|
Class 2 Device Recall PHILIPS Xper Flex Cardio Physiomonitoring System |
|
Date Initiated by Firm |
March 14, 2018 |
Create Date |
May 14, 2018 |
Recall Status1 |
Terminated 3 on April 30, 2021 |
Recall Number |
Z-1867-2018 |
Recall Event ID |
79611 |
Product Classification |
Monitor, physiological, patient (without arrhythmia detection or alarms) - Product Code MWI
|
Product |
PHILIPS Xper Flex Cardio Physiomonitoring System, Model Numbers: 453564241901, 453564483321, 453564621791, and 989803199561 (international only) The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular, disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health care providers, such as physicians, registered nurses, cardiovascular technicians, radiology technicians, and other clinical health care providers.
|
Code Information |
Model Number: 453564241901 Serial Numbers/UDI (if available): US1203010330 US1203010371 US1203010373 US1203010382 US1203010400 US1203010413 US1203010421 US1311000007 US1311000009/ (01)00884838083523(21)US1239000009 US1311000017/ (01)00884838083523(21)US1331000034 US1239000009 US1331000034 US1231000061 US1108010053 US1202010217 US1234000060 US1337000038 US1337000040 US1337000043 US1343000024 US1334000021 US1234000066 US1202010242/ (01)00884838083523(21)US1202010242 US1244000002/ (01)00884838083523(21)US1244000002 US1251000087/ (01)00884838083523(21)US1251000087 US1202010168 US1203010284 US1244000061 US1251000005 US1203010302/ (01)00884838083523(21)US1203010302 US1244000016/ (01)00884838083523(21)US1203010340 US1231000038/ (01)00884838083523(21)US1231000038 US1234000024/ (01)00884838083523(21)US1234000024 US1244000016/ (01)00884838083523(21)US1244000016 US1244000040/ (01)00884838083523(21)US1244000040 US1244000066/ (01)00884838083523(21)US1244000066 US1251000013/ (01)00884838083523(21)US1251000013 US1311000025/ (01)00884838083523(21)US1311000025 US1319000059/ (01)00884838083523(21)US1319000059 US1337000032/ (01)00884838083523(21)US1337000032 US1231000048/ (01)00884838084926(21)US1231000048 US1251000031/ (01)00884838084926(21)US1251000031 US1304000040/ (01)00884838084926(21)US1304000040 US1334000041/ (01)00884838084926(21)US1334000041 US1107010016 US1107010019 US1107010022 US1107010036 US1108010062 US1109010104 US1202010153 US1202010173 US1202010233 US1203010261 US1203010295 US1203010307 US1203010320 US1203010345 US1203010377 US1231000022 US1234000013 US1234000049 US1239000024 US1239000045 US1244000032 US1244000055 US1248000023 US1251000030 US1251000033 US1251000052 US1251000077 US1304000020 US1311000069 US1319000009 US1319000016 US1319000030 US1319000069 US1324000008 US1324000014 US1324000039 US1328000002 US1331000040 US1337000055 US1343000001 US1343000009 US1343000027 US1343000059 US1343000073 US1239000002 US1352000047 US1324000011 US1319000038 US1239000019 US1107010007 US1201010117 US1201010119 US1203010401 US1231000026 US1231000049 US1231000059 US1234000051 US1325000009 US1234000086 US1238000005 US1311000011 US1311000062 US1203010378 US1248000041/ (01)00884838083523(21)US1248000041 US1239000011 US1202010208/ 01)00884838083523(21)US1202010208 US1244000054/ (01)00884838083523(21)US1244000054 US1324000049/ (01)00884838083523(21)US1324000049 US1234000068 US1239000008 US1319000091 US1325000017 US1107010011 US1231000058 US1337000024 US1325000007/ (01)00884838083523(21)US1325000007 US1328000014 US1331000067 US1201010129 US1224000051 US1231000008 US1231000016 US1231000028 US1319000024 US1337000071 US1251000053 US1231000046 US1311000002/ (01)00884838083523(21)US1311000002 US1311000054/ (01)00884838083523(21)US1311000054 US1109010076 US1201010113 US1201010116 US1201010144 US1202010189 US1337000028 US1352000020 US1319000097 US1325000002 US1311000029/ (01)00884838083530(21)US1311000029 US1107010020 US1108010040 US1231000036 US1107010008 US1251000010 US1331000038/ (01)00884838083523(21)US1331000038 US1244000022/ (01)00884838083523(21)US1244000022 US1311000058/ (01)00884838083523(21)US1311000058 US1331000057/ (01)00884838083530(21)US1331000057 US1107010037 US1244000019 US1324000032 US1331000028 US1334000032 US1319000047 US1324000029 US1325000020 US1248000019 US1234000048 US1244000042 US1244000053 US1248000032 US1203010291 US1337000019 US1319000033/ (01)00884838083523(21)US1319000033 US1244000036/ (01)00884838084926(21)US1244000036 US1109010092 US1201010143 US1203010386 US1352000003 US1352000012 US1407000022 US1319000060 US1251000050 US1319000001 US1224000044 US1109010098 US1107010015 US1107010023 US1108010041 US1109010068 US1407000002 US1407000029 US1407000031 US1304000027 US1231000067 US1231000073 US1304000034 US1324000030/ (01)00884838083523(21)US1324000030 US1224000012 US1337000050 US1337000066 US1202010149 US1202010176 US1203010348 US1231000034 US1109010081 US1203010265 US1239000006 US1234000021 US1319000015 US1203010316 US1231000010 US1201010132 US1107010012 US1107010017 US1107010021 US1108010059 US1201010115 US1234000083 US1319000052 US1337000018 US1337000048 US1304000011 US1319000029 US1331000004 US1251000038/ (01)00884838083523(21)US1251000038 US1202010196 US1202010197 US1203010354 US1248000008 US1408000013 US1203010357 US1248000018 US1328000012 US1331000033 US1224000002 US1319000094 US1331000016 US1343000056 US1343000020 US1109010107/ (01)00884838083523(21)US1109010107 US1324000018 US1343000060 US1201010138 US1251000073 US1201010123 US1203010351 US1109010086 US1248000056 US1304000002 US1331000052 US1343000012 US1231000042 US1107010027 US1203010391 US1231000062 US1239000026 US1311000018 US1311000026 US1311000044 US1248000017 US1248000066 US1319000020 US1331000064 US1244000018 US1325000022/ (01)00884838083523(21)US1325000022 US1324000022/ (01)00884838084926(21)US1324000022 US1244000067 US1248000006 US1248000013 US1248000020 US1248000049 US1248000051 US1248000062 US1319000005 US1202010200 US1202010177 US1202010191 US1324000033 US1234000087/ (01)00884838083523(21)US1234000087 US1203010264/ (01)00884838083530(21)US1203010264 US1203010376 US1203010388 US1231000047 US1304000026 US1244000046 US1319000026 US1203010389 US1244000060 US1248000060 US1337000020 US1343000065 US1407000054 US1331000018 US1109010066 US1109010077 US1109010111 US1109010088 US1107010001 US1203010266 US1203010321 US1203010390 US1319000098 US1234000072/ (01)00884838083523(21)US1234000072 US1311000021/ (01)00884838083523(21)US1311000021 US1202010205 US1202010210 US1203010329 US1239000025 US1319000004 US1319000084 US1352000009 US1352000063 US1407000055 US1203010288/ (01)00884838083523(21)US1203010288 US1244000028/ (01)00884838083523(21)US1244000028 US1324000012 US1239000015 US1224000025 US1324000006 US1324000051 US1343000050 US1231000030 US1251000047 US1331000017 US1337000030 US1337000036 US1337000051 US1337000054 US1337000062 US1337000063 US1337000064 US1109010094 US1224000053 US1231000044 US1248000055 US1319000054 US1325000016 US1328000005 US1224000043 US1231000086 US1311000005 US1324000013 US1337000059 US1343000071 US1343000077 US1343000091 US1239000050 US1224000001 US1248000054 US1304000013 US1248000043 US1202010201 US1251000017/ (01)00884838083523(21)US1251000017 US1109010103 US1239000032 US1319000011 US1319000027 US1334000002 US1334000027 US1319000086 US1202010203 US1231000040 US1331000009 US1203010253 US1203010258 US1203010303 US1203010344 US1203010368 US1108010050 US1108010063 US1311000030 US1311000008/ (01)00884838083523(21)US1311000008 US1107010024 US1234000059 US1343000066 US1202010241 US1210000005 US1224000011 US1224000019 US1224000054 US1407000006 US1407000047 US1407000050 US1407000051 US1408000005 US1408000021 US1408000022 US1408000027 US1408000028 US1408000030 US1203010412 US1319000032 US1311000034/ (01)00884838083523(21)US1311000034 US1224000040/ (01)00884838083523(21)US1224000040 US1231000003 US1231000005 US1231000006 US1234000028 US1231000053 US1202010158 US1202010159 US1202010162 US1202010163 US1331000054/ (01)00884838084926(21)US1331000054 US1337000014/ (01)00884838084926(21)US1337000014 US1324000028 US1248000047 US1251000029 US1407000039 US1319000055 US1234000030 US1109010071/ (01)00884838083523(21)US1109010071 US1203010272/ (01)00884838083523(21)US1203010272 US1203010296/ (01)00884838083523(21)US1203010296 US1203010394/ (01)00884838083523(21)US1203010394 US1224000008/ (01)00884838083523(21)US1224000008 US1224000026/ (01)00884838083523(21)US1224000026 US1231000009/ (01)00884838083523(21)US1231000009 US1234000002/ (01)00884838083523(21)US1234000002 US1234000007/ (01)00884838083523(21)US1234000007 US1234000082/ (01)00884838083523(21)US1234000082 US1244000052/ (01)00884838083523(21)US1244000052 US1311000051/ (01)00884838083523(21)US1311000051 US1311000071/ (01)00884838083523(21)US1311000071 US1325000024/ (01)00884838083523(21)US1325000024 US1337000006/ (01)00884838083523(21)US1337000006 US1337000073/ (01)00884838083523(21)US1337000073 US1231000015/ (01)00884838083530(21)US1231000015 US1108010046 US1109010096 US1201010131 US1202010170 US1202010188 US1203010252 US1203010300 US1203010331 US1203010370 US1203010403 US1203010405 US1224000022 US1231000024 US1231000079 US1234000043 US1234000052 US1234000063 US1234000073 US1234000090 US1239000001 US1239000010 US1239000037 US1239000041 US1239000043 US1244000015 US1244000020 US1244000037 US1248000022 US1248000033 US1248000040 US1248000061 US1251000007 US1251000035 US1304000033 US1304000044 US1311000020 US1311000022 US1311000032 US1311000053 US1319000021 US1319000040 US1319000063 US1319000089 US1325000008 US1325000015 US1328000001 US1343000069 US1244000035/ (01)00884838083523(21)US1244000035 US1319000013/ (01)00884838083523(21)US1319000013 US1231000070 US1202010167 US1202010182 US1202010193 US1343000004 US1202010225/ (01)00884838083523(21)US1202010225 US1319000023 US1325000004 US1311000067 US1248000048/ (01)00884838083523(21)US1248000048 US1202010183 US1248000053 US1251000037 US1334000024 US1337000060 US1324000026-D US1202010207 US1244000049 US1201010114 US1201010118 US1319000057 US1201010125 US1203010380 US1251000028 US1324000007 US1248000027 US1203010410 US1244000001 US1244000057 US1319000041 US1337000065 US1352000029 US1352000053 US1352000058 US1311000057 US1234000089/ (01)00884838084926(21)US1234000089 US1203010356 US1201010127 US1202010211 US1203010286 US1203010310 US1203010335 US1244000004 US1244000012 US1311000039 US1331000037 US1224000015/ (01)00884838083523(21)US1224000015 US1109010067 US1109010110/ (01)00884838083523(21)US1109010110 US1248000045/ (01)00884838083523(21)US1248000045 US1325000019 US1202010157/ (01)00884838083523(21)US1202010157 US1319000045/ (01)00884838083523(21)US1319000045 US1231000055 US1239000035 US1311000024 US1319000007 US1324000003 US1324000020 US1201010121 US1203010324 US1224000038 US1234000071 US1248000016 US1203010267 US1244000064 US1311000010 US1319000066 US1251000054/ (01)00884838083523(21)US1251000054 US1352000002 US1108010055 US1201010133 US1201010137 US1202010186 US1203010290 US1203010337 US1337000009 US1324000034 US1324000038/ (01)00884838083523(21)US1324000038 US1203010385 US1224000009 US1231000014 US1234000078 US1324000010 US1324000036 US1324000035/ (01)00884838083523(21)US1324000035 US1202010238 US1248000064 US1311000056 US1412000008 US1234000041 US1107010025 US1107010077 US1334000007 US1352000011 US1352000027 US1203010417/ (01)00884838083523(21)US1203010417 US1203010404 US1352000050 US1202010209/ (01)00884838083523(21)US1202010209 US1337000049/ (01)00884838083523(21)US1337000049 US1248000078 US1251000002 US1251000078 US1319000003 US1319000028 US1325000026 US1331000061 US1107010033 US1251000055 US1337000011 US1203010342 US1108010049/ (01)00884838083523(21)US1108010049 US1234000046/ (01)00884838083523(21)US1234000046 US1319000006/ (01)00884838083523(21)US1319000006 US1319000035 US1343000058 US1311000050 US1238000009 US1331000010 US1234000039 US1234000047 US1234000061 US1239000016 US1108010042 US1202010160 US1202010214 US1231000025 US1248000009 US1203010281 US1203010313 US1203010350 US1203010355 US1324000043/ (01)00884838083523(21)US1324000043 US1352000055 US1202010213 US1202010226 US1202010231 US1234000011 US1304000025 US1231000037/ (01)00884838083523(21)US1231000037 US1203010392/ (01)00884838083523(21)US1203010392 US1244000021/ (01)00884838083523(21)US1244000021 US1201010112 US1231000051 US1337000052 US1319000010/ (01)00884838083523(21)US1319000010 US1203010381 US1231000027 US1337000015 US1244000008 US1334000017 US1244000007 US1325000010 US1331000015 US1331000030 US1334000033 US1334000035 US1334000036 US1334000040 US1334000042 US1334000044 US1337000026 US1231000020 US1248000074 US1231000002 US1324000015 US1202010219 US1202010224 US1202010227 US1202010236 US1202010239 US1202010248 US1203010270 US1224000007 US1202010199 US1304000047 US1311000004 US1328000006 US1202010194 US1203010347 US1238000004 US1319000079 US1331000019 US1331000024 US1331000031 US1331000046 US1331000065 US1234000031 US1251000067 US1107010030 US1319000078 US1109010097 US1224000032 US1239000007 US1331000029 US1234000042 US1311000006/ (01)00884838084926(21)US1311000006 US1319000051/ (01)00884838083523(21)US1319000051 US1311000077/ (01)00884838083523(21)US1311000077 US1328000004 US1251000021 US1324000017 US1343000070 US1234000045 US1324000025 US1203010418 US1334000026 US1334000028 US1203010414 US1319000061 US1331000014 US1108010051 US1108010061 US1337000034/ (01)00884838083523(21)US1337000034 US1109010105 US1203010289 US1203010326 US1319000064 US1324000021 US1311000080 US1231000063 US1109010083/ (01)00884838083523(21)US1109010083 US1107010010/ (01)00884838083523(21)US1107010010 US1244000005 US1109010065 US1248000036 US1109010084 US1202010175/ (01)00884838083523(21)US1202010175 US1201010122 US1248000079 US1325000025 US1201010126/ (01)00884838083523(21)US1201010126 US1234000025 US1343000013/ (01)00884838083523(21)US1343000013 US1244000065 US1334000030/ (01)00884838083523(21)US1334000030 US1108010058/ (01)00884838083523(21)US1108010058 US1201010120 US1202010166 US1251000044 US1324000026 US1304000007 US1234000056/ (01)00884838083523(21)US1234000056 US1331000060 US1311000013 US1202010155/ (01)00884838083523(21)US1202010155 US1251000020/ (01)00884838083523(21)US1251000020 US1337000003/ (01)00884838083523(21)US1337000003 US1337000022/ (01)00884838083523(21)US1337000022 US1337000042 US1343000078 US1337000012 US1203010285 US1224000033 US1234000069/ (01)00884838083523(21)US1234000069 US1343000021/ (01)00884838083523(21)US1343000021 US1304000005/ (01)00884838083523(21)US1304000005 US1231000029 US123100080 US1234000005 US1319000008 US1319000083 US1337000021 US1319000042/ (01)00884838083523(21)US1319000042 US1325000023 US1239000048 US1319000072 US1234000079 US1251000051 US1251000097 US1319000012 US1203010411 US1203010250 US1203010256 US1203010259 US1203010263 US1203010269 US1203010273 US1203010274 US1203010275 US1203010277 US1203010282 US1203010299 US1203010301 US1203010304 US1203010312 US1203010318 US1203010365 US1203010369 US1203010372 US1203010399 US1203010406 US1203010415 US1203010416 US1203010419 US1203010424 US1210000003 US1210000004 US1224000005 US1224000028 US1224000031 US1231000001 US1231000011 US1231000021 US1231000039 US1231000041 US1231000043 US1231000054 US1231000056 US1231000072 US1231000087 US1231000088 US1231000089 US1234000010 US1234000015 US1234000022 US1234000026 US1234000027 US1234000029 US1234000033 US1234000034 US1234000038 US1234000044 US1234000065 US1234000075 US1234000076 US1234000081 US1234000085 US1239000003 US1244000030 US1244000045 US1244000047 US1244000051 US1244000056 US1244000059 US1244000063 US1248000003 US1248000069 US1248000076 US1248000077 US1251000001 US1251000004 US1251000009 US1251000032 US1251000034 US1251000036 US1251000042 US1251000064 US1251000094 US1304000001 US1304000003 US1304000015 US1304000022 US1304000028 US1304000031 US1304000032 US1304000037 US1304000038 US1311000001 US1311000015 US1311000019 US1311000031 US1311000041 US1311000046 US1311000052 US1311000055 US1311000059 US1311000060 US1311000061 US1311000063 US1311000064 US1311000066 US1311000068 US1311000070 US1311000073 US1311000074 US1311000075 US1311000078 US1311000079 US1319000034 US1319000044 US1319000046 US1319000048 US1319000050 US1319000053 US1319000056 US1319000062 US1319000068 US1319000073 US1319000076 US1319000081 US1319000088 US1319000090 US1319000092 US1319000096 US1319000099 US1319000100 US1328000011 US1331000005 US1331000027 US1331000035 US1331000036 US1331000039 US1331000045 US1331000048 US1331000055 US1331000066 US1334000003 US1334000005 US1334000006 US1334000012 US1334000014 US1334000015 US1334000018 US1334000023 US1334000029 US1334000034 US1334000046 US1334000048 US1337000002 US1337000004 US1337000010 US1337000027 US1337000031 US1337000037 US1337000044 US1337000046 US1337000058 US1337000067 US1343000007 US1343000010 US1343000014 US1343000015 US1343000017 US1343000018 US1343000023 US1343000029 US1343000030 US1343000032 US1343000035 US1343000036 US1343000037 US1343000043 US1343000045 US1343000046 US1343000047 US1343000048 US1343000052 US1343000055 US1343000062 US1343000063 US1343000064 US1343000074 US1343000080 US1203010254 US1203010255 US1203010268 US1203010292 US1203010305 US1203010343 US1203010409 US1231000065 US1231000068 US1231000077 US1234000023 US1234000054 US1234000058 US1234000074 US1251000056 US1311000037 US1311000040 US1319000043 US1319000093 US1328000010 US1331000058 US1334000008 US1334000010 US1337000069 US1343000040 US1343000042 US1343000053 US1343000076 US1202010169 US1202010165 US1202010216 US1203010276 US1203010349 US1109010090 US1319000037 US1244000013/ (01)00884838083530(21)US1244000013 US1224000029 US1109010064 US1331000041/ (01)00884838083523(21)US1331000041 US1202010229 US1202010171 US1109010099 US1201010135 US1239000022 US1107010002 US1319000067 US1203010338 US1248000083 US1331000059/ (01)00884838083523(21)US1331000059 US1244000033/ (01)00884838083523(21)US1244000033 US1244000011 US1337000070/ (01)00884838083523(21)US1337000070 US1203010359 US1319000080 US1319000065 US1304000012 US1234000053 US1234000040 US1337000074 US1203010402 US1251000018 US1202010247 US1319000017 US1248000011/ (01)00884838083523(21)US1248000011 US1224000045 US1311000072 US1248000005 US1107010006/ (01)00884838083523(21)US1107010006 US1244000038 US1203010278/ (01)00884838083530(21)US1203010278 US1311000014 US1311000049 US1319000031/ (01)00884838083523(21)US1319000031 US1319000036 US1251000040 US1202010156 US1251000027 Model Number: 453564483321 Serial Numbers/UDI (if available): US1518000781/ (01)00884838083523(21)US1518000781 US1550001457/ (01)00884838083523(21)US1550001457 US1550001457/ (01)00884838083523(21)US1550001483 US1407000013 US1611001885 US1611001923 US1613001945 US1613001952 US1529001016/ (01)00884838083523(21)US1529001016 US1352000030/ (01)00884838083523(21)US1352000030 US1417000027 US1408000006/ (01)00884838083523(21)US1408000006 US1502000443/ (01)00884838083523(21)US1502000443 US1505000555/ (01)00884838083523(21)US1505000555 US1603001565/ (01)00884838083523(21)US1603001565 US1608001633/ (01)00884838083523(21)US1608001633 US1107010032 US1109010106 US1202010164 US1202010243 US1231000085 US1234000032 US1234000084 US1239000004 US1239000013 US1239000018 US1239000036 US1239000042 US1239000051 US1239000052 US1244000009 US1244000017 US1244000023 US1244000029 US1244000034 US1244000043 US1248000012 US1248000044 US1248000080 US1248000082 US1251000008 US1251000024 US1251000026 US1251000057 US1251000059 US1251000065 US1251000076 US1251000079 US1251000082 US1251000088 US1304000009 US1319000077 US1319000082 US1324000016 US1324000037 US1324000041 US1324000045 US1325000006 US1325000028 US1328000009 US1328000015 US1334000011 US1334000013 US1334000020 US1334000031 US1337000001 US1337000005 US1337000013 US1337000016 US1337000017 US1337000035 US1337000039 US1337000045 US1337000084 US1343000041 US1343000054 US1343000068 US1343000072 US1343000079 US1408000034 US1419000017 US1431000034 US1433000046 US1517000690 US1529001028 US1529001056 US1638002571/ (01)00884838082113(21)US1638002571 US1109010101/ (01)00884838083523(21)US1109010101 US1224000048/ (01)00884838083523(21)US1224000048 US1231000013/ (01)00884838083523(21)US1231000013 US1231000018/ (01)00884838083523(21)US1231000018 US1239000017/ (01)00884838083523(21)US1239000017 US1244000026/ (01)00884838083523(21)US1244000026 US1248000021/ (01)00884838083523(21)US1248000021 US1319000049/ (01)00884838083523(21)US1319000049 US1337000056/ (01)00884838083523(21)US1337000056 US1352000016/ (01)00884838083523(21)US1352000016 US1352000066/ (01)00884838083523(21)US1352000066 US1407000014/ (01)00884838083523(21)US1407000014 US1417000013/ (01)00884838083523(21)US1417000013 US1417000040/ (01)00884838083523(21)US1417000040 US1419000043/ (01)00884838083523(21)US1419000043 US1419000058/ (01)00884838083523(21)US1419000058 US1424000012/ (01)00884838083523(21)US1424000012 US1433000085/ (01)00884838083523(21)US1433000085 US1438000135/ (01)00884838083523(21)US1438000135 US1441000154/ (01)00884838083523(21)US1441000154 US1443000215/ (01)00884838083523(21)US1443000215 US1446000264/(01)00884838083523(21)US1446000264 US1446000301/ (01)00884838083523(21)US1446000301 US1450000347/ (01)00884838083523(21)US1450000347 US1505000538/ (01)00884838083523(21)US1505000538 US1505000554/ (01)00884838083523(21)US1505000554 US1511000603/ (01)00884838083523(21)US1511000603 US1517000662/ (01)00884838083523(21)US1517000662 US1518000747/ (01)00884838083523(21)US1518000747 US1518000756/ (01)00884838083523(21)US1518000756 US1525000989/ (01)00884838083523(21)US1525000989 US1529001019/ (01)00884838083523(21)US1529001019 US1529001047/ (01)00884838083523(21)US1529001047 US1529001075/ (01)00884838083523(21)US1529001075 US1529001081/ (01)00884838083523(21)US1529001081 US1533001114/ (01)00884838083523(21)US1533001114 US1534001202/ (01)00884838083523(21)US1534001202 US1534001205/ (01)00884838083523(21)US1534001205 US1534001231/ (01)00884838083523(21)US1534001231 US1537001273/ (01)00884838083523(21)US1537001273 US1537001307/ (01)00884838083523(21)US1537001307 US1547001368/ (01)00884838083523(21)US1547001368 US1547001390/ (01)00884838083523(21)US1547001390 US1547001432/ (01)00884838083523(21)US1547001432 US1550001495/ (01)00884838083523(21)US1550001495 US1603001547/ (01)00884838083523(21)US1603001547 US1604001578/ (01)00884838083523(21)US1604001578 US1608001659/ (01)00884838083523(21)US1608001659 US1609001690/ (01)00884838083523(21)US1609001690 US1609001705/ (01)00884838083523(21)US1609001705 US1611001829/ (01)00884838083523(21)US1611001829 US1611001834/ (01)00884838083523(21)US1611001834 US1611001852 /(01)00884838083523(21)US1611001852 US1617002011 /(01)00884838083523(21)US1617002011 US1623002178 /(01)00884838083523(21)US1623002178 US1623002184 /(01)00884838083523(21)US1623002184 US1343000003 US1352000034 US1352000037 US1352000061 US1407000037 US1412000044 US1417000005 US1419000028 US1424000007 US1424000013 US1431000016 US1433000052 US1438000091 US1443000211 US1443000223 US1502000465 US1505000530 US1511000580 US1511000602 US1511000608 US1511000614 US1511000643 US1517000658 US1518000764 US1518000796 US1525000923 US1529001044 US1534001153 US1537001286 US1547001404 US1547001429 US1604001592 US1609001723 US1610001764 US1610001782 US1611001855 US1623002127 US1623002134 US1625002340 US1644002650/ (01)00884838083523(21)US1644002650 US1547001385 US1511000645 US1517000655 US1408000035 US1518000789 US1638002590/ (01)00884838083523(21)US1638002590 US1638002596/ (01)00884838083523(21)US1638002596 US1547001421 US1547001426 US1547001433 US1450000345/ (01)00884838083523(21)US1450000345 US1511000574/ (01)00884838083523(21)US1511000574 US1523000873/ (01)00884838083523(21)US1523000873 US1511000599 US1203010423 US1438000120/ (01)00884838083523(21)US1438000120 US1529001057 US1623002123 US1505000496 US1623002137 US1550001447 US1550001466 US1517000668/ (01)00884838083523(21)US1517000668 US1108010054 US1417000019 US1623002115/ (01)00884838083523(21)US1623002115 US1107010038/ (01)00884838083523(21)US1107010038 US1518000786/ (01)00884838083523(21)US1518000786 US1518000801/ (01)00884838083523(21)US1518000801 US1525000912/ (01)00884838083523(21)US1525000912 US1108010056 US1412000011 US1611001893 US1604001580/ (01)00884838083523(21)US1604001580 US1417000032 US1417000038 US1443000177 US1518000784 US1529001088 US1533001103 US1537001310 US1537001314 US1446000299/ (01)00884838083523(21)US1446000299 US1511000604/ (01)00884838083523(21)US1511000604 US1625002284/ (01)00884838083523(21)US1625002284 US1412000020 US1419000006 US1419000047 US1419000056 US1419000059 US1511000635 US1518000768 US1523000839 US1547001357 US1547001420 US1419000031 US1502000380/ (01)00884838083523(21)US1502000380 US1537001288/ (01)00884838083523(21)US1537001288 US1608001646/ (01)00884838083523(21)US1608001646 US1523000885 US1547001383 US1547001438 US1550001446 US1608001650 US1608001657 US1523000848/ (01)00884838083523(21)US1523000848 US1537001258 US1547001416 US1547001439 US1625002257 US1610001801 US1611001860 US1518000785/ (01)00884838083523(21)US1518000785 US1625002237/ (01)00884838083523(21)US1625002237 US1623002154 US1203010314/ (01)00884838083523(21)US1203010314 US1412000046/ (01)00884838083523(21)US1412000046 US1443000169 US1446000295 US1505000537 US1505000545 US1511000592 US1517000650 US1517000673 US1517000687 US1517000709 US1529001049 US1529001059 US1331000026/ (01)00884838083523(21)US1331000026 US1534001181 US1624002222 US1511000633 US1525000921 US1525000939 US1523000828 US1502000472 US1517000735 US1525000904 US1547001397/ (01)00884838083523(21)US1547001397 US1619002060 US1619002061 US1619002102/ (01)00884838083523(21)US1619002102 US1446000257 US1529001032 US1625002238 US1502000453 US1638002614/ (01)00884838083523(21)US1638002614 US1638002547/ (01)00884838083530(21)US1638002547 US1638002577/ (01)00884838083530(21)US1638002577 US1517000683 US1517000699 US1537001298 US1518000765/ (01)00884838083523(21)US1518000765 US1502000434 US1518000783 US1518000795 US1638002559/ (01)00884838083523(21)US1638002559 US1525000922 US1525000961 US1443000222 US1604001623/ (01)00884838083523(21)US1604001623 US1352000056 US1523000849 US1611001902 US1505000506 US1505000510 US1505000550 US1319000025/ (01)00884838083523(21)US1319000025 US1517000719/ (01)00884838083523(21)US1517000719 US1537001318 US1523000809 US1626002421/ (01)00884838083523(21)US1626002421 US1419000001 US1419000002 US1419000019 US1529000992 US1529001051 US1604001616 US1625002266 US1502000462/ (01)00884838083523(21)US1502000462 US1438000116 US1143000182/ (01)00884838083523(21)US1143000182 US1511000626 US1511000640 US1534001235 US1534001238 US1534001240 US1534001241 US1537001261 US1547001415 US1550001449 US1617002020 US1611001868 US1613001927 US1517000737 US1625002240 US1625002242 US1626002401 US1626002407 US1623002132 US1623002158 US1623002186 US1623002193 US1624002212 US1502000454/ (01)00884838083523(21)US1502000454 US1443000197 US1638002615/ (01)00884838083523(21)US1638002615 US1644002543/ (01)00884838083523(21)US1644002543 US1644002675/ (01)00884838083523(21)US1644002675 US1611001872 US1611001908 US1625002298 US1638002537/ (01)00884838083523(21)US1638002537 US1638002548/ (01)00884838083523(21)US1638002548 US1638002550/ (01)00884838083523(21)US1638002550 US1638002610/ (01)00884838083523(21)US1638002610 US1412000049 US1623002109 US1623002152 US1624002213 US1625002227 US1625002234 US1625002241 US1625002244 US1625002263 US1550001482 US1550001488 US1609001716/ (01)00884838083523(21)US1609001716 US1424000036 US1417000034 US1419000021 US1424000002 US1438000115 US1523000867 US1604001582 US1604001590 US1331000020/ (01)00884838083523(21)US1331000020 US1407000052/ (01)00884838083523(21)US1407000052 US1502000467/ (01)00884838083523(21)US1502000467 US1523000864/ (01)00884838083523(21)US1523000864 US1529001064/ (01)00884838083523(21)US1529001064 US1431000037 US1433000049 US1433000050 US1505000536 US1608001628 US1517000689/ (01)00884838083523(21)US1517000689 US1547001398/ (01)00884838083523(21)US1547001398 US1412000036 US1517000667 US1611001871 US1625002275 US1625002277 US1604001600 US1604001614 US1613001936 US1613001960 US1644002624/ (01)00884838083523(21)US1644002624 US1644002638/ (01)00884838083523(21)US1644002638 US1623002133 US1623002163 US1623002171 US1624002216 US1624002219 US1408000025 US1324000009/ (01)00884838083523(21)US1324000009 US1419000009/ (01)00884838083523(21)US1419000009 US1446000240/ (01)00884838083523(21)US1446000240 US1511000637/ (01)00884838083523(21)US1511000637 US1517000686/ (01)00884838083523(21)US1517000686 US1523000820/ (01)00884838083523(21)US1523000820 US1529001037/ (01)00884838083523(21)US1529001037 US1529001045/ (01)00884838083523(21)US1529001045 US1529001054/ (01)00884838083523(21)US1529001054 US1529001013 US1529001072 US1550001475 US1533001102 US1610001805/ (01)00884838083523(21)US1610001805 US1352000043/ (01)00884838083523(21)US1352000043 US1424000049/ (01)00884838083523(21)US1424000049 US1443000198/ (01)00884838083523(21)US1443000198 US1511000596/ (01)00884838083523(21)US1511000596 US1523000837/ (01)00884838083523(21)US1523000837 US1608001631/ (01)00884838083523(21)US1608001631 US1352000005 US1412000037 US1417000014 US1505000559 US1604001569 US1604001608 US1352000033 US1604001603 US1604001615 US1446000254 US1446000256 US1446000266 US1502000397 US1529001087 US1529001089 US1603001541 US1547001366 US1608001658 US1547001414 US1525000990 US1529000997 US1529001055 US1502000474 US1534001210/ (01)00884838083523(21)US1534001210 US1623002143/ (01)00884838083523(21)US1623002143 US1505000525 US1517000674/ (01)00884838083523(21)US1517000674 US1534001221/ (01)00884838083523(21)US1534001221 US1608001671/ (01)00884838083523(21)US1608001671 US1424000042 US1518000802 US1547001407 US1534001179/ (01)00884838083523(21)US1534001179 US1625002232/ (01)00884838083523(21)US1625002232 US1502000446 US1517000652/ (01)00884838083523(21)US1517000652 US1611001847/ (01)00884838083523(21)US1611001847 US1638002581/ (01)00884838083523(21)US1638002581 US1619002050 US1523000888/ (01)00884838083523(21)US1523000888 US1511000624 US1518000745/ (01)00884838083523(21)US1518000745 US1244000039/ (01)00884838083523(21)US1244000039 US1343000031 US1408000050 US1550001494 US1604001583 US1609001692 US1609001696 US1609001699 US1609001691 US1626002431/ (01)00884838083523(21)US1626002431 US1626002454/ (01)00884838083523(21)US1626002454 US1352000067 US1408000026 US1533001098 US1331000043/ (01)00884838083523(21)US1331000043 US1511000605 US1533001100 US1610001753 US1433000074 US1433000087 US1523000872 US1502000471/ (01)00884838083523(21)US1502000471 US1417000037 US1511000584 US1511000618 US1433000054 US1518000780 US1109010102/ (01)00884838083523(21)US1109010102 US1502000459/ (01)00884838083523(21)US1502000459 US1609001698/ (01)00884838083523(21)US1609001698 US1625002292 US1529001067/ (01)00884838083523(21)US1529001067 US1511000591 US1525000946 US1550001464 US1550001504 US1234000050/ (01)00884838083523(21)US1234000050 US1352000051/ (01)00884838083523(21)US1352000051 US1408000048/ (01)00884838083523(21)US1408000048 US1613001926 US1613001973 US1517000660/ (01)00884838083523(21)US1517000660 US1550001477 US1611001819 US1611001846 US1611001856 US1617001996 US1617001999 US1534001194 US1534001208 US1431000024 US1438000145/ (01)00884838083523(21)US1438000145 US1610001772 US1533001116 US1533001140 US1231000080 US1638002573/ (01)00884838083523(21)US1638002573 US1109010095/ (01)00884838083523(21)US1109010095 US1239000005/ (01)00884838083523(21)US1239000005 US1408000010/ (01)00884838083523(21)US1408000010 US1517000729/ (01)00884838083523(21)US1517000729 US1537001324/ (01)00884838083523(21)US1537001324 US1547001364/ (01)00884838083523(21)US1547001364 US1623002168/ (01)00884838083523(21)US1623002168 US1537001260 US1537001266 US1537001300 US1547001410 US1433000088/ (01)00884838083523(21)US1433000088 US1604001570 US1604001613 US1525000905 US1623002129 US1511000623/ (01)00884838083523(21)US1511000623 US1533001111 US1533001115 US1533001118 US1533001121 US1533001125 US1533001126 US1533001132 US1533001133 US1533001134 US1534001201 US1534001203 US1534001214 US1534001216 US1412000013/ (01)00884838083523(21)US1412000013 US1231000045 US1523000811 US1443000196/ (01)00884838083523(21)US1443000196 US1529001084/ (01)00884838083523(21)US1529001084 US1604001602/ (01)00884838083523(21)US1604001602 US1408000020 US1443000194 US1518000794 US1533001137/ (01)00884838083523(21)US1533001137 US1518000791 US1518000793 US1518000797 US1518000798 US1525000943 US1525000963 US1525000966 US1626002415/ (01)00884838083523(21)US1626002415 US1626002432/ (01)00884838083523(21)US1626002432 US1626002451/ (01)00884838083523(21)US1626002451 US1626002455/ (01)00884838083523(21)US1626002455 US1626002465/ (01)00884838083523(21)US1626002465 US1502000424/ (01)00884838083523(21)US1502000424 US1502000450/ (01)00884838083523(21)US1502000450 US1502000451/ (01)00884838083523(21)US1502000451 US1533001148/ (01)00884838083523(21)US1533001148 US1523000870 US1603001518 US1603001519 US1603001551 US1603001555 US1603001557 US1604001586 US1604001597 US1604001598 US1604001599 US1604001604 US1604001621 US1613001938 US1623002128 US1623002131 US1408000023 US1424000021/ (01)00884838083523(21)US1424000021 US1505000561/ (01)00884838083523(21)US1505000561 US1424000040 US1443000191 US1617001995 US1617002001 US1619002055 US1619002062 US1619002064 US1619002068 US1446000294/ (01)00884838083523(21)US1446000294 US1443000173 US1619002058 US1619002070 US1619002082 US1619002085 US1438000108/ (01)00884838083523(21)US1438000108 US1408000031 US1608001662 US1609001693 US1609001700 US1609001715 US1604001584 US1419000012 US1419000025 US1529001014 US1624002207 US1534001182 US1534001186 US1537001299 US1644002651/ (01)00884838082113(21)US1644002651 US1502000396/ (01)00884838083523(21)US1502000396 US1626002378/ (01)00884838083530(21)US1626002378 US1626002458/ (01)00884838083530(21)US1626002458 US1604001574/ (01)00884838083523(21)US1604001574 US1623002122/ (01)00884838083523(21)US1623002122 US1319000074 US1424000009 US1608001675 US1613001954 US1613001958 US1613001977 US1331000051/ (01)00884838083523(21)US1331000051 US1419000030 US1443000209 US1511000613 US1603001550 US1604001606 US1625002243 US1537001263 US1623002120 US1619002067 US1623002179 US1623002183 US1505000492 US1505000533 US1505000535 US1625002230 US1623002108 US1625002245 US1625002250 US1433000045 US1433000047 US1547001412 US1238000006/ (01)00884838083523(21)US1238000006 US1547001379 US1547001392 US1417000035/ (01)00884838083523(21)US1417000035 US1525000909/ (01)00884838083523(21)US1525000909 US1525000959/ (01)00884838083523(21)US1525000959 US1251000069 US1517000678 US1431000003/ (01)00884838083523(21)US1431000003 US1613001965 US1419000015 US1547001427 US1625002233 US1534001227 US1431000001 US1433000076 US1610001788 US1617002006 US1617002036 US1419000029/ (01)00884838083523(21)US1419000029 US1443000167 US1611001925 US1505000503 US1518000792 US1638002498/ (01)00884838083523(21)US1638002498 US1638002511/ (01)00884838083523(21)US1638002511 US1511000646 US1517000724 US1517000730 US1517000732 US1239000023/ (01)00884838083523(21)US1239000023 US1446000262 US1450000373 US1525000911 US1525000955 US1525000979 US1431000013 US1424000017 US1502000387 US1502000392 US1502000412 US1502000390 US1603001520 US1603001526 US1603001527 US1603001532 US1603001539 US1610001803 US1611001808 US1511000620/ (01)00884838083523(21)US1511000620 US1638002582/ (01)00884838083523(21)US1638002582 US1644002648/ (01)00884838083523(21)US1644002648 US1644002669/ (01)00884838083523(21)US1644002669 US1533001106 US1613001984 US1617002000 US1617002015 US1623002114 US1623002125 US1623002126 US1623002191 US1624002202 US1624002205 US1529001024/ (01)00884838083523(21)US1529001024 US1438000146/ (01)00884838083523(21)US1438000146 US1533001094 US1534001166 US1534001168 US1534001174 US1534001211 US1352000039/ (01)00884838083523(21)US1352000039 US1547001437 US1604001581 US1608001647 US1608001666 US1617002027 US1424000008/ (01)00884838083523(21)US1424000008 US1638002603/ (01)00884838083523(21)US1638002603 US1638002604/ (01)00884838083523(21)US1638002604 US1417000016 US1517000654 US1517000693 US1525000958 US1525000964 US1547001406 US1623002194 US1624002196 US1624002217 US1624002218 US1624002221 US1610001762/ (01)00884838083523(21)US1610001762 US1626002461/ (01)00884838083523(21)US1626002461 US1638002551/ (01)00884838083523(21)US1638002551 US1525000920 US1525000930 US1525000936 US1625002264 US1625002293 US1424000014/ (01)00884838083523(21)US1424000014 US1505000479/ (01)00884838083523(21)US1505000479 US1533001150/ (01)00884838083523(21)US1533001150 US1604001620 US1608001655 US1609001695 US1609001733 US1412000003/ (01)00884838083523(21)US1412000003 US1610001755/ (01)00884838083523(21)US1610001755 US1534001192 US1617002034/ (01)00884838083523(21)US1617002034 US1547001376 US1547001393 US1319000075 US1529001086 US1534001187 US1546001355 US1638002561/ (01)00884838083523(21)US1638002561 US1638002600/ (01)00884838083523(21)US1638002600 US1644002649/ (01)00884838083530(21)US1644002649 US1644002664/ (01)00884838083530(21)US1644002664 US1352000025 US1407000026 US1407000034 US1407000036 US1407000049 US1534001180 US1424000043 US1433000042 US1433000043 US1617001993 US1617002017 US1617002039 US1617002044 US1619002084 US1417000004/ (01)00884838083523(21)US1417000004 US1408000032 US1525000985 US1438000106 US1438000134 US1443000213 US1505000543 US1438000125/ (01)00884838083523(21)US1438000125 US1617001990/ (01)00884838083523(21)US1617001990 US1343000044 US1547001381 US1550001465 US1550001486 US1511000638/ (01)00884838083523(21)US1511000638 US1603001521/ (01)00884838083523(21)US1603001521 US1441000152 US1518000790 US1450000346/ (01)00884838083523(21)US1450000346 US1613001974 US1617002007 US1617002019 US1617002024 US1617002037 US1617002042 US1623002180 US1625002271 US1625002272 US1450000354 US1450000364 US1417000036/ (01)00884838083523(21)US1417000036 US1511000579 US1517000681 US1511000598 US1517000697 US1517000704 US1625002231 US1625002239 US1625002253 US1625002255 US1625002259 US1547001365 US1603001510 US1603001514 US1625002262 US1446000286/ (01)00884838083523(21)US1446000286 US1617002004 US1617002018 US1617002023 US1617002041 US1609001734/ (01)00884838083523(21)US1609001734 US1638002505/ (01)00884838083530(21)US1638002505 US1638002543/ (01)00884838083530(21)US1638002543 US1638002568/ (01)00884838083530(21)US1638002568 US1638002594/ (01)00884838083530(21)US1638002594 US1638002606/ (01)00884838083530(21)US1638002606 US1547001440 US1547001409 US1619002047 US1619002048 US1619002057 US1619002090 US1623002167 US1623002173 US1623002182 US1234000036/ (01)00884838083523(21)US1234000036 US1638002563/ (01)00884838083523(21)US1638002563 US1529001077 US1537001283 US1537001303 US1537001305 US1537001315 US1251000071/ (01)00884838083523(21)US1251000071 US1109010089 US1224000014/ (01)00884838083523(21)US1224000014 US1431000029/ (01)00884838083523(21)US1431000029 US1441000159 US1443000174 US1537001275 US1537001308 US1550001467 US1550001498 US1529001078 US1534001185 US1547001377 US1547001402 US1502000423 US1610001778/ (01)00884838083523(21)US1610001778 US1518000740 US1424000046 US1424000048 US1603001512 US1603001522 US1417000021 US1505000515 US1505000548 US1443000188 US1547001430 US1550001503/ (01)00884838083523(21)US1550001503 US1609001688/ (01)00884838083523(21)US1609001688 US1611001817/ (01)00884838083523(21)US1611001817 US1248000038 US1407000030 US1603001511 US1517000707 US1550001489 US1603001516 US1324000050/ (01)00884838083523(21)US1324000050 US1424000023/ (01)00884838083523(21)US1424000023 US1534001173/ (01)00884838083523(21)US1534001173 US1547001389/ (01)00884838083523(21)US1547001389 US1611001809/ (01)00884838083523(21)US1611001809 US1419000039 US1424000024 US1604001619 US1525000972 US1525000944 US1443000186 US1511000644 US1644002647/ (01)00884838082113(21)US1644002647 US1638002558/ (01)00884838083523(21)US1638002558 US1608001630 US1608001684 US1609001687 US1613001959 US1625002260 US1603001506 US1603001525 US1529001042/ (01)00884838083523(21)US1529001042 US1603001515 US1626002373 US1626002443 US1626002444 US1623002156 US1624002210 US1525000906/ (01)00884838083523(21)US1525000906 US1608001635/ (01)00884838083523(21)US1608001635 US1417000031 US1450000344 US1505000489 US1505000493 US1419000020/ (01)00884838083523(21)US1419000020 US1431000025/ (01)00884838083523(21)US1431000025 US1518000788 US1525000987 US1523000825/ (01)00884838083523(21)US1523000825 US1408000008 US1412000033 US1446000247 US1511000634 US1517000705 US1525000976 US1109010074 US1224000030 US1239000530 US1234000009 US1234000019 US1343000051/ (01)00884838083523(21)US1343000051 US1352000038/ (01)00884838083523(21)US1352000038 US1352000040/ (01)00884838083523(21)US1352000040 US1417000007/ (01)00884838083523(21)US1417000007 US1424000004/ (01)00884838083523(21)US1424000004 US1424000037/ (01)00884838083523(21)US1424000037 US1443000203/ (01)00884838083523(21)US1443000203 US1446000305/ (01)00884838083523(21)US1446000305 US1450000338/ (01)00884838083523(21)US1450000338 US1502000420/ (01)00884838083523(21)US1502000420 US1518000803/ (01)00884838083523(21)US1518000803 US1518000807/ (01)00884838083523(21)US1518000807 US1523000884/ (01)00884838083523(21)US1523000884 US1529000993/ (01)00884838083523(21)US1529000993 US1547001380/ (01)00884838083523(21)US1547001380 US1550001485/ (01)00884838083523(21)US1550001485 US1611001835/ (01)00884838083523(21)US1611001835 US1626002384/ (01)00884838083523(21)US1626002384 US1234000067 US1251000066 US1331000032 US1337000061 US1352000013 US1408000037 US1412000021 US1417000026 US1417000041 US1424000005 US1424000025 US1424000027 US1424000035 US1431000017 US1433000086 US1438000129 US1443000200 US1502000448 US1502000475 US1511000581 US1518000758 US1523000886 US1523000887 US1525000938 US1525000988 US1534001164 US1534001189 US1547001378 US1547001396 US1604001571 US1604001588 US1604001601 US1608001627 US1608001685 US1609001725 US1623002121 US1624002197 US1624002225 US1441000160 US1443000204 US1443000207 US1203010367 US1517000731 US1517000734 US1523000889 US1523000894 US1523000897 US1525000919 US1525000925 US1525000934 US1417000028 US1419000037 US1419000041 US1419000050 US1443000218/ (01)00884838083523(21)US1443000218 US1443000168 US1201010145 US1529001079/ (01)00884838083523(21)US1529001079 US1547001417/ (01)00884838083523(21)US1547001417 US1502000445 US1511000586/ (01)00884838083523(21)US1511000586 US1433000053 US1518000766/ (01)00884838083523(21)US1518000766 US1417000017 US1443000220 US1523000827 US1617002035 US1626002394 US1518000804/ (01)00884838083523(21)US1518000804 US1604001593 US1604001596 US1604001605 US1412000012/ (01)00884838083523(21)US1412000012 US1626002420/ (01)00884838083523(21)US1626002420 US1626002430/ (01)00884838083523(21)US1626002430 US1438000105/ (01)00884838083523(21)US1438000105 US1502000391/ (01)00884838083523(21)US1502000391 US1610001774/ (01)00884838083523(21)US1610001774 US1502000436 US1502000455 US1502000463 US1502000473 US1505000563 US1511000578 US1511000611 US1623002175 US1624002214 US1324000001/ (01)00884838083523(21)US1324000001 US1611001922/ (01)00884838083523(21)US1611001922 US1638002565/ (01)00884838083523(21)US1638002565 US1352000010/ (01)00884838083523(21)US1352000010 US1613001983/ (01)00884838083523(21)US1613001983 US1417000022 US1424000032 US1417000024 US1604001572 US1609001732/ (01)00884838083523(21)US1609001732 US1611001882/ (01)00884838083523(21)US1611001882 US1412000048 US1529001050 US1623002107 US1517000703 US1523000813 US1450000336/ (01)00884838083523(21)US1450000336 US1534001224/ (01)00884838083523(21)US1534001224 US1638002510/ (01)00884838083523(21)US1638002510 US1534001175 US1534001212 US1550001474 US1511000577/ (01)00884838083523(21)US1511000577 US1638002601/ (01)00884838083523(21)US1638002601 US1638002618/ (01)00884838083523(21)US1638002618 US1610001773 US1611001839 US1624002203 US1412000009 US1433000068 US1446000249 US1523000900 US1525000973 US1525000978 US1443000214/ (01)00884838083523(21)US1443000214 US1517000701 US1517000708 US1407000021 US1443000176 US1619002097 US1619002104 US1623002110 US1623002119 US1624002208 US1331000042/ (01)00884838083523(21)US1331000042 US1234000035 US1505000532 US1505000558 US1517000733 US1525000917 US1533001119 US1533001128 US1638002616/ (01)00884838083523(21)US1638002616 US1644002631/ (01)00884838083523(21)US1644002631 US1431000028 US1502000460 US1610001779 US1525000910/ (01)00884838083523(21)US1525000910 US1239000029 US1511000648 US1523000859 US1525000962 US1529001029 US1529001031 US1529001036 US1604001594 US1644002636/ (01)00884838083530(21)US1644002636 US1644002642/ (01)00884838083530(21)US1644002642 US1603001531 US1603001533 US1408000044/ (01)00884838083523(21)US1408000044 US1412000043/ (01)00884838083523(21)US1412000043 US1626002426/ (01)00884838083523(21)US1626002426 US1638002502/ (01)00884838083523(21)US1638002502 US1417000025/ (01)00884838083523(21)US1417000025 US1617001986 US1419000040/ (01)00884838083523(21)US1419000040 US1517000694 US1626002404/ (01)00884838083523(21)US1626002404 US1626002411/ (01)00884838083523(21)US1626002411 US1626002412/ (01)00884838083523(21)US1626002412 US1523000883 US1518000777/ (01)00884838083523(21)US1518000777 US1604001576 US1611001806 US1611001897 US1613001956 US1319000019/ (01)00884838083523(21)US1319000019 US1424000033/ (01)00884838083523(21)US1424000033 US1518000782/ (01)00884838083523(21)US1518000782 US1529001021/ (01)00884838083523(21)US1529001021 US1518000763 US1525000940 US1525000960 US1525000968 US1529001005 US1529001010 US1533001144 US1534001177 US1623002146 US1638002560/ (01)00884838083530(21)US1638002560 US1638002597/ (01)00884838083530(21)US1638002597 US1638002598/ (01)00884838083530(21)US1638002598 US1638002612/ (01)00884838083530(21)US1638002612 US1505000491 US1505000500 US1505000516 US1505000527 US1505000549 US1529001046 US1517000695 US1517000698 US1511000615/ (01)00884838083523(21)US1511000615 US1626002408/ (01)00884838083523(21)US1626002408 US1626002422/ (01)00884838083523(21)US1626002422 US1626002427/ (01)00884838083523(21)US1626002427 US1352000026/ (01)00884838083523(21)US1352000026 US1424000050/ (01)00884838083523(21)US1424000050 US1505000560/ (01)00884838083523(21)US1505000560 US1523000863/ (01)00884838083523(21)US1523000863 US1352000018 US1613001964 US1617002008 US1617002021 US1617002026 US1610001796 US1610001804 US1417000042 US1450000322 US1608001648 US1608001649 US1608001653 US1608001663 US1608001683 US1626002397/ (01)00884838083523(21)US1626002397 US1610001785 US1613001978 US1617001992 US1617002014 US1617002033 US1617002040 US1625002294 US1417000030 US1438000103 US1438000139 US1441000151 US1438000107 US1325000011/ (01)00884838083523(21)US1325000011 US1412000027/ (01)00884838083523(21)US1412000027 US1626002387 US1626002390 US1626002417 US1626002424 US1623002166/ (01)00884838083523(21)US1623002166 US1525000974 US1446000278 US1511000647 US1517000661 US1547001371 US1529001085/ (01)00884838083523(21)US1529001085 US1431000035 US1502000466 US1505000501 US1517000659 US1547001356 US1547001388 US1419000011/ (01)00884838083523(21)US1419000011 US1547001408 US1408000003/ (01)00884838083523(21)US1408000003 US1511000590 US1511000642 US1625002291 US1412000016 US1511000585 US1517000700 US1610001781 US1610001791 US1610001797 US1610001798 US1611001815 US1611001816 US1613001955 US1525000984/ (01)00884838083523(21)US1525000984 US1419000018 US1625002261 US1625002290 US1251000061 US1533001099 US1533001104 US1533001110 US1533001117 US1533001129 US1533001130 US1534001170 US1537001301 US1613001953 US1324000024/ (01)00884838083523(21)US1324000024 US1433000083 US1511000589 US1251000058 US1517000710 US1525000942 US1609001724 US1610001748 US1610001775 US1419000036/ (01)00884838083523(21)US1419000036 US1450000369/ (01)00884838083523(21)US1450000369 US1626002418/ (01)00884838083523(21)US1626002418 US1626002457/ (01)00884838083523(21)US1626002457 US1626002466/ (01)00884838083523(21)US1626002466 US1417000020/ (01)00884838083523(21)US1417000020 US1623002149/ (01)00884838083523(21)US1623002149 US1617001994 US1617002032 US1625002281 US1224000018 US1518000741/ (01)00884838083523(21)US1518000741 US1611001919 US1613001944 US1613001972 US1613001975 US1610001771/ (01)00884838083523(21)US1610001771 US1625002256/ (01)00884838083523(21)US1625002256 US1502000402 US1502000429 US1502000432 US1311000028 US1626002385/ (01)00884838083523(21)US1626002385 US1626002393/ (01)00884838083523(21)US1626002393 US1626002402/ (01)00884838083523(21)US1626002402 US1626002409/ (01)00884838083523(21)US1626002409 US1626002435/ (01)00884838083523(21)US1626002435 US1107010026 US1450000376 US1502000399 US1441000161 US1525000975 US1529000999 US1529001000 US1529001001 US1529001004 US1433000066 US1417000002/ (01)00884838083523(21)US1417000002 US1638002554/ (01)00884838083523(21)US1638002554 US1638002576/ (01)00884838083523(21)US1638002576 US1638002589/ (01)00884838083523(21)US1638002589 US1638002592/ (01)00884838083523(21)US1638002592 US1533001093 US1638002585/ (01)00884838083523(21)US1638002585 US1638002587/ (01)00884838083523(21)US1638002587 US1644002633/ (01)00884838083523(21)US1644002633 US1441000157 US1419000005 US1505000557 US1446000273 US1446000285 US1446000291 US1644002632/ (01)00884838082113(21)US1644002632 US1626002434 US1626002459 US1419000053 US1608001634 US1609001697 US1609001726 US1609001731 US1505000573/ (01)00884838083523(21)US1505000573 US1537001321/ (01)00884838083523(21)US1537001321 US1525000956 US1547001358 US1547001370 US1547001373 US1547001391 US1547001401 US1547001435 US1617002013 US1617002016 US1638002562/ (01)00884838083523(21)US1638002562 US1638002602/ (01)00884838083523(21)US1638002602 US1534001199 US1611001883 US1502000417/ (01)00884838083523(21)US1502000417 US1408000016 US1446000274 US1502000440 US1352000007 US1412000026 US1412000030 US1523000810 US1547001423 US1623002157 US1623002161 US1624002211 US1625002251 US1625002274 US1625002287 US1239000012 US1625002312 US1625002318 US1625002320 US1626002372 US1608001680/ (01)00884838083523(21)US1608001680 US1446000259 US1450000361 US1638002488/ (01)00884838083523(21)US1638002488 US1638002497/ (01)00884838083523(21)US1638002497 US1611001875 US1613001937 US1443000217/ (01)00884838083523(21)US1443000217 US1611001870 US1611001877 US1611001901 US1613001929 US1613001948 US1438000136 US1623002124 US1231000076/ (01)00884838083523(21)US1231000076 US1534001183/ (01)00884838083523(21)US1534001183 US1502000464 US1523000817 US1537001325 US1608001639 US1417000008/ (01)00884838083523(21)US1417000008 US1610001760/ (01)00884838083523(21)US1610001760 US1625002330/ (01)00884838083523(21)US1625002330 US1613001966/ (01)00884838083523(21)US1613001966 US1419000038/ (01)00884838083523(21)US1419000038 US1505000562/ (01)00884838083523(21)US1505000562 US1608001670 US1611001910 US1611001918 US1443000190 US1523000896/ (01)00884838083523(21)US1523000896 US1529001012 US1518000778 US1424000041 US1443000199 US1446000255/ (01)00884838083523(21)US1446000255 US1446000269/ (01)00884838083523(21)US1446000269 US1537001330/ (01)00884838083523(21)US1537001330 US1412000042 US1517000714 US1517000727 US1638002492/ (01)00884838083523(21)US1638002492 US1638002506/ (01)00884838083523(21)US1638002506 US1638002521/ (01)00884838083523(21)US1638002521 US1638002525/ (01)00884838083523(21)US1638002525 US1441000156/ (01)00884838083523(21)US1441000156 US1533001096/ (01)00884838083523(21)US1533001096 US1603001508/ (01)00884838083523(21)US1603001508 US1603001529/ (01)00884838083523(21)US1603001529 US1518000755 US1438000150/ (01)00884838083523(21)US1438000150 US1534001160 US1534001218 US1319000039 US1529001006 US1502000442/ (01)00884838083523(21)US1502000442 US1505000570 US1505000571 US1505000572 US1511000587 US1511000588 US1511000593 US1604001575 US1419000032 US1625002267 US1433000062 US1248000029/ (01)00884838083523(21)US1248000029 US1443000206/ (01)00884838083523(21)US1443000206 US1604001595/ (01)00884838083523(21)US1604001595 US1412000005 US1412000025 US1417000015 US1417000018 US1450000342 US1518000771 US1523000812 US1603001538 US1304000042/ (01)00884838083523(21)US1304000042 US1608001669/ (01)00884838083523(21)US1608001669 US1352000068 US1431000023 US1611001886 US1518000770 US1523000869 US1604001617 US1604001622 US1502000422/ (01)00884838083523(21)US1502000422 US1533001146 US1533001151 US1626002399/ (01)00884838083523(21)US1626002399 US1626002383 US1626002391 US1626002396 US1626002398 US1626002410 US1626002439 US1603001528 US1603001530 US1626002381/ (01)00884838083523(21)US1626002381 US1626002419/ (01)00884838083523(21)US1626002419 US1408000024 US1617002043 US1619002054 US1248000039/ (01)00884838083523(21)US1248000039 US1533001095 US1537001281 US1537001296 US1537001323 US1505000551 US1505000568 US1505000481 US1505000486 US1505000519 US1525000916 US1626002376 US1626002433 US1550001442 US1550001443 US1550001448 US1550001450 US1550001455 US1550001460 US1617002009 US1617002029 US1638002528/ (01)00884838083523(21)US1638002528 US1525000908/ (01)00884838083523(21)US1525000908 US1502000384 US1502000395 US1304000036 US1623002153 US1625002297 US1525000970/ (01)00884838083523(21)US1525000970 US1608001632 US1608001664 US1625002249/ (01)00884838083523(21)US1625002249 US1319000070 US1443000178 US1617001998 US1617002022 US1617002030 US1431000038 US1433000070 US1433000072 US1438000144/ (01)00884838083523(21)US1438000144 US1625002282 US1625002345 US1626002386 US1626002423 US1626002447 US1611001909 US1613001930 US1613001962 US1625002247 US1424000047 US1424000052 US1424000054 US1446000267 US1611001845 US1611001821 US1611001838 US1517000711 US1525000929 US1523000901 US1525000914 US1525000949 US1623002160/ (01)00884838083523(21)US1623002160 US1537001284 US1537001340 US1511000627/ (01)00884838083523(21)US1511000627 US1604001566/ (01)00884838083523(21)US1604001566 US1533001097/ (01)00884838083523(21)US1533001097 US1251000070 US1419000014 US1609001710 US1626002377 US1537001279 US1537001294 US1537001312 US1626002382 US1626002436 US1523000882 US1523000836 US1203010387 US1443000184 US1505000522 US1352000065/ (01)00884838083523(21)US1352000065 US1533001147 US1619002098 US1623002170 US1517000706/ (01)00884838083523(21)US1517000706 US1537001297 US1450000316 US1502000419 US1419000044/ (01)00884838083523(21)US1419000044 US1610001769/ (01)00884838083523(21)US1610001769 US1625002302/ (01)00884838083523(21)US1625002302 US1407000009 US1438000118 US1547001419 US1450000330/ (01)00884838083523(21)US1450000330 US1417000003/ (01)00884838083523(21)US1417000003 US1505000553/ (01)00884838083523(21)US1505000553 US1537001287/ (01)00884838083523(21)US1537001287 US1604001618 US1623002140 US1203010332 US1352000021 US1446000263/ (01)00884838083523(21)US1446000263 US1518000805/ (01)00884838083523(21)US1518000805 US1511000629 US1443000192 US1537001306/ (01)00884838083523(21)US1537001306 US1617002012/ (01)00884838083523(21)US1617002012 US1438000113 US1417000011 US1239000047/ (01)00884838083523(21)US1239000047 US1603001509/ (01)00884838083523(21)US1603001509 US1609001745/ (01)00884838083523(21)US1609001745 US1638002575/ (01)00884838083523(21)US1638002575 US1638002591/ (01)00884838083523(21)US1638002591 US1644002662/ (01)00884838083523(21)US1644002662 US1644002668/ (01)00884838083523(21)US1644002668 US1644002682/ (01)00884838083523(21)US1644002682 US1644002685/ (01)00884838083523(21)US1644002685 US1546001353 US1523000858/ (01)00884838083523(21)US1523000858 US1547001411/ (01)00884838083523(21)US1547001411 US1626002429/ (01)00884838083523(21)US1626002429 US1424000022 US1547001374 US1547001394 US1547001395 US1201010146 US1202010179 US1244000031 US1244000058 US1505000520 US1517000739 US1407000004/ (01)00884838083523(21)US1407000004 US1443000182/ (01)00884838083523(21)US1443000182 US1505000552/ (01)00884838083523(21)US1505000552 US1604001568/ (01)00884838083523(21)US1604001568 US1609001735/ (01)00884838083523(21)US1609001735 US1611001878/ (01)00884838083523(21)US1611001878 US1611001892/ (01)00884838083523(21)US1611001892 US1644002701/ (01)00884838083523(21)US1644002701 US1604001624 US1638002619/ (01)00884838084926(21)US1638002619 US1644002625/ (01)00884838084926(21)US1644002625 US1644002635/ (01)00884838084926(21)US1644002635 US1644002684/ (01)00884838084926(21)US1644002684 US1441000153/ (01)00884838083523(21)US1441000153 US1738002727/ (01)00884838084926(21)US1738002727 US1738002739/ (01)00884838084926(21)US1738002739 US1738002743/ (01)00884838084926(21)US1738002743 US1534001213 US1609001737 US1644002697/ (01)00884838083523(21)US1644002697 US1738002736/ (01)00884838083523(21)US1738002736 US1738002749/ (01)00884838083523(21)US1738002749 US1738002756/ (01)00884838083523(21)US1738002756 US1738002759/ (01)00884838083523(21)US1738002759 US1644002629/ (01)00884838083530(21)US1644002629 US1738002733/ (01)00884838083523(21)US1738002733 US1738002751/ (01)00884838083523(21)US1738002751 US1638002570/ (01)00884838084926(21)US1638002570 US1644002683/ (01)00884838084926(21)US1644002683 US1644002698/ (01)00884838084926(21)US1644002698 US1438000122/ (01)00884838083523(21)US1438000122 US1623002189 US1623002192 US1644002634/ (01)00884838083523(21)US1644002634 US1644002644/ (01)00884838083523(21)US1644002644 US1343000038/ (01)00884838083523(21)US1343000038 US1517000675/ (01)00884838083523(21)US1517000675 US1518000800/ (01)00884838083523(21)US1518000800 US1537001309/ (01)00884838083523(21)US1537001309 US1234000017 US1424000039 US1604001612 US1505000534/ (01)00884838083523(21)US1505000534 US1638002593/ (01)00884838083523(21)US1638002593 US1638002622/ (01)00884838083523(21)US1638002622 US1644002641/ (01)00884838083523(21)US1644002641 US1644002693/ (01)00884838083523(21)US1644002693 US1638002569/ (01)00884838083523(21)US1638002569 US1644002653/ (01)00884838083523(21)US1644002653 US1644002666/ (01)00884838083523(21)US1644002666 US1644002695/ (01)00884838083523(21)US1644002695 US1511000617 US1337000047/ (01)00884838083523(21)US1337000047 US1244000014 US1431000020 US1450000362 US1224000042 US1450000355 US1604001585/ (01)00884838083523(21)US1604001585 US1319000071/ (01)00884838083523(21)US1319000071 US1407000007/ (01)00884838083523(21)US1407000007 US1505000521 US1623002172 US1424000044/ (01)00884838083523(21)US1424000044 US1518000787 US1518000762/ (01)00884838083523(21)US1518000762 US1234000018/ (01)00884838083523(21)US1234000018 US1407000019/ (01)00884838083523(21)US1407000019 US1529001058/ (01)00884838083523(21)US1529001058 US1533001136/ (01)00884838083523(21)US1533001136 US1334000016 US1408000049 US1446000282/ (01)00884838083523(21)US1446000282 US1337000072 US1517000702 US1511000631/ (01)00884838083523(21)US1511000631 US1529001008/ (01)00884838083523(21)US1529001008 US1611001899/ (01)00884838083523(21)US1611001899 US1304000016 US1625002246 US1625002236 US1625002269 US1625002276 US1626002463/ (01)00884838083523(21)US1626002463 US1407000023/ (01)00884838083523(21)US1407000023 US1529000998 US1611001880 US1419000055/ (01)00884838083523(21)US1419000055 US1443000212/ (01)00884838083523(21)US1443000212 US1107010031/ (01)00884838083523(21)US1107010031 US1517000670 US1518000753/ (01)00884838083523(21)US1518000753 US1502000389 US1433000044 US1408000014/ (01)00884838083523(21)US1408000014 US1534001162 US1244000062 US1433000041 US1505000544 US1609001717/ (01)00884838083523(21)US1609001717 US1334000009 US1202010220 US1625002228 US1604001610/ (01)00884838083523(21)US1604001610 US1523000842 US1511000597 US1610001768 US1407000042 US1533001124/ (01)00884838083523(21)US1533001124 US1419000024 US1408000033 US1610001747/ (01)00884838083523(21)US1610001747 US1529001076 US1511000612/ (01)00884838083523(21)US1511000612 US1518000776/ (01)00884838083523(21)US1518000776 US1547001413 US1604001609 US1511000622/ (01)00884838083523(21)US1511000622 US1550001462 US1623002148 US1624002224/ (01)00884838083523(21)US1624002224 US1419000033 US1518000752 US1523000861 US1625002322 US1352000031 US1505000485/ (01)00884838083523(21)US1505000485 US1608001661/ (01)00884838083523(21)US1608001661 US1529001048 US1517000657/ (01)00884838083523(21)US1517000657 US1511000609/ (01)00884838083523(21)US1511000609 US1231000066 US1408000001 US1433000080 US1438000133 US1505000547 US1511000575 US1525000947 US1547001360/ (01)00884838083523(21)US1547001360 US1443000187 US1433000059/ (01)00884838083523(21)US1433000059 US1433000059/ (01)00884838083523(21)US1518000772 US1203010306/ (01)00884838083523(21)US1203010306 US1502000416/ (01)00884838083523(21)US1502000416 US1517000692/ (01)00884838083523(21)US1517000692 US1537001248/ (01)00884838083523(21)US1537001248 US1609001728/ (01)00884838083523(21)US1609001728 US1502000430/ (01)00884838083523(21)US1502000430 US1431000012 US1511000582 US1529001011 US1623002136 US1343000081/ (01)00884838083523(21)US1343000081 US1450000327/ (01)00884838083523(21)US1450000327 US1525000954/ (01)00884838083523(21)US1525000954 US1443000216 US1523000890 US1523000898/ (01)00884838083523(21)US1523000898 US1408000007 US1505000504 US1529001053/ (01)00884838083523(21)US1529001053 US1443000210 US1611001862/ (01)00884838083523(21)US1611001862 US1529000994/ (01)00884838083523(21)US1529000994 US1625002317/ (01)00884838083523(21)US1625002317 US1244000044/ (01)00884838083523(21)US1244000044 US1523000840/ (01)00884838083523(21)US1523000840 US1424000051/ (01)00884838083523(21)US1424000051 US1518000779/ (01)00884838083523(21)US1518000779 US1547001386/ (01)00884838083523(21)US1547001386 US1304000006 US1412000010 US1431000004 US1443000170 US1450000308 US1505000526 US1511000600 US1611001911 US1419000046/ (01)00884838083523(21)US1419000046 US1626002467/ (01)00884838083523(21)US1626002467 US1638002482/ (01)00884838083523(21)US1638002482 US1638002484/ (01)00884838083523(21)US1638002484 US1638002487/ (01)00884838083523(21)US1638002487 US1638002496/ (01)00884838083523(21)US1638002496 US1638002503/ (01)00884838083523(21)US1638002503 US1638002509/ (01)00884838083523(21)US1638002509 US1638002512/ (01)00884838083523(21)US1638002512 US1638002513/ (01)00884838083523(21)US1638002513 US1638002516/ (01)00884838083523(21)US1638002516 US1638002517/ (01)00884838083523(21)US1638002517 US1638002519/ (01)00884838083523(21)US1638002519 US1638002520/ (01)00884838083523(21)US1638002520 US1638002522/ (01)00884838083523(21)US1638002522 US1638002526/ (01)00884838083523(21)US1638002526 US1638002527/ (01)00884838083523(21)US1638002527 US1638002530/ (01)00884838083523(21)US1638002530 US1638002531/ (01)00884838083523(21)US1638002531 US1638002532/ (01)00884838083523(21)US1638002532 US1638002535/ (01)00884838083523(21)US1638002535 US1638002536/ (01)00884838083523(21)US1638002536 US1638002538/ (01)00884838083523(21)US1638002538 US1638002540/ (01)00884838083523(21)US1638002540 US1638002541/ (01)00884838083523(21)US1638002541 US1638002542/ (01)00884838083523(21)US1638002542 US1638002544/ (01)00884838083523(21)US1638002544 US1638002546/ (01)00884838083523(21)US1638002546 US1638002552/ (01)00884838083523(21)US1638002552 US1638002553/ (01)00884838083523(21)US1638002553 US1638002555/ (01)00884838083523(21)US1638002555 US1638002557/ (01)00884838083523(21)US1638002557 US1638002564/ (01)00884838083523(21)US1638002564 US1638002605/ (01)00884838083523(21)US1638002605 US1638002608/ (01)00884838083523(21)US1638002608 US1638002609/ (01)00884838083523(21)US1638002609 US1638002611/ (01)00884838083523(21)US1638002611 US1638002613/ (01)00884838083523(21)US1638002613 US1638002621/ (01)00884838083523(21)US1638002621 US1644002637/ (01)00884838083523(21)US1644002637 US1644002639/ (01)00884838083523(21)US1644002639 US1644002640/ (01)00884838083523(21)US1644002640 US1644002645/ (01)00884838083523(21)US1644002645 US1644002646/ (01)00884838083523(21)US1644002646 US1644002652/ (01)00884838083523(21)US1644002652 US1644002654/ (01)00884838083523(21)US1644002654 US1644002656/ (01)00884838083523(21)US1644002656 US1644002657/ (01)00884838083523(21)US1644002657 US1644002658/ (01)00884838083523(21)US1644002658 US1644002659/ (01)00884838083523(21)US1644002659 US1644002660/ (01)00884838083523(21)US1644002660 US1644002661/ (01)00884838083523(21)US1644002661 US1644002663/ (01)00884838083523(21)US1644002663 US1644002665/ (01)00884838083523(21)US1644002665 US1644002667/ (01)00884838083523(21)US1644002667 US1644002671/ (01)00884838083523(21)US1644002671 US1644002672/ (01)00884838083523(21)US1644002672 US1644002673/ (01)00884838083523(21)US1644002673 US1644002676/ (01)00884838083523(21)US1644002676 US1644002677/ (01)00884838083523(21)US1644002677 US1644002686/ (01)00884838083523(21)US1644002686 US1644002687/ (01)00884838083523(21)US1644002687 US1644002688/ (01)00884838083523(21)US1644002688 US1644002690/ (01)00884838083523(21)US1644002690 US1644002692/ (01)00884838083523(21)US1644002692 US1644002694/ (01)00884838083523(21)US1644002694 US1644002700/ (01)00884838083523(21)US1644002700 US1343000082 US1352000049 US1407000005 US1407000008 US1407000011 US1407000017 US1407000018 US1407000025 US1407000028 US1407000032 US1407000033 US1407000035 US1407000043 US1407000044 US1407000045 US1407000057 US1407000062 US1408000004 US1408000009 US1408000012 US1408000017 US1408000019 US1408000029 US1408000038 US1408000039 US1408000040 US1408000042 US1408000043 US1408000045 US1408000046 US1408000047 US1412000001 US1412000002 US1412000004 US1412000007 US1412000014 US1412000015 US1412000017 US1412000018 US1412000022 US1412000023 US1412000028 US1412000029 US1412000034 US1412000035 US1412000039 US1412000040 US1412000047 US1412000050 US1417000001 US1417000033 US1417000039 US1419000003 US1419000008 US1419000013 US1419000016 US1419000022 US1419000034 US1419000045 US1419000048 US1419000054 US1419000057 US1419000060 US1424000006 US1424000011 US1424000016 US1424000019 US1424000026 US1424000029 US1424000030 US1424000031 US1424000034 US1424000038 US1424000045 US1424000053 US1424000055 US1431000002 US1431000007 US1431000008 US1431000009 US1431000010 US1431000015 US1431000018 US1431000019 US1431000021 US1431000022 US1431000030 US1431000031 US1431000032 US1431000033 US1431000036 US1431000039 US1433000048 US1433000051 US1433000055 US1433000056 US1433000057 US1433000058 US1433000060 US1433000061 US1433000063 US1433000069 US1433000071 US1433000075 US1433000077 US1433000078 US1433000082 US1433000084 US1433000089 US1433000090 US1438000092 US1438000093 US1438000094 US1438000095 US1438000096 US1438000097 US1438000098 US1438000099 US1438000101 US1438000104 US1438000109 US1438000112 US1438000117 US1438000123 US1438000124 US1438000127 US1438000130 US1438000131 US1438000137 US1438000138 US1438000147 US1443000175 US1443000185 US1443000189 US1443000193 US1443000231 US1443000232 US1443000234 US1446000237 US1446000238 US1446000239 US1446000250 US1446000251 US1446000252 US1446000253 US1446000258 US1446000268 US1446000270 US1446000271 US1446000272 US1446000277 US1446000280 US1446000281 US1446000283 US1446000284 US1446000288 US1446000289 US1446000290 US1446000293 US1446000296 US1446000297 US1446000298 US1446000300 US1446000302 US1446000303 US1446000304 US1446000307 US1450000309 US1450000311 US1450000312 US1450000314 US1450000317 US1450000319 US1450000321 US1450000324 US1450000328 US1450000329 US1450000331 US1450000332 US1450000333 US1450000334 US1450000335 US1450000337 US1450000340 US1450000341 US1450000343 US1450000348 US1450000350 US1450000352 US1450000353 US1450000356 US1450000358 US1450000359 US1450000360 US1450000363 US1450000365 US1450000366 US1450000367 US1450000368 US1450000370 US1450000371 US1450000372 US1450000374 US1450000375 US1450000377 US1502000378 US1502000382 US1502000383 US1502000385 US1502000386 US1502000393 US1502000394 US1502000398 US1502000403 US1502000404 US1502000406 US1502000407 US1502000408 US1502000413 US1502000415 US1502000418 US1502000421 US1502000425 US1502000438 US1502000456 US1502000457 US1502000458 US1502000468 US1502000469 US1502000476 US1505000495 US1505000539 US1511000594 US1511000601 US1511000606 US1511000607 US1511000625 US1511000628 US1511000632 US1511000636 US1511000641 US1517000653 US1517000664 US1517000718 US1517000720 US1517000721 US1517000722 US1517000723 US1517000728 US1518000748 US1518000749 US1518000750 US1518000754 US1518000757 US1518000759 US1518000760 US1518000767 US1518000769 US1518000799 US1523000808 US1523000814 US1523000818 US1523000819 US1523000821 US1523000824 US1523000826 US1523000829 US1523000830 US1523000831 US1523000832 US1523000834 US1523000835 US1523000838 US1523000843 US1523000845 US1523000846 US1523000847 US1523000850 US1523000851 US1523000852 US1523000853 US1523000856 US1523000857 US1523000860 US1523000865 US1523000868 US1523000871 US1523000874 US1523000876 US1523000877 US1523000878 US1523000880 US1523000899 US1525000913 US1525000918 US1525000928 US1525000932 US1525000937 US1525000948 US1525000951 US1525000953 US1525000957 US1525000965 US1525000967 US1525000969 US1525000971 US1525000977 US1525000980 US1525000981 US1525000982 US1525000983 US1529000996 US1529001002 US1529001003 US1529001007 US1529001009 US1529001015 US1529001018 US1529001020 US1529001022 US1529001023 US1529001026 US1529001030 US1529001034 US1529001035 US1529001039 US1529001040 US1529001041 US1529001043 US1529001060 US1529001061 US1529001062 US1529001065 US1529001066 US1529001069 US1529001070 US1529001071 US1529001074 US1529001091 US1533001101 US1533001105 US1533001107 US1533001108 US1533001109 US1533001112 US1533001113 US1533001120 US1533001122 US1533001123 US1533001131 US1533001135 US1533001138 US1533001141 US1533001142 US1533001143 US1533001145 US1533001149 US1534001152 US1534001154 US1534001156 US1534001161 US1534001176 US1534001178 US1534001191 US1534001197 US1534001198 US1534001200 US1534001206 US1534001207 US1534001219 US1534001222 US1534001225 US1534001234 US1534001239 US1534001244 US1537001249 US1537001250 US1537001251 US1537001252 US1537001253 US1537001254 US1537001255 US1537001256 US1537001257 US1537001262 US1537001264 US1537001267 US1537001269 US1537001270 US1537001271 US1537001274 US1537001278 US1537001285 US1537001290 US1537001291 US1537001292 US1537001304 US1537001313 US1537001319 US1537001322 US1537001326 US1537001327 US1537001328 US1537001329 US1537001332 US1537001333 US1537001334 US1537001335 US1537001336 US1537001337 US1537001339 US1546001352 US1546001354 US1547001359 US1547001361 US1547001363 US1547001375 US1547001382 US1547001418 US1547001422 US1547001424 US1547001425 US1547001428 US1547001434 US1547001436 US1550001444 US1550001452 US1550001456 US1550001458 US1550001459 US1550001461 US1550001481 US1550001490 US1550001492 US1550001496 US1550001497 US1550001499 US1603001507 US1603001513 US1603001517 US1603001523 US1603001524 US1603001534 US1603001535 US1603001537 US1603001540 US1603001542 US1603001543 US1603001544 US1603001545 US1603001546 US1603001548 US1603001553 US1603001554 US1603001558 US1603001559 US1603001560 US1603001561 US1603001562 US1603001563 US1603001564 US1604001567 US1604001577 US1604001579 US1604001587 US1604001589 US1608001629 US1608001660 US1608001667 US1608001674 US1608001681 US1609001686 US1609001694 US1609001711 US1609001713 US1609001720 US1609001721 US1609001730 US1609001736 US1609001740 US1609001741 US1609001742 US1609001744 US1610001746 US1610001751 US1610001752 US1610001754 US1610001756 US1610001758 US1610001759 US1610001763 US1610001765 US1610001766 US1610001767 US1610001770 US1610001777 US1610001784 US1610001790 US1610001795 US1610001799 US1610001800 US1610001802 US1611001810 US1611001812 US1611001813 US1611001818 US1611001822 US1611001823 US1611001824 US1611001825 US1611001827 US1611001830 US1611001833 US1611001836 US1611001837 US1611001840 US1611001842 US1611001849 US1611001851 US1611001853 US1611001857 US1611001859 US1611001861 US1611001864 US1611001865 US1611001866 US1611001873 US1611001874 US1611001879 US1611001881 US1611001884 US1611001887 US1611001894 US1611001898 US1611001903 US1611001905 US1611001907 US1611001912 US1611001915 US1611001921 US1613001939 US1613001941 US1613001942 US1613001946 US1617002031 US1619002046 US1619002052 US1619002053 US1619002056 US1619002063 US1619002065 US1619002066 US1619002069 US1619002072 US1619002073 US1619002074 US1619002075 US1619002086 US1619002087 US1619002088 US1619002089 US1619002091 US1619002092 US1619002093 US1619002094 US1619002095 US1619002096 US1619002100 US1619002101 US1623002106 US1623002111 US1623002130 US1623002135 US1623002138 US1623002142 US1623002145 US1623002164 US1623002165 US1623002169 US1623002174 US1623002177 US1623002181 US1623002185 US1623002187 US1623002188 US1623002190 US1623002195
US1624002200 US1624002204 US1624002215 US1624002220 US1625002229 US1625002235 US1625002248 US1625002252 US1625002254 US1625002258 US1625002265 US1625002270 US1625002283 US1625002285 US1625002288 US1625002296 US1625002299 US1625002300 US1625002301 US1625002303 US1625002304 US1625002305 US1625002306 US1625002308 US1625002309 US1625002310 US1625002313 US1625002314 US1625002316 US1625002319 US1625002321 US1625002323 US1625002325 US1625002326 US1625002327 US1625002328 US1625002329 US1625002331 US1625002332 US1625002334 US1625002335 US1625002336 US1625002337 US1625002339 US1625002341 US1625002342 US1625002343 US1626002395 US1626002400 US1626002414 US1626002425 US1626002428 US1626002437 US1626002440 US1626002441 US1626002442 US1626002445 US1626002446 US1626002448 US1626002449 US1626002450 US1626002452 US1626002453 US1626002456 US1638002539/ (01)00884838083523(21)US1638002539 US1407000001 US1407000061 US1412000019 US1412000024 US1412000031 US1424000001 US1424000028 US1431000006 US1433000073 US1438000102 US1438000114 US1443000179 US1446000244 US1446000275 US1446000306 US1450000315 US1450000339 US1502000400 US1502000410 US1502000461 US1511000630 US1523000879 US1523000891 US1525000902 US1525000931 US1537001282 US1537001317 US1550001493 US1608001668 US1608001676 US1611001811 US1611001843 US1611001854 US1611001890 US1619002051 US1625002280 US1625002295 US1450000310 US1352000004 US1523000841 US1248000001/ (01)00884838083523(21)US1248000001 US1109010093 US1109010108 US1534001155 US1534001172 US1534001236 US1611001841 US1407000010 US1417000012 US1533001139/ (01)00884838083523(21)US1533001139 US1505000483 US1239000040 US1412000006 US1424000015 US1511000595 US1352000014 US1352000008 US1537001320 US1505000505 US1319000087/ (01)00884838083523(21)US1319000087 US1407000015/ (01)00884838083523(21)US1407000015 US1603001552 US1518000761/ (01)00884838083523(21)US1518000761 US1537001289/ (01)00884838083523(21)US1537001289 US1337000053 US1505000512 US1523000833/ (01)00884838083523(21)US1523000833 US1625002307 US1511000583 US1537001316/ (01)00884838083523(21)US1537001316 US1523000875/ (01)00884838083523(21)US1523000875 US1534001228/ (01)00884838083523(21)US1534001228 US1611001889/ (01)00884838083523(21)US1611001889 US1523000822/ (01)00884838083523(21)US1523000822 US1511000621 US1534001169/ (01)00884838083523(21)US1534001169 US1352000019 US1234000077 US1446000246 US1529001017 US1502000405/ (01)00884838083523(21)US1502000405 US1610001776 US1547001431/ (01)00884838083523(21)US1547001431 US1352000064 US1511000576 US1626002392/ (01)00884838083523(21)US1626002392 US1625002289 US1244000024/ (01)00884838083523(21)US1244000024 US1511000619 US1624002201 US1523000895/ (01)00884838083523(21)US1523000895 US1529000995 US1638002620/ (01)00884838082113(21)US1638002620 US1419000051 US1511000616/ (01)00884838083523(21)US1511000616 US1604001591 US1623002139 US1446000265 US1424000020 US1523000892 US1446000292/ (01)00884838083523(21)US1446000292 US1529001025 US1603001556 US1511000610/ (01)00884838083523(21)US1511000610 US1251000074 US1441000155 US1517000688/ (01)00884838083523(21)US1517000688 US1604001611/ (01)00884838083523(21)US1604001611 US1609001689/ (01)00884838083523(21)US1609001689 US1610001761/ (01)00884838083523(21)US1610001761 US1419000010 US1419000007 US1319000058 US1611001904/ (01)00884838083523(21)US1611001904 US1419000049 US1338000004 US1547001369/ (01)00884838083523(21)US1547001369 US1523000844/ (01)00884838083523(21)US1523000844 US1424000003 US1529001038/ (01)00884838083523(21)US1529001038 US1623002162 US1523000823/ (01)00884838083523(21)US1523000823 US1626002462/ (01)00884838083523(21)US1626002462 US1446000236 US1502000411 US1517000736/ (01)00884838083523(21)US1517000736 US1604001607/ (01)00884838083523(21)US1604001607 US1343000067 US1523000862 Model Number: 453564621791 Serial Numbers/UDI (if available): US1734000898, (01)00884838084933(21)US1734000898 US1734000907, (01)00884838084933(21)US1734000907 US1734000909, (01)00884838084933(21)US1734000909 US1729000608, (01)00884838084902(21)US1729000608 US1731000696, (01)00884838084902(21)US1731000696 US1736000924, (01)00884838084902(21)US1736000924 US1736000933, (01)00884838084902(21)US1736000933 US1736000936, (01)00884838084902(21)US1736000936 US1736000939, (01)00884838084902(21)US1736000939 US1736000942, (01)00884838084902(21)US1736000942 US1736000956, (01)00884838084902(21)US1736000956 US1736000959, (01)00884838084902(21)US1736000959 US1736000962, (01)00884838084902(21)US1736000962 US1736000965, (01)00884838084902(21)US1736000965 US1736000967, (01)00884838084902(21)US1736000967 US1736000971, (01)00884838084902(21)US1736000971 US1736000973, (01)00884838084902(21)US1736000973 US1736000977, (01)00884838084902(21)US1736000977 US1736000998, (01)00884838084902(21)US1736000998 US1736001000, (01)00884838084902(21)US1736001000 US1736001004, (01)00884838084902(21)US1736001004 US1736001006, (01)00884838084902(21)US1736001006 US1736001009, (01)00884838084902(21)US1736001009 US1738001012, (01)00884838084902(21)US1738001012 US1738001058, (01)00884838084902(21)US1738001058 US1738001083, (01)00884838084902(21)US1738001083 US1738001085, (01)00884838084902(21)US1738001085 US1738001095, (01)00884838084902(21)US1738001095 US1746001211, (01)00884838084902(21)US1746001211 US1746001213, (01)00884838084902(21)US1746001213 US1746001237, (01)00884838084902(21)US1746001237 US1746001238, (01)00884838084902(21)US1746001238 US1724000514, (01)00884838084902(21)US1724000514 US1729000605, (01)00884838084902(21)US1729000605 US1729000631, (01)00884838084902(21)US1729000631 US1729000652, (01)00884838084902(21)US1729000652 US1731000702, (01)00884838084902(21)US1731000702 US1731000740, (01)00884838084902(21)US1731000740 US1734000790, (01)00884838084902(21)US1734000790 US1734000801, (01)00884838084902(21)US1734000801 US1734000863, (01)00884838084902(21)US1734000863 US1734000882, (01)00884838084902(21)US1734000882 US1734000888, (01)00884838084902(21)US1734000888 US1736000944, (01)00884838084902(21)US1736000944 US1736000983, (01)00884838084902(21)US1736000983 US1736000985, (01)00884838084902(21)US1736000985 US1736000987, (01)00884838084902(21)US1736000987 US1736000991, (01)00884838084902(21)US1736000991 US1736001003, (01)00884838084902(21)US1736001003 US1738001029, (01)00884838084902(21)US1738001029 US1738001120, (01)00884838084902(21)US1738001120 US1734000850, (01)00884838083523(21)US1734000850 US1734000899, (01)00884838083523(21)US1734000899 US1734000908, (01)00884838083523(21)US1734000908 US1736000919, (01)00884838083523(21)US1736000919 US1736000925, (01)00884838083523(21)US1736000925 US1738001020, (01)00884838083523(21)US1738001020 US1709000217, (01)00884838083523(21)US1709000217 US1716000288, (01)00884838083523(21)US1716000288 US1716000293, (01)00884838083523(21)US1716000293 US1716000306, (01)00884838083523(21)US1716000306 US1716000308, (01)00884838083523(21)US1716000308 US1716000312, (01)00884838083523(21)US1716000312 US1716000322, (01)00884838083523(21)US1716000322 US1716000342, (01)00884838083523(21)US1716000342 US1738001060, (01)00884838084933(21)US1738001060 US1738001079, (01)00884838084933(21)US1738001079 US1731000745, (01)00884838084933(21)US1731000745 US1731000753, (01)00884838084933(21)US1731000753 US1734000864, (01)00884838084933(21)US1734000864 US1738001118, (01)00884838084933(21)US1738001118 US1738001124, (01)00884838084933(21)US1738001124 US1729000646, (01)00884838084902(21)US1729000646 US1729000623, (01)00884838084926(21)US1729000623 US1729000640, (01)00884838084926(21)US1729000640 US1724000503, (01)00884838084926(21)US1724000503 US1724000511, (01)00884838084926(21)US1724000511 US1724000545, (01)00884838084926(21)US1724000545 US1724000547, (01)00884838084926(21)US1724000547 US1729000571, (01)00884838083523(21)US1729000571 US1729000616, (01)00884838083523(21)US1729000616 US1729000619, (01)00884838083523(21)US1729000619 US1734000816, (01)00884838083523(21)US1734000816 US1734000817, (01)00884838083523(21)US1734000817 US1729000632, (01)00884838084926(21)US1729000632 US1729000575, (01)00884838084926(21)US1729000575 US1731000677, (01)00884838084926(21)US1731000677 US1731000664, (01)00884838084902(21)US1731000664 US1731000732, (01)00884838083523(21)US1731000732 US1729000612, (01)00884838084926(21)US1729000612 US1731000685, (01)00884838084902(21)US1731000685 US1652000033, (01)00884838083523(21)US1652000033 US1724000470, (01)00884838083523(21)US1724000470 US1729000576, (01)00884838083523(21)US1729000576 US1729000565, (01)00884838084902(21)US1729000565 US1729000618, (01)00884838083523(21)US1729000618 US1731000682, (01)00884838083523(21)US1731000682 US1724000490, (01)00884838084902(21)US1724000490 US1734000792, (01)00884838083523(21)US1734000792 US1734000825, (01)00884838083523(21)US1734000825 US1738001133, (01)00884838084926(21)US1738001133 US1746001143, (01)00884838084926(21)US1746001143 US1746001181, (01)00884838084926(21)US1746001181 US1720000350, (01)00884838083523(21)US1720000350 US1720000365, (01)00884838083523(21)US1720000365 US1729000638, (01)00884838084902(21)US1729000638 US1652000021, (01)00884838083523(21)US1652000021 US1652000057, (01)00884838083523(21)US1652000057 US1652000071, (01)00884838083523(21)US1652000071 US1734000877, (01)00884838084902(21)US1734000877 US1738001077, (01)00884838084902(21)US1738001077 US1729000649, (01)00884838083523(21)US1729000649 US1731000675, (01)00884838083523(21)US1731000675 US1720000459, (01)00884838084926(21)US1720000459 US1709000206, (01)00884838083523(21)US1709000206 US1709000223, (01)00884838083523(21)US1709000223 US1734000866, (01)00884838083523(21)US1734000866 US1744001138, (01)00884838083523(21)US1744001138 US1709000207, (01)00884838083523(21)US1709000207 US1709000254, (01)00884838083523(21)US1709000254 US1729000609, (01)00884838084902(21)US1729000609 US1716000297, (01)00884838084926(21)US1716000297 US1716000326, (01)00884838084926(21)US1716000326 US1720000373, (01)00884838084926(21)US1720000373 US1734000879, (01)00884838083523(21)US1734000879 US1738001125, (01)00884838083523(21)US1738001125 US1738001129, (01)00884838083523(21)US1738001129 US1709000214, (01)00884838083523(21)US1709000214 US1709000232, (01)00884838083523(21)US1709000232 US1734000791, (01)00884838084933(21)US1734000791 US1734000830, (01)00884838084933(21)US1734000830 US1736000923, (01)00884838084933(21)US1736000923 US1729000630, (01)00884838084902(21)US1729000630 US1738001104, (01)00884838084926(21)US1738001104 US1746001193, (01)00884838084926(21)US1746001193 US1746001196, (01)00884838084926(21)US1746001196 US1746001197, (01)00884838084926(21)US1746001197 US1724000505, (01)00884838084902(21)US1724000505 US1734000871, (01)00884838084902(21)US1734000871 US1720000387, (01)00884838084926(21)US1720000387 US1720000389, (01)00884838084926(21)US1720000389 US1720000425, (01)00884838084926(21)US1720000425 US1724000541, (01)00884838083530(21)US1724000541 US1708000193, (01)00884838083523(21)US1708000193 US1716000289, (01)00884838083523(21)US1716000289 US1716000299, (01)00884838083523(21)US1716000299 US1709000210, (01)00884838083523(21)US1709000210 US1709000226, (01)00884838083523(21)US1709000226 US1734000775, (01)00884838083523(21)US1734000775 US1734000776, (01)00884838083523(21)US1734000776 US1734000780, (01)00884838083523(21)US1734000780 US1734000783, (01)00884838083523(21)US1734000783 US1734000785, (01)00884838083523(21)US1734000785 US1734000786, (01)00884838083523(21)US1734000786 US1734000807, (01)00884838083523(21)US1734000807 US1734000814, (01)00884838083523(21)US1734000814 US1734000836, (01)00884838083523(21)US1734000836 US1734000852, (01)00884838083523(21)US1734000852 US1734000870, (01)00884838083523(21)US1734000870 US1734000878, (01)00884838083523(21)US1734000878 US1724000554, (01)00884838083523(21)US1724000554 US1729000585, (01)00884838083523(21)US1729000585 US1729000614, (01)00884838083523(21)US1729000614 US1729000639, (01)00884838083523(21)US1729000639 US1709000287, (01)00884838083523(21)US1709000287 US1716000294, (01)00884838083523(21)US1716000294 US1716000332, (01)00884838083523(21)US1716000332 US1716000335, (01)00884838083523(21)US1716000335 US1716000339, (01)00884838083523(21)US1716000339 US1724000518, (01)00884838084926(21)US1724000518 US1724000524, (01)00884838084926(21)US1724000524 US1652000006, (01)00884838083530(21)US1652000006 US1652000058, (01)00884838083530(21)US1652000058 US1652000073, (01)00884838083530(21)US1652000073 US1652000069, (01)00884838083530(21)US1652000069 US1652000076, (01)00884838083530(21)US1652000076 US1652000094, (01)00884838083530(21)US1652000094 US1709000213, (01)00884838084926(21)US1709000213 US1736000979, (01)00884838084902(21)US1736000979 US1709000234, (01)00884838084926(21)US1709000234 US1720000360, (01)00884838084926(21)US1720000360 US1720000367, (01)00884838084926(21)US1720000367 US1720000372, (01)00884838084926(21)US1720000372 US1716000311, (01)00884838084926(21)US1716000311 US1716000323, (01)00884838084926(21)US1716000323 US1734000857, (01)00884838083523(21)US1734000857 US1734000862, (01)00884838083523(21)US1734000862 US1709000259, (01)00884838083523(21)US1709000259 US1652000020, (01)00884838084926(21)US1652000020 US1652000038, (01)00884838084926(21)US1652000038 US1734000894, (01)00884838083523(21)US1734000894 US1734000900, (01)00884838083523(21)US1734000900 US1734000901, (01)00884838083523(21)US1734000901 US1738001031, (01)00884838083523(21)US1738001031 US1738001046, (01)00884838083523(21)US1738001046 US1738001059, (01)00884838083523(21)US1738001059 US1738001072, (01)00884838083523(21)US1738001072 US1738001086, (01)00884838083523(21)US1738001086 US1731000703, (01)00884838084902(21)US1731000703 US1729000586, (01)00884838084902(21)US1729000586 US1724000478, (01)00884838084933(21)US1724000478 US1724000484, (01)00884838084933(21)US1724000484 US1720000374, (01)00884838084926(21)US1720000374 US1720000376, (01)00884838084926(21)US1720000376 US1729000589, (01)00884838083530(21)US1729000589 US1709000220, (01)00884838083523(21)US1709000220 US1709000230, (01)00884838083523(21)US1709000230 US1709000235, (01)00884838083523(21)US1709000235 US1709000237, (01)00884838083523(21)US1709000237 US1731000756, (01)00884838084926(21)US1731000756 US1734000884, (01)00884838084926(21)US1734000884 US1652000010, (01)00884838083523(21)US1652000010 US1708000107, (01)00884838083523(21)US1708000107 US1708000133, (01)00884838083523(21)US1708000133 US1708000142, (01)00884838083523(21)US1708000142 US1709000248, (01)00884838084902(21)US1709000248 US1652000019, (01)00884838083523(21)US1652000019 US1724000515, (01)00884838084933(21)US1724000515 US1724000519, (01)00884838084933(21)US1724000519 US1724000544, (01)00884838084933(21)US1724000544 US1724000553, (01)00884838084933(21)US1724000553 US1720000410, (01)00884838084926(21)US1720000410 US1720000442, (01)00884838084926(21)US1720000442 US1720000447, (01)00884838084926(21)US1720000447 US1720000453, (01)00884838084926(21)US1720000453 US1724000525, (01)00884838084933(21)US1724000525 US1724000533, (01)00884838084933(21)US1724000533 US1720000384, (01)00884838084933(21)US1720000384 US1720000391, (01)00884838084933(21)US1720000391 US1720000400, (01)00884838084933(21)US1720000400 US1720000412, (01)00884838084933(21)US1720000412 US1734000847, (01)00884838084933(21)US1734000847 US1734000849, (01)00884838084933(21)US1734000849 US1734000911, (01)00884838084933(21)US1734000911 US1736000915, (01)00884838084933(21)US1736000915 US1736000917, (01)00884838084933(21)US1736000917 US1708000119, (01)00884838083523(21)US1708000119 US1708000135, (01)00884838083523(21)US1708000135 US1708000138, (01)00884838083523(21)US1708000138 US1708000145, (01)00884838083523(21)US1708000145 US1708000159, (01)00884838083523(21)US1708000159 US1708000161, (01)00884838083523(21)US1708000161 US1729000655, (01)00884838083523(21)US1729000655 US1731000666, (01)00884838083523(21)US1731000666 US1731000727, (01)00884838083523(21)US1731000727 US1731000737, (01)00884838083523(21)US1731000737 US1731000746, (01)00884838083523(21)US1731000746 US1731000747, (01)00884838083523(21)US1731000747 US1731000750, (01)00884838083523(21)US1731000750 US1731000752, (01)00884838083523(21)US1731000752 US1734000768, (01)00884838083523(21)US1734000768 US1731000665, (01)00884838084902(21)US1731000665 US1652000027, (01)00884838083523(21)US1652000027 US1652000041, (01)00884838083523(21)US1652000041 US1652000061, (01)00884838083523(21)US1652000061 US1652000075, (01)00884838083523(21)US1652000075 US1652000081, (01)00884838083523(21)US1652000081 US1652000082, (01)00884838083523(21)US1652000082 US1652000085, (01)00884838083523(21)US1652000085 US1652000088, (01)00884838083523(21)US1652000088 US1652000090, (01)00884838083523(21)US1652000090 US1709000255, (01)00884838083523(21)US1709000255 US1652000014, (01)00884838083523(21)US1652000014 US1652000060, (01)00884838083523(21)US1652000060 US1716000290, (01)00884838084926(21)US1716000290 US1716000324, (01)00884838084926(21)US1716000324 US1729000577, (01)00884838083523(21)US1729000577 US1734000891, (01)00884838084902(21)US1734000891 US1729000567, (01)00884838083523(21)US1729000567 US1729000568, (01)00884838083523(21)US1729000568 US1729000569, (01)00884838083523(21)US1729000569 US1729000578, (01)00884838083523(21)US1729000578 US1729000599, (01)00884838083523(21)US1729000599 US1724000481, (01)00884838084902(21)US1724000481 US1720000455, (01)00884838084926(21)US1720000455 US1724000462, (01)00884838084926(21)US1724000462 US1731000711, (01)00884838084926(21)US1731000711 US1731000722, (01)00884838084926(21)US1731000722 US1652000029, (01)00884838083523(21)US1652000029 US1652000080, (01)00884838083523(21)US1652000080 US1738001055, (01)00884838083523(21)US1738001055 US1738001091, (01)00884838083523(21)US1738001091 US1724000528, (01)00884838084926(21)US1724000528 US1724000529, (01)00884838084926(21)US1724000529 US1734000859, (01)00884838084933(21)US1734000859 US1708000175, (01)00884838084902(21)US1708000175 US1709000264, (01)00884838084902(21)US1709000264 US1724000497, (01)00884838084933(21)US1724000497 US1724000475, (01)00884838083523(21)US1724000475 US1729000590, (01)00884838083523(21)US1729000590 US1729000591, (01)00884838083523(21)US1729000591 US1652000032, (01)00884838083530(21)US1652000032 US1652000051, (01)00884838083530(21)US1652000051 US1734000777, (01)00884838084933(21)US1734000777 US1734000781, (01)00884838084933(21)US1734000781 US1734000808, (01)00884838084933(21)US1734000808 US1729000657, (01)00884838084902(21)US1729000657 US1734000881, (01)00884838084933(21)US1734000881 US1736000961, (01)00884838084933(21)US1736000961 US1652000015, (01)00884838083523(21)US1652000015 US1652000040, (01)00884838083523(21)US1652000040 US1720000348, (01)00884838083523(21)US1720000348 US1720000349, (01)00884838083523(21)US1720000349 US1720000352, (01)00884838083523(21)US1720000352 US1720000417, (01)00884838083523(21)US1720000417 US1724000531, (01)00884838084902(21)US1724000531 US1652000044, (01)00884838083523(21)US1652000044 US1652000070, (01)00884838083523(21)US1652000070 US1652000072, (01)00884838083523(21)US1652000072 US1652000078, (01)00884838083523(21)US1652000078 US1652000079, (01)00884838083523(21)US1652000079 US1652000084, (01)00884838083523(21)US1652000084 US1652000086, (01)00884838083523(21)US1652000086 US1652000087, (01)00884838083523(21)US1652000087 US1652000089, (01)00884838083523(21)US1652000089 US1652000091, (01)00884838083523(21)US1652000091 US1652000092, (01)00884838083523(21)US1652000092 US1652000095, (01)00884838083523(21)US1652000095 US1652000096, (01)00884838083523(21)US1652000096 US1724000502, (01)00884838084926(21)US1724000502 US1724000508, (01)00884838084926(21)US1724000508 US1729000574, (01)00884838083523(21)US1729000574 US1729000584, (01)00884838083523(21)US1729000584 US1729000595, (01)00884838083523(21)US1729000595 US1709000225, (01)00884838084926(21)US1709000225 US1716000316, (01)00884838084926(21)US1716000316 US1720000369, (01)00884838084926(21)US1720000369 US1720000371, (01)00884838084926(21)US1720000371 US1731000734, (01)00884838084926(21)US1731000734 US1734000769, (01)00884838084926(21)US1734000769 US1734000778, (01)00884838083523(21)US1734000778 US1734000789, (01)00884838083523(21)US1734000789 US1734000833, (01)00884838083523(21)US1734000833 US1734000834, (01)00884838083523(21)US1734000834 US1734000838, (01)00884838083523(21)US1734000838 US1734000840, (01)00884838083523(21)US1734000840 US1734000843, (01)00884838083523(21)US1734000843 US1734000846, (01)00884838083523(21)US1734000846 US1734000854, (01)00884838083523(21)US1734000854 US1734000874, (01)00884838083523(21)US1734000874 US1736000964, (01)00884838084902(21)US1736000964 US1709000216, (01)00884838084926(21)US1709000216 US1709000250, (01)00884838084926(21)US1709000250 US1731000691, (01)00884838083523(21)US1731000691 US1731000720, (01)00884838083523(21)US1731000720 US1652000013, (01)00884838083530(21)US1652000013 US1652000077, (01)00884838083530(21)US1652000077 US1652000093, (01)00884838083530(21)US1652000093 US1729000634, (01)00884838084902(21)US1729000634 US1724000500, (01)00884838084902(21)US1724000500 US1729000570, (01)00884838084902(21)US1729000570 US1724000549, (01)00884838084926(21)US1724000549 US1736000966, (01)00884838084926(21)US1736000966 US1738001094, (01)00884838084926(21)US1738001094 US1738001106, (01)00884838084926(21)US1738001106 US1724000516, (01)00884838084926(21)US1724000516 US1724000534, (01)00884838084926(21)US1724000534 US1724000542, (01)00884838084926(21)US1724000542 US1709000249, (01)00884838083530(21)US1709000249 US1709000275, (01)00884838083530(21)US1709000275 US1720000357, (01)00884838083530(21)US1720000357 US1709000261, (01)00884838083523(21)US1709000261 US1709000284, (01)00884838083523(21)US1709000284 US1729000641, (01)00884838084902(21)US1729000641 US1734000763, (01)00884838084926(21)US1734000763 US1716000295, (01)00884838083523(21)US1716000295 US1716000321, (01)00884838083523(21)US1716000321 US1738001100, (01)00884838084926(21)US1738001100 US1746001142, (01)00884838084926(21)US1746001142 US1746001174, (01)00884838084926(21)US1746001174 US1746001180, (01)00884838084926(21)US1746001180 US1746001186, (01)00884838084926(21)US1746001186 US1729000629, (01)00884838084902(21)US1729000629 US1734000880, (01)00884838084902(21)US1734000880 US1720000356, (01)00884838084926(21)US1720000356 US1720000433, (01)00884838084926(21)US1720000433 US1652000054, (01)00884838083523(21)US1652000054 US1652000059, (01)00884838083523(21)US1652000059 US1652000067, (01)00884838083523(21)US1652000067 US1729000592, (01)00884838084902(21)US1729000592 US1729000596, (01)00884838084902(21)US1729000596 US1738001087, (01)00884838083523(21)US1738001087 US1652000012, (01)00884838083523(21)US1652000012 US1652000017, (01)00884838083523(21)US1652000017 US1652000023, (01)00884838083523(21)US1652000023 US1652000030, (01)00884838083523(21)US1652000030 US1652000048, (01)00884838083523(21)US1652000048 US1652000049, (01)00884838083523(21)US1652000049 US1652000055, (01)00884838083523(21)US1652000055 US1652000056, (01)00884838083523(21)US1652000056 US1652000065, (01)00884838083523(21)US1652000065 US1652000068, (01)00884838083523(21)US1652000068 US1708000103, (01)00884838083523(21)US1708000103 US1708000105, (01)00884838083523(21)US1708000105 US1708000108, (01)00884838083523(21)US1708000108 US1708000109, (01)00884838083523(21)US1708000109 US1708000111, (01)00884838083523(21)US1708000111 US1708000112, (01)00884838083523(21)US1708000112 US1708000117, (01)00884838083523(21)US1708000117 US1708000118, (01)00884838083523(21)US1708000118 US1708000122, (01)00884838083523(21)US1708000122 US1736000992, (01)00884838084902(21)US1736000992 US1708000144, (01)00884838083530(21)US1708000144 US1709000251, (01)00884838083530(21)US1709000251 US1724000540, (01)00884838083530(21)US1724000540 US1709000269, (01)00884838084926(21)US1709000269 US1708000150, (01)00884838083530(21)US1708000150 US1720000361, (01)00884838083530(21)US1720000361 US1720000424, (01)00884838083530(21)US1720000424 US1720000429, (01)00884838083530(21)US1720000429 US1720000432, (01)00884838083530(21)US1720000432 US1652000007, (01)00884838083523(21)US1652000007 US1652000024, (01)00884838083523(21)US1652000024 US1652000036, (01)00884838083523(21)US1652000036 US1652000053, (01)00884838083523(21)US1652000053 US1652000083, (01)00884838083523(21)US1652000083 US1724000522, (01)00884838083523(21)US1724000522 US1729000580, (01)00884838083523(21)US1729000580 US1729000622, (01)00884838083523(21)US1729000622 US1729000651, (01)00884838083523(21)US1729000651 US1709000212, (01)00884838084926(21)US1709000212 US1716000303, (01)00884838084926(21)US1716000303 US1720000366, (01)00884838084926(21)US1720000366 US1720000368, (01)00884838084926(21)US1720000368 US1736000954, (01)00884838083523(21)US1736000954 US1738001057, (01)00884838083523(21)US1738001057 US1738001062, (01)00884838083523(21)US1738001062 US1738001068, (01)00884838083523(21)US1738001068 US1738001090, (01)00884838083523(21)US1738001090 US1738001126, (01)00884838083523(21)US1738001126 US1708000114, (01)00884838083523(21)US1708000114 US1708000127, (01)00884838083523(21)US1708000127 US1708000132, (01)00884838083523(21)US1708000132 US1708000140, (01)00884838083523(21)US1708000140 US1708000146, (01)00884838083523(21)US1708000146 US1708000149, (01)00884838083523(21)US1708000149 US1708000152, (01)00884838083523(21)US1708000152 US1708000154, (01)00884838083523(21)US1708000154 US1708000163, (01)00884838083523(21)US1708000163 US1708000168, (01)00884838083523(21)US1708000168 US1708000170, (01)00884838083523(21)US1708000170 US1708000171, (01)00884838083523(21)US1708000171 US1708000173, (01)00884838083523(21)US1708000173 US1708000174, (01)00884838083523(21)US1708000174 US1708000183, (01)00884838083523(21)US1708000183 US1708000198, (01)00884838083523(21)US1708000198 US1708000200, (01)00884838083523(21)US1708000200 US1708000202, (01)00884838083523(21)US1708000202 US1709000205, (01)00884838083523(21)US1709000205 US1736001010, (01)00884838083523(21)US1736001010 US1738001032, (01)00884838083523(21)US1738001032 US1738001045, (01)00884838083523(21)US1738001045 US1738001049, (01)00884838083523(21)US1738001049 US1738001063, (01)00884838083523(21)US1738001063 US1738001067, (01)00884838083523(21)US1738001067 US1738001074, (01)00884838083523(21)US1738001074 US1738001076, (01)00884838083523(21)US1738001076 US1709000222, (01)00884838083523(21)US1709000222 US1720000370, (01)00884838083523(21)US1720000370 US1729000625, (01)00884838083523(21)US1729000625 US1729000628, (01)00884838083523(21)US1729000628 US1729000636, (01)00884838083523(21)US1729000636 US1729000637, (01)00884838083523(21)US1729000637 US1729000643, (01)00884838083523(21)US1729000643 US1729000644, (01)00884838083523(21)US1729000644 US1729000645, (01)00884838083523(21)US1729000645 US1724000493, (01)00884838084902(21)US1724000493 US1731000662, (01)00884838084902(21)US1731000662 US1709000219, (01)00884838083530(21)US1709000219 US1709000253, (01)00884838083530(21)US1709000253 US1746001185, (01)00884838084926(21)US1746001185 US1709000233, (01)00884838084926(21)US1709000233 US1709000236, (01)00884838084926(21)US1709000236 US1709000239, (01)00884838084926(21)US1709000239 US1709000240, (01)00884838084926(21)US1709000240 US1709000276, (01)00884838084926(21)US1709000276 US1734000867, (01)00884838083523(21)US1734000867 US1731000695, (01)00884838083523(21)US1731000695 US1731000697, (01)00884838083523(21)US1731000697 US1731000708, (01)00884838083523(21)US1731000708 US1731000710, (01)00884838083523(21)US1731000710 US1731000724, (01)00884838083523(21)US1731000724 US1652000031, (01)00884838083530(21)US1652000031 US1652000039, (01)00884838083530(21)US1652000039 US1652000074, (01)00884838083530(21)US1652000074 US1738001066, (01)00884838083523(21)US1738001066 US1734000893, (01)00884838084926(21)US1734000893 US1736000918, (01)00884838084926(21)US1736000918 US1709000209, (01)00884838083523(21)US1709000209 US1709000241, (01)00884838083523(21)US1709000241 US1709000271, (01)00884838083523(21)US1709000271 US1709000277, (01)00884838083523(21)US1709000277 US1709000278, (01)00884838083523(21)US1709000278 US1709000281, (01)00884838083523(21)US1709000281 US1709000282, (01)00884838083523(21)US1709000282 US1709000283, (01)00884838083523(21)US1709000283 US1709000285, (01)00884838083523(21)US1709000285 US1736000946, (01)00884838083523(21)US1736000946 US1736000950, (01)00884838083523(21)US1736000950 US1738001026, (01)00884838083523(21)US1738001026 US1734000902, (01)00884838084926(21)US1734000902 US1734000906, (01)00884838084926(21)US1734000906 US1736000997, (01)00884838084926(21)US1736000997 US1738001018, (01)00884838084926(21)US1738001018 US1738001051, (01)00884838084926(21)US1738001051 US1738001052, (01)00884838084926(21)US1738001052 US1709000280, (01)00884838084926(21)US1709000280 US1709000286, (01)00884838084926(21)US1709000286 US1716000296, (01)00884838084926(21)US1716000296 US1716000300, (01)00884838084926(21)US1716000300 US1716000301, (01)00884838084926(21)US1716000301 US1716000309, (01)00884838084926(21)US1716000309 US1716000310, (01)00884838084926(21)US1716000310 US1716000313, (01)00884838084926(21)US1716000313 US1716000315, (01)00884838084926(21)US1716000315 US1708000194, (01)00884838084926(21)US1708000194 US1734000822, (01)00884838083523(21)US1734000822 US1734000782, (01)00884838084902(21)US1734000782 US1731000728, (01)00884838083523(21)US1731000728 US1731000729, (01)00884838083523(21)US1731000729 US1709000228, (01)00884838083523(21)US1709000228 US1709000231, (01)00884838083523(21)US1709000231 US1729000650, (01)00884838084902(21)US1729000650 US1708000134, (01)00884838083523(21)US1708000134 US1709000265, (01)00884838083523(21)US1709000265 US1709000266, (01)00884838083523(21)US1709000266 US1724000488, (01)00884838083530(21)US1724000488 US1724000513, (01)00884838083530(21)US1724000513 US1724000550, (01)00884838083530(21)US1724000550 US1724000555, (01)00884838083530(21)US1724000555 US1734000803, (01)00884838084933(21)US1734000803 US1734000839, (01)00884838084933(21)US1734000839 US1736000993, (01)00884838084902(21)US1736000993 US1724000486, (01)00884838083523(21)US1724000486 US1724000491, (01)00884838083523(21)US1724000491 US1724000494, (01)00884838083523(21)US1724000494 US1724000509, (01)00884838083523(21)US1724000509 US1708000178, (01)00884838083523(21)US1708000178 US1708000181, (01)00884838083523(21)US1708000181 US1708000186, (01)00884838083523(21)US1708000186 US1709000272, (01)00884838083523(21)US1709000272 US1720000351, (01)00884838083523(21)US1720000351 US1720000362, (01)00884838083523(21)US1720000362 US1720000364, (01)00884838083523(21)US1720000364 US1731000712, (01)00884838084902(21)US1731000712 US1738001102, (01)00884838084902(21)US1738001102 US1731000671, (01)00884838084902(21)US1731000671 US1736000943, (01)00884838084902(21)US1736000943 US1729000626, (01)00884838084902(21)US1729000626 US1731000714, (01)00884838084902(21)US1731000714 US1724000498, (01)00884838084902(21)US1724000498 US1729000611, (01)00884838084902(21)US1729000611 US1736000952, (01)00884838084902(21)US1736000952 US1738001023, (01)00884838084902(21)US1738001023 US1709000218, (01)00884838084902(21)US1709000218 US1736000982, (01)00884838084902(21)US1736000982 US1716000341, (01)00884838084902(21)US1716000341 US1731000705, (01)00884838084902(21)US1731000705 US1652000002, (01)00884838084926(21)US1652000002 US1652000018, (01)00884838084926(21)US1652000018 US1652000022, (01)00884838084926(21)US1652000022 US1652000046, (01)00884838084926(21)US1652000046 US1652000050, (01)00884838084926(21)US1652000050 US1652000097, (01)00884838084926(21)US1652000097 US1708000104, (01)00884838084926(21)US1708000104 US1708000106, (01)00884838084926(21)US1708000106 US1708000121, (01)00884838084926(21)US1708000121 US1708000123, (01)00884838084926(21)US1708000123 US1708000124, (01)00884838084926(21)US1708000124 US1708000125, (01)00884838084926(21)US1708000125 US1708000128, (01)00884838084926(21)US1708000128 US1708000129, (01)00884838084926(21)US1708000129 US1708000130, (01)00884838084926(21)US1708000130 US1708000131, (01)00884838084926(21)US1708000131 US1708000136, (01)00884838084926(21)US1708000136 US1708000137, (01)00884838084926(21)US1708000137 US1708000139, (01)00884838084926(21)US1708000139 US1708000141, (01)00884838084926(21)US1708000141 US1708000143, (01)00884838084926(21)US1708000143 US1708000147, (01)00884838084926(21)US1708000147 US1708000148, (01)00884838084926(21)US1708000148 US1708000151, (01)00884838084926(21)US1708000151 US1708000153, (01)00884838084926(21)US1708000153 US1708000155, (01)00884838084926(21)US1708000155 US1708000156, (01)00884838084926(21)US1708000156 US1708000157, (01)00884838084926(21)US1708000157 US1708000158, (01)00884838084926(21)US1708000158 US1708000160, (01)00884838084926(21)US1708000160 US1708000162, (01)00884838084926(21)US1708000162 US1708000165, (01)00884838084926(21)US1708000165 US1708000166, (01)00884838084926(21)US1708000166 US1708000167, (01)00884838084926(21)US1708000167 US1708000169, (01)00884838084926(21)US1708000169 US1708000172, (01)00884838084926(21)US1708000172 US1708000176, (01)00884838084926(21)US1708000176 US1708000179, (01)00884838084926(21)US1708000179 US1708000180, (01)00884838084926(21)US1708000180 US1708000182, (01)00884838084926(21)US1708000182 US1708000185, (01)00884838084926(21)US1708000185 US1708000187, (01)00884838084926(21)US1708000187 US1708000188, (01)00884838084926(21)US1708000188 US1708000189, (01)00884838084926(21)US1708000189 US1708000191, (01)00884838084926(21)US1708000191 US1708000192, (01)00884838084926(21)US1708000192 US1708000195, (01)00884838084926(21)US1708000195 US1708000196, (01)00884838084926(21)US1708000196 US1708000197, (01)00884838084926(21)US1708000197 US1708000199, (01)00884838084926(21)US1708000199 US1708000201, (01)00884838084926(21)US1708000201 US1709000204, (01)00884838084926(21)US1709000204 US1709000208, (01)00884838084926(21)US1709000208 US1709000211, (01)00884838084926(21)US1709000211 US1709000215, (01)00884838084926(21)US1709000215 US1709000227, (01)00884838084926(21)US1709000227 US1709000229, (01)00884838084926(21)US1709000229 US1709000242, (01)00884838084926(21)US1709000242 US1709000243, (01)00884838084926(21)US1709000243 US1709000246, (01)00884838084926(21)US1709000246 US1709000247, (01)00884838084926(21)US1709000247 US1709000257, (01)00884838084926(21)US1709000257 US1709000262, (01)00884838084926(21)US1709000262 US1709000267, (01)00884838084926(21)US1709000267 US1709000270, (01)00884838084926(21)US1709000270 US1716000291, (01)00884838084926(21)US1716000291 US1716000292, (01)00884838084926(21)US1716000292 US1716000298, (01)00884838084926(21)US1716000298 US1716000302, (01)00884838084926(21)US1716000302 US1716000314, (01)00884838084926(21)US1716000314 US1716000317, (01)00884838084926(21)US1716000317 US1716000318, (01)00884838084926(21)US1716000318 US1716000319, (01)00884838084926(21)US1716000319 US1716000320, (01)00884838084926(21)US1716000320 US1716000325, (01)00884838084926(21)US1716000325 US1716000328, (01)00884838084926(21)US1716000328 US1716000330, (01)00884838084926(21)US1716000330 US1716000331, (01)00884838084926(21)US1716000331 US1716000334, (01)00884838084926(21)US1716000334 US1716000336, (01)00884838084926(21)US1716000336 US1716000337, (01)00884838084926(21)US1716000337 US1720000359, (01)00884838084926(21)US1720000359 US1720000375, (01)00884838084926(21)US1720000375 US1720000379, (01)00884838084926(21)US1720000379 US1720000380, (01)00884838084926(21)US1720000380 US1720000381, (01)00884838084926(21)US1720000381 US1720000383, (01)00884838084926(21)US1720000383 US1720000385, (01)00884838084926(21)US1720000385 US1720000388, (01)00884838084926(21)US1720000388 US1720000390, (01)00884838084926(21)US1720000390 US1720000392, (01)00884838084926(21)US1720000392 US1720000393, (01)00884838084926(21)US1720000393 US1720000394, (01)00884838084926(21)US1720000394 US1720000395, (01)00884838084926(21)US1720000395 US1720000396, (01)00884838084926(21)US1720000396 US1720000397, (01)00884838084926(21)US1720000397 US1720000398, (01)00884838084926(21)US1720000398 US1720000399, (01)00884838084926(21)US1720000399 US1720000401, (01)00884838084926(21)US1720000401 US1720000402, (01)00884838084926(21)US1720000402 US1720000403, (01)00884838084926(21)US1720000403 US1720000404, (01)00884838084926(21)US1720000404 US1720000406, (01)00884838084926(21)US1720000406 US1720000408, (01)00884838084926(21)US1720000408 US1720000411, (01)00884838084926(21)US1720000411 US1720000413, (01)00884838084926(21)US1720000413 US1720000415, (01)00884838084926(21)US1720000415 US1720000418, (01)00884838084926(21)US1720000418 US1720000420, (01)00884838084926(21)US1720000420 US1720000426, (01)00884838084926(21)US1720000426 US1720000438, (01)00884838084926(21)US1720000438 US1720000440, (01)00884838084926(21)US1720000440 US1720000443, (01)00884838084926(21)US1720000443 US1720000445, (01)00884838084926(21)US1720000445 US1720000446, (01)00884838084926(21)US1720000446 US1720000449, (01)00884838084926(21)US1720000449 US1720000450, (01)00884838084926(21)US1720000450 US1720000451, (01)00884838084926(21)US1720000451 US1720000452, (01)00884838084926(21)US1720000452 US1720000454, (01)00884838084926(21)US1720000454 US1720000460, (01)00884838084926(21)US1720000460 US1720000461, (01)00884838084926(21)US1720000461 US1724000465, (01)00884838084926(21)US1724000465 US1724000467, (01)00884838084926(21)US1724000467 US1724000468, (01)00884838084926(21)US1724000468 US1724000469, (01)00884838084926(21)US1724000469 US1724000471, (01)00884838084926(21)US1724000471 US1724000472, (01)00884838084926(21)US1724000472 US1724000476, (01)00884838084926(21)US1724000476 US1724000477, (01)00884838084926(21)US1724000477 US1724000479, (01)00884838084926(21)US1724000479 US1724000483, (01)00884838084926(21)US1724000483 US1724000504, (01)00884838084926(21)US1724000504 US1724000527, (01)00884838084926(21)US1724000527 US1724000532, (01)00884838084926(21)US1724000532 US1724000536, (01)00884838084926(21)US1724000536 US1724000556, (01)00884838084926(21)US1724000556 US1724000557, (01)00884838084926(21)US1724000557 US1720000457, (01)00884838084926(21)US1720000457 US1731000673, (01)00884838084902(21)US1731000673 US1734000796, (01)00884838084902(21)US1734000796 US1729000613, (01)00884838084902(21)US1729000613 US1731000715, (01)00884838084902(21)US1731000715 US1652000043, (01)00884838084902(21)US1652000043 US1716000307, (01)00884838084902(21)US1716000307 US1709000260, (01)00884838084902(21)US1709000260 US1729000647, (01)00884838084902(21)US1729000647 US1729000564, (01)00884838084902(21)US1729000564 US1734000762, (01)00884838084902(21)US1734000762 US1716000305, (01)00884838084902(21)US1716000305 US1724000520, (01)00884838084902(21)US1724000520 US1736000947, (01)00884838084902(21)US1736000947 Model Number 989803199561 Serial Numbers/UDI (if available): US1652000004, (01)00884838083523(21)US1652000004 US1716000327, (01)00884838083523(21)US1716000327 US1716000329, (01)00884838083523(21)US1716000329 US1716000333, (01)00884838083523(21)US1716000333 US1716000340, (01)00884838083523(21)US1716000340 US1734000892, (01)00884838083523(21)US1734000892 US1734000895, (01)00884838083523(21)US1734000895 US1738001036, (01)00884838083523(21)US1738001036 US1738001040, (01)00884838083523(21)US1738001040 US1738001043, (01)00884838083523(21)US1738001043 US1738001044, (01)00884838083523(21)US1738001044 US1738001050, (01)00884838083523(21)US1738001050 US1709000238, (01)00884838084926(21)US1709000238 US1709000252, (01)00884838084926(21)US1709000252 US1709000258, (01)00884838084926(21)US1709000258 US1729000579, (01)00884838084926(21)US1729000579 US1729000615, (01)00884838084926(21)US1729000615 US1729000617, (01)00884838084926(21)US1729000617 US1729000624, (01)00884838084926(21)US1729000624 US1729000642, (01)00884838084926(21)US1729000642 US1652000047, (01)00884838083523(21)US1652000047 US1652000001, (01)00884838083523(21)US1652000001 |
Recalling Firm/ Manufacturer |
Invivo Corporation 12151 Research Pkwy Ste 200 Orlando FL 32826-3286
|
For Additional Information Contact |
407-275-3220
|
Manufacturer Reason for Recall |
The real-time numeric value for ventricular end-diastolic pressure (EDP) displayed on the Live Display may be inaccurate. Because ventricular pressure monitoring is only performed in the cardiac catheterization procedure room using the FC2010 device, the FC2020 device, which is used in the Pre or Post-Op Holding Areas, is not impacted by this issue.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm initiated their recall by letter on March 14, 2018. The Field Safety Notice informed the consignee about:
" what the problem is and under what circumstances it can occur
" the actions that should be taken by the customer / user in order to prevent risks for patients
" the actions planned by Philips to correct the problem.
If further information or support concerning this problem is needed, the consignee was directed contact the local Philips representative: 800-669-1328 option 2, then option 3. |
Quantity in Commerce |
4375 units |
Distribution |
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico., and to the countries of : Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Canada, Chile, China, Cyprus, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Italy, Kazakhstan, Kenya, Malaysia, Mexico, Mauritius, Netherlands, New Zealand, Pakistan, Philippines, Poland, Romania, Russian Fed, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom, Utd. Arab Emir, Uzbekistan and Vietnam. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|