| Class 2 Device Recall Sheath Adapter | |
Date Initiated by Firm | April 11, 2018 |
Create Date | May 30, 2018 |
Recall Status1 |
Terminated 3 on May 15, 2020 |
Recall Number | Z-2029-2018 |
Recall Event ID |
79890 |
510(K)Number | K021723 |
Product Classification |
Introducer, catheter - Product Code DYB
|
Product | Sheath Adapter with Cath-Gard Catheter Contamination Shield for use with 4 - 7 Fr. Catheters; Product codes: SA-09847 |
Code Information |
Lot Numbers: 13F17B0051 13F17E0713 13F17F0767 13F17J0010 |
Recalling Firm/ Manufacturer |
Arrow International Inc 2400 Bernville Rd Reading PA 19605-9607
|
Manufacturer Reason for Recall | Product sterility may be compromised due to unsealed packaging. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | On April 11, 2018 Arrow International mailed Urgent Medical Device Recall notifications to affected customers. Customers were instructed to take the following actions:
1. If you have affected stock, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above, so that the affected products can be returned to Arrow.
2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International.
3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter.
Customers with questions may contact your local sales representative or Customer Service at (866)246-6990. |
Quantity in Commerce | 416,055 total products |
Distribution | US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDONESIA, IRELAND, ITALY, JAPAN, KOREA, NETHERLANDS, NEW ZEALAND, PANAMA, PHILIPPINES, POLAND, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED KINGDOM |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DYB
|
|
|
|