Class 2 Device Recall Leica M530 OHX Microsope System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Leica M530 OHX Microsope Systemsee related information
Date Initiated by FirmJune 25, 2018
Create DateAugust 07, 2018
Recall Status1 Terminated 3 on February 26, 2019
Recall NumberZ-2659-2018
Recall Event ID 80497
ProductLeica M530 OHX Microscope System Supplied with 100 - 120 Vac.
Code Information US: 90116001, 30118001, 51215001, 201117001, 30916001, 270216001, 100917001, 31117001, 00817001, 0316001, 31217001, 180118001, 200218001, 151217001, 21217001, 21214001, 80218001, 170118001, 01017002, 101017001, 170915001, 150117001, 150217001, 291117001, 260517001, 300915001, 0318001, 30318001, 100915001, 310717001, 140716001, 50516001, 281116001, 300517001, 0316001, 30115001, 200617001, 190117001, 170117001, 241217001, 261217001, 271217001, 90118001, 0617001, 250118001, 280115001, 191015001, 81117001, 150617001, 120417001, 140917001, 10318001, 140417001, 250717001, 80518001, 160616001, 170417001, 150417001, 11116001, 91017001, 81117002, 281117001, 20218001, 300118001, 131015001, 140318001, 110516001, 00318001, 160118001, 30518001, 200316001, 120917001, 210118001, 310717001, 30218001, 1217001, 190617001, 130617001, 280817001, 80317001, 310817001, 10917001, 60317001, 60218001, 80917001 230318001, 170318001;   Puerto Rico: 241216001, 261216001;   JAPAN: 140717001,171017001, 211117001, 291217001, 290118001, 210218001, 280218001, 020318001, 120318001, 160318001;   BRAZIL: 120416001, 220917001, 311017001, 70118001, 70318001;   COLOMBIA: 60517001;   ECUADOR: 170316001, 231216001, 190917001;   MEXICO: 31215001, 130117001, 91217001;   TAIWAN: 190815001, 220317001, 190317001, 40517001, 210717001, 210717002, 211217001 
FEI Number 3000210133
Recalling Firm/
Manufacturer
Leica Microsystems, Inc.
1700 Leider Ln
Buffalo Grove IL 60089-6622
For Additional Information ContactPhilip De Guia
800-225-8867
Manufacturer Reason
for Recall
Recall is due to a design weakness of the power insert module.
FDA Determined
Cause 2
Device Design
ActionLeica Microsystems notified customers on about 06/25/2018 via URGENT FIELD SAFETY NOTICE letter. Instructions included advising customers of the issue, informing them that Leica Microsystems representative will contact them to arrange the fiel update to the device, distribute the recall notification to all individuals that should be aware of the recall or if the device was further distributed and a response form was provided to complete and return. The notification letter also instructed customers to report any anomalies with the system to Leica Microsystems if they occur prior to the correction being made and to ensure a functional check is performed prior to each surgery as described in the user manual.
Quantity in Commerce123 devices total
DistributionUnited States distribution: AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, KY, LA, MO, NJ, NY, NV, OH, OR, PA, TX, UT, PR. International distribution to Japan, Brazil, Columbia, Ecuador, Mexico, Taiwan

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-