|
Class 2 Device Recall Dimension Vista Cardiac Troponin I (CTNI) Flex |
 |
Date Initiated by Firm |
January 02, 2018 |
Date Posted |
May 04, 2018 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1671-2018 |
Recall Event ID |
79505 |
510(K)Number |
K063756
|
Product Classification |
Immunoassay method, troponin subunit - Product Code MMI
|
Product |
Dimension Vista¿ Cardiac Troponin I (CTNI) Flex¿ reagent cartridge |
Code Information |
CTNI, immunoassay method, troponin subunit, K6421, SMN# 10445098 Lot codes: 16336BA 16347BD 16363BC 17032AD 17045BC 17054BD 17074BC 17093BD 17121BA 17145BE 17180BB 17193BD 17201BB 17223BA 17264BC |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc. 511 Benedict Ave Tarrytown NY 10591-5005
|
For Additional Information Contact |
Ms. Aarti M. Aziz 914-631-8000
|
Manufacturer Reason for Recall |
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
|
FDA Determined Cause 2 |
Error in labeling |
Action |
Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA |
Distribution |
Nationwide and Canada, Mexico, and Thailand |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = MMI and Original Applicant = DADE BEHRING, INC.
|
|
|
|