Date Initiated by Firm | February 21, 2018 |
Create Date | March 28, 2018 |
Recall Status1 |
Terminated 3 on April 06, 2020 |
Recall Number | Z-1262-2018 |
Recall Event ID |
79613 |
Product Classification |
unknown device name - Product Code N/A
|
Product | F-T3SYSTEM, T3 (Trumpf Table Top) System
The device is a surgical table accessory system used to allow prone positioning on OR table tops during spinal surgery. The system consists of a table top frame, arm supports, |
Code Information |
Serial Numbers: A166677 A201568 A311424 A333604 A353171 A385408 A166678 A201569 A311425 A339446 A357145 A385409 A170724 A201570 A314187 A339447 A359864 A385410 A179337 A209110 A319402 A348001 A362219 A386775 A182027 A209111 A320583 A348174 A364982 A386776 A183578 A209112 A320584 A348175 A369558 A386777 A188535 A209113 A322480 A348176 A369884 A386778 A194789 A282259 A322597 A348177 A370262 A390698 A194790 A282260 A325663 A348178 A372387 A390699 A197337 A282537 A325664 A348179 A378593 A390700 A197338 A282538 A326358 A352174 A378594 A390701 A197339 A309460 A331161 A352175 A378595 A390702 A197554 A309461 A331162 A352176 A378596 A390703 A198334 A309462 A331163 A352177 A378597 A395121 A198335 A309463 A333601 A353168 A384029 A395122 A198336 A309464 A333602 A353169 A384329 A395123 A201567 A309465 A333603 A353170 A384330 |
Recalling Firm/ Manufacturer |
Allen Medical Systems 100 Discovery Way Acton MA 01720-4481
|
For Additional Information Contact | Allen Medical Customer Service 800-433-5774 |
Manufacturer Reason for Recall | The recalling firm has become aware that the rail section clamps used to attach the leg support of the device may fail during use, which could result in critical harm to the patient. |
FDA Determined Cause 2 | Error in labeling |
Action | On February 2, 2018, the firm issued an Urgent Medical Device Correction notification to their distributors to make them aware of the device correction. The firm requested that distributors further disseminate an attached recall letter to consignees.
The communication addressed to consignees advised of the issue affecting the device. Consignees were asked to inform users that the device is for armboard use only, and not to be used with the product system's leg support. The notice informed consignees that the firm would provide new units upon completion of development along with revised Instructions for Use.
Customers with questions concerning this Urgent Medical Device Correction were referred to Allen Medical Customer Service: (800) 433-5774, 8:30AM-5PM (EST) |
Quantity in Commerce | 83 |
Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Australia, France, Saudi Arabia, Germany, and France. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|