| Date Initiated by Firm | February 13, 2017 |
| Create Date | August 03, 2018 |
| Recall Status1 |
Terminated 3 on November 07, 2018 |
| Recall Number | Z-2604-2018 |
| Recall Event ID |
80499 |
| 510(K)Number | K141253 K150647 |
| Product Classification |
Orthosis, spinal pedicle fixation - Product Code MNI
|
| Product | Modera Modular Pedicle Screw System Shank Insertion Instrument, 3-107-07 |
| Code Information |
11125 |
| FEI Number |
3010760160
|
Recalling Firm/ Manufacturer |
New Era Orthopaedics, LLc
1214 Research Blvd Ste 1019
Hummelstown
PA
17036-9153
|
| For Additional Information Contact | 717-585-6785 |
Manufacturer Reason for Recall | This is a retrospective recall of an instrument used in spine fusion procedures with Modera Modular Pedicle Screw system implants. The firm received a customer complaint in which excessive forces were applied to the shank insertion instrument and the laser weld failed. |
FDA Determined Cause 2 | Device Design |
| Action | On February 13, 2017, the firm emailed its distributors to notify them that their Rev. A devices would be replaced with Rev. B devices, and asked the distributors to mail back the Rev. A devices. |
| Quantity in Commerce | 6 pieces |
| Distribution | 2 distributors in CO and TX. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|