| Date Initiated by Firm | March 01, 2018 |
| Create Date | August 16, 2018 |
| Recall Status1 |
Terminated 3 on January 26, 2024 |
| Recall Number | Z-2806-2018 |
| Recall Event ID |
80460 |
| 510(K)Number | K964435 |
| Product Classification |
Stopcock, i.V. Set - Product Code FMG
|
| Product | Kit w/3 Ext Sets, 3 Drop-In MicroClave Clear |
| Code Information |
Item Number 011-AM6123. a. Lot Number 3393136, UDI Number (01) 1 0840619 09404 8 (17) 220101 (30) 20 (10) 3393136, Expiration Date 11/01/2022; b. Lot Number 3508527, UDI Number (01) 1 0840619 09404 8 (17) 220801 (30) 20 (10) 3508527, Expiration Date 08/01/2022 |
| FEI Number |
3003807466
|
Recalling Firm/ Manufacturer |
ICU Medical de Mexico, S.A. de C.V.
Avenida Cuarzo 250
Ensenada
Ejido Maneadero
Mexico
|
Manufacturer Reason for Recall | ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold. |
FDA Determined Cause 2 | Other |
| Action | ICU Medical notified customers of the recall on about 03/01/2018 via URGENT: Medical Device Recall Notification. The notification informed customers that this was an update and expansion to previously initiated recall. Instructions included to inspect inventory for affected devices, arrange for the return of affected devices, and complete and return the Urgent Medical Device Recall Notification Response Form to recall@icumed.com. |
| Quantity in Commerce | 302 pouches |
| Distribution | US distribution to AZ, MA, NC, OH, TX, VA, WA.
Worldwide distribution to Belgium, France, Germany. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = FMG
|