1925-03 16J8-07 16J8-08 16J8-09 16J8-10 16J8-11 2080-03 2085-15 2252-04 2252-03 2278-08 2079-19 2080-01 2081-04 2190-06 2190-07 2190-08 2190-09 2190-11 2190-12 2192-25 2192-32 2252-02 2253-15 2251-19 2251-20 2283-16 2296-20 27S8-10 27S4-06 2486-14 2486-15 27R3-20 27R5-01 2869-18 16Z2-18 2107-04 2485-29 2487-28 2587-12 2145-08 2290-12 2486-02 2062-02 2467-03 2066-22 2066-23 2102-36 2708-19 2719-08 27R3-05 1556-16 1861-23 2581-21 2066-35 2066-36 26G4-18 26G4-11 26G7-22 26C9-13 26C9-14 2070-15 2472-10 2481-09 2195-36 1160-01 1160-05 1183-04 1217-09 1225-11 26E4-11 2212-19 2212-20 1864-17 1864-18 2427-13 2431-26 26D6-06 14A8-02 16L5-02 16L8-01 16L8-02 16L8-03 16P4-10 27M5-15 2399-09 2399-13 2399-04 2087-08 2585-18 1459-10 1459-11 2104-04 2104-07 2472-11 2472-12 2423-18 27X6-02 27X6-03 27X6-04 27J7-10 27S2-17 27S2-20 2493-26 2212-12 2212-13 2232-13 2232-14 2232-15 26E0-19 2145-13 26B5-24 26C3-05 26C3-06 2107-30 2137-12 2137-13 2581-16 2441-04 1684-05 1572-19 2192-10 2103-29 2104-06 2356-01 27P0-08 27S0-16 1571-13 1867-19 1935-18 1908-03 0756-30 1139-02 1139-03 1139-04 1139-05 2427-19 2427-20 2497-14 2305-23 2307-04 27S2-19 2543-04 27Z0-04 2059-23 2382-05 26G6-30 2358-14 2570-29 1118-08 13A9-08 2348-20 26E4-07 1684-06 2057-28 2487-30 1873-03 2535-22 2535-21 2049-16 27S0-20 186225 2395-28 2865-17 2533-21 1910-08 27S4-01 27P2-01 2485-30 26D8-23 26H6-25 2367-10 2439-14 2486-18 2389-15 1142-17 1860-22 2389-12 1142-14 2865-19 1310-20 1871-12 2069-24 2250-08 2256-14 27S2-01 2227-14 2547-30 2548-01 1925-07 2413-15 2565-21 2565-22 2577-05 2585-22 1556-17 1843-02 26F0-29 27L8-01 27L8-08 2409-26 26A4-27 26B6-30 14G5-06 2565-23 2565-28 2073-27 2079-39 2427-23 2427-24 2544-21 1383-03 2566-25 2566-30 2486-07 2402-28 2226-16 2226-18 26D6-11 26D8-28 2585-25 27M5-11 26A4-23 26A4-29 26A5-01 2587-13 26D0-25 26D0-26 26D5-30 26D9-25 26E0-14 26G4-13 2585-20 2432-29 2432-30 2432-24 1211-19 1211-04 1211-05 1335-08 1335-09 1335-10 1335-11 1910-06 2438-14 27L1-05 2077-40 2078-21 2535-12 2535-13 2535-14 2550-12 2550-14 2565-25 2565-26 27J6-19 2550-15 2564-23 2565-24 2108-30 14B1-03 14B1-04 2423-03 26B4-22 2072-40 2107-38 2229-12 2233-20 2450-17 2451-01 2446-17 2446-19 2406-25 27Z0-12 27P6-16 27P8-01 16R0-06 16R5-06 2581-14 2581-15 2706-01 2493-25 2867-09 2871-19 27R1-16 2355-01 2485-28 1931-09 2039-11 2224-17 2473-04 2473-06 2486-17 2714-09 2714-10 2458-23 1387-16 26F3-20 26G4-12 2581-17 2437-18 27J6-17 2486-13 27Z6-18 27Z6-19 2472-08 2472-09 27S2-15 2720-16 16X4-16 26G4-10 16K3-11 16K3-12 16K3-13 26H3-25 2427-16 2427-17 26H5-29 14M7-19 2396-18 2321-23 2321-24 2362-20 1122-19 1146-01 1146-02 1146-03 16K3-10 2071-40 2219-17 2229-14 2535-06 2535-07 27W3-19 27W3-20 2398-15 2398-16 2407-06 26D4-28 26D8-22 2256-17 26D0-21 26D1-30 2061-29 1139-06 1139-07 15H2-11 15H2-12 15H2-14 2585-06 28E7-12 28F7-02 1132-05 1132-10 1137-02 2423-21 2425-13 26H0-30 14M0-20 2350-04 2408-23 2409-23 2432-25 1924-12 1935-17 2073-37 2042-03 2497-11 2140-12 2704-16 27L1-15 2220-06 2444-07 26F0-18 2080-25 2106-10 28E9-12 28F5-13 16R5-07 14B1-07 15B7-07 15B7-13 2842-14 2842-15 1853-03 1860-16 1733-16 1736-03 2070-01 2070-02 1601-03 1705-20 1208-05 1211-07 1211-08 2330-20 2409-24 2409-25 2107-18 1139-08 1139-10 1140-07 1140-10 1140-08 1156-06 1158-09 1156-07 14B3-08 14B3-09 14B3-10 2369-02 2564-21 2564-22 2227-09 1572-20 1673-01 27Y6-11 27Z0-06 27Z0-11 2413-12 2413-13 1928-01 1848-04 1925-08 1909-09 1909-10 1926-03 2567-23 2423-05 1936-10 1171-10 1211-13 1988-12 1983-06 2494-22 2054-09 1563-04 26B4-21 16K7-12 16M7-06 2856-19 2869-17 2856-20 16Y8-19 16Z2-17 16Y8-20 16L3-20 16K7-11 1866-13 2427-30 2428-01 26A5-08 26A9-03 2304-30 2305-08 2137-06 2564-29 1984-01 2384-19 26A4-24 26A4-25 2048-06 2048-29 2050-28 2072-19 26D6-07 26D6-08 26D6-10 15F0-14 15H2-06 15H2-07 15H5-01 2423-22 2867-05 2867-06 2894-07 2867-12 2894-10 2894-11 16K5-16 16K5-20 16M7-07 16M7-08 2867-13 2585-05 2585-26 2145-14 2145-15 2080-26 27X4-19 27X6-01 2398-17 2396-01 26A9-07 15H4-08 15H4-18 15H5-02 15H5-03 16K5-17 16K5-19 2069-08 2070-05 2313-17 2871-20 16R5-09 16R5-10 16R5-11 27L8-13 1992-19 1999-16 1999-17 2550-10 2550-11 1865-05 1908-10 0757-07 2867-10 16P1-06 16P1-07 16P1-08 16P1-09 2306-02 2306-03 1159-17 1159-18 27P2-05 1927-08 2219-20 16J9-20 1611-20 2312-26 2312-28 2444-08 1987-12 28F5-17 2493-08 2497-17 2422-17 26G4-16 2137-16 2896-02 27P2-04 2079-27 2079-36 2565-27 2567-21 2074-04 2074-01 2538-29 15B7-05 15B7-09 15B7-10 26B3-29 1991-07 1860-23 1871-11 2059-11 27L1-14 27Z8-13 2550-04 2896-05 2060-25 2066-39 26D5-28 16M7-19 16M7-20 1173-01 1985-06 2842-13 27Z8-04 27Z8-05 1919-07 1919-08 2560-03 13A1-02 2256-16 1919-03 1911-15 2050-31 2050-32 1909-08 2074-13 26G4-15 27Y6-15 27Y6-18 2842-17 2858-01 2188-03 26G4-17 2497-22 2538-28 1156-05 1158-10 1461-01 1461-02 1870-06 1870-14 2429-29 1263-19 2071-24 1469-01 1480-18 1925-12 1926-02 1548-17 1845-15 198703 1987-04 1159-20 15D9-20 15D9-19 2145-09 2145-10 2148-16 27Y6-10 27Z0-05 2218-14 27J7-08 2869-19 16Z2-19 2894-14 2137-18 2149-14 2585-24 2384-20 2385-01 27X6-05 2347-09 2347-10 2358-17 1118-11 26C3-14 26C3-20 2708-17 2708-18 2234-03 1156-08 1156-10 1344-09 27Y6-17 27S0-17 27Y6-13 27X4-15 26H4-29 27S2-11 27S2-12 2066-38 1867-20 1868-01 2047-09 2086-35 2369-06 2720-18 16X4-18 1987-14 1874-05 1985-08 1985-10 2494-28 2497-21 2064-06 2067-29 2472-19 2321-21 2195-03 1714-18 26B4-24 26B4-25 2495-02 2543-03 2543-08 2543-09 26E0-22 27Z0-13 2048-32 27S2-02 27S2-03 2704-18 2719-03 1208-08 1208-09 1211-02 1211-03 2367-14 2367-15 2059-10 27S2-04 27S2-08 27R5-02 27R3-18 2051-32 2896-06 2894-13 2058-32 27Y6-12 27S2-13 2102-33 2219-19 2225-01 2234-06 2224-15 2224-16 15C9-09 2229-10 2233-19 2233-13 2234-04 2496-21 1871-03 2382-07 2382-12 2403-21 2493-24 26H3-27 2848-11 26D8-19 26D8-27 2479-08 1601-02 1858-05 2317-18 2332-11 2332-12 2396-13 2398-06 2312-27 2332-20 26D0-24 1853-02 1855-10 1855-11 1858-02 1858-03 1849-02 1853-01 1691-14 1860-12 1860-13 1860-14 1860-11 2104-21 2104-03 2368-04 1128-17 2070-22 2581-19 2581-20 27J6-20 27X4-16 27X4-17 2850-07 2850-08 1146-05 1146-06 1147-04 1485-11 2067-30 2070-40 2414-10 2414-11 2414-12 2427-22 2430-28 2431-24 2431-25 26D9-29 26E4-09 2850-11 28B6-19 1611-19 2192-16 27Z8-10 27Z8-11 26G4-19 2858-05 2860-20 2368-13 2369-13 2304-12 1131-08 1132-01 1132-02 15J6-10 2144-13 2140-04 2894-09 2894-16 2894-17 1983-09 27Z8-15 2822-11 28D7-20 28F5-18 2432-27 1862-27 2719-09 2304-29 2305-09 2566-28 2566-29 2073-09 2059-40 2422-23 2422-24 2149-13 2430-30 2431-21 2431-22 2108-25 2108-26 1988-14 2494-27 2209-09 2209-10 26B6-29 14G5-05 2072-28 27J7-05 26G7-21 2073-40 27Z8-17 2848-13 26F0-30 2534-04 2534-03 2535-09 2535-10 2535-11 2405-28 2405-29 26B3-28 2544-22 2550-18 2488-04 27W5-01 1338-12 1338-13 1338-14 2202-11 2292-17 143115 2209-19 2327-26 1988-10 2056-09 2056-12 1559-13 1539-13 1539-14 1571-12 1480-19 1490-09 1490-13 1490-14 1491-01 15B7-08 28F5-14 2468-17 28E7-15 2432-22 2432-23 190520 2149-12 2104-08 2104-20 1867-16 1156-04 2444-14 1868-02 2423-04 2535-05 2192-13 2196-03 2497-15 2422-16 2137-03 2137-04 1698-05 1711-17 1991-13 1991-14 2233-17 2233-18 2543-11 2543-12 1544-06 26F1-18 26H6-28 26G4-14 2071-23 27Y6-14 2422-28 2423-20 2056-10 2056-13 2581-24 2585-23 2086-36 2108-37 2047-38 0884-03 16L7-18 16P4-01 2085-30 2087-11 26F0-28 26F1-16 27X8-20 15V4-20 2321-22 16P1-12 16P3-02 16P3-03 1865-19 1999-20 1861-20 15H7-19 15H7-20 15H8-01 15H8-09 15H9-13 15H9-14 15H9-15 15H9-16 15J0-03 15J0-04 15J0-06 15J0-08 16K4-01 16K4-02 16K4-03 16K4-04 16K5-07 16K5-08 2267-20 2262-19 2262-20 2221-03 199902 2414-09 2535-08 2385-02 2385-03 2327-12 2867-02 2427-14 2427-15 2234-08 2234-09 2236-10 1008-10 16K6-20 16K7-01 16K7-02 16K7-03 2486-12 2389-16 1142-18 1461-05 1468-18 1468-19 1468-20 2330-16 2330-18 1171-07 1171-08 1171-09 1173-10 1160-08 1171-06 1174-06 1174-07 2275-11 2278-10 2278-11 2330-19 2140-02 2262-17 2267-17 2267-18 2283-17 2330-17 2585-01 2585-02 2585-03 2585-16 2585-17 2399-05 2399-07 2233-15 2891-19 2079-29 15G9-03 15G9-04 2494-29 26G3-30 27J7-12 2305-10 1142-09 1145-10 26E0-20 26E0-21 2293-05 1140-01 2052-05 26D5-23 2196-07 1858-35 26D0-22 26D0-23 2192-09 2322-08 2107-22 2107-23 2190-38 2195-10 2402-29 2195-01 2103-36 2321-29 2066-28 1139-09 1173-08 2070-07 2310-08 1334-15 1335-13 1459-14 14B2-14 2488-03 1428-07 1182-08 1211-06 1331-13 1338-15 1338-16 27S2-14 1147-05 1149-01 26C7-28 14H8-02 1926-04 2432-26 2432-28 2566-27 26E4-13 15J6-02 15J6-08 2427-28 2037-13 2059-12 2058-18 2429-28 1263-18 2148-14 2149-08 2149-20 16P6-19 16R7-03 2256-02 2076-18 2262-16 2414-13 2891-16 2891-17 2148-09 28E7-18 2086-31 0829-20 2894-18 2894-19 1601-01 1698-03 16P6-14 2058-23 1429-13 2250-01 2307-05 1919-02 1919-04 2220-02 2220-04 1469-02 2220-07 26A5-02 2850-14 2431-23 2471-20 27K0-06 1211-20 15B7-06 27R3-09 27R3-13 1907-02 2321-20 2438-04 2409-22 2408-20 15C8-11 14B1-05 14B1-08 2332-10 2438-13 26E0-15 2037-11 1991-17 1991-16 26H6-27 2061-40 0802-01 199010 27Z0-08 15R8-20 27J6-14 27J6-13 27J7-01 27J6-18 27Z8-07 27P2-09 2479-10 2493-17 26D5-21 2850-02 27X6-06 27Y6-09 187018 1925-13 27Z8-06 2488-01 26F1-21 2708-16 27Z8-14 2042-02 2321-27 26A9-10 27L1-12 27L1-13 27L1-03 27L1-04 27K9-16 27L1-01 27L1-02 1983-08 1868-14 2288-08 2328-02 2473-01 2077-36 1928-02 1924-11 2192-11 1149-07 1149-08 1156-03 1167-02 16R0-07 16R0-08 1863-05 27J7-11 27J9-12 27L1-06 1935-16 26G4-08 15C9-08 27L1-07 27L1-08 15C8-12 15C8-13 15C8-14 15C8-15 27L1-10 27L1-11 2301-05 2308-06 2229-03 2229-07 2067-33 27S2-16 27Z8-16 1698-06 1733-17 26D5-27 26E0-17 1862-16 27J6-16 1538-10 27Z2-09 2494-23 2494-24 2494-25 2494-26 28F5-15 28F5-16 16R5-04 16R5-05 2061-34 2064-05 2137-05 2227-12 2227-13 2413-10 2422-15 2226-19 2233-14 1016-05 1017-01 2102-04 2102-05 27R3-14 27R3-16 2578-23 13B8-10 2708-20 27Z6-20 2867-04 2486-22 16L8-08 16M4-14 16M4-15 16M4-16 16M7-04 2550-16 2563-29 1905-06 2848-10 2842-04 2061-37 2061-38 0801-28 0801-29 2413-07 2422-19 |
| FEI Number |
1823260
|
Recalling Firm/ Manufacturer |
Roche Diagnostics Corporation 9115 Hague Rd Indianapolis IN 46256-1025
|
| For Additional Information Contact | 800-428-2336 |
Manufacturer Reason for Recall | Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation. |
FDA Determined Cause 2 | Software design |
| Action | Roche sent an Urgent Medical Device Correction (UMDC) sent via UPS Ground (receipt required) dated July 6, 2017, and an amended UMDC on August 31 ,2017. All consignees will be notified by Urgent Medical Device Correction sent via UPS Ground (receipt required).
Customers/consignees were instructed the following: If the Tip/Cup pick up error alarm is issued then followed by the Abnormal L2-line Movement alarm on your MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer, after which the system goes into Sampling Stop, follow the steps in this UMDC and the appropriate How to Identify Potentially Mismatched Sample Results within the enclosed document. For further questions, please call (317) 521-4343. |
| Quantity in Commerce | 2517 |
| Distribution | USA (nationwide) Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|