Class 2 Device Recall The UNICP bone fixation system

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall The UNICP bone fixation systemsee related information
Date Initiated by FirmMay 22, 2018
Date PostedJuly 16, 2018
Recall Status1 Terminated 3 on November 04, 2019
Recall NumberZ-2446-2018
Recall Event ID 80312
510(K)NumberK093914 
Product Classification Plate, fixation, bone - Product Code HRS
ProductLarge QWIX, Model Numbers: 111740S, 111745S, 111750S, 111755S, 111760S, 111765S, 111770S, 111775S, 111780S, 111785S, 111790S, 111795S, 111800S, 111805S, 111810S, 111815S, 111820S, 121740S, 121745S, 121750S, 121755S, 121760S, 121765S, 121770S, 121775S, 121780S, 121785S, 121790S, 121795S, 121800S, 121805S, 121810S, 121815S & 121820S
Code Information 111740S - Lot # F6H4, F70W, F9B8, F9MY, FAJ6, FATG, FBHN, FC9D, FD33, FD34, FD35, FDQR, FDQS, FELG, FEW0, FF8X, FGA8, FGCS, FH80 111745S - Lot # F6H5, F8S3, F9R8, FA4X, FAJ7, FB30, FBHP, FBRR, FBWE, FC0A, FE0C, FELH, FELJ, FEW1, FGCT & FHKY 111750S - Lot # F70X, F7VP, F7VQ, F7VR, F9Z0, FA4Y, FAP1, FB38, FBHQ, FC0B, FD36, FD37, FD38, FE0D, FE0E, F7VR, F9Z0, FA4Y, FAP1, FB38, FBHQ, FC0B, FD36, FD37, FD38, FE0D, FE0E, FE0F, FELK, FEW2, FF16, FF4V, FF8Y, FF8Z, FFVR, FG3M, FG3N, FH82 & FHCM 111755S - Lot # F5Q7, F6H6, F7VS, F7VT, F7VU, FAJ8, FATF, FBHR, FBRS, FCAL, FE0G, FE0H, FE0J, FELL, FELM, FF4W, FF90, FFVS, FFVT, FGCU, FH83 & FHCP 111760S - Lot # F5Q8, F6H7, F6H8, F7VV, F7VW, F9Z1, FA4Z, FAP2, FB31, FB3C, FBHS, FBWP, FD39, FD3A, FD3B, FD3C, FE0K, FE0L, FE0M, FE0N, FE0P, FEW3, FF4X, FF91-FF92, FFVU, FFVV, FFVW, FGCV & FHCR 111765S - Lot # F5Q9, F6H9, F7VX, F7VY, F7VZ, F9Z2, FA50, FAP3, FB29, FBHT, FBWF, FD3D, FD3E, FD3F, FE0Q, FE0R, FE0S, FE0T, FE0U, FF4Y, FFVX, FFVY, FHCU & FHCV 111770S - Lot # F5QA, F6QQ, F70Y, F9E9, FA51, FA5L, FB32, FB3D, FBWQ, FE0V, FE0W, FE0X, FELN, FEW4, FF4Z & FFVZ 111775S - Lot # F6HA, F70Z, F9B9, F9Z3, FA53, FB39, FBWG & FF50 111780S - Lot # F61H, F710, F8S4-F9RQ, F9Z4, FA54, FB2A, FB33 & FBWH 111785S - Lot # F61J, F6HB, F711, F8S5, F9N1, F9RR, F9Z5, FA55, FB2B, FBWJ, FBWJ/1 & FHCW 111790S - Lot # F5QB, F712, F7W0, F7W1, F7W2, F9RS, F9Z6, FA56, FB2C, FB3A & FBWK 111795S - Lot # F6HC, F7W3, F7W4, F9EA, F9RT, F9Z7, FA57, FBWL, FE0Y-FE0Z, FELP, FEW5 & FF52 111800S - Lot # F5LE, F61K, F7W5, F7W6, F9R9, F9RA, F9Z8, FA58, FAP4, FB2D, FB3B, FBWM, FDQT, FDQU, FELQ, FELR, FEW6 & FGCW 111805S - Lot # F5QC, F6HD, F8S6, F9PV, F9Z9, FA59, FAP5, FBWN, FD3G, FE10, FELS, FELT, FF55, FF93, FFW4, FH4V, FHL0 & FH4U 111810S - Lot # E62H/G, F5LF, F7W7, F9PW, FATE, FBHU, FC0C, FD3H, FD3J, FE11, FELU, FELV, FF56, FFW5, FGA9 & FGCX 111815S - Lot # F7RW, F7W8, F9EB, FAJ9, FBEA, FBRT, FD3K, FD3L, FE12, FELW, FEW7, FF57, FFW6 & FGCY 111820S - Lot # E62K/G, F5LG, F7W9, F9EC, F9RB, FBHV, FC9E, FD3M, FD3N, FDQV, FELX, FELY, FELZ & FGAA 121740S - Lot # F7WH, FAJD, FB2R, FBJ5, FD83, FE1F, FE1G, FEX3, FF9G & FGD4 121745S - Lot # F6BV, F7WJ, F9BC, FAJE, FB2S, FBJ6, FC2J, FD84, FE1H, FE1J & FFX8 121750S - Lot # F6HN, F7WK, F8YB, F9U8, FA4M, FAP7, FB2H, FBJ7, FC2K, FD85, FD86, FE1K, FE1L, FFX9, FFXA & FGD5 121755S - Lot # F6HP, F7WL, FAJF, FB2J, FBJ8, FBRV, FC2L, FD87, FE1M, FE1N, FF9H, FFXB & FHL1 121760S - Lot # F6HQ, F7WM, FAH2, FB2V, FB88, FBJ9, FD89, FE1P, FE1Q, FEX4, FHDB & FHDB/1 121765S - Lot # F6HR, F8BW, F9U9, FB2T, FBJA, FD8A, FD8B, FE1R, FE1S, FFXC & FH84 121770S - Lot # F64T, F6HS, F7WW, F9RD, FA4N, FB2W, FBWT, FE1T, FFXD, FH85 & FH85/1 121775S - Lot # F6HT, F8YC, F9PX, F9UA, F9ZD, FA4P, FBWU, FDQW, FDQX, FDQY, FDQZ, FE1U & FFXF 121780S - Lot # F6HU, F7JQ, F9UB, FA4Q, FAP8, FBJB, FBWV, FC2M, FDR0, FDR1, FDR2 & FDR2/1 121785S - Lot # F60U, F6HV, F7WN, FAJG, FBJC, FD8C, FD8D, FD8E, FE1V, FEX5, FH54 & FH54/1 121790S - Lot # F8YD, FB2X, FCAJ, FE1W, FEX6, FFXG & FH86 121795S - Lot # F6HW, F7BH, FBRW, FCCC, FE1X, FE1Y, FF9K, FGAC & FHDE 121800S - Lot # F6AR, FCAM, FD8F, FE1Z, FF9L, FF9M, FH55, FH87 & FH87/1 121805S - Lot # F6HX, F9FN, FCAK, FE20, FF9N & FGD6 121810S - Lot # F5LK, F7WP, F9RE, FD8G, FE21, FEX7, FF9P & FFXH 121815S - Lot # F7WQ, FBJD, FCCD, FD8H, FE22, FF9Q, FF9R & FGAD 121820S - Lot # F9BD, FBJE, FD8J, FD8K, FE23, FF9S & FG3P 
FEI Number 3002952164
Recalling Firm/
Manufacturer
NewDeal SA
Immeuble Sequoia 2 97 A
97 Allee Alexandre Borodineimmeuble Sequoia 2 97 Allee Alexandre Borodine
St Priest France
Manufacturer Reason
for Recall
Use of the impacted product may cause a superficial infection requiring PO antibiotics and wound care or deep infection requiring IV antibiotics and device removal.
FDA Determined
Cause 2
Packaging change control
ActionOn May 22, 2018 Newdeal SAS, a company of Integra LifeSciences, issued URGENT VOLUNTARY MEDICAL DEVICE RECALL notices to their customers. On October 22, 2018 URGENT VOLUNTARY MEDICAL DEVICE RECALL notices were issued to additional customers. Customers were advised to take the following actions: 1. Examine your inventory to determine if you have any affected lots in original product packaging identified within the notice. 2. If you do have any of the affected lots in original product packaging, stop using the product immediately, remove them from service and place in quarantine until the product can be returned. 3. Complete the Acknowledgement and Return Form and check the box: I do have affected products on the list and record the lot number. 4. Or, complete the attached Acknowledgement and Return Form and check the box: I do not have affected products. 5. Complete the other information on the form as indicated on the form. Keep a copy of the form for your records. Return the completed Acknowledgement and Return Form by email or fax indicating your receipt and review of the notification. Customers with questions regarding the instructions, please contact Customer Service at 1-800-654-2873 Monday to Friday 8:00 AM  8:00 PM EST or custsvcnj@integralife.com .
DistributionCA, CO, CT, ID, KS, MA, MD, MN, MO, MS, NC, NY, OH, OR, PA, SD, VA, WA & WI
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HRS
-
-