Class 2 Device Recall Leica M320 Microscope Systems

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Leica M320 Microscope Systemssee related information
Date Initiated by FirmJune 04, 2018
Create DateJuly 26, 2018
Recall Status1 Terminated 3 on March 25, 2021
Recall NumberZ-2555-2018
Recall Event ID 80229
Product Classification Microscope, surgical - Product Code EPT
ProductLeica Microsystems products labeled as the following: Leica M320 Microscope Systems with Article Numbers 10448419, 10448420, 10448485, 10448486 Product Usage: Intended use of this device is an optical instrument for magnifying and illuminating specimens. It can be applied for observation and documentation and for human and veterinary medical treatment. Not for use in Eye Operations.
Code Information A. Article Number 10448419 (Serial Numbers 160714001, 10814001, 200814001, 200814002, 90914001, 021014001, 21214001, 31214001, 81214001, 190115001, 190115002);   B Article Number 10448420 (Serial Numbers 180514001, 180514002, 190514001, 190514002, 190514003, 190514004, 200514001, 200514002, 200514003, 200514004, 210514001, 210514002, 210514003, 210514004, 220514001, 220514002, 220514003, 260514001, 260514002, 270514001, 270514002, 270514003, 270514004, 270514005, 280514001, 280514002, 280514003, 290514001, 300514001, 300514002, 300514003, 20614001, 40614001, 50614001, 50614002, 50614003, 60614001, 90614001, 90614002, 100614001, 100614002, 100614003, 120614001, 120614002, 120614003, 130614001, 130614002, 130614003, 130614004, 90514002, 160614001, 160614002, 160614003, 160614004, 170614001, 190614001, 200614001, 200614002, 200614003, 230614001, 230614002, 230614003, 240614001, 250614001, 250614002, 250614003, 250614004, 250614005, 270614001, 300614001, 300614002, 300614003, 10714001, 10714002, 10714003, 10714004, 20714001, 20714002, 20714003, 30714001, 30714002, 40714001, 40714002, 40714003, 70714001, 70714002, 70714003, 80714001, 80714002, 80714003, 90714001, 100714001, 110714001, 140714001, 160714001, 160714002, 170714001, 170714002, 180714001, 180714002, 180714003, 210714001, 210714002, 210714003, 220714001, 220714002, 220714003, 220714004, 230714001, 250714001, 250714002, 250714003, 250714004, 250714005, 250714006, 290714001, 290714002, 290714003, 290714004, 300714001, 300714002, 300714003, 300714004, 310714001, 310714002, 310714003, 310714004, 10814001, 10814002, 040814001, 040814002, 040814003, 040814004, 050814001, 050814002, 050814003, 050814004, 060814001, 060814002, 070814001, 070814002, 070814003, 070814004, 080814001, 080814002, 120814001, 120814002, 130814001, 130814002, 130814003, 130814004, 140814001, 140814002, 140814003, 140814004, 150814001, 150814002, 150814003, 160814001, 160814002, 160814003, 180814001, 180814002, 180814003, 190814001, 190814002, 190814003, 190814004, 190814005, 190814006, 200814001, 200814002, 210814001, 210814002, 210814003, 210814004, 220814001, 220814002, 220814003, 250814001, 250814002, 270814001, 270814002, 270814003, 270814004, 270814005, 270814006, 280814001, 280814002, 280814003, 280814004, 290814001, 290814002, 10914001, 10914002, 10914003, 10914004, 20914001, 20914002, 30914001, 30914002, 30914003, 80714002, 70814003, 40914001, 40914002, 50914001, 50914002, 80914001, 80914002, 80914003, 80914004, 80914005, 90914001, 90914002, 90914003, 100914001, 100914002, 100914003, 100914004, 110914001, 110914002, 110914003, 120914001, 120914002, 120914003, 120914004, 120914005, 130914001, 130914002, 130914003, 150914001, 160914001, 160914002, 160914003, 160914004, 160914005, 170914001, 170914002, 170914003, 170914004, 170914005, 180914001, 180914002, 180914003, 180914004, 180914005, 190914001, 190914002, 190914003, 200914001, 200914002, 200914003, 200914004, 200914005, 220914001, 220914002, 220914003, 220914004, 230914001, 230914002, 90614001, 90614002, 240914001, 240914002, 240914003, 250914001, 250914002, 100614002, 100614003, 260914001, 260914002, 260914003, 260914004, 290914001, 290914002, 290914003, 300914001, 300914002, 300914003, 021014001, 021014002, 031014001, 031014002, 031014003, 031014004, 41014001, 41014002, 41014003, 41014004, 41014005, 41014006, 41014007, 61014001, 61014002, 71014001, 71014002, 71014003, 71014004, 81014001, 81014002, 81014003, 81014004, 81014005, 91014001, 91014002, 101014001, 101014002, 101014003, 101014004, 121014001, 121014002, 121014003, 121014004, 121014005, 121014006, 131014001, 141014001, 141014002, 141014003, 141014004, 151014001, 151014002, 151014003, 151014004, 161014001, 161014002, 161014003, 171014001, 171014002, 171014003, 171014004, 181014001, 181014002, 181014003, 181014004, 191014001, 191014002, 191014003, 191014004, 201014001, 201014002, 201014003, 211014001, 211014002, 211014003, 211014004, 231014001, 231014002, 231014003, 241014001, 241014002, 241014003, 241014004, 241014005, 281014001, 281014002, 281014003, 281014004, 281014005, 291014001, 291014002, 291014003, 291014004, 291014005, 291014006, 301014001, 301014002, 311014001, 311014002, 311014003, 311014004, 11114001, 11114002, 11114003, 11114004, 11114005, 31114001, 31114002, 31114003, 31114004, 31114005, 41114001, 41114002, 41114003, 51114001, 51114002, 51114003, 51114004, 81114001, 81114002, 81114003, 81114004, 101114001, 101114002, 101114003, 111114001, 111114002, 121114001, 121114002, 121114003, 121114004, 131114001, 131114002, 141114001, 141114002, 141114003, 161114001, 161114002, 161114003, 161114004, 161114005, 161114006, 171114001, 171114002, 171114003, 171114004, 181114001, 181114002, 181114003, 191114001, 191114002, 191114003, 201114001, 201114002, 201114003, 201114004, 201114005, 211114001, 211114002, 211114003, 211114004, 241114001, 241114002, 241114003, 241114004, 251114001, 251114002, 251114003, 251114004, 261114001, 261114002, 261114003, 261114004, 271114001, 271114002, 281114001, 281114002, 11214001, 11214002, 21214001, 31214001, 31214002, 31214003, 41214001, 41214002, 41214003, 51214001, 51214002, 71214001, 71214002, 81214001, 81214002, 81214003, 91214001, 91214002, 91214003, 91214004, 101214001, 101214002, 101214003, 111214001, 111214002, 111214003, 121214001, 121214002, 121214003, 121214004, 151214001, 151214002, 151214003, 151214004, 151214005, 151214006, 151214007, 151214008, 161214001, 161214002, 171214001, 171214002, 171214003, 181214001, 181214002, 181214003, 181214004, 191214001, 191214002, 191214003, 191214004, 191214005, 191214006, 211214001, 221214001, 221214002, 241214001, 241214002, 241214003, 261214001, 261214002, 261214003, 261214004, 271214001, 271214002, 271214003, 281214001, 281214002, 281214003, 20115001, 20115002, 20115003, 20115004, 50115001, 50115002, 60115001, 60115002, 60115003, 60115004, 60115005, 60115006, 70115001, 70115002, 80115001, 80115002, 90115001, 90115002, 90115003, 90115004, 110115001, 110115002, 120115002, 120115003, 120115004, 130115001, 130115002, 130115003, 130115004, 140115001, 140115002, 140115003, 140115004, 150115001, 150115002, 160115001, 160115002, 190115001, 190115002, 200115001, 200115002, 200115003, 200115004, 210115002, 210115004, 130215004, 110515005, 120115004, 130115003, 50915001);   C. Article Number 10448485 (Serial Numbers 180514001, 180514002, 230514001, 230514002, 230514003, 230514004, 260514001, 260514002, 280514001, 290514001, 20614001, 20614002, 20614003, 30614001, 30614002, 40614001, 40614002, 60614001, 60614002, 90614001, 90614002, 110614001, 110614002, 170614001, 170614002, 170614003, 170614004, 170614005, 180614001, 190614001, 190614002, 190614003, 190614004, 190614005, 200614001, 230614001, 300614001, 20714001, 30714001, 30714002, 40714001, 80714001, 90714001, 90714002, 100714001, 100714002, 110714001, 110714002, 140714001, 140714001, 170714001, 180714001, 210714001, 080814001, 080814002, 120814001, 120814002, 150814001, 220814001, 260814001, 260814002, 260814003, 260814004, 290814001, 290814002, 20914001, 20914002, 50914001, 50914002, 110914001, 130914001, 150914001, 150914002, 230914001, 240914001, 250914001, 250914002, 290914001, 300914001, 011014001, 011014002, 011014003, 011014004, 021014001, 41014001, 61014001, 91014001, 91014002, 101014001, 121014001, 151014001, 211014001, 231014001, 241014001, 241014002, 301014001, 301014002, 301014003, 311014001, 11114001, 11114002, 11114003, 41114001, 81114001, 101114001, 111114001, 111114002, 191114001, 211114001, 271114001, 271114002, 281114001, 11214001, 21214001, 21214002, 51214001, 71214001, 71214002, 71214003, 101214001, 111214001, 161214001, 161214002, 171214001, 191214001, 191214002, 221214001, 221214002, 231214001, 231214002, 231214003, 241214001, 251214001, 251214002, 251214003, 70115001, 80115001, 80115002, 150115001, 150115002, 160115001, 160115002);   D. Article Number 10448486 (Serial Numbers 110614001, 110614002, 120614001, 310114001, 060814001, 060814002, 160814001, 61014001, 121014001, 281014001, 131114001, 131114002, 181114001, 231214001)      
FEI Number 3000210133
Recalling Firm/
Manufacturer
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove IL 60089-6622
For Additional Information ContactPhilip De Guia
847-291-1125
Manufacturer Reason
for Recall
Possible failure of screws which hold the gas spring fixation bracket in place. Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionLeica Microsystems notified customers on about 06/04/2018 via URGENT FIELD SAFETY NOTICE letter. The letter informed customers that they would be contacted by a Leica Microsystems representative to arrange for the field update in the next 12 months. Customers were instructed that they could continue to use the system pending the update if they ensure a functional check is performed prior to surgery and follow user manual directions. If any anomalies are noticed prior to the correction, customers should contact their Leica representative. Customers were also instructed to provide the notification to all appropriate personnel , notify customers if the unit was transferred, and complete and return the Medical Device Correction Form. A follow-up recall notification was sent to customers on or about 06/07/2018. The follow up letter discusses the functional check prior to use and provides additional compliant information.
Quantity in Commerce8,564 Total Units
DistributionWorldwide Distribution US Nationwide in the states of : CA, CO, CT, FL, HI, ID, IL, KS, KY, LA, MD, MI, MN, MO, NC, NJ, NY, OH, PA, SD, TN, TX. International Distribution: Canada, AUSTRALIA, BELGIUM, Bolivia, BRAZIL, BULGARIA, CHILE, CHINA, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, HONGKONG, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KUWAIT, MALAYSIA, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, TAIWAN,THAILAND, TURKEY, UK
Total Product Life CycleTPLC Device Report