Class 2 Device Recall JPlasma(R) Handpiece

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall JPlasma(R) Handpiece see related information
Date Initiated by Firm February 08, 2018
Create Date March 21, 2018
Recall Status1 Terminated 3 on September 17, 2020
Recall Number Z-1155-2018
Recall Event ID 79398
510(K)Number K170188  
Product Classification Electrosurgical, cutting & coagulation & accessories - Product Code GEI
Product J-Plasma(R) Handpiece, Catalog Numbers: GS-018C, GS-270C, BVX-270B
Code Information All lots manufactured since 01/01/2014
Recalling Firm/
Manufacturer
Bovie Medical Corporation
5115 Ulmerton Rd
Clearwater FL 33760-4004
For Additional Information Contact Dr. Topaz J. Kirlew, MBA, MT(ASCP)
727-803-8617
Manufacturer Reason
for Recall
Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.
FDA Determined
Cause 2
Nonconforming Material/Component
Action The firm initiated their recall on 02/08/2108 by email to distributors, and letter to the remaining consignees. Distributors were directed as follows: "We ask that you immediately: 1. Quarantine your inventory of the J-Plasma Handpieces o Provide us with a list of catalog numbers and quatities that you have on your possession using the attached Inventory List template. o Email the information to me at Rubiela.Maldonado@boviemed.com Note: as soon as we receive the information from you we will ship the box labels and insert stuffer notices for the stock that you have on your possession. o After receiving these labels, please follow the attached instructions on how to add each of the insert stuffer notice and the outer box label to each of the boxes that you have on your possession. o Document the number of labels/stuffers added to each of the products on the Inventory List provided, sign and date to confirm this was performed. o Return the signed document to me via email at Rubiela.maldonado@boviemed.com. 2. Inform your customers about the correction: o Identify and contact your customers that have received the product affected by this correction. o Provide them with a copy of the Urgent Medical Device Correction Template attached to this letter. o Track and document the customer responses to ensure 100% effectiveness. o Provide us a summary of the effectiveness of this correction (number of customers impacted and number of responses received)." The remaining consignees (users of the device) were instructed as follows: "ACTIONS: At the end of each procedure, inspect the handpiece cable plug and generator receptacle for the presence of fluids. If fluids are present, immediately discontinue use of the generator. In addition to following the J-Plasma¿ handpiece Instructions for Use, immediately begin to follow the supplemental instructions provided below to ensure that the J-Plasma¿ handpiece cable plug is properly and co
Quantity in Commerce 5416 units
Distribution worldwide
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = GEI and Original Applicant = Bovie Medical Corporation
-
-