| | Class 2 Device Recall Leica M530 OHX Microsope System |  |
| Date Initiated by Firm | June 25, 2018 |
| Create Date | August 07, 2018 |
| Recall Status1 |
Terminated 3 on February 26, 2019 |
| Recall Number | Z-2659-2018 |
| Recall Event ID |
80497 |
| Product | Leica M530 OHX Microscope System Supplied with 100 - 120 Vac. |
| Code Information |
US: 90116001, 30118001, 51215001, 201117001, 30916001, 270216001, 100917001, 31117001, 00817001, 0316001, 31217001, 180118001, 200218001, 151217001, 21217001, 21214001, 80218001, 170118001, 01017002, 101017001, 170915001, 150117001, 150217001, 291117001, 260517001, 300915001, 0318001, 30318001, 100915001, 310717001, 140716001, 50516001, 281116001, 300517001, 0316001, 30115001, 200617001, 190117001, 170117001, 241217001, 261217001, 271217001, 90118001, 0617001, 250118001, 280115001, 191015001, 81117001, 150617001, 120417001, 140917001, 10318001, 140417001, 250717001, 80518001, 160616001, 170417001, 150417001, 11116001, 91017001, 81117002, 281117001, 20218001, 300118001, 131015001, 140318001, 110516001, 00318001, 160118001, 30518001, 200316001, 120917001, 210118001, 310717001, 30218001, 1217001, 190617001, 130617001, 280817001, 80317001, 310817001, 10917001, 60317001, 60218001, 80917001 230318001, 170318001; Puerto Rico: 241216001, 261216001; JAPAN: 140717001,171017001, 211117001, 291217001, 290118001, 210218001, 280218001, 020318001, 120318001, 160318001; BRAZIL: 120416001, 220917001, 311017001, 70118001, 70318001; COLOMBIA: 60517001; ECUADOR: 170316001, 231216001, 190917001; MEXICO: 31215001, 130117001, 91217001; TAIWAN: 190815001, 220317001, 190317001, 40517001, 210717001, 210717002, 211217001 |
| FEI Number |
3000210133
|
Recalling Firm/ Manufacturer |
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove
IL
60089-6622
|
| For Additional Information Contact | Philip De Guia 800-225-8867 |
Manufacturer Reason for Recall | Recall is due to a design weakness of the power insert module. |
FDA Determined Cause 2 | Device Design |
| Action | Leica Microsystems notified customers on about 06/25/2018 via URGENT FIELD SAFETY NOTICE letter. Instructions included advising customers of the issue, informing them that Leica Microsystems representative will contact them to arrange the fiel update to the device, distribute the recall notification to all individuals that should be aware of the recall or if the device was further distributed and a response form was provided to complete and return. The notification letter also instructed customers to report any anomalies with the system to Leica Microsystems if they occur prior to the correction being made and to ensure a functional check is performed prior to each surgery as described in the user manual. |
| Quantity in Commerce | 123 devices total |
| Distribution | United States distribution: AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, KY, LA, MO, NJ, NY, NV, OH, OR, PA, TX, UT, PR.
International distribution to Japan, Brazil, Columbia, Ecuador, Mexico, Taiwan |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|